PHAROS is a dedicated PET scanner intended to obtain Positron Emission Tomography (PET) images of parts of human body that fit in the patient aperture (brain, breast, arms and legs) to detect abnormal patterns of distribution of radioactivity after injection of a positron emitting radiopharmaceutical. This information can assist in diagnosis, therapeutic planning and therapeutic outcome assessment.
Device Story
PHAROS is a dedicated PET scanner for imaging brain, breast, and extremities (arms/legs). It detects 511 keV gamma rays resulting from positron annihilation using a ring-shaped gantry of pixelated lutetium-based scintillators coupled to silicon photomultipliers. The system includes a variable patient chair and gantry for multiple scan positions (sitting/lying). Data is processed via GPU-based reconstruction algorithms, including Time-of-Flight (TOF) and scatter correction (tail-fitting/single scatter simulation). The system connects to a workstation via optical fiber and Ethernet for image acquisition, reconstruction, and DICOM/PACS integration. Operated by clinicians in hospital settings, the device provides 3D tomographic visualizations of biochemical/metabolic processes. Output assists physicians in diagnosis, treatment planning, and monitoring therapeutic outcomes. Benefits include high-resolution imaging of specific anatomical regions with adjustable detector positioning for patient comfort.
Clinical Evidence
Bench testing performed per NEMA NU2:2018 and NEMA NU4:2008 standards, demonstrating compliance with spatial resolution (<2.3 mm), scatter fraction (<35%), sensitivity, energy resolution (<18%), and time resolution (<275 ps). Clinical effectiveness was supported by five clinical images (patients and normal controls) assessed by a nuclear medicine physician for clinical acceptability. No comparative clinical trial data provided.
Technological Characteristics
PET scanner; pixelated lutetium-based scintillators; silicon photomultiplier detectors; circular gantry configuration; GPU-based reconstruction; Time-of-Flight (TOF) enabled; connectivity via DICOM/PACS; non-sterile; electrical safety per ANSI/AAMI ES60601-1:2005/A2:2021; EMC per IEC 60601-1-2:2014/A1:2020.
Indications for Use
Indicated for adults and adolescents requiring PET imaging of the brain, breast, arms, or legs to detect abnormal radioactivity distribution following radiopharmaceutical injection. Assists in diagnosis, therapeutic planning, and outcome assessment.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 15, 2025
Brightonix Imaging, Inc.
c/o Lee Strong
Medical Device Regulatory Consultant
510K FDA Inc.
156 E. Granada Blvd.
Ormond Beach, FL 32176
Re: K250170
Trade/Device Name: PHAROS
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: KPS
Dated: July 16, 2025
Received: July 16, 2025
Dear Lee Strong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250170 – Lee Strong
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250170 – Lee Strong
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
Magnetic Resonance and Nuclear Medicine Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250170 | |
| Device Name PHAROS | |
| Indications for Use (Describe) PHAROS is a dedicated PET scanner intended to obtain Positron Emission Tomography (PET) images of parts of human body that fit in the patient aperture (brain, breast, arms and legs) to detect abnormal patterns of distribution of radioactivity after injection of a positron emitting radiopharmaceutical. This information can assist in diagnosis, therapeutic planning and therapeutic outcome assessment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary K250170 – PHAROS
August 13, 2025
1 Applicant
Brightonix Imaging, Inc.
#B203, #409, Seongsu AK valley 76, Yeonmujang-gil, Seongdong-gu,
Seoul, 04784, Republic of Korea
Phone: +82-2-7668745
Contact: Radcheck Yang, (Radcheck.yang@brtnx.com)
Consultant
510K FDA Inc.
156 E. Granada Blvd
Ormond Beach, FL 32176
Phone: (386)846-6037
Contact: Lee Strong, (lee@510kfda.com)
2 Device Identification
Trade/Model Names: PHAROS
Regulation Name: Emission computed tomography system
Regulation Number: 21 CFR 892.1200
Primary Product Code: KPS
Classification Name: System, tomography, computed, emission
Regulatory Class: Class II
510k Review Panel: Radiology
3 Predicate Device
Predicate #: K210450
Predicate Trade Name: BBX-PET Scanner
Product Code: KPS
Classification Name: System, tomography, computed, emission
Regulation: 21 CFR 892.1200
Classification: Class II
Panel: Radiology
4 Device Description
PHAROS is a specialized high-sensitivity and high-resolution PET system designed for imaging specific organs, such as the brain, breast, arms and legs.
Positron emission tomography (PET) captures images by detecting the distribution of internal radioactivity in human organs, utilizing radioactive pharmaceuticals. This technology reconstructs the body's internal biochemical and metabolic processes, producing high-resolution 3D visualizations. The method involves measuring a pair of
{5}
simultaneous gamma rays, each with an energy of $511\mathrm{keV}$ , resulting from the annihilation of positrons. By labeling the positron emitter with a tracer and using a ring-shaped gamma ray detector, the spatial location of positron-emitting nuclides within the body is visualized.
PHAROS features four different scanning modes, each tailored for specific types of imaging:
1) Brain Scan Mode (Sitting Position):
This mode is designed for brain imaging while the patient is seated.
2) Brain Scan Mode (Lying Position):
This mode is designed for brain imaging while the patient lies down on a bed.
3) Breast Scan Mode:
This mode is designed for breast imaging while the patient lies in a prone position.
4) Periphery Scan Mode:
This mode is designed for imaging the periphery of the body, including the arms, hands, legs, and knees.
For both upper and lower extremity imaging, the height of detector head can be adjusted to ensure optimal patient comfort and accurate positioning. Aside from the physical height adjustment of the detector head, there is no difference in image acquisition method or image generation algorithm between upper and lower extremity scans.
# 5 Intended Use/Indications for Use
PHAROS is a dedicated PET scanner intended to obtain Positron Emission Tomography (PET) images of parts of human body that fit in the patient aperture (brain, breast, arms and legs) to detect abnormal patterns of distribution of radioactivity after injection of a positron emitting radiopharmaceutical. This information can assist in diagnosis, therapeutic planning and therapeutic outcome assessment.
# Indications for Use Comparison
The subject device indications for use include examples of parts of human body that fit in the patient aperture (brain, breast, arms and legs).
6 Technological Comparison
| Characteristic | Predicate Device | Subject Device | Comparison |
| --- | --- | --- | --- |
| Device name | BBX-PET Scanner | PHAROS | |
| Device Photograph | | | |
| 510(k) | K210450 | K250170 | |
| Manufacturer | Prescient Imaging, LLC | Brightonix Imaging Inc. | |
| Product Code | KPS | KPS | Same as Predicate |
| Intended Use | Dedicated Positron Emission Tomography | Dedicated Positron Emission Tomography | Same as Predicate |
{6}
| Characteristic | Predicate Device | Subject Device | Comparison |
| --- | --- | --- | --- |
| Indications for Use | BBX-PET is a scanner intended to obtain Positron Emission Tomography (PET) images of parts of the human body that fit in the patient aperture (e.g., head) to detect abnormal patterns of distribution of radioactivity after injection of a positron-emitting radiopharmaceutical. This information can assist in research, diagnosis, therapeutic planning, and therapeutic outcome assessment. | PHAROS is a dedicated PET scanner intended to obtain Positron Emission Tomography (PET) images of parts of human body that fit in the patient aperture (brain, breast, arms and legs) to detect abnormal patterns of distribution of radioactivity after injection of a positron emitting radiopharmaceutical. This information can assist in diagnosis, therapeutic planning and therapeutic outcome assessment. | Similar |
| Principle of detection | Positron Emission Tomography (PET) system to image the distribution of injected positron emitting radiopharmaceuticals into live humans or animals. The BBX-PET Scanner produces images that represent the internal distribution of radioactivity in the head. | PHAROS generates tomographic images that represent the internal distribution of radioactivity in the patient's head and brain, Extremity, Breast | Similar |
| Scintillator configuration / Scanner | Double-layer staggered Silicate Lutetium Fine pixelated crystals (13 x 13 and 14 x 14 arrays, 1.76-mm pitch) coupled to light detector solid state silicon photomultiplier. One hundred twenty-eight blocks positioned in a circular shape make up the gantry, with bore diameter of 288mm, and 250mm and 100mm transaxial and axial FoVs. | Pixelated lutetium-based scintillator coupled to silicon photomultipliers. A single detector module consists of 24 x 32 scintillators with 1.92 x 1.92 x 15 mm3. 2(B480D)/3(B720D)/4(B960D) rings of 20 detector modules positioned in circular shape. Bore diameter is 320 mm and useful FOV is 300 mm transaxial and 130/196/262 mm (B480D/B720D/B960D) axial depending on type. | Both use pixelated scintillation crystals coupled to silicon photomultipliers in circular shape. PHAROS uses single-layer crystal, but predicate device uses dual-layer crystal. PHAROS has bigger bore size and axial field-of-view. These differences do not affect the indications, only the performance. |
| Target population | Adults, young adults | Adults and adolescent | Same as Predicate |
| Anatomical site | Parts of the human body that fit in the patient aperture | Parts of the human body that fit in the patient aperture | Same as Predicate |
| Where used | Hospital | Hospital | Same as Predicate |
| Energy used and/or delivered | Detects distribution of radioactivity after injection of a positron emitting radiopharmaceutical. No energy delivered. | Detects distribution of radioactivity after injection of a positron emitting radiopharmaceutical. No Energy Delivered | Same as Predicate |
| Human factors | PET detection system in a movable cart. Its gantry can move up to allow brain imaging while the patient is seated. It can also move down to image the breast without compression, while a patient is lying on a biopsy table or rotate and | The PET detection system in a movable gantry and includes variable patient chair. Gantry and patient chair can be moved automatically to make various scan position. Gantry moves up to allow brain imaging while the patient is | Both systems can acquire images of brain, breast, hand and leg in lying and seated position. |
| | in the patient's head and brain. The patient is lying on a biopsy table or rotate and | in the patient's head and brain. The patient is lying on a biopsy table or rotate and | |
| | in the patient's head and brain. The patient is lying on a biopsy table or rotate and | in the patient's head and brain. The patient is lying on a biopsy table or rotate and | |
{7}
| Characteristic | Predicate Device | Subject Device | Comparison |
| --- | --- | --- | --- |
| | allow imaging the breast, hand, or leg in a seated position. | seated. And it can also move down to image the brain and breast in lying position and allow imaging the leg and knee in a seated position. | Main difference is that the variable patient chair is part of the system in PHAROS. The basic safety of the patient chair has been verified through the IEC60601-1 test, and the mechanical movement of the chair does not affect the image performance and effectiveness. |
| Design | The BBX-PET Scanner is comprised of two parts; the Gantry containing detectors and electronics, and the Universal Console that the computer contains workstation. These two parts are connected to each other using optical fiber and a USB cable. | PHAROS is comprised of three parts; the Main body containing detectors and electronics, variable gantry, patient chair; and the control box for communication between operator and patient and device control; workstation including software and performing data acquisition and reconstruction. Main body and workstation are connected though a dedicated optical cable. Main body, control box and workstation are also connected though a dedicated ethernet network. Workstation is connected to hospital’s network with DICOM Worklist support and PACS connectivity. | Similar.
PHAROS includes an additional control box for user convenience and the predicate device uses an extra USB cable.
These differences do not affect the indications for use, safety or effectiveness. |
| Performance Data (Specifications) | Spatial resolution in FWHM at the center: 2.2mm
Spatial resolution in FWHM at 10 cm: Not published
Transverse resolution: 2.2mm
Axial resolution: 2.2mm
Energy resolution: Not Published
System sensitivity: 1.1%
Coincidence timing window: Not published
Scatter fraction: Not published
Scatter correction method: Not published
Slice thickness: Depends on reconstruction (2mm/4mm)
Count rate sensitivity: 10 cps/kBq | Spatial resolution in FWHM at the center: <2.3 mm
Spatial resolution in FWHM at 10 cm: <3.5 mm
Transverse resolution: <2.3 mm
Axial resolution: <2.3 mm
Energy resolution: < 18%
System sensitivity: > 4% (B480D), > 5.5% (B720D), > 6.5% (B960D)
Coincidence timing window: 3.2 ns
Scatter fraction: < 35%
Scatter correction method: tail-fitting / Single scatter simulation
Slice thickness: Depends on reconstruction (1 mm / 2 mm)
Count rate sensitivity: > 3 cps/kBq (B480D), > 7 cps/kBq (B720D), | Similar.
PHAROS is equipped with Time-of-Flight (TOF) functionality, a feature not present in the predicate device.
This does not alter the indications for use, safety or effectiveness. |
{8}
| Characteristic | Predicate Device | Subject Device | Comparison |
| --- | --- | --- | --- |
| | Isolation of the detector from background: Not applicable for PET Intrinsic spatial resolution: Not applicable for PET Attenuation: Calculated method Time-of-flight: Not applicable Computer: GPU | >10 cps/kBq (B960D) Isolation of the detector from background: Not applicable for PET Intrinsic spatial resolution: Not applicable for PET Attenuation: Calculated method, Template-based method Time-of-flight: <275 ps Computer: GPU | |
| Materials and biocompatibility | Standard electronics and medical grade materials | Standard electronic and medical grade materials | Same as Predicate |
| Compatibility with the environment and other devices | Complies with standard IEC 60601-1-2 ed 4.0 (2014-02) Medical equipment electrical Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances - Requirements and tests for EMC. | Complies with standard IEC 60601-1-2:2014/A1:2020. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility Requirements and tests for EMC. | Same as Predicate |
| Sterility | The product is not sterile and has not to be sterilized by the user. Cleaning procedure devices. Cleaning of standard medical devices | The product is not sterile and has not to be sterilized by the user. Cleaning procedure devices. Cleaning of standard medical devices | Same as Predicate |
| Mechanical and electrical safety | Complies with standard IEC (Third 60601-:2005 Edition) + COOR.1:2006 + CORR.2:2007 + A1:2012 Medical equipment electrical Part 1: General requirements for basic safety and essential performance for electrical safety | Complies with standard ANSI/AAMI ES60601-1:2005/A2:2021. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Same as Predicate |
| Clinical effectiveness | 3 clinical images are provided from President BBX-PET to demonstrate the image capability and the fulfillment with its predetermined specification. | 5 clinical images are provided from PHAROS to demonstrate the image capability and the fulfillment with its predetermined specification. | Similar |
{9}
7 Performance Data
7.1 Electrical safety and electromagnetic compatibility (EMC)
The device passed these tests to the standards IEC 60601-1 and IEC 60601-1-2.
7.2 Performance Testing – Bench
The device conforms to all applicable portions of "Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems", document issued on: December 3, 1998.
The device was evaluated in accordance with the NEMA NU2:2018 and NEMA NU4:2008 standards.
| Item | Pass/Fail Criteria | | | Data Analysis and Result | | | | Result |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Spatial resolution | Spatial resolution
< 2.3 mm @ 1 cm offset | Type | B480D-X | B720D-X | B960D-X | Pass | | |
| | | | | Spatial resolution @ 1 cm | 2.23 mm | | 2.21 mm | 2.09 mm |
| | | | | Spatial resolution @ 10 cm | 3.34 mm | | 3.23 mm | 3.32 mm |
| Scatter fraction @ peak NECR | Scatter fraction
< 35% for all type | B480D-X | B720D-X | B960D-X | | Pass | | |
| | | | | 25.93% | | | 26.43% | 27.12% |
| | | | | | | | | |
| Peak NECR | Peak NECR | | | B480D-X | B720D-X | B960D-X | | Pass |
| | B480D-X | B720D-X | B960D-X | 33.9 kcps | 71.1 kcps | 109.9 kcps | | |
| | > 30 kcps | > 60 kcps | > 90 kcps | | | | | |
| Sensitivity | | | | | | | | Pass |
| | B480D-X | B720D-X | B960D-X | B480D-X | B720D-X | B960D-X | | |
| | > 3 cps/kBq | > 7 cps/kBq | > 10 cps/kBq | 3.46 cps/kBq | 7.61 cps/kBq | 13.3 cps/kBq | | |
| Energy resolution | Energy resolution < 18% | B480D-X | B720D-X | B960D-X | | Pass | | |
| | | | | 13.2% | | | 13.8% | 13.4% |
| Time resolution | Time resolution < 275 ps | B480D-X | B720D-X | B960D-X | | Pass | | |
| | | | | 249 ps | | | 245 ps | 247 ps |
6 / 7
{10}
7 / 7
## 7.3 Clinical Evaluation Summary & Conclusions
To evaluate the clinical performance of PHAROS, a total of five images were obtained, including those from both patients and a normal control group. The images were then assessed by a nuclear medicine physician for clinical acceptability.
## 7.4 Sterilization and Shelf Life
Sterilization does not apply to this device as the device is non-sterile when used. Shelf life is not applicable to this device; however, instructions for cleaning and disinfection are listed in the operator manual.
## 7.5 Biocompatibility
The patient-contacting components are evaluated according to ISO 10993-1.
## 7.6 Software Verification and Validation Testing
Software Verification and Validation was completed according to IEC 62304.
The Risk Management report was completed according to ISO 14971.
## 8 Substantial Equivalence Summary
The similarities and differences between the devices are summarized below:
- The subject and predicate devices are PET nuclear medical scanning systems which employ similar design, construction, materials, energy source, operating principles, and technology.
- Slight differences in performance originate from design decisions including detector configuration.
- The subject device includes a patient chair. The basic safety of the patient chair has been verified.
- The subject and predicate have some of the same performance tests. The subject device passed all performance tests listed.
As shown in the comparison table above:
- Both devices have the same classification and product code.
- Both devices have similar indications for use and similar intended use.
- Both devices have similar technological characteristics.
- Both devices have similar performance.
The comparison analysis demonstrates the subject device is at least as safe and effective as the legally marketed predicate. Only minor differences were noted of the subject and predicate devices due to the shape, size and design; none raise new issues of safety or effectiveness.
Performance test data support the safety of the device and the hardware and software verification and validation demonstrate that the PHAROS should perform as intended in the specified use conditions. The sample images from five clinical cases supported the clinical effectiveness of the PHAROS. Based upon performance data, PHAROS is substantially equivalent to the predicate device.
## 9 Conclusion
The performance data provided demonstrate that PHAROS performs comparably to the predicate device and has a safety and effectiveness profile that is similar to the predicate device. Brightonix believes PHAROS is substantially equivalent to the legally marketed BBX-PET Scanner.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.