K250389 · Changzhou Xin Neng Yuan Medical Stapler Co.,Ltd · KNT · Apr 10, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K250389
Device Name
XNY Disposable Gastric Calibration Tube
Applicant
Changzhou Xin Neng Yuan Medical Stapler Co.,Ltd
Product Code
KNT · Gastroenterology, Urology
Decision Date
Apr 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.
Device Story
Disposable gastric calibration tube used in gastric and bariatric surgery; facilitates stomach decompression, fluid drainage, and gastric pouch sizing. Device consists of medical-grade silicone tubing; available in single or dual-lumen configurations; includes aspiration holes and optional inflatable balloon for calibration. Operated by surgeons or clinical staff during procedures; connects to external suction source via integrated adapters. Provides physical sizing guide and suction pathway; assists in determining residual gastric capacity and removing gastric contents. Benefits include accurate pouch sizing and efficient fluid/gas removal.
Clinical Evidence
Bench testing only. Performance data includes balloon firmness, suction port/catheter connection strength, aspiration function (500ml/min), inflation valve integrity, balloon leakage, biocompatibility (ISO 10993: no cytotoxicity, sensitization, or irritation), dimensional accuracy, scale mark clarity, bend/break resistance, and sealability.
Technological Characteristics
Medical-grade silicone construction. Dual or single lumen configurations. Features distal tip with aspiration holes, optional inflatable balloon (≥100cc/min capacity), and suction adapters. Dimensions range 10.5mm to 14.5mm diameter; lengths 700mm or 900mm. Sterile, single-use. Mechanical operation via suction and manual inflation.
Indications for Use
Indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, and size the gastric pouch.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2025
Changzhou Xin Neng Yuan Medical Stapler Co., Ltd.
% Jenny Xia
Director
LSI International Inc
504E Diamond Ave., Suite H
Gaithersburg, Maryland 20877
Re: K250389
Trade/Device Name: XNY Disposable Gastric Calibration Tube
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube And Accessories
Regulatory Class: Class II
Product Code: KNT
Dated: February 9, 2025
Received: February 11, 2025
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250389 - Jenny Xia
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K250389 - Jenny Xia
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony C. Lee Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250389 | |
| Device Name XNY Disposable Gastric Calibration Tube | |
| Indications for Use (Describe) XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K250389 Page 1 of 4
# 510k Summary
# K250389
# 1. General Information
| Applicant Information | Changzhou Xin Neng Yuan Medical Stapler Co., Ltd. No.51 Shuishan Rd, Zhonglou Economic Development District Changzhou, Jiangsu, China www.xnywhq.com whq@xnywhq.com |
| --- | --- |
| Contact Person | Jenny Xia LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, MD 20877 jxia@lsi-consulting.org 301-525-6856 |
| Date Summary Prepared | February 9, 2025 |
| Device Common Name | Disposable Gastric Calibration Tubes |
| Device Trade Name | XNY Disposable Gastric Calibration Tubes |
| FDA Product Code | KNT |
| Classification Name | 876.5980 Gastrointestinal tube and accessories |
| Classification | Class II |
| Predicate Device | Gastrointestinal Boundary Identifier (GIBI HD™), K221898 |
# 2. Device Description
The XNY Disposable Gastric Calibration Tubes are indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. The device has the advantages of accurately determining residual gastric capacity and easy positioning, and it can be connected to a suction device to extract the gas or liquid in the patient's stomach.
# 3. Indication for Use
XNY Disposable Gastric Calibration Tube is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch.
# 4. Comparison of Technological Characteristics with the Predicate Device
| Feature | Predicate Device Calibration Tube, K221898 | Subject Device Calibration Tubes K250389 | Effect on Substantial Equivalence |
| --- | --- | --- | --- |
| Indications for Use | The GIBI HDTM is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide. | Product is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid and size the gastric pouch. | None |
| Outer Diameter | B-2032: 10.5mm +0/-0.3mm | JZGA/B-32FR: 10.5mm±0.5mm | Outer diameters are similar and no impact to |
| | | | them. |
| Diameter | B-2032: 10.5mm | JZGA/B-32FR: 10.5mm±0.5mm | Dilatation is not significant. |
{6}
K250389 Page 2 of 4
| | B-2036: 11.9mm +0/-0.3mm
B-2040: 13.2mm +0/-0.3mm | JZGA/B-34FR: 11.2mm±0.5mm
JZGA/B-36FR: 11.9mm±0.5mm
JZGA/B-38FR: 12.5mm±0.5mm
JZGA/B-40FR: 13.2mm±0.5mm
JZGA/B-42FR: 13.9mm±0.5mm
JZGA/B-44FR: 14.5mm±0.5mm | substantial equivalence |
| --- | --- | --- | --- |
| Working Length | 823.7mm±13.2mm | JZGA type: 700mm±10mm
JZGB type: 900mm±20mm | No impact to substantial equivalence with the similar lengths. |
| Tubing | Dual lumen | JZGA: Dual lumen
JZGB: single lumen | Dual lumen and single lumen only choose different models according to different clinical needs, and no impact to substantial equivalence. |
| Distal Tip | Includes a tip distal to the balloon with 12 aspiration holes. | Includes a tip distal to the balloon with 3 aspiration holes. | The number of suction holes has no substantial effect on the actual use. No impact to substantial equivalence |
| Connector for Suction | The catheter includes an adapter for room suction | There is a matching adapter, which can be adapted to various connectors | None |
| Balloon + inflation Valve | The balloon has an inflation capacity ≥ 100 cc/min. The catheter is attached to a 6-inch tubing with a stopcock for filling the balloon. | JZGA type balloon with inflation capacity ≥ 100cc/min.
JZGB has no balloon | None |
| Tubing Material | Silicone | Medical grade silicone | None |
| Markings | Indication marks at 30, 35, 40, 45, and 50 centimeters. | A type indication marks at 52, 53, 54, 55, 56, 57, 58, 59, 60 centimeters;
B type indication marks at 30, 31, 32, 33...89, 90 centimeters | Marks are included and no impact to Substantial equivalence |
| Connector | Luer taper, tape adapter | Normal joint, tape joint | Similar adapters and no impact to substantial equivalence |
| Sterility | Non-sterile, disposable, single patient use | Sterile, single-use, single patient use | None |
{7}
K250389 Page 3 of 4
# 5. Performance Data
| Test Performed | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Firmness of the balloon | The purpose of this test is to evaluate the firmness of the balloon of Disposable gastric calibration tube. | The firmness of the balloon is greater than 60N. | Pass |
| The firmness between the suction port and the catheter | The purpose of this test is to evaluate the firmness between the suction port and the catheter of Disposable gastric calibration tube. | The firmness between the suction port and the catheter is greater than 5N. | Pass |
| The catheter aspiration function | The purpose of this test is to evaluate catheter aspiration function of Disposable gastric calibration tube. | The catheter should be able to draw out 500ml of liquid within 1 minute. | Pass |
| The firmness between suction port 2 (adapter) and catheter | The purpose of this test is to evaluate the firmness between suction port 2 and catheter of Disposable gastric calibration tube. | The firmness between suction port 2 and catheter is greater than 30N. | Pass |
| The firmness of the inflation valve | The purpose of this test is to evaluate the firmness of the inflation valve of Disposable gastric calibration tube. | The firmness of the inflation valve is greater than 20N. | Pass |
| Balloon leakage test | The purpose of this test is to evaluate balloon leakage. | Can tolerate double the working pressure of 80 ml gas without leakage or damage. | Pass |
| Biocompatibility Testing | The purpose of this test is to evaluate the device to meet ISO 10993 on biocompatibility for Biological evaluation of medical devices. | No cytotoxicity
No sensitization
No irritation. | Pass |
| Balloon Inspection | The purpose of this test is to evaluate the appearance of the balloon of Disposable gastric calibration tube. | The balloon should be colorless and transparent.
The balloon should be tightly connected to the tube and the connection between the balloon and the tube should be homogeneous and without twisting. | Pass |
| Tube Dimension | The purpose of this test is to evaluate the tube dimension of Disposable gastric calibration tube. | The tube length and diameter meet the required measurements. | Pass |
| Tube scale mark inspection | The purpose of this test is to evaluate the appearance of the tube scale marks on Disposable gastric calibration tube. | The scale marks on the tube should be clear and not faded. | Pass |
{8}
K250389 Page 4 of 4
| Tube resistance to bend and break | The purpose of this test is to evaluate the Disposable gastric calibration tube’s resistance to bending and breaking. | Hold the tube at both ends and bend the tube so that the 1cm of long tube at both ends are parallel and in contact, hold for 15 seconds, then release, ensuring that no creases form at the bend of the catheter. | Pass |
| --- | --- | --- | --- |
| Firmness of insertion tip and tube | The purpose of this test is to evaluate the firmness of the insertion tip and tube of Disposable gastric calibration tube. | The connection between the Insertion tip and the tube should be able to withstand a static axial pull of 15N for 15S without breaking or falling off. | Pass |
| The firmness between suction port 1(Adapter 1) and suction port 2(Adapter 2) is greater than 5N. | The purpose of this test is to evaluate the firmness between adapter 1 and adapter 2 | The firmness between adapters is greater than 5N. | Pass |
| Pipe Clip Sealability Test. | The purpose of this test is to evaluate the Pipe Clip sealability of disposable gastric calibration tubes. | When the Pipe Clip is closed, no bubbles are generated at the catheter port. | Pass |
## 6. Basis for Substantial Equivalence
Substantial equivalence of the XNY Disposable Gastric Calibration Tube to the predicate device was established through an evaluation of the indications for use, principle of operation, device design, materials of construction, and an assessment of safety, and effectiveness via bench studies.
The data presented in this summary demonstrates the technological similarity and equivalency of the XNY Disposable Gastric Calibration Tube to the predicate device (GIBI HD, K221898).
The devices:
- have the same intended use,
- use the same principle of operation,
- incorporate the same basic design,
- use similar construction and material
## 7. Conclusion
Based on the non-clinical assessments, the comparison of the device classification, indications for use, operating principle, technological characteristics, and biocompatibility that the subject device, the XNY Disposable Gastric Calibration Tube, is substantially equivalent to the predicate device, the GIBI HD, K221898. Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.