K243608 · Smith & Nephew, Inc. · JDS · Jun 12, 2025 · Orthopedic
Device Facts
Record ID
K243608
Device Name
TRIGEN Stable Lock Nut & Washer
Applicant
Smith & Nephew, Inc.
Product Code
JDS · Orthopedic
Decision Date
Jun 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject device can be used with the TRIGEN META-TAN Trochanteric Antegrade Nailing System, TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System, and the TRIGEN META-NAIL Retrograde Femoral Nail System therefore the indications for use for all three systems are applicable to the subject TRIGEN Stable Lock Nut & Washer. TRIGEN META-TAN Trochanteric Antegrade Nailing System The TRIGEN META-TAN Trochanteric Antegrade Nail is indicated for fixation of fractures that occur in and between the proximal third and distal fourth of the femur including simple long bone fractures, severely comminuted, spiral, and segmental fractures, and supracondylar fractures. In addition, TRIGEN META-TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; and intracapsular fractures. TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted spiral, large oblique and segmental fractures; and nonunions. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures and intertrochanteric fractures. TRIGEN META-NAIL Retrograde Femoral Nail System The TRIGEN META-NAIL Retrograde Femoral Nail is indicated for fractures of the femur including stable and unstable distal metaphyseal fractures, diaphyseal fractures, intra-articular fractures, and peri-prosthetic fractures.
Device Story
TRIGEN Stable Lock Nut & Washer acts as an accessory to TRIGEN intramedullary nailing systems; provides compression and rotational stability for femoral fracture fixation. Device is implanted by orthopedic surgeons in clinical settings. Input consists of mechanical torque during insertion; output is secure fixation of bone screws within the nail construct. Benefits include stabilization of fractures to aid normal healing. Device is single-use and gamma-sterilized.
Clinical Evidence
No clinical tests were performed. Substantial equivalence is supported by bench testing, including theoretical failure analysis, torque to failure/insertion/extraction testing, and MR compatibility testing per ASTM F2182-19e2, ASTM F2052-15, ASTM F2213-2017, and ASTM F2119-07(2013).
Technological Characteristics
Material: Ti-6Al-4V. Form factor: Nut and washer assembly. Sterilization: Gamma. Connectivity: None. Sensing/Actuation: Mechanical bone fixation appliance.
Indications for Use
Indicated for patients requiring fixation of femoral fractures, including simple, comminuted, spiral, segmental, supracondylar, subtrochanteric, intertrochanteric, intracapsular, diaphyseal, intra-articular, and peri-prosthetic fractures, as well as nonunions, when used with compatible TRIGEN nailing systems.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
TRIGEN META-NAIL Retrograde Femoral Nail (K061019, K230761)
TRIGEN META-TAN Trochanteric Antegrade Nail (K243364)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 12, 2025
Smith & Nephew, Inc.
Jessica Roach
Regulatory Affairs Specialist II
1450 E Brooks Road
Memphis, Tennessee 38116
Re: K243608
Trade/Device Name: TRIGEN Stable Lock Nut & Washer
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/multiple component metallic bone fixation appliances and accessories
Regulatory Class: Class II
Product Code: JDS, HSB
Dated: May 20, 2025
Received: May 20, 2025
Dear Jessica Roach:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243608 - Jessica Roach
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243608 - Jessica Roach
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Farzana Sharmin -S
Digitally signed by
Farzana Sharmin -S
Date: 2025.06.12
11:45:02 -04'00'
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243608
Device Name
TRIGEN Stable Lock Nut & Washer
Indications for Use (Describe)
The subject device can be used with the TRIGEN META-TAN Trochanteric Antegrade Nailing System, TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System, and the TRIGEN META-NAIL Retrograde Femoral Nail System therefore the indications for use for all three systems are applicable to the subject TRIGEN Stable Lock Nut & Washer.
TRIGEN META-TAN Trochanteric Antegrade Nailing System
The TRIGEN META-TAN Trochanteric Antegrade Nail is indicated for fixation of fractures that occur in and between the proximal third and distal fourth of the femur including simple long bone fractures, severely comminuted, spiral, and segmental fractures, and supracondylar fractures.
In addition, TRIGEN META-TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; and intracapsular fractures.
TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted spiral, large oblique and segmental fractures; and nonunions. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures and intertrochanteric fractures.
TRIGEN META-NAIL Retrograde Femoral Nail System
The TRIGEN META-NAIL Retrograde Femoral Nail is indicated for fractures of the femur including stable and unstable distal metaphyseal fractures, diaphyseal fractures, intra-articular fractures, and peri-prosthetic fractures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
K243608- Page 1 of 2
| 510(k) #: K243608 | 510(k) Summary | Prepared on: 2025-06-11 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Smith & Nephew, Inc. | |
| Applicant Address | 1450 E Brooks Road Memphis TN 38116 United States | |
| Applicant Contact Telephone | 978-474-6318 | |
| Applicant Contact | Ms. Jessica Roach | |
| Applicant Contact Email | jessica.roach@smith-nephew.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | TRIGEN Stable Lock Nut & Washer | |
| Common Name | Single/multiple component metallic bone fixation appliances and accessories | |
| Classification Name | Nail, Fixation, Bone | |
| Regulation Number | 888.3030 | |
| Product Code(s) | JDS, HSB | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K983942 | Stainless-Steel Stable Lock Nut & Washer | JDS |
| K111025 | TRIGEN Low Profile Screw | HSB |
| K061019, K230761 | TRIGEN META-NAIL Retrograde Femoral Nail | JDS |
| K243364 | TRIGEN META-TAN Trochanteric Antegrade Nail | HSB |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The device in scope of this submission is the TRIGEN Stable Lock Nut & Washer.
The TRIGEN Stable Lock Nut & Washer is intended to be used as an aid to normal fracture healing for fractures that occur in and between the proximal third and distal fourth of the femur including simple long bone fractures, severely comminuted, spiral, and segmental fractures, and supracondylar fractures; simple long bone fractures; severely comminuted spiral, large oblique and segmental fractures; nonunions; for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated; and for fractures of the femur that include stable and unstable distal metaphyseal fractures, diaphyseal fractures, intra-articular fractures, and peri-prosthetic fractures. The TRIGEN Stable Lock Nut & Washer is made from Ti-6Al-4V material. The device is Gamma sterilized and intended for single-use only. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The subject device can be used with the TRIGEN META-TAN Trochanteric Antegrade Nailing System, TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System, and the TRIGEN META-NAIL Retrograde Femoral Nail System therefore the | | |
{6}
K243608- Page 2 of 2
indications for use for all three systems are applicable to the subject TRIGEN Stable Lock Nut & Washer.
TRIGEN META-TAN Trochanteric Antegrade Nailing System
The TRIGEN META-TAN Trochanteric Antegrade Nail is indicated for fixation of fractures that occur in and between the proximal third and distal fourth of the femur including simple long bone fractures, severely comminuted, spiral, and segmental fractures, and supracondylar fractures.
In addition, TRIGEN META-TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; and intracapsular fractures.
TRIGEN TAN FAN Trochanteric Antegrade and Femoral Antegrade Nailing System
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted spiral, large oblique and segmental fractures; and nonunions. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures and intertrochanteric fractures.
TRIGEN META-NAIL Retrograde Femoral Nail System
The TRIGEN META-NAIL Retrograde Femoral Nail is indicated for fractures of the femur including stable and unstable distal metaphyseal fractures, diaphyseal fractures, intra-articular fractures, and peri-prosthetic fractures.
Indications for Use Comparison 21 CFR 807.92(a)(5)
The subject TRIGEN Stable Lock Nut & Washer indications fall within the predicate devices' indications for use.
Technological Comparison 21 CFR 807.92(a)(6)
The overall device design is similar to that of the predicate Smith & Nephew system cleared under the premarket notification Stainless-Steel Stable Lock Nut & Washer (K983942, SE 12/04/1998) except for the device material, and threads. The subject device material and threads are similar to that of the predicate Smith & Nephew system cleared under the premarket notification TRIGEN Low Profile Screw (K111025, SE 07/01/2011).
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
- Theoretical Failure Analysis
- Torque to Failure/Insertion/Extraction Torque Testing and Analysis
- MR Compatibility per ASTM F2182-19e2, ASTM F2052-15, ASTM F2213-2017, and ASTM F2119-07(2013)
No clinical tests were performed to support safety and effectiveness of the subject devices.
The performance bench tests were used as a basis for the determination of substantial equivalence. The results of the equivalence testing demonstrate that the differences in design of the subject device does not raise any difference in safety or effectiveness as compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.