K251226 · Aqua Medical, Inc. · KNS · Aug 8, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K251226
Device Name
Aqua Medical RF Vapor Ablation System
Applicant
Aqua Medical, Inc.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Aug 8, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.
Device Story
Catheter-based system uses RF energy to heat saline, producing water vapor for tissue coagulation/ablation. System comprises software-controlled RF generator, disposable 7F/10.5F catheters, and saline delivery cassette. Operated by physician via GUI and foot pedal during endoscopic procedures. Generator controls saline flow via syringe pump and RF energy delivery to catheter. Vapor directed to target tissue to achieve therapeutic effect. System intended for use in clinical settings for GI tract lesions. Benefits include controlled thermal ablation of bleeding/non-bleeding sites.
RF-based vapor ablation system. Components: RF generator, disposable 7F/10.5F catheters, saline cassette. Catheters compatible with 3.7mm endoscope working channels. Energy source: RF. Software-controlled saline delivery via syringe pump. Sterilization: Validated for single-use catheters.
Indications for Use
Indicated for coagulation and ablation of bleeding/non-bleeding GI tract sites (e.g., esophagus) in adults. Specific conditions: Esophageal Ulcers, Mallory-Weiss tears, AVMs, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, GAVE, and Radiation Proctitis.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 8, 2025
Aqua Medical, Inc.
% Bosmat Friedman-Cox
Regulatory Consultant
ProMedoss, Inc.
6026 Beech Cove Ln.
Charlotte, North Carolina 28269
Re: K251226
Trade/Device Name: Aqua Medical RF Vapor Ablation System
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic electrosurgical unit and accessories
Regulatory Class: Class II
Product Code: KNS, GEI
Dated: July 11, 2025
Received: July 11, 2025
Dear Bosmat Friedman-Cox:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251226 - Bosmat Friedman-Cox
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K251226 - Bosmat Friedman-Cox
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony Lee, PhD, MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
Aqua Medical RF Vapor Ablation System
Page 8 of 49
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251226 | ? |
| Please provide the device trade name(s). | | ? |
| Aqua Medical RF Vapor Ablation System | | |
| Please provide your Indications for Use below. | | ? |
| The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.
Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
K251226
Page 1 of 3
| | Aqua Medical RF Vapor Ablation System | |
| --- | --- | --- |
| | Traditional 510(k); | 510(k) Summary |
1. **SUBMITTER**
Applicant’s Name:
Aqua Medical, Inc.
6634 Owens Drive
Pleasanton, CA 94588
Phone: 949-233-5172
Primary Contact:
Bosmat Friedman
Regulatory Affairs Consultant
6026 Beech Cove Ln.
Charlotte, NC 28269
Phone: 980-308-1636
bosmat.f@promedoss.com
Date Prepared:
July 11, 2025
2. **DEVICE**
Trade Name:
Aqua Medical RF Vapor Ablation System
Classification Name: Endoscopic electrosurgical unit and accessories
Product Code: KNS
Regulation No: 876.4300
Class: 2
Review Panel: Gastroenterology/Urology
Classification Name: Electrosurgical cutting and coagulation device and accessories
Product Code: GEI
Regulation No: 878.4400
Class: 2
Review Panel: General & Plastic Surgery
3. **PREDICATE DEVICES**
Primary Predicate:
Aqua Medical RF Vapor System, by Aqua Medical Inc., Product code KNS, GEI cleared Under: K213627, K211282, K183595 and K241271.
4. **DEVICE DESCRIPTION**
The Aqua RF Vapor Ablation System is a catheter-based device that uses radiofrequency (RF) energy to produce heated water vapor for the coagulating and ablating tissue. The RF energy
Page 1 of 3
{6}
K251226
Page 2 of 3
| | Aqua Medical RF Vapor Ablation System | |
| --- | --- | --- |
| | Traditional 510(k); | 510(k) Summary |
heats liquid as it flows through the catheter, creating steam that is directed to the target tissue. The system includes the following components:
- Aqua RF Vapor Generator: A software-controlled RF generator is operated through a graphical user interface (GUI) and incorporates a syringe pump that delivers saline to the catheter. The catheter connects, fluidically and electrically, to the generator. A foot pedal is used to control the RF energy and vapor delivery.
- Aqua RF Vapor Catheters: disposable, sterile, single-use catheters with a diameter of 7F (2.3 mm) with an outer catheter sheath that is 10.5F and a length between 145 cm and 210 cm. Catheters are designed to be used through an endoscope working channel diameter of 3.7mm. This submission included only the modified circumferential catheter.
- Cassette (Syringe and Tubing): Tubing and syringe (60cc) provide a means of delivering saline to the catheter during treatment.
## 5. INDICATIONS FOR USE
The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.
Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.
## 6. SUBSTANTIAL EQUIVALENCE
The reason for this submission includes modifications to the generator, modifications to the Cassette (syringe and tubing), and design modifications to the circumferential catheter. To support the requested modifications the company submitted supporting non-clinical data. The proposed modifications do not introduce new safety and effectiveness concerns and support the company's substantial equivalency claim.
## 7. PERFORMANCE DATA
In order to support the proposed modifications to the generator, catheter and Cassette the company conducted the following:
**Generator:**
- EMC & Electrical Safety
- Software validation
- Generator Hardware Verification
**Catheter:**
- Hardware Verification Testing
- Simulated-use tissue validation (Lean Beef Testing)
**Cassette:**
- Biocompatibility
Page 2 of 3
{7}
K251226
Page 3 of 3
| | Aqua Medical RF Vapor Ablation System | |
| --- | --- | --- |
| | Traditional 510(k); | 510(k) Summary |
- Shelf-Life assessment
- Sterilization Validation
## 8. CONCLUSION
Aqua Medical has demonstrated that the Aqua RF Vapor Ablation System is substantially equivalent to the predicate device. Differences between the proposed Aqua RF Vapor Ablation System and the predicate device do not raise new questions of safety or effectiveness. The requested modifications have been adequately supported with bench testing and scientific rationales.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.