The Precision Flow™ is intended to be used for adding warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.
Device Story
Precision Flow™ provides heated, humidified respiratory gas to patients; inputs include external medical air and oxygen sources. Main unit uses mass flow sensors and proportional solenoid valves to blend gases; integral oxygen analyzer monitors mixture integrity. Vapor transfer cartridge uses hollow fiber technology to humidify gas via vapor diffusion without direct water-gas contact. Triple-lumen heated delivery tube maintains temperature and minimizes rain-out. Operated by clinicians or patients in hospital, sub-acute, or home settings. Firmware monitors gas pressure, water levels, and temperatures; triggers alarms for out-of-range parameters. Internal battery provides 15-minute backup. Output is humidified gas delivered via nasal cannula; assists in respiratory therapy by providing controlled heat and moisture to breathing gases.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (LLNA, intracutaneous reactivity, MEM elution), VOC and particulate matter analysis, EMC/emissions testing, and safety testing per IEC 60601-1, 60601-1-4, and 60601-1-8. Software verified and validated per FDA guidance. 30-day comparative use testing showed no detectable bacteria in water condensation samples.
Technological Characteristics
System comprises main unit and disposable module (water reservoir, vapor transfer cartridge, heated delivery tube). Gas pathway uses hollow fiber polymer membranes for vapor diffusion. Sensors monitor gas flow, pressure, water level, and temperature. Connectivity includes internal battery backup. Software-controlled electronic blender with proportional solenoid valves. Sterilization/biocompatibility per ISO/ASTM standards. Electrical safety per IEC 60601-1.
Indications for Use
Indicated for neonate, infant, pediatric, and adult patients requiring warmed, humidified respiratory gas administration via nasal cannula at flow rates of 1 to 40 liters per minute in hospital, sub-acute, or home settings.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
K072845
## 510(k) SUMMARY
## Vapotherm, Inc.'s Precision Flow™
JUL 1 7 2008
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Vapotherm, Inc. 198 Log Canoe Circle Stevensville, Maryland 21666
Phone: 410-604-3977 Facsimile: 410-604-3978
Contact Person: Gregory A. Whitney
April 15, 2008 Date Prepared:
### Name of Device and Name/Address of Sponsor
Precision Flow™ Vapotherm, Inc. 198 Log Canoe Circle Stevensville, Maryland 21666
#### A.. Common or Usual Name
Humidifier, Respiratory Gas (Direct Patient Interface)
#### B. Classification Name
Humidifier, Respiratory Gas (Direct Patient Interface) Anesthesiology Panel (868.5450) Class II
#### C. Product Code
BTT
#### D. Predicate Devices
Vapotherm. Inc. Medtec Bird Products
2000i& 2000h Oxygen Sensor Air-Oxygen Blender
K000401, K0142245, K042245 K063488 K911962
{1}------------------------------------------------
### E. Intended Use / Indications for Use
The Precision Flow™ is intended to be used for adding warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.
#### F. Technological Characteristics
The Precision Flow™ consists of two parts:
The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection.
The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.
# 1. Main unit:
- · The flow of oxygen and air are measured by mass flow sensors. The operating software calculates the required flow of each needed to reach the target flow and oxygen percentage set by the operator. The system controls gas flows accordingly by adjusting proportional solenoid valves on the gas lines. An oxygen sensor monitors the gas mixture and signals any discrepancy between target and measured percentage. The oxygen sensor is automatically calibrated with oxygen at power-up and every 24 hours.
- · Firmware running in the main unit uses sensors to monitor gas pressure, water level and water temperature, and to detect air leaks into the water circuit (bubble detector) and water leaks into the gas circuit (droplet detector). Alarms are displayed if any parameters are outside the normal range. Other indicators show low charge in the backup battery, and the type of cartridge installed. An internal battery backup will maintain the set flow and oxygen blend for at least 15 minutes without AC power.
{2}------------------------------------------------
### Disposable components: 2.
- · Vapor Transfer Cartridge. In the cartridge blended gas passes through the lumens of hundreds of parallel hollow fibers made of a specially developed polymer. Warm water circulates around the fibers and diffuses as vapor through the fiber material into the gas stream flowing through each fiber. Unlike most humidifiers, there is no direct contact between the water and gas streams. The gas stream leaves the cartridge essentially saturated with vapor at the set temperature.
- · Triple-lumen heated delivery tube. The warmed humidified gas passes through the center lumen. The center lumen is surrounded by two outer lumens carrying warmed water to maintain the temperature of the inner lumen and to minimize rain-out. A proprietary short nasal cannula is connected to the end of the delivery tube and passes the humidified breathing gas to the patient's nose.
- · Disposable module. The module houses a water reservoir, pump, connections for the cartridge and delivery tube, and sensor interfaces to the main unit. Water is pumped past a heater plate through the outer lumens of the delivery tube. Returning water passes through the outer jacket of the Vapor Transfer Cartridge where some water is lost as vapor to the gas stream. There is no direct contact between water and gas flows. The water then returns to the pump reservoir. Heater power is adjusted continuously to maintain the set temperature. Water flows into the circuit from the water bag to replace evaporative losses in the Vapor Transfer Cartridge. Air is purqed to atmosphere from the circulation via a hydrophobic filter membrane.
#### 3. Summary of Non-Clinical Performance Data
The following table lists the non-clinical Performance Testing Data that was performed on the Precision Flow™:
| Number | Test Description |
|--------|------------------------------------------------------------------------------------------------------------------------|
| 1 | Biocompatibility – Murine Local Lymph Node Assay (LLNA) (Sensitization) |
| 2 | Biocompatibility - ISO Intracutaneous Reactivity Test (Irritation) |
| 3 | Biocompatibility – MEM Elution (Cytotoxicity) |
| 4 | Volatile Organic Compounds (VOC), EPA Compendium Method TO-15 |
| 5 | Particulate Matter, NIOSH Method 0500 |
| 6 | EMC Test |
| 7 | Emissions Test |
| 8 | Safety Inspection Test |
| 9 | IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety |
| 10 | IEC 60601-1-4, Medical Electrical Equipment Part 1-4 - Collateral Standard:<br>Programmable Electrical Medical Systems |
{3}------------------------------------------------
| (<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| |
| Number | Test Description |
|--------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 11 | IEC 60601-1-8, Medical electrical equipment, General requirements for basic safety and<br>essential performance - Collateral standard: General requirements, tests and guidance<br>for alarm systems in medical electrical equipment and medical electrical systems |
| 12 | Bio Burden, 30 Day Comparative Use Testing |
| 13 | Ximedica TRP 1097, “Vapotherm Precision Flow Hyperthermic Humidification System,<br>System Software Verification & Validation Test Plan" |
| 14 | TRP 1071 "System Software Test" (Black Box) |
| 15 | Unit Test Case Listing - 3VAP1004/TRP 1092 (White Box) |
| 16 | Thermal Stability |
| 17 | Blender Comparison Performance |
- Bio Burden, 30 Day Comparative Use Testing: After a thirty day . comparative use testing, there were no detectible bacteria in any water condensation samples from the Precision Flow® devices either initially (3 days) or after 30 days of operation.
- The Precision Flow™ software was verified and validated in . accordance with applicable FDA guidance documents.
- . The Volatile Organic Components and Particulate Mater tests were performed to recognized standards.
### Substantial Equivalence
The Precision Flow™ is as safe and effective as the Vapotherm 2000i and 2000h, the Bird Mircoblender, and the Medtec Oxygen Sensor. The Precision Flow™ has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the Precision Flow™ and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Precision Flow™ is as safe and effective as the predicate devices. Thus, the Precision Flow™ is substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. In the center of the circle is a stylized eagle emblem, with its wings spread and head turned to the side. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vapotherm, Incorporated C/O Mr. Jonathan S. Kahan Hogan & Hartson L.L.P. 555 Thirteenth Street, NW Washington DC 20004
JUL 1 7 2008
Re: K072845
Trade/Device Name: Precision FlowTM Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: June 27, 2008 Received: June 30, 2008
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mcan that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to; registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your tevice to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Quars
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for use
510(k) Number (if known):
Device Name: Precision Flow™
Indications for Use:
Precision Flow™ is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.
Precision Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Will Maly for M. Husband Page _ of _
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 1< 072 845
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.