MARIGEN WOUND DRESSING

K132343 · Kerecis Limited · KGN · Oct 23, 2013 · SU

Device Facts

Record IDK132343
Device NameMARIGEN WOUND DRESSING
ApplicantKerecis Limited
Product CodeKGN · SU
Decision DateOct 23, 2013
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K132343 · Oct 23, 2013MARIGEN WOUND DRESSINGKerecis LimitedCommercial clinical use dataThe sponsor utilized data from limited commercial clinical use (post-CE mark) to support the safety profile of the MariGen Wound Dressing.Commercial use; Clinical experience; Safety monitoring

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Limited clinical use; Observational / Commercial clinical experience; Study Period: 2011-2013Patients with wounds; Sample Size: Limited number of subjectsNot applicable for this studySafety / Adverse events

Indications for Use

MariGen Wound Dressing is indicated for the management of wounds including: Partial- and full-thickness wounds Pressure ulcers venous ulcers Chronic vascular ulcers Diabetic ulcers Trauma wounds (abrasions, lacerations, second-degree burns, skin tears) Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), and Draining wounds.

Device Story

MariGen Wound Dressing is a processed fish dermal matrix composed of fish collagen. Supplied as sterile intact or meshed sheets. Used for management of various wound types including ulcers, trauma, and surgical wounds. Applied by clinicians to the wound site to support healing. Device acts as a collagen-based scaffold. Benefits include wound management and tissue support. Single-patient, one-time use.

Clinical Evidence

Bench testing included biocompatibility per ISO 10993-12 (cytotoxicity, sensitization, irritation, subchronic toxicity, genotoxicity) with favorable results. Animal studies performed to support safety. Limited clinical use data from CE-marked commercial availability since 2011 reported no serious adverse events.

Technological Characteristics

Processed fish dermal matrix composed of fish collagen. Supplied as sterile intact or meshed sheets in sizes 3x3.5 cm, 3x7 cm, and 7x10 cm. Biocompatibility tested per ISO 10993-12.

Indications for Use

Indicated for management of partial/full-thickness wounds, pressure ulcers, venous/chronic vascular ulcers, diabetic ulcers, trauma wounds (abrasions, lacerations, second-degree burns, skin tears), surgical wounds (donor sites/grafts, post-Mohs, post-laser, podiatric, dehiscence), and draining wounds in patients requiring wound care.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2 3 2013 # 510k Summary #### SUBMITTER/510K HOLDER 1 | Address: | Kerecis Itd.<br>Eyrargata 2<br>400 Isafjordur<br>Iceland | |-----------------|----------------------------------------------------------| | Contact Person: | Gudmundur Fertram Sigurjonsson<br>Executive Chairman | | Telephone: | 011 354 562 2601 | | Date Prepared: | October 1, 2013 | #### 2 DEVICE NAME | Proprietary Name: | MariGen Wound Dressing | |----------------------|-------------------------| | Common/Usual Name: | MariGen Wound | | Classification Name: | Collagen Wound Dressing | | 510k Number: | K132343 | #### PREDICATE DEVICES 3 - . Mesynthes Endoform™ Dermal Template (K092096) - ACell MatriStem™ (K092926) . - Integra™ Wound Matrix (Integra Life Sciences) (K021792) . - . LifeCell Corporation's LTM Wound Dressing (K082103) - . TEl Biosciences Inc. PriMatrix (K083440) - . HemCon Chitoflex Surgical Dressing (K071519) - Cook Biotech, Oasis (K061711) . Traditional 510k, K132343 {1}------------------------------------------------ #### DEVICE DESCRIPTION 4 The Kerecis MariGen Wound Dressing is processed fish dermal matrix composed of fish collagen and is supplied as a sterile intact, or meshed sheet ranging in size from 3 x 3.5 cm (10.5 cm2), 3 x 7 cm (21 cm2) and 7 x 10 cm (70 cm2). The device is intended for a single patient, one time use only. #### INTENDED USE 5 MariGen Wound is indicated for the management of wounds including: - . Partial and full-thickness wounds - . Pressure ulcers - . Venous ulcers - . Chronic vascular ulcers - . Diabetic ulcers - . Trauma wounds (abrasions, lacerations, second-degree burns, skin tears) - . Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence). - . Draining wounds ### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL 6 EQUIVALENCE The Kerecis MariGen Wound Dressing is substantially equivalent to the predicate devices with respect to material composition, device characteristics and intended use. #### MECHANICAL TESTS AND BIOCOMPATIBILITY TESTS 7 The Kerecis MariGen Wound Dressing has been subjected to extensive testing to assess the biocompatibility and the performance of the device. The Kerecis MariGen Wound Dressing was shown to be usable and biocompatible. {2}------------------------------------------------ All biocompatibility tests that are required according to ISO 10993-12 "Biological Evaluation of Medical Devices" where performed with favorable results. Tests were done in the following categories; cytotoxicity, sensitization, irritation, subchronic toxicity and genotoxicity. #### ANIMAL TESTING 8 Several animal studies have been performed on the device. The results of the tests support the safety of the device. #### CLINICAL TESTING ਰੇ MariGen Wound is CE marked and has been in commercial use since 2011 The product has been studied on a limited number of subjects and this use supports the safety of the device. No serious adverse events have been noted in limited clinical use. ### 10 CONCLUSIONS The proposed and predicate devices have the same general intended use and principles of operation. The overall design of the proposed and predicate devices is similar in that they are all comprised of collagen. The differences between these devices are limited to the animal source of collagen. This difference is minor and does not raise new issues of safety or effectiveness. The MariGen Wound Dressing meets all defined acceptance criteria for the non-clinical testing required by the risk analysis. The design validation activities confirmed that the device as designed functions to meet specified user requirements. The information provided confirms that the MariGen Wound Dressing device is safe and effective for its intended use and performs as intended. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the Department of Health and Human Services (USA). The seal features an eagle-like bird with three stylized wing feathers, representing health, services, and people. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" which is arranged in a circular fashion around the bird. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WOop-G609 Silver Spring MD 20993-0002 October 23, 2013 Kerecis LTD. Gudmundur F. Sigurjonsson Executive Chairman Eyrargata 2 400 Isafjordur Iceland Re: K132343 Trade/Device Name: MeriGen Wound Dressing Regulatory Class: Unclassified Product Code: KGN Dated: July 19. 2013 Received: August 7, 2013 Dear Sigurionsson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA`s issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that 1197 has Intacted and regulations administered by other Federal agencies. You must or any with all the Act's requirements, including, but not limited to: registration and listing (2 l CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ ### Page 2 - Gudmundur F. Sigurjonsson forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please. contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notilication" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.lda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincerely yours. ## Mark N. Melkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ### 510(k) Number (if known) K132343 #### Device Name MariGen Wound Dressing Indications for Use (Describe) MariGen Wound Dressing is indicated for the management of wounds including: - Partial- and full-thickness wounds - Pressure ulcers - venous ulcers - Chronic vascular ulcers - Diabetic ulcers - Trauma wounds (abrasions, lacerations, second-degree burns, skin tears) - Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), and - Draining wounds. Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...