Arm Blood Pressure Monitor (ARM-30H); Arm Blood Pressure Monitor (ARM-30J); Arm Blood Pressure Monitor (ARM-30K); Arm Blood Pressure Monitor (ARM-90B)
K250116 · Shenzhen AOJ Medical Technology Co., Ltd. · DXN · Apr 24, 2025 · Cardiovascular
Device Facts
| Record ID | K250116 |
| Device Name | Arm Blood Pressure Monitor (ARM-30H); Arm Blood Pressure Monitor (ARM-30J); Arm Blood Pressure Monitor (ARM-30K); Arm Blood Pressure Monitor (ARM-90B) |
| Applicant | Shenzhen AOJ Medical Technology Co., Ltd. |
| Product Code | DXN · Cardiovascular |
| Decision Date | Apr 24, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1130 |
| Device Class | Class 2 |
Intended Use
The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.
Device Story
Battery-powered automatic non-invasive blood pressure monitor; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Input: pressure oscillations from inflatable upper-arm cuff (22-42 cm). Operation: device automatically inflates/deflates cuff; processes pressure signals to calculate BP and pulse; displays results (mmHg or Kpa) on integrated screen. Used in medical facilities or home settings by patients or clinicians. ARM-90B model includes 4G network connectivity for data transmission. Output: visual display of BP, pulse, date, time, and WHO classification bar. Benefits: enables convenient, routine BP monitoring for hypertension management.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018/AMD1:2020. Two studies conducted: ARM-90B (n=100, ages 19-77) and ARM-30H (n=100, ages 18-80). Both studies utilized the Same Arm Sequential Method. Results demonstrated mean error and standard deviation of differences for systolic and diastolic pressure within ISO 81060-2 limits. No adverse effects or complications reported.
Technological Characteristics
Battery-powered (3.7V Li-ion) automatic sphygmomanometer. Oscillometric sensing principle. Cuff size 22-42 cm. Connectivity: 4G (ARM-90B) or standalone. Biocompatibility per ISO 10993-1, -5, -10, -23. Electrical safety per IEC 60601-1, IEC 60601-1-11, ISO 80601-2-30. EMC per IEC 60601-1-2. Wireless coexistence per ANSI C63.27 and AAMI TIR69.
Indications for Use
Indicated for adult patients requiring non-invasive measurement of systolic and diastolic blood pressure and pulse rate via oscillometric technique at medical facilities or home. No known contraindications.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
- Arm Blood Pressure Monitor, model AOJ-30B (K222125)
Related Devices
- K193627 — Automatic Upper Arm Blood Pressure Monitor · Dongguan E-Test Technology Co., Ltd. · Jul 12, 2020
- K182495 — Electronic Blood Pressure Monitor · Shenzhen Jumper Medical Equipment Co., Ltd. · Mar 3, 2019
- K191180 — Electronic Blood Pressure Monitor · Shenzhen AOJ Medical Technology Co., Ltd. · Aug 23, 2019
- K150430 — Arm Blood Pressure Monitor, Wrist Blood Pressure Monitor · Fudakang Industrial Co., Ltd. · Apr 17, 2015
- K200716 — Electronic Blood Pressure Monitor · Shenzhen Pango Electronic Co., Ltd. · Dec 4, 2020
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2025
Shenzhen AOJ Medical Technology Co., Ltd
% Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong 518067
China
Re: K250116
Trade/Device Name: Arm Blood Pressure Monitor (ARM-30H); Arm Blood Pressure Monitor (ARM-30J); Arm Blood Pressure Monitor (ARM-30K); Arm Blood Pressure Monitor (ARM-90B)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: January 16, 2025
Received: March 21, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250116 - Jie Yang
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250116 - Jie Yang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250116 | |
| Device Name Arm Blood Pressure Monitor (ARM-30H);Arm Blood Pressure Monitor (ARM-30J); Arm Blood Pressure Monitor (ARM-30K);Arm Blood Pressure Monitor (ARM-90B) | |
| Indications for Use (Describe) The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250116
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025/01/17
## 1. Submission sponsor
Name: Shenzhen AOJ Medical Technology Co., Ltd.
Address: Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126, Shenzhen, China
Contact person: Jack Wang
Title: Deputy Chief
TEL: 86 755-27786026
## 2. Submission correspondent
Name: Chonconn Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
## 3. Subject Device Information
| Trade/Device Name | Arm Blood Pressure Monitor |
| --- | --- |
| Model | models ARM-30H, ARM-30J, ARM-30K and AOJ-90B |
| Common Name | Automatic Blood Pressure Monitor |
| Regulatory Class | Class II |
| Product Code | DXN |
| Submission type | Traditional 510(K) |
## 4. Predicate Device
Manufacturer: Shenzhen AOJ Medical Technology Co., Ltd.
Device name: Arm Blood Pressure Monitor, model AOJ-30B
510(K) Number: K222125
## 5. Device Description
Arm blood pressure monitor, models (models ARM-30H, ARM-30J, ARM-30K and AOJ-90B), are designed as a battery driven automatic no-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.
{5}
K250116
All the models included in this submission follow the similar software, same measurement principle and same specifications. All the models can be used with one cuff size $22\sim 42$ cm (8.6~16.5 inches). AOJ-90B has extra 4G network except for other models.
# 6. Intended use & Indication for use
The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.
7. Comparison to the Predicate Device
| ITEM | Proposed Device ARM-30H, ARM-30J, ARM-30K and AOJ-90B/K250116 | Predicate Device AOJ-30B/K222125 | Comparison Result |
| --- | --- | --- | --- |
| Manufacturer | Shenzhen AOJ Medical Technology Co., Ltd. | Shenzhen AOJ Medical Technology Co., Ltd. | Same |
| Intended Use/Indications for Use | The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | Same |
| Contraindications | Not known | Not Known | Same |
| Application scenario | Medical Facilities and Home Use | Medical Facilities and Home Use | Same |
| Operational Specifications | | | |
| Principle | Oscillometric | Oscillometric | Same |
| Measurement Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same |
| Patient population | Adult | Adult | Same |
| Measurement site | Upper arm | Upper arm | Same |
| Blood pressure measurement range | 30-255 mmHg | 30 - 255 mmHg | Same |
{6}
K250116
| Accuracy | ± 3 mmHg | ± 3 mmHg | Same |
| --- | --- | --- | --- |
| Heart rate measurement range | 40-199 bpm | 40-199 bpm | Same |
| Accuracy | ± 5% | ± 5% | Same |
| Cuff size | 22 - 42 cm | 22 - 42 cm | Same |
| Display | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification
Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification
Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Same |
| Auto shutdown | YES | YES | Same |
| Operating environment | Temperature: 5°C~40°C
Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPa | Temperature: 5°C~40°C
Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPa | Same |
| Storage environment | Ambient Temperature: -20°C to 55°C
Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPa | Ambient Temperature: -20°C to 55°C
Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPa | Same |
| Battery type | Lithium-ion battery, d.c.3.7V | 6Vdc (4*1.5VAAA batteries) | Similar |
| Weight | ARM-90B: About 263g
ARM-30J: About 281g
ARM-30K: About 280g
ARM-30H: About 195g | (262 ±5)g without battery | Similar |
| Dimensions | ARM-90B:130mm×100mm×60mm
ARM-30J: 130mm×100mm×60mm
ARM-30K: 129mm×112mm×58.5mm
ARM-30H: 12 mm×89.8mm×30.8mm | 127mm×93mm×73mm | |
| Patient Contacting | Surface-contacting, Less than 24 h | Surface-contacting, Less than 24 h | Same |
3 / 5
{7}
K250116
| Biocompatibility evaluation | Cytotoxicity, skin sensitization and irritation | Cytotoxicity, skin sensitization and irritation | Same |
| --- | --- | --- | --- |
| Electrical safety | IEC 60601-1
IEC 60601-1-11
ISO 80601-2-30 | IEC 60601-1
IEC 60601-1-11
ISO 80601-2-30 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Biocompatibility | ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-1
ISO 10993-5
ISO 10993-10 | Equivalent |
| Data transmission | ARM-90B:4G
ARM-30H, ARM-30J, ARM-30K:N/A | Bluetooth | Similar |
## 8. Non-clinical Data
### Biocompatibility testing
The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
### Bench testing
The device has been tested according to the following standards:
- IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
- IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 80601-2-30: Medical electrical equipment – Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
- IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- FDA Guidance for Non-Automated Sphygmomanometer.
{8}
K250116
Wireless testing:
- ANSI C63.27: 2017: American National Stand for Evaluation of Wireless Coexistence.
- AAMI TIR69: 2017 Risk management of radio-frequency wireless coexistence for medical devices and systems.
- Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013)
## 9. Clinical data
Two studies were performed on ARM-30H and ARM-90B separately according to ISO 81060-2 Third edition 2018-11 [Including AMD1:2020] Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type [Including: Amendment 1 (2020)]. The clinical validation data on ARM-90B can cover ARM-30K and ARM-30J. The Same Arm Sequential Method was chosen for both studies.
ARM-90B study included 100 adult subjects (61 females, 39 males) with an age range of 19 to 77 years.
ARM-30H study included 100 adult subjects (54 females, 46 males) with an age range of 18 to 80 years
All data’s mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]. No adverse effect and/or complication is found in this study.
## 10. Conclusion
It is concluded from the non-clinical and clinical tests that demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed predicate device identified above.