K250276 · Embolization, Inc. · KRD · May 15, 2025 · Cardiovascular
Device Facts
Record ID
K250276
Device Name
Nitinol Enhanced Device (NED)
Applicant
Embolization, Inc.
Product Code
KRD · Cardiovascular
Decision Date
May 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nitinol Enhanced Device (NED) is intended for arterial and venous embolization in the peripheral vasculature. This device is not intended for neurovascular use.
Device Story
NED system consists of polymer-based framing and packing implants, delivery system, and transfer/removal tool; used for embolization in 3-6 mm peripheral vessels. Implants feature UHMWPE braid and nylon 6,6 tails to scaffold thrombus formation; mechanical obstruction combined with thrombus results in vessel occlusion. Framing implant deployed first to create backing; packing implants deployed subsequently per physician discretion. Delivery system includes super-elastic nitinol pusher with platinum/iridium radiopaque marker; allows repositioning and withdrawal prior to detachment. Transfer/removal tool facilitates implant transfer into 0.027" microcatheter. Used in clinical settings by physicians; output is physical vessel occlusion. Benefits include controlled, repositionable, and retrievable embolization for peripheral vascular procedures.
Clinical Evidence
No human clinical testing performed. Substantial equivalence supported by non-clinical bench testing and GLP animal studies (acute and subchronic ovine models). Bench testing included biocompatibility (ISO 10993-1), corrosion testing (ASTM F2129, F3044), tensile strength, deployment/retraction force, detachment mechanism testing, and MRI/CT compatibility.
Indicated for arterial and venous embolization in the peripheral vasculature. Not for neurovascular use.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Interlock-18 Fibered Interlocking Detachable Coil (IDC) Occlusion System (K132578)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 15, 2025
Embolization Inc.
% Melissa Brookshier
Directory of Quality and Regulatory Affairs
Boulder BioMed, Inc.
5421 Western Ave.
Boulder, Colorado 80301
Re: K250276
Trade/Device Name: Nitinol Enhanced Device (NED)
Regulation Number: 21 CFR 870.3300
Regulation Name: Vascular Embolization Device
Regulatory Class: Class II
Product Code: KRD
Dated: May 2, 2025
Received: May 5, 2025
Dear Melissa Brookshier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250276 - Melissa Brookshier
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K250276 - Melissa Brookshier
Page 3
Sincerely,
Misti L. Malone -S
Misti Malone, PhD
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250276 | |
| Device Name Nitinol Enhanced Device (NED) | |
| Indications for Use (Describe) The Nitinol Enhanced Device (NED) is intended for arterial and venous embolization in the peripheral vasculature. This device is not intended for neurovascular use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K250276
Page 1 of 8
# 510(K) SUMMARY NITINOL ENHANCED DEVICE (NED)
## 1. SUBMISSION SPONSOR
Company Name: Embolization, Inc.
Address: 5421 Western Avenue, Boulder, CO 80301
Contact: James Frank Kasic, II
Title: CEO
## 2. SUBMISSION CORRESPONDENT
Boulder iQ, Inc.
5421 Western Avenue
Boulder, CO 80301
Office Phone: 303-531-1238 x726
Contact: Melissa Brookshier
Title: Directory of Quality and Regulatory Affairs
## 3. DATE PREPARED
April 1, 2025
## 4. DEVICE IDENTIFICATION
Trade/Proprietary Name: Nitinol Enhanced Device (NED)
Common/Usual Name: Vascular embolization device
Classification Name: Vascular embolization device
Regulation Number: 21CFR §870.3300
Product Code: KRD
Device Class: Class II
Classification Panel: Cardiovascular
{6}
K250276
Page 2 of 8
## 5. LEGALLY MARKETED PREDICATE / REFERENCE DEVICES
Predicate Device: K132578 - Interlock-18 Fibered Interlocking Detachable Coil (IDC) Occlusion System (Boston Scientific Corporation)
Reference Device: K141378 - IDC Interlocking Detachable Coil (Boston Scientific Corporation)
## 6. INDICATIONS FOR USE
The Nitinol Enhanced Device (NED) system is intended for arterial and venous embolization in the peripheral vasculature. This device is not intended for neurovascular use.
## 7. DEVICE DESCRIPTION
The NED system is comprised of polymer-based occlusion devices intended for embolization procedures in 3-6 mm vessels of the peripheral vasculature. It is intended to be used with 0.027" labeled microcatheters. The NED consists of three components: the NED implants (framing and packing), the delivery system, which is supplied engaged with the implant, and the transfer/removal tool.
The NED framing and packing implants are permanently implanted and include a 0.016" diameter polymer filament which compacts upon delivery. The implants work together to obstruct blood flow in the vessel. The framing implant is deployed first to form the initial pack and create a backing for additional implants to be deployed in accordance with physician discretion. The deployment of the framing implant is followed by the deployment of packing implants, the quantity of which is determined by the physician based on vessel size and occlusion. The physician also has the option to finish the implant pack with another framing implant. The UHMWPE braid on the implants and the nylon 6,6 tails on the packing implant provide a scaffold for thrombus to form and attach. The combination of the mechanical obstruction of blood flow and the formation of thrombus result in vessel occlusion.
The delivery system is comprised of two components: the implant pusher and the introducer. The NED delivery system comes with the NED implant pre-loaded in its package, sterile and ready for use. The pusher is a 180 cm long 0.022" diameter single super-elastic nitinol wire with a 30 cm PTFE jacket and a platinum/iridium radiopaque marker near its distal end. The nitinol pusher coupler is located on the distal end of the pusher, which as described above, serves to engage with the coupler on the implant's proximal end. The NED pusher allows the physician to reposition the NED to ensure preferred vessel positioning prior to detachment. Further, the detachment mechanism provides the physician the ability to completely withdraw the NED before detachment, should a condition result that necessitates its removal. The introducer is a PTFE tube used to protect the implant and maintain coupling between the implant and pusher prior to use (as described in Section 2.1). Once the implant and pusher are transferred into the delivery microcatheter, the introducer is discarded.
The transfer/removal tool is a disposable accessory included with the device that attaches to the hub of the delivery microcatheter prior to inserting the implant. The purpose of the transfer/removal tool is to facilitate the transfer of the implant from the device introducer
{7}
K250276
Page 3 of 8
into the delivery microcatheter hub and to facilitate the removal of the implant from the delivery microcatheter if necessary. The transfer/removal tool is designed to fit 0.027” labeled microcatheters. The tool is comprised of an ABS and PC Luer adapter assembly bonded with medical grade Dymax 1162-M UV adhesive. The tool utilizes a spring-loaded 300 series stainless steel tube to form the lumen.
{8}
K250276
Page 4 of 8
# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
The following table compares the Nitinol Enhanced Device (NED) to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
| | Embolization, Inc.
NED
(Subject Device) | Boston Scientific
Interlock-18 Fibered
IDC Occlusion System
(Predicate) | Boston Scientific IDC
Interlocking
Detachable
Coil (Reference
Device) | Comparison to
Predicate |
| --- | --- | --- | --- | --- |
| 510(k) Number | K250276 | K132578 | K141378 | N/A |
| Regulation Number | 21 CFR §870.3300 | 21 CFR §870.3300 | 21 CFR §870.3300 | Identical to Predicate
and Reference Devices |
| Product Code | KRD | KRD | KRD | Identical to Predicate
and Reference Devices |
| Classification | Class II | Class II | Class II | Identical to Predicate
and Reference Devices |
| Intended Use | Embolization in the
peripheral
vasculature | Embolization in the
Peripheral
vasculature | Embolization in the
peripheral
vasculature | Identical to Predicate
and Reference Devices |
| Indications for
Use | The NED is intended
for arterial and
venous embolization
in the peripheral
vasculature. This
device is not
intended for
neurovascular use. | The Interlock-18
Fibered ID Occlusion
System is indicated to
obstruct or reduce
the rate of blood flow
in the peripheral
vasculature. This
device is not
intended for
neurovascular use. | The IDC Coil is a
modified interlocking
detachable coil
indicated to obstruct
or reduce rate of
blood flow in the
peripheral
vasculature. This
device is not
intended for
neurovascular use. | Substantially
Equivalent to
Predicate Device |
{9}
K250276
Page 5 of 8
| | Embolization, Inc. NED (Subject Device) | Boston Scientific Interlock-18 Fibered IDC Occlusion System (Predicate) | Boston Scientific IDC Interlocking Detachable Coil (Reference Device) | Comparison to Predicate |
| --- | --- | --- | --- | --- |
| **Components** | Implants (Framing and Packing), Delivery System and Transfer/ Removal Tool | Implant and Delivery System | Implant (to supplement previously deployed Interlock-18 implants in same procedure) | Substantially Equivalent to Predicate Device For Use with IDC Coil Reference Device |
| **Implant Material** | Framing implant: Radiopaque polymer, UHMWPE external braid, nitinol, PET atraumatic tip
Packing implant: Radiopaque polymer with thrombogenic tails, UHMWPE external braid, PET atraumatic tip | Platinum with thrombogenic tails | Platinum-tungsten alloy | Substantially Equivalent to Predicate Device |
| **Thrombogenic Fibers?** | Yes | Yes | No | Identical to Predicate Device |
| **Device Length** | Framing implant: 6, 10, and 15 cm
Packing implant: 14 cm | 2.3, 4, 4.1, 5.8, 6, 8, 10, 12, 15, 20, 30, 50, 60 cm | 2-12cm, various diameters | Substantially Equivalent to Predicate Device |
| **Target Vessel/ Implant Curl Diameter** | 3-4, 4-5, 5-6 mm | 2, 3, 4, 5, 6, 8, 10, 12, 14, 18, 20, 22 mm | N/A | Substantially Equivalent to Predicate Device |
{10}
K250276
Page 6 of 8
| | Embolization, Inc. NED (Subject Device) | Boston Scientific Interlock-18 Fibered IDC Occlusion System (Predicate) | Boston Scientific IDC Interlocking Detachable Coil (Reference Device) | Comparison to Predicate |
| --- | --- | --- | --- | --- |
| Delivery Microcatheter/ Sheath ID | 0.027" | 0.021" | 0.021" | Substantially Equivalent to Predicate Device |
| Detachment Method | Interlocking arms detach when past catheter tip | Interlocking arms detach when past catheter tip | Interlocking arms detach when past catheter tip | Identical to Predicate and Reference Devices |
| Single-Use? | Yes | Yes | Yes | Identical to Predicate and Reference Devices |
| MR Cond Up to 3.0 Tesla? | Yes | Yes | Unknown | Identical to Predicate Device |
{11}
K250276
Page 7 of 8
# 9. NON-CLINICAL PERFORMANCE DATA
The NED meets all the requirements for overall design, sterilization, and biocompatibility.
The NED passed all the testing in accordance with internal requirements, FDA Guidance, national standards, and international standards shown below to support substantial equivalence of the subject device:
- Nitinol Corrosion Testing (ASTM F2129 and ASTM F3044)
- Product Evaluation for Sharp Edges
- Chemical Characterization of Materials, including Metals Analysis by Inductively Coupled Plasma Mass Spectrometry, Semi-Volatile and Amenable Non-Volatile Organic Compounds by Gas Chromatography/ Mass Spectrometry, Non-Volatile Organic Compounds Semi-Quantitative Scan by Liquid Chromatography/Mass Spectrometry, and Volatile Organic Compounds by Headspace Gas Chromatography/ Mass Spectrometry, and Toxicological Risk Assessment
- Biocompatibility evaluation per ISO 10993-1
- Successful Deployment
- Ease of Delivery
- Delivery and Deployment Force Testing
- Retraction Force Testing
- Detachment Mechanism Testing
- Vessel Distension Measurement
- Compatibility with Delivery Catheters
- Visual Inspection Post-deployment
- End-to-End Tensile Strength
- Radiopacity Test
- Implant with Delivery System Removal Test
- Nitinol Corrosion Test
- Implant Retrieval Test
- Fiber Pull-Out Force
- MRI Compatibility
- CT Artifact Evaluation
- GLP Safety Study (Acute Ovine Study)
- GLP Subchronic Ovine Product Validation Study
- Packaging Integrity
- Shelf Life Validation (packaging and device)
# 10. CLINICAL PERFORMANCE DATA
No human clinical testing was required to support the NED as the indications for use are equivalent to the predicate device the Interlock-18 Fibered Interlocking Detachable Coil (IDC)
{12}
K250276
Page 8 of 8
Occlusion System. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
## 11. STATEMENT OF SUBSTANTIAL EQUIVALENCE
The Nitinol Enhanced Device (NED), as designed and manufactured, is determined to be substantially equivalent to the predicate device, the Interlock-18 Fibered Interlocking Detachable Coil (IDC) Occlusion System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.