Chlorine Sentinel II

K243898 · Nelson Environmental Technologies, Inc. · MSY · Apr 11, 2025 · Gastroenterology, Urology

Device Facts

Record IDK243898
Device NameChlorine Sentinel II
ApplicantNelson Environmental Technologies, Inc.
Product CodeMSY · Gastroenterology, Urology
Decision DateApr 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5665
Device ClassClass 2

Intended Use

The Chlorine Sentinel II is intended for use by hemodialysis professionals as a secondary chlorine device to provide continuous monitoring of Combined Chlorine. It is not intended to replace the primary method of monitoring chlorine as part of the hemodialysis water treatment system. It functions completely independent of the water treatment system does not come into direct contact with feed water used to prepare dialysate.

Device Story

Chlorine Sentinel II is a secondary monitoring system for combined chlorine in hemodialysis water treatment. It attaches to the drain of a water treatment sample port. The device uses a polarographic (amperometry) sensor with a polymeric membrane to detect dissolved combined chlorine. It performs automated daily self-tests using chlorinated city water. Users configure sampling intervals (5-20 minutes) and daily schedules via a touch screen. The system features two alarm set-points: a user-defined warning (0.01-0.09 PPM) and a factory-set alarm (0.10 PPM). Alarms trigger visual indicators and buzzers. The device includes a water shut-off feature and logs data. It is operated by hemodialysis professionals in a water room setting, with a remote module located at the nurse's station. The output provides real-time chlorine presence and status alerts, allowing clinicians to proactively manage water quality before reaching AAMI standard limits, thereby protecting patients from chlorine exposure during dialysis.

Clinical Evidence

Bench testing only. Performance verified via linearity studies (CLSI EP06-A) and Limit of Blank/Quantitation/Detection studies (CLSI EP17-A2). Electrical safety and EMC testing performed per IEC 61010-1 and IEC 60601-1-2:2020.

Technological Characteristics

Polarographic (amperometric) sensor with polymeric membrane. Enclosure: V0 fire-rated ABS, IP66 rated. PLC-based control (Siemens 1200) with 7" color touch screen. Connectivity: MODBUS TCP, SD card data logging. Power: 24 VDC (110-240 VAC line power). Dimensions: 38.0 x 56.0 x 18.5 cm (controller).

Indications for Use

Indicated for hemodialysis professionals as a secondary monitor for combined chlorine in hemodialysis water treatment systems. Not for use as a primary chlorine detection device.

Regulatory Classification

Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION April 11, 2025 Nelson Environmental Technologies, Inc. Quochuy (Huy) Nguyen Finance Manager / New Product Development Manager 813 E. Fir Avenue McAllen, Texas 78501 Re: K243898 Trade/Device Name: Chlorine Sentinel II Regulation Number: 21 CFR 876.5665 Regulation Name: Water Purification System For Hemodialysis Regulatory Class: Class II Product Code: MSY, PSX Dated: March 18, 2025 Received: March 18, 2025 Dear Quochuy (Huy) Nguyen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243898 - Quochuy (Huy) Nguyen Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K243898 - Quochuy (Huy) Nguyen Page 3 Sincerely, Maura Rooney -S Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243898 Device Name Chlorine Sentinel II Indications for Use (Describe) The Chlorine Sentinel II is intended for use by hemodialysis professionals as a secondary chlorine device to provide continuous monitoring of Combined Chlorine. It is not intended to replace the primary method of monitoring chlorine as part of the hemodialysis water treatment system. It functions completely independent of the water treatment system does not come into direct contact with feed water used to prepare dialysate. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} NELSON ENVIRONMENTAL TECHNOLOGIES, INC. K243898 Page 1 of 11 NELSON WATER SYSTEMS FOR HEMODIALYSIS FDA-Cleared Systems for Hemodialysis 510k) # 993877 FDA-Cleared Systems for Nelson Sentinel Series Basic Chlorine Sentinel 510k) # 193169 ISO 13485:2016 Certified Certificate# C2023-05290 # 6. Summary of 510(k) This 510(k) Summary is in conformance with 21CFR 807.92 Submitter: Nelson Environmental Technologies, Inc. 813 E. Fir Avenue McAllen, TX 78501 Phone: 956-618-0375 Primary Contact: QuocHuy (Huy) Nguyen Email: admin@nws.bz Phone: 956-618-0375 Fax: 956-618-4330 Date Prepared: 2024 December 18 # Device Name and Classification Trade Name: Chlorine Sentinel II Common Name: Water Purification System for Hemodialysis Classification: Class II Regulation Number: 21 CFR 876.5665 Water Purification System for Hemodialysis Classification Panel: Gastroenterology/Urology Product Code: MSY, PSX # Predicate Device: | | Predicate Device | | --- | --- | | Trade Name | Nelson Basic Chlorine Sentinel Basic Chlorine Sentinel | | Common Name | Basic Chlorine Sentinel | | 510(k) Submitter / Holder | Nelson Environmental Technologies, Inc. | | 510(k) Number | K193169/S001 | | Regulation Number | 21 CFR 876.5665 Water purification system for hemodialysis | | Classification Panel | Gastroenterology/Urology | | Product Code | PSX, FIP | # NELSON ENVIRONMENTAL # TECHNOLOGIES, INC P.O. Box 5026, McAllen, TX 78502-5026 10518 Cash Rd., Stafford, TX 77477 Tel: 281-265-0666 / Fax: 832-539-6607 1312 Corporate Dr. E Suite F, Arlington, TX 76006 Tel: 817-652-1152 / Fax: 817-470-4975 http://www.nws.bz 813 E Fir Ave, McAllen, TX 78501 Tel: 956-618-0375 / Fax: 956-618-0375 {5} K243898 Page 2 of 11 # Device Description Chlorine Sentinel II is a complete system that attaches to the drain of a hemodialysis water treatment system sample port. It detects dissolved combined chlorine in concentrations at or above 10 PPB by primary detection technology. It self-monitors for maintenance-required conditions and loss of electrical power. The Chlorine Sentinel II is self-testing. It uses chlorinated city water to automatically test the chlorine detecting functionality of the device once a day and records if the test passed, failed to detect chlorine, or didn't rinse out after detecting chlorine. The device is designed to use as little water as possible. The sampling water is only used for the length of time that it takes for the chlorine test to be completed. The customer determines how frequently the test needs to be taken from every 5 minutes to every 20 minutes. The user also sets the start time and end time for each day of the week. The Chlorine Sentinel II has two combined chlorine concentration set points. The user can set a warning set-point at a value of 0.01 PPM to 0.09 PPM. Once the chlorine concentration exceeds this value, the device will alarm at 2-HZ to warn the user of chlorine present at that set-point. The second set-point is factory set at 0.10 PPM. A chlorine concentration equal to or above the 0.10 PPM set-point will activate the buzzers and the red chlorine alarm indicators at 5 HZ for as long as the condition exists. The buzzers can only be muted for 3 minutes (then automatically reset) for as long as the condition exists. The alarms can be by-passed. When by-passed the pulsing lights will turn solid and the buzzer will double click over 15 minutes. The front cover of the device has two screens. The screen of the chlorine monitor indicates the presence of combined chlorine when the value is not 0.000. The touch screen is used to enter user data and show the statuses of the device. # Indications for Use The Chlorine Sentinel II is intended for use by hemodialysis professionals as a secondary chlorine device to provide continuous monitoring of Combined Chlorine. It is not intended to replace the primary method of monitoring chlorine as part of the hemodialysis water treatment system. It functions completely independent of the water treatment system does not come into direct contact with feed water used to prepare dialysate. # Risk Analysis Method Chlorine Sentinel II was assessed to determine risks to health associated with the use of the device, a risk analysis was conducted in accordance with ISO 14971: 2007, Medical devices – Application of risk management to medical devices. 6-2 {6} Substantial Equivalence K243898 Page 3 of 11 The Chlorine Sentinel II is essentially the same as the Nelson Basic Chlorine Sentinel with the following differences: 1. A Touch Screen and PLC replace the electrical components of the Basic Chlorine Sentinel; 2. It is inside one enclosure instead of two; 3. User can set a warning set-point for chlorine concentration; 4. User can set the start and stop times for each day; 5. User can set the sampling rate times from 5 minutes to 20 minutes; 6. The Touch Screen highlights alarm statuses; 7. Includes water Shut-off Device. The table below provides a detailed comparison of Nelson Sentinel Series Basic Chlorine Sentinel to the predicate and reference devices. 6-3 {7} K243898 Page 4 of 11 Detailed Comparison of the Subject and Predicate Devices | Item | Subject Device | Predicate Device | Reference Device | Comparison to Predicate Device (Noted reference device comparison for items that are closer to it) | | --- | --- | --- | --- | --- | | | Chlorine Sentinel II | Nelson Sentinel Series Basic Chlorine Sentinel | Hach CM130 Chlorine Monitor | | | Intended Use/Indications for Use | The Chlorine Sentinel II is an instrument that is intended for use by hemodialysis professionals as a secondary chlorine monitor to indicate the presence of combined chlorine in feed water used in hemodialysis systems. It can be used as a secondary chlorine test. It is not intended to be used as a primary chlorine detection device | Nelson Sentinel Series Basic Chlorine Sentinel is indicated for detection of Total Chlorine in Water. Nelson Sentinel Series Basic Chlorine Sentinel is indicated for testing water used to prepare dialysate. | Hach CM130 is intended for use in hemodialysis Water treatment systems to read Total Chlorine concentration. | Same as Predicate | | Provides Exact Chlorine Value? | No; Intended to indicate chlorine presence above 0.01 PPM and reference value. | Approximate values | Yes; presents an exact chlorine value on its display intended to be used to make clinical decisions | Same as Predicate | | Provides Exact Combined Chlorine Value? | Yes; presents an exact combined chlorine value on its display | Yes | No; measures total chlorine which may be free, combined dissolved, and suspended chlorinated solids | Same as Predicate | | Detection Chemistry/Technology | Direct measuring polarographic sensor utilizing a special polymeric membrane. | Direct measuring polarographic sensor utilizing a special polymeric membrane. | DPD Colorimetric Method | Same as Predicate | | Chlorine Sensor Type | Polarography (Amperometry) | Polarographic sensor | Photometric sensor | Same as Predicate | | Interference with other compounds in potable water | None | None | Chlorinated organics/suspended solids | Same as Predicate | | Reagents | None (Ammonium Sulfate if used with municipal free chlorine disinfected water) | Ammonium sulfate | Indicator solution (DPD indicator with potassium iodide), buffer solution | Same as Predicate | 6-4 {8} K243898 Page 5 of 11 | Reagent Delivery | N/A | Peristaltic pump | Peristaltic pump | Same | | --- | --- | --- | --- | --- | | Type of Chlorine Detected | Dissolved Combined Chlorine, free chlorine to a lesser degree | Dissolved Combined Chlorine, free chlorine to a lesser degree | Total Chlorine (free chlorine plus total chloramines) | Same as Predicate | | Detection Range | > 0.01 mg/L | 0.05 mg/L Cl₂ – 0.15 mg/L Cl₂ | 0.03 mg/L Cl₂ – 0.20 mg/L Cl₂ | Same as Predicate | | Chlorine Level Alarm | 0.1 mg/L Cl₂ | 0.1 mg/L Cl₂ | 0.1 mg/L Cl₂ | Same as Predicate | | Warning Chlorine Set-point | 0.01PPM – 0.09PPM | NA | NA | Different. Warning set-point is a customer requested feature to allow them to be proactive when the presence of combined chlorine is detected versus reactive when the AAMI standard chlorine concentration is reached. | | Self-Testing | Automatic testing daily with test passed notification, manual reset. Manual testing for maintenance and calibration | Automatic testing daily with test passed notification, manual reset. Manual testing for maintenance and calibration | None | Same as Predicate | | Sampling | The chlorine sensor is located on the downstream side of a water treatment test sample port. The sampled water is then discharged down the drain. | The chlorine sensor is located on the downstream side of a water treatment test sample port. The sampled water is then discharged down the drain. | Feed water is sampled. Water that flows into the CM130 device exits the device to a drain. The water samples, once tested, are also discarded to a drain. | Same as Predicate | | Water Sample Conditioning and Delivery | Automatic | Automated | Automated | Same as Predicate | | Measurement Interval | 5 -30 Minutes | Continuous | 5 minutes | Different. Chlorine Sentinel II only uses water during the sampling cycle. This significantly reduces the amount of water used over the life of the device | 6-5 {9} K243898 Page 6 of 11 | Alarms/ Alerts | Controller: Multi-color Touch Screen highlighted Colors (green for power indication, Red for Chlorine over set-points, yellow for maintenance notification, orange for past chlorine detected, blue for auto-test passed) and a buzzer Remote Module: Four colors (green for power indication, red for chlorine alerts, yellow for maintenance notification, blue for auto test passed) and a buzzer | Controller: Three colors (green for power indication, 2 yellow, and red for alerts) and a buzzer Remote Module: Three color lights (green for power indication, 2 yellow, and a red for alerts) and a buzzer | Analyzer: Two colors (amber and red) on the display and three sounds for notifications. Remote Indicator: Two lights show the analyzer status (blue, amber, or red). Two lights show the chlorine status (blue, amber, or red). Three different sounds: hardware alarm, medium-high chlorine alert, and high chlorine alarm. | Different. Auto-test successful light is blue instead of yellow (chlorine Detected) in the predicate device Lights and buzzers pulse at two different rates depending on user-set set-point and 0.10 mg/L set-point for combined chlorine concentration. | | --- | --- | --- | --- | --- | | Sounds | Piezoelectric buzzer providing 5 Hz pulses for chlorine over 0.10 mg/L, 2 Hz chlorine over user-defined Warning Set-Point or for maintenance notification, solid if PLC failure. 15-minute double chirp if any alarms have been by-passed | | Analyzer: Two different sounds identify a hardware alarm or high chlorine alarm. Remote Indicator: Three different sounds identify a power-loss alarm, hardware alarm or high chlorine alarm. | Different. Predicate device has solid buzzer. Chlorine Sentinel II has added two pulse frequencies to indicate Chlorine over set-points and action-required notifications. | | Measurement Log | Unlimited | | 3000 measurements maximum | Different. Predicate device has no log. | | Measurement Point | Between the primary and secondary carbon tanks (filters) | After the secondary carbon filter, before the reverse osmosis (RO) machine. | Between the primary and secondary carbon tanks (filters). | Same as Predicate | | Device Location | Remote Module is located in the nurse’s station; all other components are located in the water room | Remote Module is located in the nurse’s station; all other components are located in the water room | Remote indicator is located in the patient room; all other components are located in the water room | Same as Predicate | 16-6 {10} K243898 Page 7 of 11 | Enclosure | V0 Fire-rated ABS, IP66 rated with door closed. Remote Indicator Material: PC/ABS | V0 Fire-rated ABS, IP66 rated with door closed. Remote Indicator Material: PC | Enclosure Rating: IP52 rated with the door closed. Enclosure Material: PC/ABS case, PC door, PC hinges & latches, 316 SST hardware. Remote Indicator Material: PC/ABS | Same as Predicate | | --- | --- | --- | --- | --- | | External Drain Sample Port | On right side wall of enclosure designed for easy access of sample bottles. | On right side wall of enclosure designed for easy access of sample bottles. | | Same as Predicate | | Enclosure to protect electronics from splashing or spills | Physical splash barrier | NA | Yes | Predicate had separate enclosures to separate electronics from water. Substantially equivalent. Similar to CM130 | | Water Treatment Required | No | No | No | Same as Predicate | | Device Components | Analyzer, Remote Indicator | Controller, Probe/Tester, Remote Module | Analyzer, Remote Indicator | Substantially equivalent to Predicate | | Control Mechanism | PLC – Siemens 1200 Touch Screen - Mitsubishi | Microcontroller | Microcontroller | Different. Substantially equivalent to Reference | | Maintenance | Quarterly | Quarterly | Monthly | Same as Predicate | | Data Logger | SD Card | No data logger | 3000 measurements maximum | Different. Substantially equivalent to Reference | | Status Outputs | 1. Combined chlorine over set-point (Default 0.100) 2. Analyzer Self-Diagnostic system failures or maintenance required. 3. Auto-Test Passed notification 4. Loss of analyzer signal 5. Loss of analyzer electrical power 6. Auto-Test Failed | 5 water treatment status outputs - RO shut-down, flood, loss of electricity, device maintenance, and high chlorine | Two levels of chlorine, power statuses | Substantially equivalent to Predicate | 6-7 {11} K243898 Page 8 of 11 | Data Outputs | MODBUS TCP | | MODBUS TCP | | | --- | --- | --- | --- | --- | | Data History | The measurement history is available through the Ethernet connection. Measurement History & Event History can be copied to an SD card as two separate files. | NA | The measurement history is available through the Ethernet connection. Measurement History & Event History can be copied to an SD card as two separate files. | Different. Substantially equivalent to Reference | | Display | 17.9 cm (7”) Color Touch Screen Monitor: Backlit LCD: 1.91 cm (0.75 in.) high 4-digit main display, 0.76 cm (0.3 in.) 5x7 dot matrix 12-digit secondary display | Monitor: Backlit LCD: 1.91 cm (0.75 in.) high 4-digit main display, 0.76 cm (0.3 in.) 5x7 dot matrix 12-digit secondary display | 10.9 cm (4.3 in.) color display, backlit LCD< WQVGA | Different. Predicate has no Touch Screen. Substantially equivalent to Reference | | System Power | 24 VDC | 24VDC | 24 VDC | Same as Predicate | | Power Requirement | 110 – 240 VAC line power | 110 – 240 VAC line power | 100 – 240 VAC, 50/60 Hz, 1 A max | Same as Predicate | | Mounting | Water Room: Wall Remote Indicator: Wall | Water Room: Wall Remote Indicator: Wall | Analyzer: Wall Remote Indicator: 1.1 kg (2.5 lbs.) | Same as Predicate | | Dimensions | Controller: 38.0 x 56.0 x 18.5 cm Remote Module: 14 x 11 x 5 cm | Controller: 32.0 x 42.0 x 16.5 cm (12.5 x 16.5 x 6.5 in.) Probe/Tester: 18.0 x 28.0 x 18.5 cm (7.1 x 11.0 x 7.3 in.) Remote Module: 20.0 x 14.5 x 10.0 cm (7.9 x 5.7 x 3.9 in.) | Analyzer: 37.0 x 53.8 x 22.6 cm (14.6 x 21.2 x 8.9 in.) Remote indicator: 12.3 x 19.4 x 10.7 cm (4.8 x 11.6 x 4.2 in.) | Different. Predicate has 2 enclosures. Chlorine Sentinel II has the components of both predicate devices in 1 enclosure. | 6-8 {12} K243898 Page 9 of 11 | Weight | Controller: 6.5 kg (14.3 lbs.) Remote Module: 0.25 kg (0.5 lb.) | Controller: 6.5 kg (14.3 lbs.) Probe/Tester: 3 kg (6.6 lbs.) Remote Module: 1 kg (2.2 lbs.) | Analyzer: 8.9 kg (19.6 lbs.) Remote Indicator: 1.1 kg (2.5 lbs.) | Substantially equivalent to Predicate except only 1 enclosure | | --- | --- | --- | --- | --- | | Materials | ABS Plastic enclosure, PVC, Stainless steel, aluminum, polyethylene, electronic components, 3-D printed material: PLA | Controller: PVC and stainless-steel monitor, ABS plastic enclosure Probe/Tester: ABS plastic enclosure Remote Module: ABS plastic enclosure | Analyzer: polycarbonate/ABS plastic case, polycarbonate door, polycarbonate hinges and latches, 316 SST (stainless steel) hardware Remote indicator: polycarbonate/ABS plastic | Same as Predicate | | Sample Flow Rate | 250/350 mL/min minimum | 250 mL/min minimum | 250 mL/min minimum | Same as Predicate | | Sampling Schedule | Site set between 00:00 and 23:59 for M-W-F, T-T-S, and Sunday | NA | Continuous if not manually put into “Idle Mode” and manually taken out of “Idle Mode” | Different. Not available on Predicate | | Sampling Duration | 1-2 Minute | Continuous 24/7 | 5 Minutes | Different. Not available on Predicate | | Sample Pressure | 270 kPa – 530 kPa (40 psi – 80 psi) | 270 kPa – 530 kPa (40 psi – 80 psi) | 2.76 bar – 6.89 bar (40 psi – 100 psi) nominal; pressure spikes 8.27 bar (120 psi) or less | Same as Predicate | | Sample Temperature | 15 – 40° (60 – 105°F) | 15 – 40°C (60 – 105°F) | 15 – 30°C (60 – 85°F) | Same as Predicate | | Sampling Sediment Filter | City water cartridge sediment Filter: 5” x 2.5” x 1-μm | | Requires less than 15 μm | Same as Predicate | | Includes water Shut-off Device | Yes | No | No | Different. Customer requested assessory | | Operating Temperature | 15 – 40°C (60 – 105°F) | 15 – 40°C (60 – 105°F) | 15 – 30°C (60 – 85°F) | Same as Predicate | | Operating Humidity | 35 – 85% RH | 35 - 85% RH | 5 – 90% non-condensing at 30°C (85°F) maximum | Same as Predicate | 6-9 {13} K243898 Page 10 of 11 # Testing The following testing has been performed for Chlorine Sentinel II: ## Electrical Safety / Electromagnetic Compatibility Testing was performed in accordance with the following standards: - IEC 61010-1 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements - IEC 60601-1-2:2020 (Ed. 4.1) Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests ## Shelf-Life Shelf-life testing has been performed for the ammonium sulfate reagent used in the predicate device. ## Performance – Bench Bench Testing has been performed to verify the performance of the Chlorine Sentinel II, which includes: - Linearity Study to CLSI standard EP06-A - Limit of Blank, Limit of Quantitation, Limit of Detection Study to CLSI standard EP17-A2 ## Software Software validation has been performed for the software component of the Chlorine Sentinel II. ## Substantial Equivalence Conclusions In conclusion, the Chlorine Sentinel II indications for use are the same as the predicate device Nelson Sentinel Series Basic Chlorine Sentinel, related to the detection of Chlorine in Water and testing water used to prepare dialysate. The performance characteristics and testing demonstrate that the Chlorine Sentinel II are substantially equivalent to the predicate device the Nelson Sentinel Series Basic Chlorine Sentinel. The technological characteristics for the detection of chlorine are the same as the predicate device, the Nelson Sentinel Series Basic Chlorine Sentinel. It is different in that the mechanical push buttons and switches are replaced with a PLC and color touch screen. In this aspect it is closer the Hach CM130 Chlorine Monitor. As both the predicate device and the reference device are FDA certified, the Chlorine Sentinel II is as safe and effective as they are. {14} K243898 Page 11 of 11 # Conclusion The 510(k) Pre-market Notification for Chlorine Sentinel II contains adequate information and data to determine that Chlorine Sentinel II is as safe and effective as the legally marketed predicate device. 6-11
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...