K243843 · Kerecis Limited · OWY · Jun 4, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243843
Device Name
Tendon Protect (50242)
Applicant
Kerecis Limited
Product Code
OWY · General, Plastic Surgery
Decision Date
Jun 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Kerecis® Tendon Protect™ is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
Device consists of solid North Atlantic Cod fish skin sheets (≥0.4 mm thick). Applied by surgeons during tendon repair procedures to cover the repaired tendon. Acts as a supportive scaffold for host tissue deposition and provides a barrier to facilitate sliding of the tendon within surrounding tissue. Biomechanical strength is provided by primary surgical sutures or bone anchors; device is not intended to provide primary structural repair. Fully bioresorbable and integrates into host tissue over time. Supplied sterile for single-use. Benefits include protection of the repair site and support for natural healing processes.
Clinical Evidence
No human clinical data provided. Evidence includes extensive biocompatibility testing (ISO 10993 series), bench testing (ball burst, suture retention, tensile strength per FDA surgical mesh guidance), and a GLP animal study (chicken flexor tendon model). Animal study compared subject device to predicate (K053655) and sham control across early, mid-term, and late time points, evaluating histopathology, macroscopic appearance, and biomechanical function; results met all acceptance criteria.
Technological Characteristics
Material: North Atlantic Cod fish skin (non-crosslinked, bioresorbable). Form factor: Solid rectangles (3x5 cm, 6x9 cm). Sterilization: Terminal sterilization. Shelf life: 3 years. Biocompatibility: ISO 10993 compliant. Mechanical properties: Tensile strength, suture retention, and ball burst strength tested per FDA surgical mesh guidance.
Indications for Use
Indicated for management and protection of tendon injuries without substantial loss of tendon tissue.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 4, 2025
Kerecis Limited
Skuli Magnusson
Vice President Regulatory Affairs, Quality & Clinical Affairs
Sundstraeti 38, 400 Isafjordur
P.O. Box 151
Isafjordur, 400
Iceland
Re: K243843
Trade/Device Name: Tendon Protect (50242)
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: OWY, FTM
Dated: April 28, 2025
Received: April 28, 2025
Dear Skuli Magnusson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243843 - Skuli Magnusson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243843 - Skuli Magnusson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K243843
Device Name
Tendon Protect (50242)
Indications for Use (Describe)
Kerecis Tendon Protect is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
K243843 - Page 1 of 1
{4}
Kerecis
# K243843
# Kerecis Tendon Protect 510(k) Summary
Premarket Notification Submission (510(k) Summary)
## Submitter Information
Sponsor Name: Kerecis Limited
Sponsor Address: Eyrargata 2 – PO Box 151, 400 Isafjordur, Iceland
Sponsor Telephone: +011 354-419-8000
Establishment Registration: 301060025
Applicant Name and Contact Person: Skuli Magnusson
Title: VP of Regulatory Affairs, QA & Clinical Affairs
Email Direct: sm@kerecis.com
Date Summary Prepared: May 29, 2025
## Device Information
Trade Name (Proprietary): Kerecis® Tendon Protect
Common (Usual): Surgical mesh
Classification Name: Mesh, Surgical, Absorbable, Orthopaedics, Reinforcement of Tendon (Primary); Mesh, Surgical
Device Code: OWY (Primary), FTM
Device Classification: Class II
Regulation Number: 21 CFR 878.3300
## Primary Predicate Device
Company Name: Integra Lifesciences Corporation
Device Name (Proprietary): Tendon Wrap Tendon Protector
Device 510(k): K053655
## Additional Predicate Device
Company Name: Kerecis Limited
Device Name (Proprietary): Kerecis® Reconstruct™
Device 510(k): K202430
## Reference Device
Company Name: Kerecis Limited
Device Name (Proprietary): MariGen Wound Dressing
Device 510(k): K132343
K243843 - Page 1 of 4
{5}
KERECIS
# Device Description
Kerecis® Tendon Protect is a medical device made from solid fish skin (North Atlantic Cod) sheets of thickness no less than 0.4 mm. The product is intended to provide a covering around the newly repaired tendon for the management and protection of tendon injuries in which there has been no substantial loss of tissue. The device with natural structure and proteins from fish skin is applied around the newly repaired tendon. The structure of the fish skin provides a supportive environment for new host tissue deposition as part of wound healing while still providing an initial barrier facilitating the sliding of the repaired tendon within the surrounding tissue during tendon use. The biomechanical strength for the tendon itself is provided separately by sutures used to repair the tendon tear and/or by sutures or bone anchors used to attach the tendon tissue to the bone.
The fish skin is fully integrated into the surrounding tissue over time, with corresponding new host tissue deposition. The device is biocompatible, non-crosslinked, bioresorbable, strong, and pliable. Its tensile strength supports fixation by sutures customary for tendon protection surgical procedures. The device is intended for one-time use and supplied sterile in labelled Tyvek® pouch packaging.
The subject device is available in following sizes
- 3x5 cm size
- 6x9 cm size
Kerecis® Tendon Protect is a standalone, single use medical device made from fish skin that is technologically similar to the predicate device Tendon Wrap Tendon Protector (K053655) (Substantial Equivalence) and identical to the Kerecis® Reconstruct™ (K202430).
# Intended Use / Indication For Use
Kerecis® Tendon Protect™ is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
# Technological Characteristics
Kerecis® Tendon Protect™ and its predicates are discussed in the table below.
| | Tendon Protect (subject device) | Tendon Wrap Tendon Protector (primary predicate device) | Kerecis Reconstruct (additional predicate device) | Discussion |
| --- | --- | --- | --- | --- |
| Manufacturer | Kerecis | Integra Life Sciences | Kerecis | No discussion necessary |
| 510k number | K243843 | K053655 | K202430 | No discussion necessary |
| Indications for use | Kerecis® Tendon Protect™ is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. | Tendon Wrap Tendon Protector is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. | For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery. | Identical to primary predicate |
K243843 - Page 2 of 4
{6}
KERECIS
| | | | | |
| --- | --- | --- | --- | --- |
| Material source/ Chemical composition | Fish skin (North Atlantic Cod) | Bovine / Porous matrix of cross-linked bovine Type I collagen and glycosaminoglycan | Fish skin (North Atlantic Cod) | Identical to additional predicate |
| Design | Solid rectangles | Solid rectangles | Solid rectangles | Identical to primary and additional predicate |
| Sterilization | Terminally sterilized | Terminally sterilized | Terminally sterilized | Identical to primary and additional predicate |
| Shelf life | 3 years | 2 years | 3 years | Identical to additional predicate |
| Dimensions | 3 x 5 cm
6 x 9 cm | 5 x 5 cm
10 x 12.5 cm | 4x7 cm
7x10 cm
7x20 cm | Size range similar to the primary and additional predicate |
## Non-Clinical Performance Testing
The Kerecis Tendon Protect device was subjected to a range of biological and physical tests demonstrating physical and tensile characteristics. All of the tests below were conducted and passed successfully according to the test specifications:
## Biocompatibility Testing:
- Cytotoxicity (ISO 10993-5)
- Sensitization (10993-10)
- Irritation or intracutaneous reactivity (ISO 10993-10)
- Material mediated pyrogenicity (ISO 10993-11)
- Implantation Effects (ISO 10993-6)
- Genotoxicity (ISO 10993-3)
- Acute systemic toxicity (ISO 10993-11)
- Subacute/subchronic toxicity (ISO 10993-11)
## Bench Testing:
- Tendon Summative Usability Evaluation. As part of the product development process, the Kerecis Tendon product was subjected to usability engineering activities to evaluate its safety and effectiveness as a medical device per ISO 62366-2. Through these activities and analyses, it was determined that the Kerecis Tendon product is safe and effective for use as it met all the pre-defined acceptance criteria (detailed below).
- To support the performance requirements of a surgical mesh, testing was completed per FDA Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh. Ball burst testing, suture retention testing and tensile strength testing was performed. The results met the
K243843 - Page 3 of 4
{7}
KERECIS
acceptance criteria.
## Animal Testing:
### Tendon GLP Chicken Study.
Testing was performed to evaluate the Kerecis Tendon Protect in a chronic chicken flexor tendon model that evaluated an early, mid-term and late time point corresponding to collagen mesh materials. The performance of the subject device was compared to both a control (K053655) and a negative control (sham surgery group). The primary and secondary endpoints included histopathology evaluation, macroscopic evaluations, tendon biomechanical function assessment and overall animal health and the study demonstrated that acceptance criteria were met. The testing demonstrated that the subject device, with different technological characteristics (fish skin derived), exhibited similar overall performance to the marketed predicate device in terms of tendon healing.
## Conclusion
Kerecis® Tendon Protect is safe and effective under the proposed conditions of use, and substantially equivalent to its predicate devices. Kerecis® Tendon Protect has similar composition, indications for use, intended use, and technological characteristics as compared to the predicate device. The predicate device Integra Tendon Wrap device (K053655) has been legally marketed, in sizes from 2-inch x 2-inch to 5 inch and has been well-received in the wound management market. There are no significant differences in the safety and effectiveness between the subject device and the predicate device. Safety and efficacy are supported through usability testing, animal testing, biocompatibility, and physical properties tested.
K243843 - Page 4 of 4
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.