CEC101 and EE0101: Microcurrent Facial Device is intended for facial stimulation for over-the-counter aesthetic use. EEI101: Microcurrent Facial Device is intended for facial, neck and body skin stimulation for over-the-counter aesthetic use.
Device Story
Portable, reusable microcurrent facial device; models CEC101, EE0101, EEI101. Input: internal rechargeable lithium-ion battery power. Operation: device delivers bi-phase square-wave electrical pulses to skin via electrodes; controlled by internal microprocessor; user selects intensity levels and modes. Output: transcutaneous electrical stimulation for aesthetic purposes. Used in home healthcare environment by consumers. Healthcare provider does not operate device; output is intended for aesthetic benefit. Device includes automatic shut-off and user override controls. Benefits: aesthetic skin stimulation.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-11, -2-10), EMC (IEC 60601-1-2), and battery safety (IEC 62133-2). Software verification and validation confirmed all specifications met and hazards mitigated.
Indicated for over-the-counter aesthetic facial stimulation (CEC101, EE0101) or facial, neck, and body skin stimulation (EEI101).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 10, 2025
Shenzhen Dachi communication Co., Ltd.
% Jilan Luo
RA Specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center,
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K244004
Trade/Device Name: Microcurrent Facial Device (CEC101, EE0101, EEI101)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: June 20, 2025
Received: June 20, 2025
Dear Jilan Luo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K244004 - Jilan Luo
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K244004 - Jilan Luo
Page 3
Sincerely,
Tushar Bansal -S
- Tushar Bansal, PhD
- Acting Assistant Director, Acute Injury Devices Team
- DHT5B: Division of Neuromodulation and Physical Medicine Devices
- OHT5: Office of Neurological and Physical Medicine Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K244004
Device Name
Microcurrent Facial Device (CEC101, EE0101, EEI101)
Indications for Use (Describe)
☐ CEC101 and EE0101:
Microcurrent Facial Device is intended for facial stimulation for over-the-counter aesthetic use.
EEI101:
Microcurrent Facial Device is intended for facial, neck and body skin stimulation for over-the-counter aesthetic use.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
# 510(k) Summary of K244004
"510(k) Summary" as required by 21 CFR Part 807.92.
Date: 2025-07-10
## I. Submitter
Shenzhen Dachi communication Co., Ltd.
Room 201, 202, 203, Building 3, Phase I, Huangtian Yangbei Industrial Zone, Huangtian Community, Hangcheng Street, Bao'an District, Shenzhen, China.
Post code: 518101
Tel.: +86 755 83142102
Xiao Chunhua
Title: SCM Manager
Tel.: +86 130 6873 9296
Email: pod@dachitel.com
## II. Device
Name of Device: Microcurrent Facial Device
Model(s): CEC101, EE0101, EEI101
Common or Usual Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes
Classification Name: Transcutaneous electrical nerve stimulator for pain relief
Regulatory Class: II
Product Code: NFO
Regulation Number: 21 CFR 882.5890
## III. Predicate Device
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| BELEGA Co., Ltd. | BEAGANK 4T PLUS | K233010 | November 21, 2023 |
| Bio-Medical Research Ltd. | BMR Face | K103031 | NOV 10, 2011 |
## IV. Device Description
The Microcurrent Facial Device is a portable, non-sterile, reusable device designed to achieve the aesthetic effect. It consists of main unit, charging cable, membrane cloth, conductive gel and user
Page 3 of 11
{5}
Shenzhen Dachi communication Co.,Ltd.
510(k)s -510(k) Summary
manual. The device is supplied by internal rechargeable lithium battery, which can be recharged by external charger through the Type-C charging cable. The device is un-usable when charging.
## V. Indications for Use
CEC101 and EE0101: Microcurrent Facial Device is intended for facial stimulation for over-the-counter aesthetic use.
EEI101: Microcurrent Facial Device is intended for facial, neck and body skin stimulation for over-the-counter aesthetic use.
## VI. Comparison of Technological Characteristics With the Predicate Device
Microcurrent Facial Device is compared with the following Predicate Devices in terms of intended use, design, specifications, materials and performance:
Page 4 of 11
{6}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K244004 | K233010 | K103031 | / |
| Device Name and Model | Microcurrent Facial Device
Models: CEC101, EE0101, EEI101 | BEAGANK 4T PLUS | BMR Face | / |
| Regulation number | 882.5890 | 882.5890 | 882.5890 | Same |
| Classification name | Transcutaneous Electrical Nerve Stimulator for Pain Relief | Transcutaneous Electrical Nerve Stimulator for Pain Relief | Transcutaneous Electrical Nerve Stimulator for Pain Relief | |
| Product code | NFO | NFO | NFO | Same |
| Device classification | Class II | Class II | Class II | Same |
| OTC or prescription | OTC | OTC | OTC | Same |
| Intended Use | CEC101 and EE0101:
Microcurrent Facial Device is intended for facial stimulation for over-the-counter aesthetic use.
EEI101:
Microcurrent Facial Device is intended for facial, neck and body skin stimulation for over-the-counter aesthetic use. | The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation. | BMR Face is intended for facial stimulation and is indicated for Over the Counter Cosmetic Use. | Same |
| Method of Line Current Isolation | Type BF Applied Part | Type BF Applied Part | Not publicly available | Same |
Page 3 of 11
{7}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | | K244004 | K233010 | K103031 | / |
| Patient Leakage Current | NC | DC: < 1μA
AC: < 1μA | DC: < 1μA
AC: < 1μA | Not publicly available | Same |
| | SFC | DC: < 1μA
AC: < 1μA | DC: < 1μA
AC: < 1μA | Not publicly available | |
| Average DC current through electrodes when device is on but no pulses are being applied | | < 1μA | < 1μA | Not publicly available | Same |
| Number of Output Channels | | 1 output channel | 1 output channel | 2 channels | Same |
| Number of Output Modes | | CEC101: 2 modes
EE0101: 2 modes
EEI101: 3 modes | 4 modes | 3 modes | Similar
**Note 1** |
| Output Intensity Level | | CEC101: 10 intensity levels
EE0101: 4 intensity levels
EEI101: 4 intensity levels | Not publicly available | 0.0 - 100.0 scale | Similar
**Note 1** |
| Synchronous or Alternating? | | NA-1 output channel | Not publicly available | Not publicly available | Same |
Page 4 of 11
{8}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | | K244004 | K233010 | K103031 | / |
| Method of Channel Isolation | | N/A-1 output channel | N/A | Not publicly available | Same |
| Regulated Current or Regulated Voltage? | | Voltage Control | Voltage Control | Not publicly available | Same |
| Software/Firmware/Microprocessor Control? | | Yes | Yes | Not publicly available | Same |
| Automatic Overload Trip | | No | Not required due to circuit design | Not publicly available | Same |
| Automatic No-Load Trip | | No | Not required due to circuit design | Not publicly available | Same |
| Automatic Shut Off | | Yes | Yes | Yes | Same |
| User Override Control | | Yes | Yes | Yes | Same |
| Indicator Display | On/Off Status | Yes | Not publicly available | Yes | Same |
| | Low Battery | Yes | Yes | Yes | Same |
Page 5 of 11
{9}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | | K244004 | K233010 | K103031 | / |
| | Voltage/Current Level | Yes | Not publicly available | Not publicly available | Same |
| Dimensions | | CEC101:193.68*68*60mm
EE0101: 90*65*39mm
EEI101: 202.6*69.32*62mm | Not publicly available | 6.0 x 8.0 x 2.1 (cm) | Different
**Note 2** |
| Weight | | CEC101: 266g
EE0101: 120g
EEI101: 230g | 136g | 63g | Different
**Note 2** |
| Treatment area | | EEI101: Facial, neck and body skin
CEC101: Facial skin
EE0101: Facial skin | Face, neck and body | Facial skin | Same |
| Software control | | YES | YES | Yes | Same |
| Handheld | | YES | YES | Wearable | Same |
| Power source | | Input:5V==1A; Output:3.7V
CEC101: 3.7V/2600mAh
EE0101: 3.7V/800mAh
EEI101: 3.7V/2600mAh | Internal rechargeable Lithium-ion battery | 3.6V rechargeable batteries | Similar
**Note 3** |
Page 6 of 11
{10}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K244004 | K233010 | K103031 | / |
| Waveform Type | Bi-phase square-wave pulse | Rectangle, biphasic asymmetric | biphasic symmetrical | Same |
| Maximum Output Voltage | CEC101:
3.77 V±10% @ 500 Ω
11.85 V±10% @ 2K Ω
33.2 V±10% @ 10K Ω
EE0101:
3.57 V±10%@ 500 Ω
11.25 V±10% @ 2K Ω
31.25 V±10% @ 10K Ω
EEI101:
3.54 V±10%@ 500 Ω
12.3 V±10% @ 2K Ω
34 V±10% @ 10K Ω | Mode 1:
0.12V@ 500 Ω
0.51V@ 2K Ω
1.83V@ 10K Ω
Mode 2:
0.11V@ 500 Ω
0.50V@ 2K Ω
1.87V@ 10K Ω
Mode 3:
16.6V@ 500 Ω
21.0V@ 2K Ω
22.5V@ 10K Ω
Mode 4:
15.3V@ 500 Ω
21.2V@ 2K Ω
22.9V@ 10K Ω | 15 V±10% @ 500 Ω
60 V±10% @ 2K Ω
32 V±10% @ 10K Ω | Similar Note 4 |
| Maximum Output Current | CEC101:
7.54 mA±10% @ 500 Ω
5.93 mA±10% @ 2K Ω
3.32 mA±10% @ 10K Ω
EE0101:
7.14 mA±10%@ 500 Ω
5.63 mA±10%@ 2K Ω | Mode 1:
0.19mA@ 500 Ω
0.18mA@ 2K Ω
0.16mA@ 10K Ω
Mode 2:
0.45mA@ 500 Ω
0.39mA@ 2K Ω | 30 mA±10%@ 500 Ω
30 mA±10%@ 2K Ω
3.2 mA±10%@ 10K Ω | |
Page 7 of 11
{11}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K244004 | K233010 | K103031 | / |
| | 3.13 mA±10%@ 10K Ω
EEI101:
7.08 mA±10%@ 500 Ω
6.15 mA±10%@ 2K Ω
3.4 mA±10%@ 10K Ω | 0.24mA@ 10K Ω
Mode 3:
30.8mA@ 500 Ω
8.52mA@ 2K Ω
2.26mA@ 10K Ω
Mode 4:
31.4mA@ 500 Ω
8.60mA@ 2K Ω
2.15mA@ 10K Ω | | |
| Pulse Width | CEC101: 174μs
EE0101: 174μs
EEI101: 145μs | Mode 1: 265μs
Mode 2: 265μs
Mode 3: 274μs
Mode 4: 610μs | 160-200μs | Similar
**Note 4** |
| Output Frequency | 40Hz, 97.8Hz | Mode 1: 3.80kHz
Mode 2: 3.80kHz
Mode 3: 3.65kHz
Mode 4: 1.64kHz | 70-80Hz | Similar
**Note 4** |
| Net charge | 0 μC @ 500Ω | Mode 1: 0.025μC @ 500Ω
Mode 2: 0.025μC @ 500Ω
Mode 3: -1.3μC @ 500Ω
Mode 4: 0 μC @ 500Ω | 0 μC @ 500Ω | Same |
| Maximum Current Density | CEC101: 0.09mA/cm²@500Ω
EE0101: 0.45mA/cm²@500Ω
EEI101: 0.15mA/cm²@500Ω | Mode 1: 0.15~1.36 mA/cm²@500Ω
Mode 2: 0.10~3.21 mA/cm²@500Ω
Mode 3: 4.26~220 mA/cm²@500Ω | 1mA/cm²@500Ω | Similar
**Note 4** |
Page 8 of 11
{12}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K244004 | K233010 | K103031 | / |
| | | Mode 4: 3.13~224.3 mA/cm²@500Ω | | |
| Maximum Power Density | CEC101: 0.045mW/cm²@500Ω
EE0101: 0.21mW/cm²@500Ω
EEI101: 0.065mW/cm²@500Ω | Mode 1: 8.51~11.39 μW/cm²@500Ω
Mode 2: 17.46~33.87 μW/cm²@500Ω
Mode 3: 1.124~106.6 mW/cm²@500Ω
Mode 4: 0.56~51.78 mW/cm²@500Ω | 3.91W/cm²@500Ω | |
| Treatment frequency | CEC101: 3-8 minutes
EE0101: 5-8 minutes
EEI101: 5-8 minutes | 5 minutes | 5/10/20 minutes | Similar |
| Main Materials | ABS, Stainless steel | ABS | Plastic case built of abs, electrodes of skin conductive through adhesive hydrogel layer. | Same |
| Biocompatibility | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10 | Same |
| Safety and EMC | IEC 60601-1
IEC 60601-2-10
IEC 60601-1-11
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-2-10
IEC 60601-1-2 | Same |
Comparison in Detail(s):
{13}
Shenzhen Dachi communication Co., Ltd.
510(k)s -510(k) Summary
## Note 1:
Though the number of Output Modes and Output Intensity Level are little different from the predicate devices, this difference is insignificant and do not raise any safety or effectiveness problems. Also, we have tested the product waveform parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not lead to safety and effectiveness problems.
## Note 2:
Though the dimension and weight are little different from the predicate devices, this difference is insignificant and do not raise any safety or effectiveness problems.
## Note 3:
Although the "Power source" is slightly different from the predicate devices, they are both powered by the internal lithium battery. The lithium battery of the subject device complies with the IEC 62133-2 standard, and the device complies with IEC 60601-1 and IEC 60601-1-2 requirements, so this difference will not raise any safety or effectiveness issue.
## Note 4:
Although the "Maximum Output Voltage", "Maximum Output Current", "Maximum Current Density(r.m.s)", and "Maximum Power Density" of subject device are little different from the predicate devices, but their maximum and minimum are all in the range of predicate devices; "Pulse Width" and "Output frequency" of subject device are similar with the predicate device 1, we also have tested the product waveform parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not raise safety and effectiveness problems.
Page 10 of 11
{14}
Shenzhen Dachi communication Co.,Ltd.
510(k)s -510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recognized by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation
### 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
- IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems.
- IEC 60601-2-10 Medical electrical equipment – Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
### 3) Software Verification and Validation
Software documentation consistent with **Basic Documentation** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.
## VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design and performance, it can be concluded that the Microcurrent Facial Device is as safe, as effective and performs as well as the legally marketed predicate devices.
Page 11 of 11
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.