Standalone performance testing: 17,885 fetal ultrasound images from 7 clinical sites in the United States, France and Germany
—
Indications for Use
Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
Device Story
Sonio Detect is a SaaS-based concurrent reading aid for fetal ultrasound examinations. It processes fetal ultrasound images and clips (Trimesters 1-3) acquired from GE, Samsung, or Canon ultrasound systems. An edge software application installed on a local network server receives DICOM instances from the ultrasound machine and uploads them to the cloud. The device uses AI, biometric analysis, and colorimetry to automatically detect specific fetal views, identify anatomical structures, and verify image quality criteria (e.g., presence of structures, zoom levels). The system highlights unverified items in yellow for the clinician. Clinicians review, edit, or override the software's outputs in real-time or post-exam. By ensuring protocol compliance and image completeness, the device assists healthcare professionals in performing routine fetal ultrasound exams, potentially improving diagnostic consistency and workflow efficiency.
Clinical Evidence
Bench testing only. No clinical studies were conducted. Performance was validated on an independent dataset of 17,885 fetal ultrasound images from 7 clinical sites. Results showed high sensitivity for 3D (0.980) and Doppler (0.963) image detection, and high sensitivity/specificity for anatomical structure detection and quality criteria verification (e.g., brain structures sensitivity 0.857, specificity 0.963; heart structures sensitivity 0.900, specificity 0.982). Subgroup analysis included ultrasound manufacturer, BMI, maternal age, gestational age, and ethnicity.
Technological Characteristics
SaaS cloud-based software with local edge server component. Compatible with GE, Samsung, and Canon ultrasound systems. Uses AI, biometric analysis, and colorimetry for image processing. Software verification and validation performed per FDA guidance. Connectivity via DICOM. Intended for professional prenatal ultrasound environments.
Indications for Use
Indicated for use by qualified healthcare professionals (sonographers, MFMs, OB/GYNs, fetal surgeons) in professional prenatal ultrasound environments to assist in the acquisition and interpretation of fetal ultrasound images during trimesters 1, 2, and 3 (gestational age 11-37 weeks).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SonoLyst feature embedded in the GE Medical device Voluson SWIFT (K201828)
Submission Summary (Full Text)
{0}------------------------------------------------
July 25, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sonio % Florian Akpakpa Head of Regulatory Affairs and Quality Assurance 24 Rue du Faubourg Saint Jacques Paris. FR-75014 FRANCE
### Re: K230365
Trade/Device Name: Sonio Detect Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, QIH Dated: June 26, 2023 Received: June 27, 2023
### Dear Florian Akpakpa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
| Submission Number (if known) | K230365 |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sonio Detect |
| Indications for Use (Describe) | Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.<br>The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images. |
| Type of Use (Select one or both, as applicable) | <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Sonio. The logo consists of a stylized, abstract symbol on the left, resembling a curved shape with a dot above it. To the right of the symbol is the word "sonio" in lowercase letters. The color of both the symbol and the text is a shade of blue.
# K230365
# 510(k) Summary
In accordance with 21 CFR 807.92 the 510(k) summary for Sonio Detect is provided below.
# I. Submitter
| Applicant: | Sonio<br>24 rue du Faubourg Saint Jacques,<br>75014, Paris France |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Florian Akpakpa<br>Head of Regulatory Affairs and Quality Assurance<br>Sonio<br>Phone: +33 6 19 38 71 45<br>Email: florian.akpakpa@sonio.ai |
| Secondary Contact Person: | Donna-Bea Tillman<br>Senior Consulting<br>Biologics Consulting<br>Phone: +1 (410) 531-6542 - Direct<br>Email: dtillman@biologicsconsulting.com |
| Date Prepared: | July 25, 2023 |
### II. Device
| Device Trade Name: | Sonio Detect |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System<br>21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System<br>21 CFR 892.2050 - Medical image management and processing system |
| Regulatory Class: | Class II |
| Product Codes: | IYN (Primary)<br>IYO, QIH (Secondary) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sonio. The logo consists of a stylized blue icon resembling a sound wave or a stylized letter 'S', followed by the word 'sonio' in a sans-serif font, also in blue. The logo is simple and modern, with a clean design.
# III. Predicate Device
SonoLyst feature embedded in the GE Medical device Voluson SWIFT, K201828.
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
### IV. Device Description
Sonio Detect is a Software as a Service SaaS solution that aims at helping sonographers, OB/GYNs, MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP) to perform their routine fetal ultrasound examinations in real-time. Sonio Detect can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 1, Trimester 2 and Trimester 3 of the fetus (Gestational Age: from 11 weeks to 37 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol.
Sonio Detect requires the following:
- Edge Software (described below) to install on a server on the same network as the ● Ultrasound Machine;
- SaaS accessibility from internet browser.
Sonio's Edge Software is a light-weight application that runs on a server (computer) connected to the same network as the Ultrasound Machine. Sonio Edge Software is installed on the HCP server (computer) and network and the main purpose is to receive DICOM instances from the Ultrasound Machine and upload them to Sonio's Cloud to be used by Sonio Detect.
Sonio Detect receives fetal ultrasound images and clips from the ultrasound machine, that are submitted through the edge software by the performing healthcare professional, in real-time and performs the following:
- . Automatically detect views;
- Automatically detect anatomical structures within the supported views; .
- Automatically verify quality criteria of the supported views by checking whether they . conform to standardized quality criteria.
Quality criteria are related to:
- the presence or absence of an anatomical structure; ●
- the zoom level for some views.
Sonio Detect then automatically associates the image to its detected view. It also highlights in yellow the view and/or the corresponding quality criteria if there are unverified items : quality criteria not verified or view not detected.
The end user can interact with the software to override the Sonio Detect's outputs (reassign the image to another view or unassign it or assign it if it was not assigned, change the status of a
{5}------------------------------------------------
# Sonio
510(k) Premarket Notification Submission
quality criteria from verified to unverified or from unverified to verified). The user has the ability to review and edit/override the matching at any time during or at the end of the exam.
The list of views, anatomical structures and quality criteria that can be automatically detected and verified by the software are detailed in the tables 1, 2 and 3 below.
| Table 1: List of views that can be automatically detected by the software | |
|--------------------------------------------------------------------------------------------------|------------------------------|
| View | Gestational Age of the fetus |
| Transthalamic or Cavum septum Pellucidum or Midline falx /<br>Transventricular or Choroid plexus | T1 and T2/T3 |
| Profile or Nuchal translucency | T1 and T2/T3 |
| Crown rump length | T1 |
| Sagittal spine | T2/T3 |
| Abdominal circumference | T2/T3 |
| Long bone | T2/T3 |
| Transcerebellar view | T2/T3 |
| Upper lip, nose and nostrils | T2/T3 |
| Four chambers | T2/T3 |
| Left ventricular outflow tract | T2/T3 |
| Right ventricular outflow tract | T2/T3 |
| | Table 2:List of anatomical structures that can be automatically detected by the software | | | | |
|--|------------------------------------------------------------------------------------------|--|--|--|--|
| | | | | | |
| View | Structure |
|-------------------------------------|---------------------------|
| Brain views and structures at T2/T3 | |
| Transthalamic view | • Midline falx |
| Transventricular view | • Cavum septum pellucidum |
| Transcerebellar view | • Cerebellum |
| Heart views and structures at T2/T3 | |
| Four Chambers | • Aorta |
| Three vessels | • Apex heart |
| LVOT | • Ascending aorta |
| | • Descending aorta |
| | • Interatrial septum |
| | • Interventricular septum |
| | • Left atrium |
| | • Left ventricle |
| | • Pulmonary trunk |
| | • Right atrium |
| | • Right ventricle |
| | • Superior vena cava |
{6}------------------------------------------------
| Quality criteria of the brain views | |
|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| For the transthalamic view, Sonio<br>Detect automatically evaluates the<br>following criteria: | Presence of the cavum septum pellucidum Absence of the cerebellum Brain occupies more than half of the width of the<br>ultrasound image |
| For the transcerebellar view, Sonio<br>Detect automatically evaluates the<br>following criteria: | Presence of the cerebellum Presence of the cavum septum pellucidum Brain occupies more than half of the width of the<br>ultrasound image |
| For the transventricular view, Sonio<br>Detect automatically evaluates the<br>following criteria: | Presence of the cavum septum pellucidum Brain occupies more than half of the width of the<br>ultrasound image |
| Quality criteria of the heart views | |
| For the 4 chambers view, Sonio<br>Detect automatically evaluates the<br>following criteria: | Presence of the right ventricle Presence of the left ventricle Presence of the left atrium Presence of the right atrium Presence of the interventricular septum Presence of the interauricular septum Presence of the apex of the heart Presence of the descending aorta |
| For the 3 vessels and 3 vessels and<br>trachea views, Sonio Detect<br>automatically evaluates the following<br>criteria: | Presence of the pulmonary trunk Presence of the ascending aorta Presence of the superior vena cava |
| For the LVOT view, Sonio Detect<br>automatically evaluates the following<br>criteria: | Presence of the left ventricle Presence of the aorta Presence of the right ventricle Presence of the left atrium Presence of the apex of the heart |
# Table 3: List of quality criteria that can be automatically verified by the software
{7}------------------------------------------------
### V. Indications for Use
Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.
The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
Sonio Detect and the predicate device have the same intended use for automatic detection of views as well as automatic detection of anatomical structures within the views and verification of quality criteria of the views by comparing them to standardized quality criteria. The Indications for Use statement for Sonio Detect is not identical to the predicate device: however. the difference does not affect the safety and effectiveness of the device relative to the predicate, and so does not constitute a new intended use.
### VI. Comparison of Technological Characteristics with the Predicate Device
Table 4 provides a comparison of the Technological Characteristics of Sonio Detect to the predicate SonoLyst.
| Items | Predicate device 1: SonoLyst feature in<br>Voluson SWIFT - K201828 | Proposed device: Sonio Detect |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer name | GE Medical | Sonio |
| Device name | SonoLyst feature embedded in the device<br>Voluson SWIFT | Sonio Detect |
| Regulation Number | - Ultrasonic Pulsed Doppler Imaging System. 21 CFR 892.1550, 90-IYN;<br>-Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO;<br>-Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX | - Accessory to Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550<br>- Accessory to Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560<br>- Medical image management and processing system, 21 CFR 892.2050 |
| Product codes | IYN (primary)<br>IYO<br>ITX (secondary) | IYN (Primary)<br>IYO, QIH (Secondary) |
| Clinical outcome | - Images labeled with correct view<br><br>- Quality criteria are identified as “found” when detected and “not found” when not detected | - Images labeled with correct view<br><br>- Quality criteria are identified as “Verified” when detected and “Not verified” when not detected |
#### Table 4: Comparison of technological characteristics
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Sonio. The logo consists of a blue abstract symbol on the left and the word "sonio" in blue on the right. The symbol appears to be a stylized representation of a sound wave or a stylized letter S.
#### Sonio 510(k) Premarket Notification Submission
| Items | Predicate device 1: SonoLyst feature in<br>Voluson SWIFT - K201828 | Proposed device: Sonio Detect |
|--------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Users | General purpose radiology evaluation and<br>specialized for OB/GYN | Qualified and trained healthcare professional<br>personnel in a professional prenatal<br>ultrasound (US) imaging environment (this<br>includes sonographers, MFMs, OB/GYN,<br>and Fetal surgeons) |
| Clinical<br>applications | Fetal/Obstetrics | Fetal/Obstetrics |
| Algorithm<br>Methodology | Artificial Intelligence | Artificial Intelligence<br>Lecture of biometrics<br>Colorimetry for 3D and Doppler |
| Platform | Embedded in the ultrasound equipment | Secure cloud-based and stand-alone software<br>compatible with ultrasound system from GE<br>Medical, Samsung and Canon |
Sonio Detect's intended users, clinical outcome and clinical applications are similar to those of the predicate device, SonoLyst.
Sonio Detect differ to SonoLyst in the following:
- Algorithm methodology: Sonio Detect algorithm technology is based on Artificial ● Intelligence, Lecture of biometrics on the image and Colorimetry identification for 3D and Doppler while the Predicate SonoLyst's algorithm technology is only based on Artificial Intelligence:
- Platform: Sonio Detect is a stand-alone cloud based software that can be used with . different ultrasound systems while the Predicate SonoLyst is embedded in the GE ultrasound system.
However, the differences in algorithm methodology and platform do not raise different questions of safety and effectiveness of the device.
### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Design Reviews ●
- Software Development Lifecycle ●
- Algorithm Verification (Algorithm internal validation)
- Software verification ●
- Simulated use testing (Validation) ●
- Performance testing (Verification)
{9}------------------------------------------------
### Bench Testing
Sonio conducted a standalone performance testing on a dataset of 17885 fetal ultrasound images from 7 clinical sites in the United States, France and Germany representative of the intended use population. This dataset was independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points.
The results of the standalone performance testing demonstrate that Sonio Detect:
- Automatically detects 3D fetal ultrasound images with high sensitivity (0.980: 95% . Wilson's Confidence Interval: 0.930, 0.994) and Doppler fetal ultrasound images with high sensitivity (0,963; 95% Wilson's Confidence Interval: 0.908, 0.985);
- . Automatically detects fetal ultrasound views through reading of annotations on images with high proportions of annotations read correctly (0,923; 95% Wilson's Confidence Interval: 0.905, 0.938);
- Automatically detects some T1 fetal ultrasound views with high sensitivity (0,942; ● Point estimate).
- Automatically detects some T2/T3 fetal ultrasound views with high sensitivity (0,919; ● Point estimate).
- Automatically detects some fetal brain anatomical structures in some T2/T3 brain views with high sensitivity (0.857; Point estimate) and high specificity (0.963; Point estimate), and so automatically verifies the corresponding quality criteria;
- Automatically detects some fetal heart anatomical structures in the some T2/T3 heart views with high sensitivity (0,900; Point estimate) and high specificity (0,982; Point estimate) and so automatically verifies the corresponding quality criteria.
- . Automatically verifies the zoom level for some brain views with high sensitivity (0.952; 95% Wilson's Confidence Interval: 0.909-0.976) and high specificity (0.906; 95% Wilson's Confidence Interval: 0.758-0.968).
Additionally the performance was also validated for subgroups including: Ultrasound machine manufacturer, BMI, maternal age, gestational age and ethnicity when appropriate.
Sonio Detect was validated only with GE, Canon, and Samsung ultrasound devices and is intended only to be used with these ultrasound vendors.
The results of verification and performance testing demonstrate the safe and effective use of Sonio Detect.
### Clinical Study
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Sonio. The logo consists of a stylized blue icon resembling a curved shape with a dot, followed by the word "sonio" in blue, lowercase letters. The logo has a clean and modern design.
## VIII. Conclusions
Sonio Detect's intended users, clinical outcome and clinical applications are similar to those of the predicate device SonoLyst.
The technological characteristics differences identified and discussed in Section VI do not raise different questions of safety and effectiveness of the device.
Furthermore, results of successful verification and validation activities and additional bench performance testing do not raise any new issue regarding the safety and effectiveness of the device.
Thus, Sonio Detect is substantially equivalent to its predicate device SonoLyst (K201828).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.