The HYALEX® MTP Hemiarthroplasty Implant is intended to be implanted to replace the distal metatarsal surface of the great toe for treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus, hallux limitus, or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is intended for single use and for use without bone cement.
Device Story
Implant replaces distal metatarsal surface of great toe; relieves pain from degenerative/post-traumatic arthritis. System includes implant and instrumentation. Implant features HYALEX bearing material (polyurethane/polyacrylic acid) bonded to titanium coating (Ti6Al4V/commercially pure titanium) via proprietary bonding compound (PMMA/UDMA); porous titanium surface facilitates bone ingrowth. Implanted by trained foot and ankle surgeons; uncemented; utilizes interference fit for stability. Supplied sterile in 10mm and 12mm diameters. Benefits include restoration of articular surface while retaining surrounding cartilage/bone and preserving sesamoid grooves.
Clinical Evidence
Bench testing included implant insertion assessment, anatomic fatigue, micromotion, wear, contact pressure, biocompatibility, and MRI compatibility. Design validation performed in cadavers by foot and ankle surgeons. One-year large animal study demonstrated in vivo stability and durability of bearing material.
Technological Characteristics
Implant materials: polyurethane, polyacrylic acid, Ti6Al4V, commercially pure titanium, PMMA, UDMA. Form factor: 10mm and 12mm diameter caps. Fixation: uncemented, interference fit. Sterilization: radiation (SAL 10⁻⁶). Instrumentation: reusable, steam sterilized.
Indications for Use
Indicated for patients with degenerative or post-traumatic arthritis of the first MTP joint with good bone stock, including hallux valgus, hallux limitus, or hallux rigidus, and unstable or painful MTP joint.
Regulatory Classification
Identification
A toe joint phalangeal (hemi-toe) polymer prosthesis is a device made of silicone elastomer intended to be implanted to replace the base of the proximal phalanx of the toe.
BioPoly Great Toe Hemiarthroplasty Implant (K203634)
Submission Summary (Full Text)
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
April 3, 2026
Hyalex Orthopaedics, Inc.
Chris Cain
VP of Clinical and Regulatory Affairs
9 Hayden Ave. Bldg. D, Suite 340
Lexington, Massachusetts 02421
Re: K250074
Trade/Device Name: HYALEX® MTP Hemiarthroplasty Implant
Regulation Number: 21 CFR 888.3730
Regulation Name: Toe joint phalangeal (hemi-toe) polymer prosthesis
Regulatory Class: Class II
Product Code: KWD
Dated: January 2, 2026
Received: January 2, 2026
Dear Chris Cain:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250074 - Chris Cain
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250074 - Chris Cain
Page 3
Sincerely,
# STANLEY GOLISH-S
for RDML Raquel Peat, PhD, MPH, USPHS
Director
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K250074
Device Name
HYALEX® MTP Hemiarthroplasty Implant
Indications for Use (Describe)
The HYALEX® MTP Hemiarthroplasty Implant is intended to be implanted to replace the distal metatarsal surface of the great toe for treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus, hallux limitus, or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is intended for single use and for use without bone cement.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K250074, Page 1 of 3
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
# I. Submitter Information
Manufacturer: Hyalex Orthopaedics, Inc.
99 Hayden Ave., Bldg. D, Ste. 340
Lexington, MA 02421 USA
Phone: (781) 778-3232
Contact: Chris Cain
Hyalex Orthopaedics, Inc.
Phone: (781) 778-3232
chris.cain@hyalex.com
Date Prepared: 3 April 2026
# II. Name of Device
Device Trade Name: HYALEX® MTP Hemiarthroplasty Implant
Common Name: Toe Prosthesis
Regulation Name: Toe Joint Phalangeal (Hemi-Toe) Polymer Prosthesis
Regulation Number: 21 CFR 888.3730
Product Code: KWD
Device Class: Class II
Panel Identification: Orthopedics
# III. Predicate Devices
| Predicate Type | Device Name | Manufacturer | 510(k) No. |
| --- | --- | --- | --- |
| Predicate Device (Primary) | HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant | ArthroSurface, Inc. | K131377 |
| Reference Device | BioPoly Great Toe Hemiarthroplasty Implant | BioPoly, LLC | K203634 |
Page 1 of 3
{5}
K250074, Page 2 of 3
#### IV. Device Description
The system is composed of the HYALEX Implant and instrumentation. The implant restores the damaged portion of the metatarsal head to relieve pain from degenerative or post-traumatic arthritis.
The implant consists of the HYALEX bearing material (polyurethane and polyacrylic acid) bonded with a titanium coating (Ti6Al4V and commercially pure titanium) using HBC (PMMA and UDMA), a proprietary bonding compound. The device has porous titanium (commercially pure titanium) to allow bone ingrowth. The implant is supplied sterile in two sizes with a cap diameter of 10 and 12mm.
The implant is intended for single use only and should not be reused or re-sterilized. It should only be implanted by a trained and qualified foot and ankle surgeon.
#### V. Intended Use
The HYALEX MTP Hemiarthroplasty Implant is intended for use in the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock.
#### VI. Indications for Use
The HYALEX® MTP Hemiarthroplasty Implant is intended to be implanted to replace the distal metatarsal surface of the great toe for treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus, hallux limitus, or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is intended for single use and for use without bone cement.
#### VII. Comparison of Technological Characteristics with the Predicate Device
The subject and predicate device have the same intended use and indications for use. Both devices replace the articular surface of the metatarsal head while retaining surrounding cartilage and bone and preserving the sesamoid grooves. Both implants utilize a similar surgical technique, are uncemented and utilize an interference fit with the bone for implant stability. The articular surface of the HYALEX MTP Hemiarthroplasty Implant is made from polyurethane and polyacrylic acid while the predicate device has a CoCr surface. The testing described in the prior section has shown that the known questions of safety and effectiveness concerning wear, durability and biocompatibility of the articular surface of the HYALEX MTP Hemiarthroplasty Implant performs as required per recognized acceptance criteria. The HYALEX MTP Hemiarthroplasty Implant is substantially equivalent to the predicate device.
Page 2 of 3
{6}
K250074, Page 3 of 3
### VIII. Performance Testing Summary
Performance testing was conducted on the HYALEX MTP Hemiarthroplasty Implant, including implant insertion assessment, anatomic fatigue, micromotion assessment, implant wear assessment, contact pressure assessment, biocompatibility and MRI compatibility. Design validation testing was conducted in cadavers by practicing foot and ankle surgeons.
A one-year large animal study was also conducted to illustrate the stability and durability of the bearing material in vivo. Results from this study provided evidence of the safety of the implant material.
The HYALEX MTP Hemiarthroplasty Implant is sterilized by radiation to a Sterility Assurance Level (SAL) of 10⁻⁶ in accordance with a validated process. The instrumentation is reusable and is cleaned and sterilized between uses according to a validated cleaning and steam sterilization process.
### IX. Substantial Equivalence Summary
The HYALEX MTP Hemiarthroplasty Implant is substantially equivalent to the primary predicate device, HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant, on the basis of the following:
- The subject device has the SAME intended use as the predicate device; AND
- The subject device is as safe and effective as the predicate device; AND
- The subject device has some differences in technological characteristics compared to the predicate, but those differences do not raise different questions of safety and effectiveness.
The data presented in this 510(k) supports these assertions. Therefore, the HYALEX MTP Hemiarthroplasty Implant is substantially equivalent to the predicate device and meets the requirements for 510(k) clearance.
### X. Conclusion
The HYALEX MTP Hemiarthroplasty Implant is substantially equivalent to the predicate devices; and does not raise any different questions of safety or effectiveness. To address the possibility of device dissociation, a 522 Order for a Post Market Surveillance Study will be issued for this device.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.