Post-market real-world data from US patients; Retrospective survey-study published in literature
RWE was used to support the removal of specific precaution statements in the device labeling regarding long-term chronic use and use during pregnancy, based on clinical experience and survey data.
Post-market surveillance; Retrospective survey; Labeling revision; Real-world data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Synowiec A, et. al. Adv Ther. 2024 Jan;41(1):170-18; Post-market study
US patients using Nerivio
Not applicable for this study
Safety/effects of long-term chronic use
Peretz A, et. al. Headache. 2023;63(7):968-970; Retrospective survey-study
Pregnant women using Nerivio
Not applicable for this study
Safety/use during pregnancy
Indications for Use
The Nerivio / Nerivio Infinity is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment.
Device Story
Wearable, battery-powered Remote Electrical Neuromodulation (REN) device; delivers low-energy electrical pulses to upper arm for 45-minute sessions. Controlled via mobile application (Android/iOS) using Bluetooth; user adjusts stimulation intensity (0-40 mA) to level just below pain threshold. Hardware includes microcontroller, electrodes (hydrogel-covered), and armband. Nerivio Infinity includes NFC tag for electrode authentication. Used by patients at home for acute migraine relief or every other day for prevention. Output allows patients to manage migraine symptoms without pharmacological intervention; mobile app provides status, treatment tracking, and intensity control.
Clinical Evidence
No new bench or clinical testing performed for this submission. Labeling revisions regarding long-term use and pregnancy supported by published literature: real-world data study (Synowiec A, et al. 2024) and retrospective survey-study (Peretz A, et al. 2023).
Indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. Prescription use, self-administered in home environment.
Regulatory Classification
Identification
A trunk and limb electrical stimulator to treat headache is a device intended to treat headache through the application of electrical stimulation anywhere on the body of the patient apart from the patient's head or neck through electrodes placed on the skin. The stimulation may be provided transcutaneously or percutaneously.
Special Controls
In combination with the general controls of the FD&C Act, the trunk and limb electrical stimulator to treat headache is subject to the following special controls: 1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. This testing must include: a) Characterization of the electrical stimulation, including the following: waveforms; output modes; maximum output voltage and maximum output current (at 500Ω , 2kQ. and 10k≤2 loads); pulse duration; frequency; net charge per pulse; and maximum phase charge, maximum current density, maximum average current, and maximum average power density (at 500Ω); b) Characterization of the impedance monitoring system; and c) Characterization of the electrode performance including the electrical performance, adhesive integrity, shelf-life, reusability, and current distribution of the electrode surface area. 2) The patient-contacting components of the device must be demonstrated to be biocompatible. 3) Performance testing must demonstrate electromagnetic compatibility and electrical, mechanical and thermal safety in the intended use environment. 4) Software verification, validation, and hazard analysis must be performed. 5) Labeling must include the following: a) Instructions for use, including the typical sensations experienced during treatment: b) A detailed summary of the electrical stimulation output, and the device technical parameters, including any wireless specifications; c) A shelf life for the electrodes and reuse information; and d) Instructions on care and cleaning of the device.
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. This testing must include:
(i) Characterization of the electrical stimulation, including the following: Waveforms; output modes; maximum output voltage and maximum output current (at 500Ω, 2kΩ, and 10kΩ loads); pulse duration; frequency; net charge per pulse; and maximum phase charge, maximum current density, maximum average current, and maximum average power density (at 500Ω);
(ii) Characterization of the impedance monitoring system; and
(iii) Characterization of the electrode performance including the electrical performance, adhesive integrity, shelf-life, reusability, and current distribution of the electrode surface area.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate electromagnetic compatibility and electrical, mechanical, and thermal safety in the intended use environment.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) Instructions for use, including the typical sensations experienced during treatment;
(ii) A detailed summary of the electrical stimulation output, and the device technical parameters, including any wireless specifications;
(iii) A shelf life for the electrodes and reuse information; and
(iv) Instructions on care and cleaning of the device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2025
Theranica Bio-Electronics Ltd
% Janice Hogan
Partner
Hogan Lovells US LLP
1735 Market St. Floor 23
Philadelphia, Pennsylvania 19103
Re: K250405
Trade/Device Name: Nerivio; Nerivio Infinity
Regulation Number: 21 CFR 882.5899
Regulation Name: Trunk And Limb Electrical Stimulator To Treat Headache
Regulatory Class: Class II
Product Code: QGT
Dated: February 12, 2025
Received: February 13, 2025
Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250405 - Janice Hogan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250405 - Janice Hogan
Page 3
Sincerely,
Jitendra V. Virani -S
CDR Jitendra Virani, MS, MBA
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250405
Device Name
Nerivio;
Nerivio Infinity
Indications for Use (Describe)
Nerivio is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment.
Nerivio Infinity is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250405
Page 1 of 6
# 510(k) SUMMARY
## Theranica Bio-Electronics Ltd.'s Nerivio and Nerivio Infinity
**Submitter:**
Theranica Bio-Electronics Ltd.
4 Ha-Omanut St. Netanya 4250438, Israel
Phone: +972-72-3909755
Facsimile: +972-72-3909762
Contact Person: Dagan Harris
**Date Prepared:** May 14, 2025
**Name of Device:** Nerivio; Nerivio Infinity
**Common or Usual Name:** Nerivio; Nerivio Infinity
**Regulation Name:** Trunk and limb electrical stimulator to treat headache
**Regulation Number:** 21 CFR 882.5899
**Product Code:** QGT
**Regulatory Class:** Class II
**Predicate Device:** Nerivio; NerivioInfinity, Theranica Bio-Electronics Ltd. (K241756)
**Predicate Device Regulation Number:** 21 CFR 882.5899
**Predicate Device Product Code:** QGT
**Predicate Device Regulatory Class:** Class II
## Device Description
The Nerivio and Nerivio Infinity are wearable, battery-powered, Remote Electrical Neuromodulation (REN) devices that are controlled by a mobile application. The systems deliver low energy electrical pulses to the upper arm for 45 minutes per treatment, after which the device turns off automatically.
The devices are composed of:
- A pair of UltraStim® electrodes (K130987) covered with hydrogel and removable protective film. In NerivioInfinity, the electrodes are embedded in a disposable, replaceable pad. In Nerivio, the electrodes are not disposable or replaceable.
- An electronic circuitry that includes microcontroller with firmware, LED indicator, a power button for activating the device, and Bluetooth radio for wireless connection to Android and iOS mobile platforms. The circuitry also includes a battery, which is rechargeable in NerivioInfinity and not rechargeable in Nerivio.
- An armband that is wrapped over the device to secure the device position on the user's arm is also included.
{5}
K250405
Page 2 of 6
The devices are operated and controlled via mobile application software that is installed and run on the user's personal mobile device, such as a mobile phone or tablet. The device's hardware communicates with the mobile application through Bluetooth protocol. This mobile application software allows the user to control the stimulation intensity from 0 to 100% (representing intensity levels of 0-40 mA), to start, pause or stop the stimulation program, and to view device status such as the device's connection state, stimulation duration, battery level (when connected to Nerivio Infinity), remaining number of treatments (when connected to Nerivio), and user notifications.
The user is instructed to adjust the intensity to the strongest stimulation level just below the perceived pain level. Treatments with Nerivio and Nerivio Infinity are intended to be self-administered by the user immediately after the onset of migraine headache or aura or every other day for migraine prevention.
## Intended Use / Indications for Use
Nerivio and Nerivio Infinity have identical indications for use, as follows:
The [Nerivio / Nerivio Infinity] is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment.
There have been no changes to the indications for use compared to the predicate device, other than a minor change in the name of Nerivio Infinity (from one word to two words).
## Summary of Technological Characteristics
The technological characteristics of the subject Nerivio and Nerivio Infinity are nearly identical to those of the predicate Nerivio and Nerivio Infinity (K241756), and the principles of operation are the same. Both the subject devices and the predicate devices function as remote electrical neuromodulation (REN) devices that utilize electrostimulation that relieves migraine headache, and/or reduces the number of migraine days (depends on the treatment regime), using equivalent output parameters. The basic pulse structure is biphasic, with symmetrical interleaving phases and rectangular shape. The amplitude shift signal alternates between a nominal maximum and a nominal minimum of the amplitude signal. The maximal output current is 40mA. The assumed impedance is 1K ohm +/- 500 ohms.
There are a few minor technological differences between the subject and predicate devices:
- The subject Nerivio Infinity includes a near field communication (NFC) tag for authentication of the electrodes prior to treatment initiation.
- The mobile application was updated to enhance the user interface, incorporate support for the new NFC tag, support new versions of mobile operating systems, remove support of old operating systems, and implement bug fixes.
- An alternate coating for the armband was introduced for supply chain resiliency.
These minor technological changes could not significantly effect the safety or effectiveness of the device, and do not raise different questions of safety or effectiveness compared to the predicates. Table 1 provides a comparison between the key features of the subject devices and predicate devices.
{6}
K250405
Page 3 of 6
Table 1. Comparison between Subject and Predicate Devices
| Characteristic | Subject Device | | Predicate Device (K241756) | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nerivio | Nerivio Infinity | Nerivio | NerivioInfinity | |
| Indications for Use | Nerivio is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment. | Nerivio Infinity is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment. | Nerivio is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment. | NerivioInfinity is indicated for acute and/or preventive treatment of migraine with or without aura in patients 8 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura for acute treatment, or every other day for preventive treatment. | Same (minor change to name of Nerivio Infinity model) |
| Prescription or OTC | Prescription | | Prescription | | Same |
| Number of channels | 1 | | 1 | | Same |
| Electrical waveform | Biphasic rectangular, modulated | | Biphasic rectangular, modulated | | Same |
| Max output voltage 500 Ω
2 KΩ
10 KΩ | 20V (measured)
60V (measured)
60V (measured) | | 20V (measured)
60V (measured)
60V (measured) | | Same |
| Max output current 500 Ω
2 KΩ
10 KΩ | 40 mA
30 mA
6 mA | | 40 mA
30 mA
6 mA | | Same |
| Maximum phase charge (500Ω) | 8 μC | | 8 μC | | Same |
{7}
K250405
Page 4 of 6
| Characteristic | Subject Device | | Predicate Device (K241756) | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nerivio | Nerivio Infinity | Nerivio | NerivioInfinity | |
| Maximum average current (500Ω) | 1.76 mA | | 1.76 mA | | Same |
| Maximum current density (peak) (500Ω) | 1.6 mA/cm² | | 1.6 mA/cm² | | Same |
| Maximum current density (r.m.s) (500Ω) | 0.34 mA/cm² | | 0.34 mA/cm² | | Same |
| Maximum average current density (abs value) (500Ω) | 0.07 mA/cm² | | 0.07 mA/cm² | | Same |
| Maximum average power density (500Ω) | 1.41 mW/cm² | | 1.41 mW/cm² | | Same |
| Frequency | Range: 100-120 Hz
Average: 110 Hz (measured) | | Range: 100-120 Hz
Average: 110 Hz (measured) | | Same |
| Primary phase duration | 200 μSec | | 200 μSec | | Same |
| Pulse duration | 400 μSec | | 400 μSec | | Same |
| Electrode area | 25 cm² | | 25 cm² | | Same |
| Treatment location | Upper arm | | Upper arm | | Same |
| Treatment duration | 45 minutes | | 45 minutes | | Same |
| Reusable | Yes | | Yes | | Same |
| # of treatments per device | 18 treatments per one entire device (device body and electrodes). | Up to 18 treatments for each pair of electrodes (“refill unit”). No limitation for device body. | 18 treatments per one entire device (device body and electrodes). | Up to 18 treatments for each pair of electrodes (“refill unit”). No limitation for device body. | Same as respective device model. |
| Power source | LiMnO2 cell battery (non-rechargeable) | Li-Ion cell battery (rechargeable) | LiMnO2 cell battery (non-rechargeable) | Li-Ion cell battery (rechargeable) | Same as respective device model. |
| On/off button | Power push-button | | Power push-button | | Same |
{8}
K250405
Page 5 of 6
| Characteristic | Subject Device | | Predicate Device (K241756) | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nerivio | Nerivio Infinity | Nerivio | NerivioInfinity | |
| Dimensions | Device (including the electrodes): 12.0 x 7.5 x 1.5 cm
Armband: 48.0 x 10.0 x 0.3 cm | Device body: 7.9 x 3.5 x 1.7 cm
Refill unit: 13.7 x 7.6 x 5.3 cm
Armband: 48.0 x 10.0 x 0.3 cm | Device (including the electrodes): 12.0 x 7.5 x 1.5 cm
Armband: 48.0 x 10.0 x 0.3 cm | Device body: 7.9 x 3.5 x 1.7 cm
Refill unit: 13.7 x 7.6 x 5.3 cm
Armband: 48.0 x 10.0 x 0.3 cm | Same as respective device model. |
| Weight | Device (including the electrodes): 50 g
Armband: 33 g | Device body: 38 g
Refill unit: 15 g
Armband: 33 g | Device (including the electrodes): 50 g
Armband: 33 g | Device body: 38 g
Refill unit: 15 g
Armband: 33 g | Same as respective device model. |
| Mobile application software | Yes | | Yes | | Same |
| Biocompatibility | Yes | | Yes | | Same |
| Sterile | No | | No | | Same |
| Processor control | Yes | | Yes | | Same |
| Wireless control | Yes | | Yes | | Same |
| Wireless technology | Bluetooth | Bluetooth and near field communication (NFC) | Bluetooth | Bluetooth | Subject Nerivio Infinity uses NFC for authentication of the refill unit. No different questions raised. |
| Automatic overload trip | Yes | | Yes | | Same |
| Automatic no load trip | Yes | | Yes | | Same |
| Automatic shutdown | Yes | | Yes | | Same |
| Stimulation intensity control | Yes | | Yes | | Same |
| Remote (over-the-air) firmware update option | No | Yes | No | Yes | Same as respective device model. |
{9}
K250405
Page 6 of 6
# Performance Data
## Non-Clinical Tests
Non-clinical testing was not needed to support this new 510(k) notice.
## Clinical Information
The subject device includes revisions to the precautions in the device labeling regarding the effects of long-term use of Nerivio and Nerivio Infinity, and regarding device use during pregnancy. The revision to the precaution about long term use to remove a precaution statement that the long term effects of chronic use of the device are unknown was supported by post market studies of real world data on use of Nerivio collected from US patients (Synowiec A, et. al. Adv Ther. 2024 Jan;41(1):170-18). The revision to the precaution about the use of Nerivio during pregnancy to remove the statement that the device has not been evaluated for use in pregnant women and add a new precaution that users should consult their HCP if they are pregnant or intend to get pregnant prior to using the device was supported by a retrospective survey-study published in literature (Peretz A, et. al. Headache. 2023;63(7):968-970). The data reported on Nerivio is applicable to Nerivio Infinity due to the high degree of similarity and identical therapeutic outputs and controls between Nerivio and Nerivio Infinity.
## Conclusions
The subject device has the same intended use, indications for use, and core technological characteristics and principles of operation as the previously cleared predicate device. The only differences between the subject and predicate devices are the change in labeling precautions regarding long-term use and use during pregnancy, and minor technological updates to limit the ability of users to overuse electrodes, provide greater flexibility for manufacturing the armband in the event of supply chain issues, and updates to the mobile app software for a smoother user experience, compatibility with the latest operating systems, and routine bug fixes. None of these changes results in a change in the device's overall purpose of treating and preventing migraine, nor do they raise different questions of safety or effectiveness. Thus, the subject device is substantially equivalent to cited predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.