The GIBI HD™ is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide.
Device Story
GIBI HD is a flexible, dual-lumen gastric catheter used during gastric and bariatric surgeries. One lumen facilitates drainage, suction, and irrigation; the second lumen manages balloon inflation/deflation. The device provides visible and tactile delineation of the stomach antrum and serves as a sizing guide for the gastric pouch. It is operated by surgeons in a clinical/surgical setting. The device includes a 32-inch suction tube and a 6-inch fill tube with a stopcock. By providing a calibrated sizing guide and drainage capability, it assists the surgeon in performing bariatric procedures, potentially improving surgical outcomes.
Clinical Evidence
Bench testing only. Performance data included balloon leakage testing (100 cc retention), fluid removal testing (500 cc in <1 minute), and mechanical bond strength testing for the balloon, distal tip, stopcock, fill tube, and adapter components (all meeting minimum lbf requirements). Visual and measurement verification confirmed catheter length (75 cm), indicator mark placement (30-60 cm), and outer diameter sizing (32, 36, and 40 Fr).
Technological Characteristics
Dual-lumen flexible catheter; material: silicone; dimensions: 32 Fr, 36 Fr, and 40 Fr outer diameters; working length ~32.43 inches; includes integrated balloon, suction tubing, and stopcock-controlled fill tube. Non-sterile. No electronic or software components.
Indications for Use
Indicated for use in gastric and bariatric surgical procedures for stomach decompression, gastric fluid drainage/removal, irrigation, and gastric pouch sizing.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 28, 2022
ReShape Lifesciences Michelle Ravert Regulatory Affairs Manager 1001 Calle Amanecer San Clemente, CA 92673
Re: K221898
Trade/Device Name: Gastrointestinal Boundary Identifier (GIBI HDTM) Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: KNT Dated: June 28, 2022 Received: June 30, 2022
Dear Michelle Ravert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the
{2}------------------------------------------------
Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
April Marrone, Ph.D., MBA Assistant Director (acting) DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221898
Device Name
Gastrointestinal Boundary Identifier (GIBI HDTM)
Indications for Use (Describe)
The GIBI HD™ is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for ReShape Lifesciences. The logo consists of an orange circle that is not fully connected, with the text "ReShape" in purple next to it. Below the word "ReShape" is the word "Lifesciences" in gray, with a trademark symbol next to it.
# 510(k) Summary
#### I. Basic Information
| 510(k) Owner | ReShape Lifesciences<br>1001 Calle Amanecer<br>San Clemente CA 92673<br>Phone: (844) 937-7374<br>Establishment Registration No: 3013508647 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Michelle Ravert, Regulatory Affairs Manager |
| Date summary<br>Prepared | July 26, 2022 |
| Trade Name | Gastrointestinal Boundary Identifier (GIBI HD™) |
| Common Name | Introduction/Drainage Catheter |
| Classification name | Gastrointestinal tube and accessories (21 CFR 876.5980) |
| Regulatory Class | Class II |
| Product Code | KNT |
| Predicate Device | Lap-Band® System Calibration Tube, K220455 |
| Reference Device | ViSiGi 3D®, K130483 |
#### DEVICE DESCRIPTION II.
The Gastrointestinal Boundary Identifier, or GIBI HD™, is a flexible gastric tube designed to be used in gastric and bariatric surgical procedures. The catheter provides visible and tactile delineation of the antrum of the stomach along with the ability to decompress the stomach, drain and remove gastric fluid and size a gastric pouch.
The dual lumen GIBI HD utilizes one lumen for drainage, suction and irrigation and the second lumen to inflate/deflate the balloon. The catheter is attached to a 32-inch suction tubing and a 6inch tubing with a stopcock for filling the balloon.
#### III. INDICATIONS FOR USE
The GIBI HD is indicated for use in gastric and bariatric surgical procedures to decompress the stomach, drain and remove gastric fluid, irrigate, and act as a sizing guide.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH IV. THE PREDICATE DEVICE
| Technical Element | GIBI HD<br>B-2032, B-2036, B-2040 | Predicate Device<br>Calibration Tube, K220455 |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Outer Diameter<br>(OD), inches | B-2032 = 0.420" +0.000" / -0.012"<br>B-2036 = 0.472" + 0.000" / -0.012"<br>B-2040 = 0.525" + 0.000" / -0.012" | 0.380" +0.000" / -0.012" |
| OD, French Size | B-2032 = 32 F<br>B-2036 = 36 F<br>B-2040 = 40 F | 29F |
| Working length<br>(distal end of Y -<br>connector to distal<br>end of distal tip) | 32.43 ± 0.52" | 24.73" ± 0.52" |
| Tubing Material | Silicone | Silicone |
#### V. PERFORMANCE DATA
| Test Performed | Purpose | Acceptance Criteria | Results |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|---------|
| Balloon leakage<br>test | The balloon component of the GIBI HD shall<br>inflate to a minimum volume of 100 cc without<br>exhibiting fluid leak or air egress. | Samples retained 100 cc of fluid at<br>least 1 minute without leakage. | Pass |
| Fluid removal test | The GIBI HD shall remove a minimum of 500<br>cc of saline or clean water in under 1 minute. | Samples evacuated a minimum of<br>500 cc of fluid in under 1 minute. | Pass |
| Balloon bond joint<br>strength test | The mated joint between the GIBI HD 's Balloon<br>and Tubing shall withstand a minimum of 13 lbf<br>prior to separation. | Samples did not exhibit balloon<br>separation under 13 lbf. | Pass |
| Distal tip bond<br>joint strength test | The mated joint between the GIBI HD 's distal<br>tip and Tubing shall withstand a minimum of<br>19.5 lbf prior to separation. | Samples did not exhibit sensor tip<br>separation under<br>19.5 lbf. | Pass |
| Stopcock<br>separation test | The stopcock component of the GIBI HD shall<br>withstand a minimum of 1 lbf prior to<br>separation from the fill tube. | Samples did not exhibit stopcock<br>separation under 1 lbf. | Pass |
| Fill tube<br>separation test | The fill tube attached to the stopcock of the<br>GIBI HD shall withstand a minimum of 1 lbf<br>prior to separation from the y connector. | Samples did not exhibit fill tube<br>separation under 1 lbf. | Pass |
| Adapter<br>tubing<br>separation<br>test | The proximal tubing of the GIBI HD shall<br>withstand a minimum of 1 lbf prior to<br>separation from the y connector. | Samples did not exhibit adapter<br>tube separation under 1 lbf. | Pass |
| Adapter<br>separation test | The adapter of the GIBI HD shall withstand a<br>minimum of 1 lbf prior to separation from the<br>adapter tube. | Samples did not exhibit adapter<br>separation under 1 lbf. | Pass |
{6}------------------------------------------------
| Test Performed | Purpose | Acceptance Criteria | Results |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------|
| Visual Inspection<br>and Measurement<br>verification test | The dual lumen catheter component of the<br>GIBI HD, from the distal edge of the Y<br>connector to the end of the dual lumen<br>catheter, shall have a minimum length of 75<br>cm. | Minimum length of measured<br>sample is 75 cm. | Pass |
| | The dual lumen catheter component of the GIBI<br>HD shall include<br>indication marks on the surface of the tube at 30,<br>35, 40, 45, 50, 55 and 60 cm with respect to the<br>distal portion of the component. | Indicator marks are located 30, 35,<br>40, 45, 50, 55,<br>and 60 cm with respect to distal<br>portion of the component. | Pass |
| | The dual lumen catheter component of the<br>GIBI HD shall have three size configurations:<br>0.42" (32 Fr), 0.47" (36 Fr) and 0.53" (40 Fr). | Outer diameter of 32Fr, 36Fr and<br>40Fr samples are 0.42", 0.47" and<br>0.53", respectively. | Pass |
#### BASIS FOR SUBSTANTIAL EQUIVALENCE VI.
Substantial equivalence of the GIBI HD to the predicate device (Lap-Band® System Calibration Tube, K220455) was established through an evaluation of the indications for use, principle of operation, device design, materials of construction, and an assessment of usability, safety, and effectiveness via bench studies.
The data presented in this summary demonstrates the technological similarity and equivalency of the GIBI HD to the predicate device (Lap-Band® System Calibration Tube, K220455).
The devices:
- · have the same intended use,
- · use the same principle of operation,
- · incorporate the same basic design,
- · use similar construction and material, and
- · are provided non-sterile.
In addition, the GIBI HD and the reference device (ViSiGi 3D, K130483) have the same outer
diameter sizes and are substantially equivalent with respect to working length.
In summary, the Gastrointestinal Boundary Identifier (GIBI HD™) device described in this submission is substantially equivalent to the predicate device, the LAP-BAND System Calibration tube (K220455).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.