F&P Optiflow Air/Oxygen Flow Source

K243917 · Fisher & Paykel Healthcare, Ltd. · BZR · Jul 14, 2025 · Anesthesiology

Device Facts

Record IDK243917
Device NameF&P Optiflow Air/Oxygen Flow Source
ApplicantFisher & Paykel Healthcare, Ltd.
Product CodeBZR · Anesthesiology
Decision DateJul 14, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5330
Device ClassClass 2
AttributesTherapeutic

Intended Use

This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. This product is not intended for apneic ventilation.

Device Story

F&P Optiflow Air/Oxygen Flow Source is a respiratory high flow therapy device; generates air/oxygen mixture for hospital use. Inputs: ambient air and external oxygen source (40–87 psi). Operation: electronic flow and FiO2 control; user interface via touch screen display. Outputs: high flow respiratory gases (up to 70 L/min; 21–100% oxygen concentration). Operated by qualified healthcare professionals in hospitals. Device monitors flow and FiO2 output; includes visual/audible alarm system for supply pressure, FiO2, power failure, disconnection, and obstruction. Output assists clinicians in managing respiratory gas delivery; benefits patients by providing controlled, high-flow respiratory support.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety (ANSI/AAMI ES 60601-1), alarm systems (IEC 60601-1-8), software lifecycle (IEC 62304), usability (IEC 62366-1), electromagnetic immunity (AIM 7351731), respiratory high-flow equipment requirements (ISO 80601-2-90), and gas mixer performance (ISO 11195:2018).

Technological Characteristics

Mains-powered respiratory gas mixer. Electronic flow and FiO2 control. Flow range 0–70 L/min; FiO2 21–100%. Touch screen interface. Visual/audible alarm system. Complies with ISO 80601-2-90 and ISO 11195:2018. Software developed per IEC 62304.

Indications for Use

Indicated for delivery of high flow respiratory gases to patients in hospital settings by qualified healthcare professionals. Not for patients with bypassed upper airways, apneic ventilation, or life-sustaining/supporting use.

Regulatory Classification

Identification

A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION July 14, 2025 Fisher &amp; Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand Re: K243917 Trade/Device Name: F&amp;P Optiflow Air/Oxygen Flow Source Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR Dated: June 12, 2025 Received: June 12, 2025 Dear Reena Daken: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243917 - Reena Daken Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K243917 - Reena Daken Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Bradley Q. Quinn -S Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243917 Device Name F&amp;P Optiflow Air/Oxygen Flow Source Indications for Use (Describe) This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. This product is not intended for apneic ventilation. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} F&amp;P Optiflow Air/Oxygen Flow Source - K243917 # 510(k) Summary As Required by 21 CFR 807.92 ## I. SUBMITTER | Company Name and Address | Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place East Tamaki Auckland 2013, New Zealand Telephone: +64 9 574 0100 | | --- | --- | | Prepared and Submitted by | Danica Tung Regulatory Affairs Market Manager | | Contact Person | Reena Daken Regulatory Affairs Manager Telephone: +64 9 574 0100 Email: reena.daken@fphcare.co.nz | | Date Prepared | 14 July 2025 | ## II. DEVICE | Name of Device | F&P Optiflow™ Air/Oxygen Flow Source | | --- | --- | | Common/Usual Name | Breathing Gas Mixer | | Classification | 21 CFR §868.5330 | | Regulatory Class | II | | Product Code | BZR | ## III. PREDICATE DEVICES Primary Predicate Device: MaxBlend 2 and MaxBlend Lite, K161718 Secondary Predicate Device: Airvo 2 Humidifier and myAirvo 2 Humidifier, K131895 {5} F&amp;P Optiflow Air/Oxygen Flow Source - K243917 ## IV. DEVICE DESCRIPTION The F&amp;P Optiflow Air/Oxygen Flow Source is a respiratory high flow therapy device designed to generate a flow of air and/or oxygen. The device allows selectable flow rates up to 70 L/min with selectable oxygen concentrations ranging from 21% to 100%. The subject device is a multi-patient use prescription only device, provided in a non-sterile state. It operates at flow ranges between 0 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals. ## V. INDICATIONS FOR USE This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. This product is not intended for apneic ventilation. {6} F&amp;P Optiflow Flow Diverter - K243917 VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | Classification | | | | | | Device Classification | 21 CFR 868.5330 Class II | 21 CFR 868.5330 Class II | 21 CFR 868.5450 Class II | Identical | | Classification Panel | Anesthesiology | Anesthesiology | Anesthesiology | Identical | | Product Code | BZR | BZR | BTT | Identical | | Indications for use and intended use | | | | | | Indications for Use | This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. | The MaxBlend 2 and MaxBlend Lite are designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, sub-acute, and nursing-care facilities where the delivery and | The Airvo 2 Humidifier is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2 - 60 L/min depending on the patient interface. The Airvo 2 Humidifier is for patients in hospitals and long-term care facilities. | Equivalent | {7} F&amp;P Optiflow Flow Diverter - K243917 | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | | This product is not intended for apneic ventilation. | monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device. | | | | Availability | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Identical | | Intended Use Environment | Hospital | Professional healthcare settings, i.e., hospital, sub-acute, and nursing-care facilities, where the delivery and monitoring of air/oxygen mixtures is required. Not for use in MRI environments | Hospitals and long-term care facilities | Equivalent | | Users | Qualified healthcare professionals | Qualified trained personnel, under the direction of a physician | Healthcare professionals | Identical to secondary predicate. | | Contraindications | No contraindications | No contraindications | No contraindications | Identical | | Technological Characteristics | | | | | | Power/Energy Source | Mains power | Battery (4 x AA Alkaline batteries) - Power source for the oxygen monitor only. | Mains power | Identical to secondary predicate | | Control & Monitoring | Electronic flow and FiO2control. | Mechanical flow and FiO2control. Monitoring FiO2output. | Electronic flow control. | Equivalent | {8} F&amp;P Optiflow Flow Diverter - K243917 | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | | Monitoring of flow and FiO2 output. | | Monitoring of flow and FiO2 output. | | | Device User Interface | Display with touch screen | Monitoring system has a display and push buttons for user input. Flow control is via rotary dial and a ball float for indicator | Display with push buttons | Different Similar to the primary and secondary predicates, the subject device has a display screen. Input and control for the subject device is via the touch screen, whereas both predicates utilise push buttons and a rotary dial (primary predicate only). The difference in device user interface does not raise new questions of safety and effectiveness. | | Performance Specifications | | | | | | Flow Range | 0 - 70 L/min | 0 - 100 L/min | 2 - 60 L/min | Equivalent | | Flow Meter Accuracy | Accuracy of +/- 10% | Accuracy of +/- 10% | N/A | Identical | | Operating Temperature Range | 64°F to 79°F (18 °C to 26 °C) | 59°F to 122°F (15°C to 50°C) | 64°F to 79°F (18 °C to 26 °C) | Equivalent | {9} F&amp;P Optiflow Flow Diverter - K243917 | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | % Oxygen Control | 21 – 100% | 21 – 100% | 21 – 100% | Identical | | % Oxygen Control Accuracy | +/- 5% | +/- 3% | Not applicable | Different The subject device's % oxygen control accuracy limit is based on ISO 11195:2018, that specifies the requirements for the performance and safety of stand-alone gas mixers (blenders). The difference in oxygen control accuracy does not raise new questions of safety and effectiveness. | | Alarms | | | | | | Alarm method | Visual and audible alarm system. Mute button | Visual and audible alarm system. Mute button | Visual and audible alarm system. Mute button | Identical | | Oxygen Supply Alarm | Issues alarm if the oxygen supply pressure is < 40psi | Air/oxygen mixer feature: Pressure supply differential alarm - Air/oxygen pressure must be < 20 psi an alarm sounds | Not Applicable | Equivalent | {10} F&amp;P Optiflow Flow Diverter - K243917 | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | FiO2 Alarm | Device alarms if it is unable to achieve set FiO2FiO2: 21 - 100% | Oxygen monitor features: High alarm range - 16 - 100% | Issues alarm if oxygen fraction as measured by oxygen sensor is above or below user adjustable high/low limit. | Equivalent | | Power-fail Alarm | Power out audible alarm for time, t > 2 minutes or until power is restored, or capacitor charge is depleted. | Unknown | Power out audible alarm for time, t > 2 minutes or until paused by user pushing alarm mute button, power is restored (non-latching alarm), or capacitor charge is depleted. | Equivalent | | Disconnection | Disconnection alarm activated in < 5 seconds when no connection/disconnection has been detected | Unknown | ‘Check tube’ and ‘Check for leaks’ alarms are activated in < 5 seconds if the breathing tube is not plugged in correctly | Identical to secondary predicate | | Obstruction Detected/Flow Alarm | Obstruction alarm will be activated in < 5 seconds when an obstruction has been detected | Unknown | ‘Check for blockages’ alarm in <10 seconds when the device has detected a blockage | Equivalent | | General | General alarm warning is activated when internal faults, such as hardware, visual alarm, audio alarm and alarm systems has been detected. | Unknown | Fault alarm is activated when the unit has detected an internal fault and has shut itself down. | Equivalent | {11} F&amp;P Optiflow Flow Diverter - K243917 | Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments | | --- | --- | --- | --- | --- | | Other | | | | | | Gas Supply Type & Pressure | Room air: Ambient pressure External oxygen source : 40 – 87 psi | Air/Oxygen: 30 – 75 psi | Room air: Ambient pressure External oxygen source (optional) | Equivalent | | Materials in Gas Pathway | Externally communicating – tissue contact, permanent duration. No exposure to humidified gases. VOC and PM2.5 | Externally communicating – tissue contact, permanent duration. No exposure to humidified gases. VOC and PM2.5 | Indirect contact - External communicating - tissue, bone, dentin | Identical | {12} F&amp;P Optiflow Flow Diverter - K243917 # VII. PERFORMANCE DATA # Summary of Non-Clinical Tests Performance testing of the Flow Source was completed and confirms the subject device does not raise new questions of safety and effectiveness. The testing provided demonstrates substantial equivalence of the subject device to the predicate devices. The Flow Source has been tested to the applicable requirements to the following standards: - ANSI/AAMI ES 60601-1:2005/A2:2021 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety and Essential Performance - Amendments 2 - IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems - IEC 62304:2006 + A1:2015 Medical device software - Software life cycle processes - IEC60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability - ISO 17664-2:2021 Processing of health care products – Information to be provided by medical device manufacturer for the processing of medical devices. Part 2: Non-critical medical devices - IEC 62366-1:2015 + A1:2020 Medical devices - Part 1: Application of usability engineering to medical devices - AIM 7351731 Rev 3:2021 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers - ISO 80601-2-90:2021 Medical electrical equipment - Part 2-90: Particular requirements for basic safety and essential performance of respiratory high-flow therapy equipment - ISO 11195:2018 Gas mixers for medical use - Stand-alone gas mixers - ANSI/AAMI ST98:2022 Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices. # VIII. CONCLUSION The Flow Source is substantially equivalent to the predicates based on intended use, technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate device.