2290 clinically acquired images and frames from video clips from 579 fetal ultrasound examinations.
>1 (sonographer) + >1 (MFM specialist)
Indications for Use
The BrightHeart View Classifier device is intended to analyze fetal ultrasound images and video clips using machine learning techniques to automatically detect standard views during fetal heart scanning. The BrightHeart View Classifier device is intended to be used as an adjunct to the acquisition and interpretation of fetal anatomic ultrasound examinations at the second or third trimester of pregnancy performed with transabdominal probes.
Device Story
Cloud-based SaMD using AI to detect standard views during fetal heart scanning; inputs 2D grayscale ultrasound images and video clips; transforms inputs via neural network to classify views, extract example frames, and assess protocol compliance; used in professional prenatal ultrasound environments by sonographers, MFMs, OB/GYNs, and fetal surgeons; output provided as annotated DICOM files; assists clinicians in acquisition and interpretation of fetal anatomic exams; benefits include standardized documentation and improved efficiency in identifying fetal heart views.
Clinical Evidence
Bench testing only. Validated on 2,290 images/frames from 579 fetal ultrasound exams (18+ weeks gestation) across 8 centers. Reference standard derived from independent truthing by sonographer and MFM specialist. Sensitivity: 0.939 (95% CI: 0.917-0.960); Specificity: 0.984 (95% CI: 0.973-0.996). Performance consistent across subgroups; limited data for Asian/Black mothers noted.
Technological Characteristics
Cloud-based SaMD; neural network architecture; inputs 2D grayscale ultrasound (GE, Philips, Samsung, Siemens); output annotated DICOM files. Software life cycle per IEC 62304:2016. Includes Predetermined Change Control Plan (PCCP).
Indications for Use
Indicated for pregnant women at 18 weeks gestation or later undergoing fetal anatomic ultrasound examinations using transabdominal probes. Intended for use by qualified healthcare professionals including sonographers, MFMs, OB/GYNs, and fetal surgeons.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
HeartAssist feature in the Samsung V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System (K223387)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
BrightHeart
Christophe Gardella
CTO
7-11 boulevard Haussmann
Paris, 75009
France
May 7, 2025
Re: K243684
Trade/Device Name: BrightHeart View Classifier
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management and Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: March 25, 2025
Received: March 25, 2025
Dear Christophe Gardella:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243684 - Christophe Gardella
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device
{2}
K243684 - Christophe Gardella
Page 3
Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, PhD
Assistant Director,
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243684 | |
| Device Name BrightHeart View Classifier | |
| Indications for Use (Describe) The BrightHeart View Classifier device is intended to analyze fetal 2D ultrasound images and video clips using machine learning techniques to automatically detect standard views during fetal heart scanning. The BrightHeart View Classifier device is intended to be used as an adjunct to the acquisition and interpretation of fetal anatomic ultrasound examinations at the second or third trimester of pregnancy performed with transabdominal probes. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary - K243684
Page 1 of 9
# 1. SUBMITTER
| Applicant: | BrightHeart
7-11 boulevard Haussmann
Paris 75009, France |
| --- | --- |
| | |
| Contact: | Christophe Gardella
Chief Technical Officer
Tel. +0033686543950
Email. christophe@brightheart.fr |
| | |
| Submission Correspondent: | Christophe Gardella |
| | |
| Date Prepared: | May 5, 2025 |
# 2. DEVICE
| Device Trade Name: | BrightHeart View Classifier |
| --- | --- |
| Device Common Name: | Fetal ultrasound view classifier |
| Classification Name | 21 CFR §892.2050, Medical image management and processing system |
| Product Code(s): | QIH |
| Regulatory Class: | Class II |
# 3. PREDICATE DEVICE
Predicate Device: Sonix Health [K230209]
Reference Device: HeartAssist feature in the Samsung V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System [K223387]
# 4. DEVICE DESCRIPTION
BrightHeart View Classifier is a cloud-based software-only device which uses artificial intelligence (AI) to detect standard views during fetal heart scanning in fetal ultrasound images and video clips.
BrightHeart View Classifier is intended to be used by qualified, trained healthcare professional personnel in a professional prenatal ultrasound (US) imaging environment (this includes sonographers, MFMs, OB/GYN, and Fetal surgeons), to help fetal ultrasound examination acquisition and interpretation of 2D grayscale ultrasound by providing automatic classification of video clips and images into standard views, by automatically extracting example frames of standard views from video clips, and by automatically assessing whether the documentation of
{5}
510(k) Summary - K243684
Page 2 of 9
each standard view in video clips and images satisfies an acquisition protocol defined by the center. Annotated DICOM files generated by the device cannot be modified by the user.
## 5. INTENDED USE/INDICATIONS FOR USE
The BrightHeart View Classifier device is intended to analyze fetal ultrasound images and video clips using machine learning techniques to automatically detect standard views during fetal heart scanning.
The BrightHeart View Classifier device is intended to be used as an adjunct to the acquisition and interpretation of fetal anatomic ultrasound examinations at the second or third trimester of pregnancy performed with transabdominal probes.
## 6. SUBSTANTIAL EQUIVALENCE
### Technological Comparisons
The table below compares the key technological features of the subject device to the predicate device, Sonix Health.
Technological Comparison:
| | Subject Device
BrightHeart View
Classifier | Predicate Device
Sonix Health | Reference Device
HeartAssist feature in
the Samsung V8
Diagnostic
Ultrasound System,
V7 Diagnostic
Ultrasound System. |
| --- | --- | --- | --- |
| 510(k) Number | TBD | K230209 | K223387 |
| Applicant | BrightHeart | Ontact Health | Samsung |
| Classification
Regulation | 21 CFR §892.2050 | 21 CFR §892.2050 | 21 CFR §892.2150
21 CFR §892.1560
21 CFR §892.1570 |
| Product Code | QIH | QIH, LLZ | IYN, IYO, ITX |
| Device Type | SaMD | SaMD | SaMD |
| Software algorithm | Machine Learning
Model | Machine Learning
Model | Machine Learning
Model |
{6}
510(k) Summary - K243684
Page 3 of 9
| | Subject Device
BrightHeart View Classifier | Predicate Device
Sonix Health | Reference Device
HeartAssist feature in the Samsung V8 Diagnostic Ultrasound System, V7 Diagnostic Ultrasound System. |
| --- | --- | --- | --- |
| **Imaging Modality** | Fetal Ultrasound | Adult Echocardiography | The Samsung V8 and V7 are general purpose, mobile, software controlled, diagnostic ultrasound systems, including fetal ultrasound. |
| **Model Inputs** | Ultrasound images and video clips | Ultrasound images and video clips | Ultrasound images |
| **Model method** | Neural networks | Neural networks | Neural networks |
| **Model trained to identify** | Classification of ultrasound images into standard views of fetal heart. | Classification of ultrasound images into standard views of adult heart, determination cardiac measurements. | Classification of ultrasound images into standard views of fetal heart scanning, and detection of calipers used for measurements |
| **Model Output** | Identifies standard views of the heart and abdomen in images and video clips. | Identifies standard views of the heart in images and video clips, detection of cardiac measurements. | Identifies standard views of the heart and abdomen in images |
| **PCCP** | Included | Not included | Not included |
The BrightHeart View Classifier and Sonix Health devices differ in that the BrightHeart View Classifier device uses as input fetal ultrasounds, whereas the Sonix Health device uses as input adult echocardiograms. The PCCP in the subject device includes proposed modifications related to modification of training and/or validation datasets, modification of data input sources regarding ultrasound machine make, addition of standard view, and modification of data input sources regarding gestational age.
The subject BrightHeart View Classifier and the predicate Sonix Health devices are both intended to be used to analyze ultrasound images to determine standard views during heart scanning. Both the subject and the predicate are Software as a Medical device (SaMD) applications that rely on a trained neural network to identify the standard views. The Sonix Health device also includes a measurement functionality which is not included in the BrightHeart
{7}
510(k) Summary - K243684
Page 4 of 9
device. Both devices are intended to be used as an adjunct to professional decision-making by trained healthcare providers.
Similarities of the device with the predicate are maintained following PCCP modifications, and PCCP modifications do not incur additional differences with the predicate device.
In summary, the subject and predicate devices share the same fundamental technological characteristics, namely the use of software to assist healthcare professionals who are interpreting ultrasound images of the heart to determine standard views. The differences in implementations described above do not raise different questions of safety and effectiveness, so the Sonix Health device can be used as a predicate device for the BrightHeart View Classifier device.
## 7. PERFORMANCE DATA
### Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." and in accordance with IEC 62304:2016, Medical device software - Software life cycle processes.
Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
### Bench Testing
The following bench testing was performed to demonstrate substantial equivalence:
The device performance for standard view classification was validated with a dataset of 2290 clinically acquired images and frames from video clips from 579 fetal ultrasound examinations, performed with transabdominal probes at 18 weeks of gestation or later, from 8 centers.
**Demographic distribution:**
- gender: female
- age: 18 years or older
- ethnicity/country: U.S.A. and France.
**Information about how the reference standard was derived from the dataset:**
- The reference standard was derived from the dataset through a truthing process in which a sonographer and an MFM specialist with experience in fetal echocardiography determined the presence or absence of standard views on fetal ultrasound images.
- The truthing process was conducted independently of the BrightHeart View Classifier device.
**Description of how the independence of test data from training data was ensured:**
- The ultrasound examinations used for training and validation are entirely distinct from the examinations used in performance testing.
{8}
510(k) Summary - K243684
Page 5 of 9
The performance testing demonstrated that BrightHeart View Classifier identifies standard views with a mean standard view recognition sensitivity of 0.939 (95% CI, 0.917 ; 0.960) and a mean standard view recognition specificity of 0.984 (95% CI, 0.973 ; 0.996).
Stratified analysis by geographical region, by ultrasound machine make, by gestational age, by mother's BMI, and age, and by record type indicated that performance was consistent across subgroups, while analysis by mother's race showed that the specificity 95% CI lower bound was slightly lower for Asian and Black mothers, possibly due to large confidence intervals and small sample size.
BrightHeart View Classifier was validated only with General Electric, Philips, Samsung and Siemens ultrasound devices and is intended only to be used with these ultrasound vendors.
BrightHeart View Classifier was validated with pregnancies at 18 weeks of gestation or later recorded with transabdominal probes and is intended to be used only for such examinations.
## 8. PREDETERMINED CHANGE CONTROL PLAN
The BrightHeart View Classifier device is powered by a neural network performing standard view classification in fetal ultrasound images.
Modifications to the BrightHeart View Classifier device will be made in accordance with its Predetermined Change Control Plan (PCCP). The PCCP provides a description of the device's planned modifications, as well as procedures to implement, evaluate and deploy the modifications. These procedures ensure the continued safety and effectiveness of the device, mitigating risks associated with modifications to the BrightHeart View Classifier neural network to not adversely impact the device's performance, safety, or effectiveness associated with its indications for use, and an impact assessment of the planned modifications.
A summary of the planned modifications is provided in the table below. In line with the PCCP, all changes to the BrightHeart View Classifier will undergo development, evaluation and validation before release. Information on the deployment and post market surveillance of the algorithm are also provided for the proposed changes. Implemented modifications to the BrightHeart View Classifier will be communicated to users by email with a release note detailing all expected changes in the new release and user manual.
Summary of changes to BrightHeart View Classifier per the PCCP:
| Modification | Rationale | Testing Methods | Impact Assessment |
| --- | --- | --- | --- |
| BrightHeart View Classifier | High performance, high performance, low cost, low cost | PCCP | High |
| BrightHeart View Classifier | High performance, high cost, low cost | PCCP | High |
{9}
510(k) Summary - K243684
Page 6 of 9
| Modification of training and/or validation datasets to update model weights and detection thresholds | Improvements in the view classification performance of the device. | Re-training of the BrightHeart View Classifier with new data to optimize its performance followed by performance testing and a comparison of the original BrightHeart View Classifier to the modified BrightHeart View Classifier (using performance metrics) and verification and validation. | Improved performance metrics of modified BrightHeart View Classifier
**Benefit-Risk Analysis:**
Benefits: Improved performance; generalization for diverse cases.
Risks: Overfitting; unintended bias.
**Risk Mitigation:**
Testing data sequestration and testing on new data will ensure proper evaluation and mitigate risks of overfitting. |
| --- | --- | --- | --- |
{10}
510(k) Summary - K243684
Page 7 of 9
| Modification of data input sources: ultrasound machine make (Canon, Fujifilm, SIUI) | Improvement in the view classification performance with ultrasound machine makes not previously validated | Re-training of the BrightHeart View Classifier with data from new ultrasound machine makes to optimize its performance followed by performance testing and a comparison of the original BrightHeart View Classifier to the modified BrightHeart View Classifier (using performance metrics) and verification and validation. | Improved performance metrics of modified BrightHeart View Classifier on new ultrasound make
**Benefit-Risk Analysis:**
Benefits: Improved performance; generalization for diverse cases.
Risks: Overfitting; unintended bias.
**Risk Mitigation:**
Testing data sequestration and testing on new data will ensure proper evaluation and mitigate risks of overfitting. |
| --- | --- | --- | --- |
{11}
510(k) Summary - K243684
Page 8 of 9
| Addition of standard views (short-axis view at the level of the ventricles and short-axis view at the level of the atrioventricular valves) | Training of the machine learning model to achieve sufficient view classification performance for additional standard views, which will assist users in the identification of these views during fetal heart ultrasound scanning. | Re-training of the BrightHeart View Classifier with data from new standard views to optimize its performance followed by performance testing and a comparison of the original BrightHeart View Classifier to the modified BrightHeart View Classifier (using performance metrics) and verification and validation. | Improved performance metrics of modified BrightHeart View Classifier on additional standard views
**Benefit-Risk Analysis:**
Benefits: Improved performance; generalization for diverse cases.
Risks: Overfitting; unintended bias.
**Risk Mitigation:**
Testing data sequestration and testing on new data will ensure proper evaluation and mitigate risks of overfitting. |
| --- | --- | --- | --- |
{12}
510(k) Summary - K243684
Page 9 of 9
| Modification of data input sources: gestational age | Enables the device to achieve sufficient view classification performance to assist users in the identification of standard views for examinations corresponding to gestational ages not previously validated. | Re-training of the BrightHeart View Classifier with data from new gestational ages to optimize its performance followed by performance testing and a comparison of the original BrightHeart View Classifier to the modified BrightHeart View Classifier (using performance metrics) and verification and validation. | Improved performance metrics of modified BrightHeart View Classifier on new gestational age
**Benefit-Risk Analysis:**
Benefits: Improved performance; generalization for diverse cases.
Risks: Overfitting; unintended bias.
**Risk Mitigation:**
Testing data sequestration and testing on new data will ensure proper evaluation and mitigate risks of overfitting. |
| --- | --- | --- | --- |
## 9. CONCLUSION
The results of the testing described above demonstrate that the BrightHeart View Classifier is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.