QGT · Distal Transcutaneous Electrical Stimulator For Treatment Of Acute Migraine

Neurology · 21 CFR 882.5899 · Class 2

Overview

Product CodeQGT
Device NameDistal Transcutaneous Electrical Stimulator For Treatment Of Acute Migraine
Regulation21 CFR 882.5899
Device ClassClass 2
Review PanelNeurology

Identification

A trunk and limb electrical stimulator to treat headache is a device intended to treat headache through the application of electrical stimulation anywhere on the body of the patient apart from the patient's head or neck through electrodes placed on the skin. The stimulation may be provided transcutaneously or percutaneously.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the trunk and limb electrical stimulator to treat headache is subject to the following special controls: 1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. This testing must include: a) Characterization of the electrical stimulation, including the following: waveforms; output modes; maximum output voltage and maximum output current (at 500Ω , 2kQ. and 10k≤2 loads); pulse duration; frequency; net charge per pulse; and maximum phase charge, maximum current density, maximum average current, and maximum average power density (at 500Ω); b) Characterization of the impedance monitoring system; and c) Characterization of the electrode performance including the electrical performance, adhesive integrity, shelf-life, reusability, and current distribution of the electrode surface area. 2) The patient-contacting components of the device must be demonstrated to be biocompatible. 3) Performance testing must demonstrate electromagnetic compatibility and electrical, mechanical and thermal safety in the intended use environment. 4) Software verification, validation, and hazard analysis must be performed. 5) Labeling must include the following: a) Instructions for use, including the typical sensations experienced during treatment: b) A detailed summary of the electrical stimulation output, and the device technical parameters, including any wireless specifications; c) A shelf life for the electrodes and reuse information; and d) Instructions on care and cleaning of the device.

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. This testing must include: (i) Characterization of the electrical stimulation, including the following: Waveforms; output modes; maximum output voltage and maximum output current (at 500Ω, 2kΩ, and 10kΩ loads); pulse duration; frequency; net charge per pulse; and maximum phase charge, maximum current density, maximum average current, and maximum average power density (at 500Ω); (ii) Characterization of the impedance monitoring system; and (iii) Characterization of the electrode performance including the electrical performance, adhesive integrity, shelf-life, reusability, and current distribution of the electrode surface area. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Performance testing must demonstrate electromagnetic compatibility and electrical, mechanical, and thermal safety in the intended use environment. (4) Software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) Instructions for use, including the typical sensations experienced during treatment; (ii) A detailed summary of the electrical stimulation output, and the device technical parameters, including any wireless specifications; (iii) A shelf life for the electrodes and reuse information; and (iv) Instructions on care and cleaning of the device.

Recent Cleared Devices (7 of 7)

RecordDevice NameApplicantDecision DateDecision
K250405Nerivio; Nerivio InfinityTheranica Bio-Electronics, Ltd.May 14, 2025SESE
K241756Nerivio; NerivioInfinityTheranica Bio-Electronics, Ltd.Oct 8, 2024SESE
K232152NerivioInfinityTheranica Bioelectronics , Ltd.Nov 8, 2023SESE
K223169NerivioTheranica Bioelectronics , Ltd.Feb 6, 2023SESE
K203181Nerivio, FGD000075-4.7Theranica Bioelectronics , Ltd.Jan 22, 2021SESE
K201824NerivioTheranica Bio-Electronics, Ltd.Oct 23, 2020SESE
DEN180059Nerivio MigraTheranica Bioelectronics , Ltd.May 20, 2019DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...