HEMICAP MTP RESURFACING HEMI-ARTHROPLASTY IMPLANT

K131377 · Arthrosurface, Inc. · KWD · Nov 19, 2013 · Orthopedic

Device Facts

Record IDK131377
Device NameHEMICAP MTP RESURFACING HEMI-ARTHROPLASTY IMPLANT
ApplicantArthrosurface, Inc.
Product CodeKWD · Orthopedic
Decision DateNov 19, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3730
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K131377 · Nov 19, 2013HEMICAP MTP RESURFACING HEMI-ARTHROPLASTY IMPLANTArthrosurface, Inc.Retrospective clinical patient dataClinical data from over 100 patients with up to 5 years of follow-up was used to demonstrate the safety and effectiveness of the device and support substantial equivalence to predicate devices.Retrospective clinical data; Long-term follow-up; Substantial equivalence

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective clinical cohort; Follow-up/Duration: Up to 5 yearsPatients with degenerative and post-traumatic arthritis in the metatarsal joint; Sample Size: >100 patientsNot applicable for this studySafety and effectiveness

Indications for Use

Hemi-Arthroplasty Implant for the metatarsophalangeal joint for use in the treatment of patients with degenerative and post-traumatic arthritis in the metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is a single use implant intended to be used with bone cement or without bone cement.

Device Story

HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant is a single-use orthopedic prosthesis for the metatarsophalangeal (MTP) joint. It consists of an articular resurfacing component and a cancellous taper post component that mate via a taper interlock to provide stable, immobile fixation at the bone/prosthetic interface. The device is implanted by a surgeon to replace damaged articular surfaces in patients with arthritis or joint instability. It can be used with or without bone cement. The implant provides stress-bearing contact to restore joint function and reduce pain. It is designed for patients with sufficient bone stock.

Clinical Evidence

Clinical data provided for over 100 patients with follow-up ranging up to 5 years. Results demonstrated safety and effectiveness of the implant.

Technological Characteristics

Articular resurfacing component: Cobalt-Chromium alloy (Co-Cr-Mo) with Titanium (CP Ti) surface coating. Taper post component: Titanium alloy (Ti-6Al-4V). Fixation via taper interlock. Single-use. Anatomically curved design.

Indications for Use

Indicated for patients with degenerative or post-traumatic arthritis of the metatarsophalangeal (MTP) joint, hallux valgus, or hallux rigidus, presenting with good bone stock and an unstable or painful MTP joint.

Regulatory Classification

Identification

A toe joint phalangeal (hemi-toe) polymer prosthesis is a device made of silicone elastomer intended to be implanted to replace the base of the proximal phalanx of the toe.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # NOV 1 9 2013 #### Section 5 510(k) Summary | 510(k) Owner: | Arthrosurface, Inc.<br>28 Forge Parkway<br>Franklin, MA 02038<br>Tel: 508.520.3003<br>Fax: 508.528.4604 | |------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Dawn Wilson<br>VP, Quality & Regulatory | | Date of Preparation: | May 13, 2013 | | Trade Name: | HemiCAP® MTP Resurfacing Hemi-Arthroplasty<br>Implant | | Common Name: | MTP hemi-toe prosthesis | | Device:<br>Regulation Number:<br>Regulation Name:<br>Device Class:<br>Review Panel:<br>Product Code: | Prosthesis, Toe, Hemi-, Phalangeal<br>21 CFR 888.3730<br>Toe joint phalangeal (hemi-toe) polymer prosthes<br>Class II<br>Orthopedic<br>KWD | ### Device Intended Use Hemi-Arthroplasty Implant for the metatarsophalangeal joint for use in the treatment of patients with degenerative and post-traumatic arthritis in the metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is a single use implant intended to be used with bone cement or without bone cement. ### Device Description The HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant incorporates an articular resurfacing component and a cancellous taper post component that mate together via a taper interlock to provide stable and immobile fixation of the implant and stress bearing contact at the bone/prosthetic interface. {1}------------------------------------------------ ## Materials Articular Resurfacing Component: Cobalt-Chromium Alloy (Co-Cr-Mo) Surface Coating: Titanium (CP Ti) Taper Post Component: Titanium Alloy (Ti-6Al-4V) The HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant is a single use implant with updated indications for use with or without bone cement to be consistent with predicates, and updated general terminology for MTP. The implant is otherwise identical to the sponsor's previously cleared and commercially marketed implant. Referencing predicate devices, similar indications have been cleared since 2007. Such history of use demonstrates the differences in indications are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, nor do they affect the safety and effectiveness of the device when used as labeled. ## Substantial Equivalency: The intended use, materials, design features and application of the Proposed Device are substantially equivalent to the following previously cleared and commercially marketed devices: | • CAP™ Great Toe Resurfacing Hemi-Arthroplasty Implant | K031859 | |---------------------------------------------------------|---------| | • Solana Surgical LLC, Metatarsal Decompression Implant | K113752 | - Vilex Cannulated Metallic Hemi Implant . K102401 The fundamental scientific technology of the proposed device has not changed relative to the predicate devices: - Intended for the metatarsophalangeal joint . - Similar indications for use . - . Similar device designs - o Same Cobalt and Titanium implant materials - Implants available anatomically curved . - Similar fixation options . Clinical data was provided in support of the substantial equivalency. Data was obtained from over 100 patients with follow-up ranging out to 5 years. Results demonstrated the safety and effectiveness of the HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant is adequately supported by the substantial equivalence information provided within this Premarket Notification. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a bird or abstract human figure, composed of three curved lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 19, 2013 Arthrosurface, Incorporated Ms. Dawn J. Wilson Vice President of Quality and Regulatory 28 Forge Parkway Franklin, Massachusetts 02038 Re: K131377 Trade/Device Name: HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant Regulation Number: 21 CFR 888.3730 Regulation Name: Toe joint phalangeal (hemi-toe) polymer prosthesis Regulatory Class: Class II Product Code: KWD Dated: November 5, 2013 Received: November 6, 2013 Dear Ms. Wilson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - Ms. Dawn J. Wilson forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Lori A. Wigqins for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Section 4 Indications for Use Statement 510(k) Number (if known): K131377 Device Name: HemiCAP® MTP Resurfacing Hemi-Arthroplasty Implant #### Indications for Use: Hemi-Arthroplasty Implant for the metatarsophalangeal joint for use in the treatment of patients with degenerative and post-traumatic arthritis in the metatarsal joint in the presence of good bone stock along with the following clinical conditions: hallux valgus or hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The device is a single use implant intended to be used with bone cement or without bone cement. Prescription Use _____________________________________________________________________________________________________________________________________________________________ ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ΠΠ Casey L-Hanley, Ph.D. ١٦٢٦ 10/11/ Division of Orthopedic Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...