K251246 · Spark Biomedical, Inc. · PZR · Aug 25, 2025 · Neurology
Device Facts
Record ID
K251246
Device Name
Sparrow Ascent
Applicant
Spark Biomedical, Inc.
Product Code
PZR · Neurology
Decision Date
Aug 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5896
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.
Device Story
Transcutaneous auricular neurostimulation (tAN) system; delivers mild electrical stimulation to nerves around auricle (ABVN). Components: disposable Earpiece, Cable, External Pulse Generator (EPG). Used in clinical environments (clinics, hospitals) or home; clinician sets stimulation parameters; patient adjusts intensity (amplitude). Input: electrical pulses; Output: transcutaneous nerve stimulation. Mechanism: stimulates auricular branch of vagus nerve (ABVN) via mastoid electrode (modified from predicate cymba concha site). Output affects central nervous system to aid in reduction of opioid withdrawal symptoms. Benefits: non-invasive symptom management.
Clinical Evidence
Sham-controlled, crossover fMRI study (n=24 healthy adults, mean age 34.3 ± 11.2 years) compared brain activation between mastoid and cymba concha stimulation sites. Primary endpoint: spatial distribution of brain activity. Results: activation maps significantly correlated (r=0.61, p<0.001), confirming functional equivalence of mastoid site for ABVN activation.
Technological Characteristics
Transcutaneous nerve field stimulator; battery-operated. Stimulation: ABVN via mastoid electrode. Max output: 5.0 mA, 2.5-50 V (depending on impedance), 750 μs pulse width, 150 Hz frequency. Connectivity: standalone. Materials: patient-contacting materials equivalent to predicate. Software: embedded firmware for pulse generation.
Indications for Use
Indicated for adult patients experiencing opioid withdrawal symptoms, used in conjunction with standard symptomatic medications and other therapies under clinical supervision.
Regulatory Classification
Identification
A percutaneous nerve stimulator for substance use disorders is a device that stimulates nerves percutaneously to aid in the reduction of withdrawal symptoms associated with substance use disorders.
Special Controls
In combination with the general controls of the FD&C Act, the percutaneous nerve stimulator for substance use disorders is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the specified shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 25, 2025
Spark Biomedical, Inc.
% Erin Gontang, PhD
Senior Consultant
Rqm+
2790 Mosside Blvd, Suite 800
Monroeville, Pennsylvania 15146
Re: K251246
Trade/Device Name: Sparrow Ascent
Regulation Number: 21 CFR 882.5896
Regulation Name: Percutaneous Nerve Stimulator For Substance Use Disorders
Regulatory Class: Class II
Product Code: PZR
Dated: April 22, 2025
Received: July 25, 2025
Dear Dr. Gontang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251246 - Erin Gontang, PhD
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for
{2}
K251246 - Erin Gontang, PhD
Page 3
more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ROBERT KANG -S
for Pamela Scott
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251246 | |
| Device Name Sparrow Ascent | |
| Indications for Use (Describe) The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
SPARK
BIOMEDICAL
# 510(k) Summary
K251246
## DATE PREPARED
August 25, 2025
## MANUFACTURER AND 510(k) OWNER
Spark Biomedical, Inc.
18208 Preston Road, Suite D9 531, Dallas, TX 75252
Telephone: +1(844) 654-7775
Official Contact: Brent Croft, VP of Quality & Regulatory
## REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., RAC
Erin A. Gontang, Ph.D.
RQM+
Telephone: +1 (412) 816-8253
Email: akomiyama@rqmplus.com, egontang@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: Sparrow Ascent
Common Name: Percutaneous Nerve Stimulator For Opioid Withdrawal
Classification Name: Percutaneous nerve stimulator for substance use disorders
Regulation Number: 21 CFR 882.5896
Class: Class II
Product Code: PZR
Premarket Review: Neurological and Physical Medicine Devices (OHT5)
Neuromodulation and Physical Medicine Devices (DHT5B)
Review Panel: Neurology
## EXISTING DEVICE IDENTIFICATION
The Sparrow Ascent is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Predicate/Reference | Product Code / Reg Number |
| --- | --- | --- | --- |
| K230796 | Sparrow Ascent / Spark Biomedical, Inc. | Primary Predicate | PZR / 21 CFR 882.5896 |
| K233166 | NET Device / Net Recovery | Reference Device | PZR / 21 CFR 882.5896 |
Neither the predicate nor the reference device have been subject to a design related recall.
{5}
510(k) Summary
SPARK BIOMEDICAL
# DEVICE DESCRIPTION
The Sparrow Ascent is a transcutaneous auricular neurostimulation (tAN) system intended to provide non-invasive, transcutaneous stimulation of the nerves around the auricle (ear). The device is indicated as an aid in the reduction of opioid withdrawal symptoms in adult patients. The Sparrow Ascent is a battery operated, prescription device that delivers mild electrical stimulation to the nerves around the auricle (ear), which carry information to the central nervous system. The Sparrow Ascent is to be used in clinical environments (e.g., doctor's office, clinics, rehab centers, and hospitals) and/or at home. Users of the subject device include adults experiencing opioid withdrawal symptoms. Stimulation parameters (i.e., the strength of stimulation) are set by the user's clinician, and users can only adjust stimulation intensity (amplitude) at home. The system consists of three main components 1) a disposable Earpiece, 2) a Cable, and 3) the External Pulse Generator (EPG).
# INDICATIONS FOR USE
The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel.
# COMPARISON OF INDICATIONS FOR USE STATEMENT
The Indications for Use statements of the subject device and the predicate device are identical. The intended use of both devices, to reduce the signs and symptoms associated with opioid withdrawal, is the same.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Spark Biomedical, Inc. believes that the Sparrow Ascent is substantially equivalent to the predicate device based on the information summarized here:
The subject device is identical to that cleared in K230796, with the exception of a modification to the device Earpiece. While both the subject and predicate devices have an electrode that stimulates the auricular branch of the vagus nerve (ABVN), the predicate device does so via an electrode within the cymba concha, whereas the subject device does so via an electrode on the mastoid. The electrodes provide identical output, thereby providing equivalent stimuli to the ABVN. The change in the Sparrow Ascent Earpiece design does not change the device principle of operation, and there has been no change to the materials used to manufacture the earpiece. Furthermore, although the overall size of the earpiece has been reduced, an assessment of the overall ratios of the patient contacting materials in the revised Sparrow Ascent Earpiece relative to the predicate device earpiece were determined to be equivalent.
Spark Biomedical also evaluated functional magnetic resonance imaging (fMRI) data to show that the spatial distribution of brain activity following mastoid and cymba concha stimulation of
{6}
510(k) Summary
SPARK BIOMEDICAL
the ABVN was the same. The sham-controlled study looked at 24 healthy adults (mean age: $34.3 \pm 11.2$ years), and the brain activation maps showing mastoid and cymba concha stimulation were significantly correlated
$(r = 0.61, p < 0.001)$ . The results indicated consistent central responses between the two stimulation sites, and the mastoid site demonstrated reliable stimulation-induced brain activation across participants.
The subject device is similar to the reference device (NET Device; cleared in K233166), which uses a similar ABVN stimulation site and has similar outputs to support the intended use of reducing the signs and symptoms associated with opioid withdrawal. The current density (i.e., the output related to nerve activation) is similar between the subject and the reference devices and helps demonstrate equivalent effectiveness. Additionally, power and charge density (i.e., the output related to the impact on the stimulation site) is similar between the subject, the predicate, and the reference devices, and helps demonstrate equivalent safety. This comparison helps demonstrate that the subject device's equivalence to the predicate is supported by the reference.
{7}
510(k) Summary
SPARK
BIOMEDICAL
| | Spark Biomedical Sparrow Ascent K251246 | Spark Biomedical Sparrow Ascent K230796 | Net Recovery NET Device K233166 | Comparison to Predicate |
| --- | --- | --- | --- | --- |
| | Subject Device | Primary Predicate | Reference Device | |
| Indications for Use | The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel. | The Sparrow Ascent is a transcutaneous nerve field stimulator that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel. | The NET Device is a transcutaneous alternating current stimulator (tACS) that is intended to be used in patients experiencing opioid withdrawal in conjunction with standard symptomatic medications and other therapies for opioid withdrawal symptoms under the supervision of trained clinical personnel. | Identical. |
| Stimulation Sites (electrode placement) | Trigeminal & Mastoid | Trigeminal & Cymba Concha | Mastoid | Similar. The subject device now stimulates the ABVN via the mastoid as opposed to the cymba concha. Clinical data shows equivalent brain activation following stimulation at both sites. The reference device also |
{8}
510(k) Summary
SPARK BIOMEDICAL
| | | | | stimulates the ABVN via the mastoid. The cleared reference device shares the same IFU and the same electrode position as the subject device. |
| --- | --- | --- | --- | --- |
| Product Code / Regulation | PZR / 21 CFR 882.5846 | PZR / 21 CFR 882.5846 | PZR / 21 CFR 882.5846 | Identical. |
| Patient Population | Adult | Adult | Adult | Identical. |
| Maximum Voltage (V) | 2.5 @ 500 Ω
10 @ 2K Ω
50 @ 10K Ω | 2.5 @ 500 Ω
10 @ 2K Ω
50 @ 10K Ω | 6.8 @ 500 Ω
25 @ 2K Ω
77 @ 10K Ω | Identical. |
| Maximum Current (mA) | 5.0 @ 500 Ω
5.0 @ 2K Ω
5.0 @ 10K Ω | 5.0 @ 500 Ω
5.0 @ 2K Ω
5.0 @ 10K Ω | 14.1 @ 500 Ω
12.4 @ 2K Ω
7.7 @ 10K Ω | Identical. |
| Maximum Pulse Width (μs) | 750 | 750 | 750 | Identical. |
| Maximum Frequency (Hz) | 150 | 150 | 3,000 | Identical. |
| Area of Smallest Electrode (cm²) | 1.181 | 0.853 | 2.080 | Similar. The observed parameter differences do not affect the substantial equivalence of the subject device to the predicate. |
| Power Density (rms W/cm²) | 0.002 @ 500 Ω
0.007 @ 2K Ω
0.051 @ 10K Ω | 0.003 @ 500 Ω
0.01 @ 2K Ω
0.07 @ 10K Ω | Not Reported as RMS | Similar. The observed parameter differences do not affect the substantial equivalence of the subject device to the predicate. |
| Max Average Power Density (W/cm² @ 4Hz and 750 μs) | Reported as RMS | Reported as RMS | 0.57 @ 2K Ω
0.68 @ 10K Ω | Identical. |
{9}
510(k) Summary
SPARK BIOMEDICAL
# SUMMARY OF NON-CLINICAL TESTING
In addition to testing performed to support the clearance of K230796 (e.g., biocompatibility), which remains applicable to the subject device, verification testing of the revised Sparrow Ascent Earpiece was conducted. Additionally, extended service life testing was completed for the subject device to verify a service life of 3 years. Non-clinical performance testing verified that, like the predicate, the subject device operates as expected when used as intended, and both meets and satisfies product requirements. As such, it is concluded that the subject device can safely and effectively deliver transcutaneous peripheral nerve stimulation when used as intended.
# SUMMARY OF CLINICAL TESTING
To demonstrate similar spatial distribution of brain activity between mastoid and cymba concha stimulation sites, a functional magnetic resonance imaging (fMRI) study was conducted to compare central nervous system activation patterns during stimulation at each site. The fMRI study looked at 24 healthy adults (mean age: 34.3 ± 11.2 years) and was a sham-controlled, crossover study with three stimulation targets: the cymba concha, mastoid, and earlobe. Each subject underwent concurrent transcutaneous auricular vagus nerve stimulation (taVNS) and fMRI during three 8-minute stimulation blocks per site, randomized and counterbalanced, using a 30-second ON/OFF stimulation paradigm. The subject's brain responses were evaluated using general linear modeling (GLM), and functional activation maps were generated and compared across conditions.
The fMRI study showed that both mastoid and cymba concha stimulation elicited overlapping spatial patterns of brain activity. One-sample t-test analyses were performed across the 24 subjects, and activation maps showing mastoid and cymba concha stimulation were significantly correlated (r = 0.61, p < 0.001), indicating consistent central responses between the two stimulation sites. The findings of the fMRI study support the anatomical and functional equivalence of the mastoid site to the cymba concha for auricular branch of the vagus nerve (ABVN) activation. As such, the modification to the subject device Earpiece constitute a minor modification that does not change the intended use of the device, nor does it introduce different questions of safety or effectiveness.
# CONCLUSION
Based on non-clinical and clinical testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate device. The identical indications for use, similar technological characteristics, and equivalent performance characteristics for the proposed Sparrow Ascent Earpiece demonstrate the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.