The VCFix Spinal System is indicated for use in the reduction of vertebral compression fractures that may result from osteoporosis or trauma (fracture types A according to Magerl/AO Spine classification) with or without posterior instrumental fixation, and compression fractures that result from malignant lesions (myeloma or osteolytic metastasis). The VCFix Spinal System is intended to be used in combination with Teknimed F20 bone cement, and to be placed, using a transpedicular approach, in a fractured vertebra from level T9 to L5.
Device Story
Cannulated, titanium alloy (Ti6Al4V ELI) screw-shaped implant with expandable plate structure; used for vertebral compression fracture reduction. Surgeon inserts implants bilaterally via transpedicular approach into fractured vertebra (T9-L5); implants expanded in situ using jack mechanism to restore vertebral height; Teknimed F20 bone cement injected through/around implant to fixate. Used in clinical/surgical settings. Output is restored vertebral height and stabilized fracture. Benefits include minimally invasive fracture reduction and stabilization.
Clinical Evidence
Bench testing only. Includes lift force, ASTM F2077 (static/dynamic compression and shear), ASTM F2193 (torsion/driving torque), functionality with cement, MR safety, ISO 10093-1 biological safety, and additive manufacturing validation.
Technological Characteristics
Cannulated titanium alloy (Ti6Al4V ELI, ASTM F3001/F136) implant. Expandable plate structure via jack mechanism. Sterile, single-use. Bilateral transpedicular placement. Compatible with Teknimed F20 bone cement. Toolset includes stainless steel shafts and plastic handles.
Indications for Use
Indicated for adults with vertebral compression fractures from osteoporosis, trauma (Magerl/AO Spine type A), or malignant lesions (myeloma/osteolytic metastasis) in thoracic/lumbar vertebrae T9-L5. Used with Teknimed F20 bone cement; optional posterior instrumental fixation.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Amber Implants
Banafsheh Sajadi
Chief Executive Officer
Prinses Beatrixlaan 546
The Hague, 2595 BM
Netherlands
Re: K250637
Trade/Device Name: VCFix Spinal System
Regulation Number: 21 CFR 888.3027
Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement
Regulatory Class: Class II
Product Code: NDN
Dated: February 14, 2025
Received: March 3, 2025
Dear Banafsheh Sajadi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250637 - Banafsheh Sajadi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250637 - Banafsheh Sajadi
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE MUIR
-S
Digitally signed by JESSE
MUIR-S
Date: 2025.05.30
14:18:04 -04'00"
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative, Repair
and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250637
Device Name
VCFix Spinal System
Indications for Use (Describe)
The VCFix Spinal System is indicated for use in the reduction of vertebral compression fractures that may result from osteoporosis or trauma (fracture types A according to Magerl/AO Spine classification) with or without posterior instrumental fixation, and compression fractures that result from malignant lesions (myeloma or osteolytic metastasis). The VCFix Spinal System is intended to be used in combination with Teknimed F20 bone cement, and to be placed, using a transpedicular approach, in a fractured vertebra from level T9 to L5.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250637
# amberimplants
# 510(k) Summary
Device Trade Name: VCFix Spinal System
Common Name: Implantable Fracture Reduction System
Manufacturer: Amber Implants
Prinses Beatrixlaan 546,
2595 BM The Hague,
The Netherlands
Contact: Banafsheh Sajadi
Chief Executive Officer
Phone: +31 6 11 27 37 31
Email: b.sajadi@amberimplants.com
Prepared by: Justin Eggleton
MCRA, LLC
803 7th Street, NW
Washington, DC 20001
Office: 202.552.5800
Email: jeggleton@mcra.com
Date Prepared: May 30, 2025
Classifications: 21 CFR §888.3027, Polymethylmethacrylate (PMMA) bone cement
Class: II
Product Codes: NDN, Cement, Bone, Vertebroplasty
# Indications For Use:
The VCFix Spinal System is indicated for use in the reduction of vertebral compression fractures that may result from osteoporosis or trauma (fracture types A according to Magerl/AO Spine classification) with or without posterior instrumental fixation, and compression fractures that result from malignant lesions (myeloma or osteolytic metastasis). The VCFix Spinal System is intended to be used in combination with Teknimed F20 bone cement, and to be placed, using a transpedicular approach, in a fractured vertebra from level T9 to L5.
# Device Description:
The VCFix Spinal System is a cannulated screw-shaped implant with an expandable plate structure and is available in multiple sizes to accommodate varying patient anatomy. A pair of implants are inserted into the pedicles and manipulated with the toolset, resulting in the expansion of the device
{5}
K250637
to restore the vertebral height. The implant is made of titanium alloy (Ti6Al4V ELI). During the procedure, the surgeon inserts the implants bilaterally through each pedicle of the fractured vertebra and adjusts the height and angle of the device (to correct local kyphosis) based on the unique vertebral fracture of the patient.
The VCFix Spinal System includes sterile, single-use implantation kits and a preparation kit. The kits include a pair of implants and all tools necessary for preparing the implant site and vertebral body expansion. The positioning and expansion of the implant must imperatively be followed by the injection of Teknimed F20 bone cement (K103433) to fixate the restored vertebral body. The bone cement and its delivery system are sold separately.
## Predicate Devices:
Amber Implants submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA’s regulation of medical devices, VCFix Spinal System is substantially equivalent in indications, design principles, and performance to the following legally marketed predicate devices:
**Primary Predicate:** Stryker SpineJack System (K223294)
**Reference Devices:** Injection Pin (K221697) by SLK Ortho LLC
Teknimed F20 bone cement (K103433)
## Summary of Technological Characteristics:
The subject and predicate devices have the same intended use with similar technological characteristics:
- Intended Use: Intended for the adult population to treat fractures in the thoracolumbar vertebrae caused by osteoporosis, trauma, and tumors, with or without posterior instrumental fixation.
- Principle of Operation: Devices are placed with a minimally invasive, bilateral transpedicular approach, and are inserted through the pedicles into the site in a collapsed form, expanded in the cranio-caudal direction (subject and primary predicate), and then FDA-cleared bone cement is injected through and around the implant to maintain the fracture reduction. Supplemental posterior fixation is placed if necessary.
- Material: Manufactured from titanium alloy.
- Design: Cannulated devices which are offered in a variety of implant sizes to accommodate varying patient anatomy. There are minor differences in the overall geometry of the device, however, the major dimensions fall within the range of previously cleared sizes for the intended use. Compared to the primary predicate, different features of the subject device (e.g., perforated expansion plates, length, diameter) are supported by the reference device and performance testing.
- Sterility: Sterile, for single use only.
Detailed information is provided in the table below.
{6}
K250637
| Characteristics | Subject Device VCFix Spinal System | Primary Predicate Device Stryker SpineJack System (K223294) | Reference Device SLK Ortho LLC, Injection Pin (K221697) | Comparison |
| --- | --- | --- | --- | --- |
| Classification | Class II | Class II | Class II | Identical |
| Regulation | 21 CFR 888.3027 | 21 CFR 888.3027 | 21 CFR 888.3027 | Identical |
| Product Code | NDN | NDN | NDN, LOD | Identical |
| Panel | Orthopedic | Orthopedic | Orthopedic | Identical |
| Intended Use | Intended for the reduction of vertebral compression fractures | Intended for the reduction of vertebral compression fractures | Intended for the reduction of vertebral compression fractures | Identical |
| Indications for Use | The VCFix Spinal System is indicated for use in the reduction of vertebral compression fractures that may result from osteoporosis or trauma (fracture types A according to Magerl/AO Spine classification) with or without posterior instrumental fixation, and compression fractures that result from malignant lesions (myeloma or osteolytic metastasis). The VCFix Spinal System is intended to be used in combination with Teknimed F20 bone cement, and to be placed, using a transpedicular approach, in a fractured vertebra from level T9 to L5. | The SpineJack Expansion Kit is indicated for use in the reduction of painful osteoporotic vertebral compression fractures, traumatic vertebral compression fractures (Type A fractures according to the AO/Magerl classification) with or without posterior instrumental fixation, and compression fractures that result from malignant lesions (myeloma or osteolytic metastasis). It is intended to be used in combination with Stryker Vertaplex and Vertaplex HV bone cement. | Indicated for use in fractures caused by severe osteoporosis, trauma, and tumors of the thoracic/lumbar spine from T9-L5. Injection Pin is indicated for use in combination with PMMA bone cement (Teknimed F20) for the treatment of fractures caused by trauma, osteoporosis, or tumors in the thoracic/lumbar spine from T9-L5. | Same as primary predicate; specified level T9 to L5. Modified the compatible bone cement trade name which is the same as the reference devices: Teknimed F20 bone cement (K103433). |
| Target Population | Adult | Adult | Adult | Identical |
| Anatomical Site | Thoracic and lumbar vertebrae, transpedicular approach | Thoracic and lumbar vertebrae, transpedicular approach | Thoracic and lumbar vertebrae | Identical |
| Contact | Implant: implant device - permanent contact (> 30 days) with bone/tissue
Toolset: external communicating device - limited contact (< 24 hours) with bone/tissue | Implant: implant device - permanent contact (> 30 days) with bone/tissue
Toolset: external communicating device - limited contact (< 24 hours) with bone/tissue | Implant: implant device - permanent contact (> 30 days) with bone/tissue | Identical |
| System Components | VCFix Spinal System consists of:
• VCFix Preparation kit (to prepare the access to the vertebral body)
• VCFix Implantation kit (implant and implant inserter/expander) | SpineJack System consists of:
• SpineJack Preparation kit (to prepare the access to the vertebral body)
• SpineJack Expansion kit (implant and implant inserter/expander) | The Injection Pin system consists of:
• Injection Pin implant
• Injection Pin Screwdriver | Same |
| Presentation of the Device | Sterile, single use devices | Sterile, single use devices | Sterile, single use devices | Identical |
{7}
K250637
| Characteristics | Subject Device VCFix Spinal System | Primary Predicate Device Stryker SpineJack System (K223294) | Reference Device SLK Ortho LLC, Injection Pin (K221697) | Comparison |
| --- | --- | --- | --- | --- |
| Materials | • Implant: Ti-6Al-4V (ASTM F3001, ASTM F136)
• Toolset: stainless steel shafts + plastic handles (except for inner dilator made of PPSU) | • Implant: Ti-6Al-4V (ISO 5832-3)
• Toolset: stainless steel shafts + plastic handles | • Implant: Ti-6Al-4V (ISO 5832-3, ASTM F136) | Same-Implant
Similar-Instruments |
| Implant Sizes | 7 sizes:
**Major dimensions:**
• Insertion Diameter: ○ 5mm – 6.5mm
• Total Length in VB: ○ 41mm – 51.2mm
• Expanding Plate Length: ○ 14.4mm – 21mm
• Max Expansion Height: ○ 11.3mm – 16mm | 3 sizes:
**Major dimensions:**
• Insertion Diameter (A): ○ 4.2mm, 5mm, and 5.8mm
• Total Length (B) in VB: ○ 20mm, 25mm, 28mm
• Expanding Plate Length (B): ○ 14mm, 19mm, 20mm
• Max Expansion Height (C): ○ 12.5mm, 17mm, 20mm | Multiple sizes:
• Pin Diameters: ○ 5 and 6mm
• Total Length: ○ 31 to 61mm (3mm increments) | Similar and within range of predicates. |
| Insertion Feature | Spherical head with internal hexagon for mating with inserter | Mates with implant inserter | Spherical head with internal hexagon for insertion screwdriver | Same |
| Cement Features | Cannulated with perforated structure on the expansion plates | Cannulated with cut-through features | Cannulated and fenestrated screw | Same – all devices are cannulated and allow for cement injection through the length of the device. |
| Expansion Mechanism | The implant expands in the cranio-caudal direction using a jack mechanism. | The implant expands in the cranio-caudal direction using a jack mechanism. | The implant does not expand. | Similar, both the subject and primary predicate implants gradually extend in the cranio-caudal direction by use of an implant extender instrument to expand. |
| Anchorage Mechanism | Bone cement and supplemental posterior fixation. Pedicle anchorage to share axial loading between anterior and posterior columns. | Bone cement and supplemental posterior fixation | Bone cement and pedicle anchorage to share axial loading between anterior and posterior columns with bone cement. | Identical |
{8}
K250637
| Characteristics | Subject Device
VCFix Spinal System | Primary Predicate Device
Stryker SpineJack System
(K223294) | Reference Device
SLK Ortho LLC,
Injection Pin
(K221697) | Comparison |
| --- | --- | --- | --- | --- |
| **Principle of Operation** | VCFix is implanted into the vertebra after access is established and the site is prepared for implants insertion. Vertebral body access is achieved through bilateral transpedicular positioning of the trocar, followed by guidewire, soft tissue dilation, drill, and template to clean the implant’s site.
The implant is then inserted into the vertebra via the pedicle in a collapsed form, and expanded in situ, to achieve reduction of vertebral compression fractures. Once inserted and expanded, PMMA bone cement is injected into the space within and around VCFix to maintain the fracture reduction. Supplemental posterior fixation is placed if necessary. | The SpineJack is implanted into the vertebra after access is established and the site is prepared for implants insertion. Vertebral body access is achieved through bilateral transpedicular positioning of the trocar, followed by guidewire, drill with working cannula to dilate soft tissue, and template to clean the implant’s site.
The implant is then inserted into the vertebra via the pedicle in a collapsed form, and expanded in situ, to achieve reduction of vertebral compression fractures. Once inserted and expanded, PMMA bone cement is injected into the space within and around the SpineJack to maintain the fracture reduction. Supplemental posterior fixation is placed if necessary. | The Injection Pin is a cannulated fenestrated screw implanted in the spine through a minimally invasive approach. The device allows for injection of PMMA (Teknimed F20) bone cement. | Identical |
| **Surgical Technique Overview** | 1. VB access and guidewire insertion
2. Dilate soft tissue
3. Drill advanced into VB
4. Template to verify implant length
5. Insert unexpanded VCFix implants into VB
6. VCFix implants expanded
7. Cement injection
8. Supplemental posterior fixation placed (if needed) | 1. VB access and guidewire insertion
2. Dilate soft tissue
3. Reamer advanced into VB
4. Template to verify implant length
5. Insert unexpanded SpineJack implants into VB
6. SpineJack implants expanded
7. Cement injection
8. Supplemental posterior fixation placed (if needed) | Implant inserted with injection pin screwdriver followed by cement injection. | Identical to primary predicate |
| **Instruments** | Provided in sterile kit:
Preparation Kit:
• K-wire
• Two-part dilator
• Drill
• Template
Implantation Kit:
• VCFix Implant
• Implant Inserter
• Implant Expander | Provided in sterile kit:
Preparation Kit:
• Guidewire
• Access cannula
• Working cannula
• Reamer
• Template
Expansion Kit:
• Implant expander tube | Injection pin screwdriver
Compatible with previously cleared Teknimed F20 cement injection tools. | Same |
{9}
K250637
| Characteristics | Subject Device VCFix Spinal System | Primary Predicate Device Stryker SpineJack System (K223294) | Reference Device SLK Ortho LLC, Injection Pin (K221697) | Comparison |
| --- | --- | --- | --- | --- |
| | Compatible with previously cleared Teknimed F20 cement injection tools. | • Handles
• SpineJack Implants
• Quick release pin Cement Injection Tools | | |
| Performance Testing | Mechanical and functional testing was performed to verify the design of the implant, including:
• Lift Force Testing
• ASTM F2077 Static Compression
• ASTM F2077 Dynamic Compression
• ASTM F2077 Static Shear
• ASTM F2077 Dynamic Shear
• ASTM F2193 Static Torsion Test
• ASTM F2193 Screw Driving Torque
• Functionality Tests of VCFix Implant with Cement | Mechanical and functional testing was performed to verify the design of the implant, including:
• Pushing and Recompression Testing (i.e., Lift Force Testing)
• Crimp Force Testing
• Traction Resistance Testing
• Torsion Testing
• Functional Tests of SpineJack Implant with Cement | Unknown | Same – systems demonstrate sufficient strength for the intended use. |
The different technological characteristics were found to raise no new questions of safety or effectiveness.
## Performance Testing Summary:
All necessary testing has been performed for the worst-case configuration of the VCFix Spinal System to assure substantial equivalence to its predicates and to demonstrate the subject devices perform as intended.
The performance of the VCFix Spinal System in its final, finished configuration are supported by the following:
- Lift Force Testing
- ASTM F2077 Static Compression Testing
- ASTM F2077 Dynamic Compression
- ASTM F2077 Static Shear
- ASTM F2077 Dynamic Shear
- ASTM F2193 Static Torsion and Driving Torque Testing
- Functionality Testing
- Magnetic Resonance Safety Testing to support MR Conditional Labeling
- Biological safety evaluation per ISO 10093-1
- Validation activities to support additive manufacturing processes
- Packaging, sterility, and shelf-life validation activities
{10}
K250637
## Substantial Equivalence Conclusion:
The subject device and the predicate devices have the same intended use/indications for use, sufficiently similar performance, fundamental scientific technology and principles of operation. The data included in this submission demonstrate there are no new risks raised by the subject system and supportive of substantial equivalence. VCFix Spinal System is as safe, as effective, and performs as well as, or better, than the predicate devices.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.