K183595 · Aqua Medical, Inc. · KNS · Jun 21, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K183595
Device Name
Aqua Medical RF Vapor System
Applicant
Aqua Medical, Inc.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Jun 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aqua Medical RF Vapor System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.
Device Story
System uses RF energy to generate heated water vapor for tissue coagulation; avoids direct RF application to tissue. Components: RF generator with GUI/syringe pump; disposable 7F (2.3mm) catheter; saline delivery tubing. Used in clinical settings by physicians via standard endoscopy (minimum 2.8mm working channel). Physician advances catheter through endoscope to target site; engages tissue with endoscopic cap. Generator delivers bipolar RF to electrode in catheter lumen while saline flows over electrode, creating vapor. Vapor delivered to tissue; treatment depth controlled by application time. Benefits: effective coagulation of GI bleeding/lesions.
Clinical Evidence
No human clinical data. Evidence based on bench testing (biocompatibility, electrical safety/EMC per IEC 60601-1/1-2, hardware/system validation) and a porcine animal study (n=6). Animal study evaluated acute and subacute (48h) histopathological thermal coagulation depth in the esophagus. Results showed equivalent tissue injury depth to predicate without perforation or hemorrhage.
Technological Characteristics
System includes RF generator, disposable 7F (2.3mm) catheter, and saline delivery system. Sensing/actuation: Bipolar RF energy converts saline to heated water vapor. Connectivity: Foot pedal or catheter-integrated button. Software: Major level of concern. Sterilization: Disposable catheter is sterile, single-use.
Indications for Use
Indicated for coagulation of bleeding and non-bleeding GI tract sites (e.g., esophagus) in adults. Specific conditions: Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, GAVE, and Radiation Proctitis.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Halo 90 System by Barrx Medical, Inc. (K060169, K093008)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 21, 2019
Aqua Medical, Inc. % Bosmat Friedman Regulatory Affairs Consultant ProMedoss, Inc. 3521 Hatwynn Road Charlotte, NC 28269
Re: K183595
> Trade/Device Name: Aqua Medical RF Vapor System Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS, GEI Dated: May 17, 2019 Received: May 20, 2019
Dear Bosmat Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K183595
Device Name Aqua Medical RF Vapor System
The Aqua Medical RF Vapor System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.
Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
# 510(K) SUMMARY [as required by section 807.92(c)] Aqua Medical RF Vapor System 510(k) Number K183595
### 1. SUBMITTER
#### Applicant's Name:
Aqua Medical, Inc. 191 West Second St. Santa Ana, CA 92701 Phone: 949-233-5172
### Contact Person:
Bosmat Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 647-975-3974 bosmat.f@promedoss.com
Date Prepared:
June 20, 2019
## 2. DEVICE
Trade Name: Aqua Medical RF Vapor System
Classification Name: Endoscopic electrosurgical unit and accessories Product Code: KNS Regulation No: 876.4300 Class: 2 Review Panel: Gastroenterology/Urology
Classification Name: Electrosurgical cutting and coagulation device and accessories Product Code: GEI Regulation No: 878.4400 Class: 2 Review Panel: General & Plastic Surgery
## 3. PREDICATE DEVICE
#### Primary predicate:
- · Halo 90 System by Barrx Medical, Inc., cleared under K060169 and K093008
## 4. DEVICE DESCRIPTION
The Aqua RF Vapor System uses a catheter to deliver heat to the tissue surface undergoing treatment. The heat, which is created by the application of RF energy
{5}------------------------------------------------
to a bipolar electrode pair located inside the catheter tip, is delivered to the tissue through heated water vapor generated inside the catheter. Application of heat to the tissue results in coagulation of the tissues being treated. This technique avoids the need for direct RF energy application to heat the tissue.
The Aqua Medical RF Vapor System consists of the following components:
- Aqua RF Vapor Generator: A software-controlled RF generator is operated through a graphical user interface (GUI) and incorporates a syringe pump that delivers saline to the catheter. Delivery of bipolar RF energy to the catheter is controlled either by an accessory foot pedal or a button incorporated in the disposable catheter.
- -Aqua RF Vapor Catheter: a disposable, sterile, single-use catheter with a diameter of 7 F (2.3 mm) and a length of 145 cm. It includes an integrated cable for attachment to the generator and a Luer-connector that facilitates connection to the saline delivery tubing.
- -Saline Delivery Tubing and Syringe: Tubing and syringe (60cc) that provide a means of delivering saline to the catheter during treatment.
## Mode of Operation
Access to the target tissue is gained using standard endoscopic techniques. An endoscope with the FDA cleared Olympus endoscope distal attachment cap is advanced to the treatment location. The treatment catheter is inserted through the working channel of the endoscope with a minimum of a 2.8 mm working channel, and advanced to the distal end of the endoscope until the distal tip is completely visible. The endoscope and catheter are advanced to the treatment site and the target tissue is engaged with the endoscopic cap. Bipolar RF energy is then delivered to the bipolar electrode pair in the catheter, while saline is flowing through the catheter lumen and generates heated water vapor to treat the targeted tissue.
The user chooses the power via the generator's touch-screen interface. The generator creates heated water vapor by delivering a constant bipolar RF output to the bipolar electrode contained within the lumen of the disposable catheter. The software commands the integrated syringe pump to dispense saline at a flow rate proportional to the user selected power setting. Heated water vapor, generated as saline flows over the energized electrode, is delivered to the tissue through the distal tip of the catheter. The time of application of heated water vapor determines the depth of treatment.
## 5. INDICATIONS FOR USE
The Aqua Medical RF Vapor System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.
Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only
{6}------------------------------------------------
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the Aqua System and the HALO® predicate are comprised of a multiple use generator with optional footswitch and disposables. However, the Aqua System achieves its intended use by utilizing a slightly different technology from its primary predicate. While the HALO® is used to treat the esophagus by applying bipolar RF energy (creating heat) directly to the tissue, the Aqua System converts the bipolar RF energy to heated vapor which is then used to treat the tissue.
To bridge the technological gap between the two devices, comparative bench and animal testing were conducted. The results of the comparative testing demonstrate that the HALO" and the Aqua Medical RF Vapor System produce substantially equivalent results.
# 7. PERFORMANCE DATA
Below is a list of the tests that have been performed and successfully completed for the Aqua Medical RF Vapor System.
# Biocompatibility:
The following biocompatibility tests were conducted:
- Cytotoxicity -
- -Hemolysis
- Systemic Toxicity -
- -Intracutaneous
- Sensitization -
- -Pyrogenicity
# Electrical safety and electromagnetic compatibility (EMC):
The system was tested and found to comply with IEC 60601-1:2005/EN 60601-1:2006 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance 3rd Ed. and IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
# Software Verification and Validation Testing:
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered as a "major" level of concern.
# Bench:
- -HW Verification Testing
- -System Validation Testing
- Distal Assembly Pull Test -
- Lesion Comparative testing -
{7}------------------------------------------------
#### Animal:
In order to demonstrate substantial equivalence and bridge the technological gap between the Aqua Medical RF Vapor System and its primary predicate, the company conducted an animal study including both acute and subacute animals. A total of six animals were utilized in the study; three animals were treated and evaluated acutely (Day = 0) and three animals were treated and evaluated subacutely (48 hours). Each animal's esophagus received multiple test and control treatments that were applied using two different treatment settings. The histopathological evaluation assessed acute thermal coagulation through the crosssectional layers of the esophagus. There were no pathological complications associated with the test or control treatments in any animal. Both the test and control treatments demonstrated the ability to create thermal coagulative tissue injury that penetrates in depth to the external muscularis of the esophagus in healthy swine. Neither test nor control treatments were associated with esophageal perforation, significant hemorrhage or significant clinical complication. The results of the study demonstrate that the Aqua Medical RF Vapor System and the HALO® are substantially equivalent.
#### Conclusion:
The Aqua Medical RF Vapor System has the same indications for use as the HALO® System and is substantially equivalent in technological and performance characteristics to the HALO® predicate. The results obtained from comparative testing further support our substantial equivalency claim.
Consequently, it is clear that the Aqua Medical RF Vapor System is as safe and effective as its predicate devices without raising any new safety and/or effectiveness concerns. Any differences have been addressed by extensive testing and validations and therefore negate any safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.