BMR Face is intended for facial stimulation and is indicated for over the counter cosmetic use
Device Story
BMR Face (Type 371/372) is a two-channel, battery-operated facial stimulator for cosmetic use. Device consists of a rechargeable electronic control unit connected to an applicator; applicator paddles feature replaceable twin conductive gel pads. Device delivers TENS-based electrical impulses to facial area, specifically positioned anterior to ears and over cranial nerve #7. User selects from 3 programs with treatment times ranging from 10 to 20 minutes. Power provided by internal 3.6V NiMH rechargeable battery; unit cannot be charged during treatment. Internal connections are over-molded to prevent moisture ingress. Device is intended for home use by consumers. Output is controlled by user via the unit interface. Stimulation is intended for cosmetic purposes.
Clinical Evidence
No clinical studies submitted. Substantial equivalence supported by bench testing, including electrical safety, electromagnetic compatibility (IEC 60601-1-2), and biocompatibility testing of gel pads (ISO 10993-5, ISO 10993-10).
Indicated for over-the-counter cosmetic facial stimulation in healthy adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
K230506 — PureLift Pro Edition · Xtreem Pulse · Jun 21, 2023
K142794 — Everyway Facial MENS · Everyway Medical Instruments Co.,Ltd · Jun 2, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 1 0 2011
K10303/
Image /page/0/Picture/2 description: The image shows the logo for "bmr bio-medical research ltd.". The letters "bmr" are in a bold, sans-serif font, with a curved line running through the letters. Below the letters, the words "bio-medical research ltd." are written in a smaller, sans-serif font.
# Bio-Medical Research Ltd.
· Parkmore Business Park West, Galway, Ireland Tel: +353 (0)91 774300 - Fax: +353 (0)91 774301
This 510k Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
# 1. Contact Details
| Name: | Anne-Marie Keenan |
|------------|------------------------------|
| Address: | Bio-Medical Research Ltd., |
| | Parkmore Business Park, West |
| | Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774302 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | November 8, 2011 |
#### 2. Device Name
| Trade Name of Device: | BMR Face, Type 371 & Type 372* |
|-------------------------|-------------------------------------------------------------|
| Common Name: | Facial Stimulator |
| Regulation Number: | 21 CFR 882.5890 |
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief. |
| Product Code: | NFO |
| Device Class: | 2 |
#### 3. Identification of Equivalent Legally Marketed Device
| 510(k) Number: | K011935 |
|----------------|---------------------|
| Manufacturer: | Salton, Inc. |
| Trade Name: | Rejuvenique® System |
| Product Codes: | NFO and GYB |
#### 4. Description of Device
BMR Face. Type 371/372* (* Type 372 is simply a black version of Type 371) is a twochannel, battery operated cosmetic device which is intended for use on the face. The device consists of a rechargeable electronic device which connects to an applicator. Each side
{1}------------------------------------------------
paddle of the applicator contains a set of replaceable twin conductive gel pads which deliver electrical impulses to the face.
The BMR Face device operates by applying Transcutaneous Electrical Nerve Stimulation (TENS) to the facial area. Low-level electrical impulses are sent from the stimulator unit via an applicator to gel pads that are positioned anterior to each ear and strategically over the cranial nerve #7 of the face,
3 programs are available to the user and these vary in treatment times from 10 to 20 minutes. Power is derived from a 3.6V NiMH rechargeable battery pack that is pre-installed in the unit and there is no capability for charging the unit whilst any treatment is ongoing. All internal connections of the unit are over molded to prevent moisture ingress.
# 5. Statement of Intended Use/Indications for Use
BMR Face is intended for facial stimulation and is indicated for over the counter cosmetic use
#### 6. Summary of Technological Characteristics
There are no new technological characteristics that could affect safety or effectiveness of the BMR Face device. Substantial Equivalence has been demonstrated as part of this 510k submission. Table 5.1 provides a summary comparison of technological characteristics of BMR Face versus that of the predicate "Rejuvenique® System".
{2}------------------------------------------------
| Table 5.1 | BMR Face, Type 371 (New Device) | Rejuvenique System (Predicate) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same | Stimulation of the Face |
| Indications for Use | BMR Face is intended for facial stimulation and is indicated for<br>over the counter cosmetic use. | Cosmetic Use (OTC) |
| Target Population | Healthy adults | Healthy adults |
| Anatomical Site | Same | Face |
| Where Used | Same | Portable, may be used at home |
| Energy Used/Delivered | 3.6V (rechargeable battery). Symmetric. Pulsed, Biphasic | 9V battery (standard or rechargeable) |
| Human Factors | Validation Testing conducted in accordance with the Draft<br>Guidance Document Applying Human Factors and Usability<br>Engineering to Optimize Medical Device Design, and the AAMI's<br>HE75 Human factors engineering-design of medical devices | Ref: K011935 |
| Materials | Same for unit. Applicator constructed of PC-141R. 2 applicator<br>paddles contain carbon impregnated silicone rubber. The pads are<br>manufactured by Axelgaard Manufacturing Company, Fallbrook,<br>California and are composed of a skin conductive adhesive<br>hydrogel layer (Axelgaard MultiStick® MG-1500) which contains<br>glycerin, water and a polyacrylate co-polymer. Each gel pad is<br>supplied with protective liners for both skin and the paddle<br>adhesive faces. The charger is constructed of PC/ABS. | Unit constructed of ABS Plastic, Face Mask is a PBC mask<br>with 26 fixed-position, gold-plated brass electrodes. |
| Biocompatibility | Gel pads comply to ISO 10993-5 Biological evaluation of medical<br>devices - Part 5: Tests for in vitro cytotoxicity & ISO 10993-10 | Ref: K011935 |
#### Table 5.1 Summary Comparison of Technological Characteristics of BMR Face Versus Rejuvenique® System (Predicate)
.
·
:
.
{3}------------------------------------------------
| Table 5.1 | BMR Face. Type 371 (New Device) | Rejuvenique System (Predicate) |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Biological evaluation of medical devices - Part 10: Tests for<br>irritation and delayed-type hypersensitivity | |
| Compatibility with the<br>environment/other devices | Complies to IEC 60601-1-2 Medical electrical equipment - Part 1-<br>2: General requirements for safety - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests | Ref: K011935 |
| Sterility | N/A | N/A |
| Electrical Safety | Complies to IEC 60601-1 Medical electrical equipment - Part 1:<br>General requirements for safety & IEC 60601-2-10 Medical<br>electrical equipment - Part 2-10: Particular requirements for the<br>safety of nerve and muscle stimulators | Ref: K011935 |
| Mechanical Safety | Complies to IEC 60601-1 & IEC 60601-2-10 | Ref: K011935 |
| Chemical Safety | MSDS Sheet provided for conductive gel pads | Ref: K011935 |
| Design | Device contains a control unit, applicator, gel pads, charger and<br>instructions for use. | Device contains a control unit, facial mask, adjustable<br>headband, connecting cable, toning gel, and instructions for<br>use. |
| | The unit may be recharged (charger supplied).<br>12 twin conductive gel pads used<br>Applicator used to apply stimulation (gel pads placed on<br>two paddles of applicator) | Device may not be charged, rechargeable batteries<br>may be used however.<br>Toning gel supplied<br>Facial mask used to apply stimulation (toning gel<br>used on the gold-plated brass contact points). |
| User Interface | The user interfaces with the control unit, applicator and gel pads.<br>The user attaches the gel pads to the two paddles of the applicator | The user interfaces with the control unit, face mask and<br>conductive gel. The user applies a pea sized amount of the |
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
.
ﺗﺮ
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
.
.
Page 4 of 8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{4}------------------------------------------------
Image /page/4/Figure/0 description: This image shows Table 5.1, which compares the BMR Face, Type 371 (New Device) and the Rejuvenique System (Predicate). The table includes images of the control units for each device, with labels pointing to different features such as the display, buttons, and cable connection. Additionally, the table provides information on the modes of operation for the BMR Face device, including the treat time, frequency, pulse width, ramp-up, contr, ramp-down, and relax settings for three different programs. The Rejuvenique System has one program, and the intensity may be adjusted on the control unit.
Page 5 of 8
{5}------------------------------------------------
| Table 5.1 | BMR Face, Type 371 (New Device) | Rejuvenique System (Predicate) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Performance Testing | Comparison basic unit characteristics and output specifications<br>tables, using the "Guidance Document on Powered Muscle<br>Stimulator 510(k)s. 1999"and the draft guidance document "Class<br>Special Controls Guidance Document: Transcutaneous<br>Electrical Stimulator with Limited Output for Aesthetic Purposes"<br>April 5, 2010, have been included as part of this submission. | Ref: K011935 |
| Thermal Safety | Complies to IEC 60601-1 & IEC 60601-2-10 | Ref: K011935 |
| Radiation Safety | N/A | N/A |
| Standards Met | IEC 60601-1:1988 +A1:91, A2:95, Corrigendum 95<br>EN 60601-1-2:2007<br>EN 60601-2-10:2000 + A1:2001 (IEC 60601-2-10:1987/A1:2001)<br>Charger: UL 60950 & CSA-C22.2 No. 60950-00<br>ISO 14971:2007 | Ref: K011935 |
| Software | Completed based on the "Guidance for the Content of<br>Premarket Submissions for Software Contained in Medical<br>Devices" 2005. | Ref: K011935 |
.
. . . .
.
.
.
.
and the state of the state of the state of the states of the states and
. .
:
.
Page 6 of 8
:
{6}------------------------------------------------
# 7. Clinical and Non-Clinical Tests
Clinical Tests: No new clinical studies have been submitted as part of this Premarket Notification.
Non-Clinical Tests: BMR Face has been designed and independently tested to the following requirements:
- . EN 60601-1: 1990 + A1: 93 + A11: 93 ++ A12: 93, A2:95, A13: 96, Corrigendum 94 Medical Electrical Equipment - Part 1: General Requirements for Safety.
- IEC 60601-1:1988 +A1:91, A2:95, Corrigendum 95 Medical Electrical Equipment -. Part 1: General Requirements for Safety
- . EN 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (IEC 60601-1-2:2007 Modified)
- . EN 60601-2-10:2000 + A1:2001 Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators (IEC 60601-2-10:1987/A1:2001
In addition, the power supply unit complies with the following:
- . UL 60950 - Information Technology Equipment Safety - Part 1: General Requirements
- CSA-C22.2 No. 60950-00 Information Technology Equipment Safety Part 1: . General Requirements
### 8. Safety and Effectiveness
Bio-Medical Research Ltd. ("BMR") has over 30 years experience in the research. design, manufacture and marketing of medical grade products for both muscle strengthening and pain relief. BMR has two divisions - Slendertone, which develops and markets a range of consumer health and fitness products and Neurotech, which provides a range of neuromuscular stimulators for pain management and rehabilitation.
Bio-Medical Research Ltd. complies with 21 CFR 820 and is registered to I.S. EN ISO 13485:2003 Medical Device Quality Management System for the design, manufacture and distribution of electro-medical devices.
- A risk management plan was carried out to EN ISO 14971:2007. .
{7}------------------------------------------------
- Independent EMC and Electrical Safety testing has been carried out. .
.
.
- In Europe (EU), Slendertone Face has been CE marked and complies with the Medical . Device Directive 93/42/EEC.
.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Bio-Medical Research Ltd. c/o Ms. Anne-Marie Keenan Quality and Regulatory Engineer Parkmore Business Park West Galway Ireland El
NOV 1 0 2011
Re: K103031
Trade/Device Name: BMR Face Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NFO Dated: October 27, 2011 Received: October 31, 2011
Dear Ms. Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Ms. Anne-Marie Keenan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kesia Alexander
for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known):
_K103031
Device Name:
BMR Face, Type 371 & Type 372
Indications for Use:
BMR Face is intended for facial stimulation and is indicated for over the counter
cosmetic use
Prescription Use _ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K103031
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.