Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761); Codman Libertís™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762)
K243552 · Integra Lifesciences Production Corporation · JXG · Aug 6, 2025 · Neurology
Device Facts
Record ID
K243552
Device Name
Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761); Codman Libertís™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762)
Applicant
Integra Lifesciences Production Corporation
Product Code
JXG · Neurology
Decision Date
Aug 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.
Device Story
The Codman Libertis™ EVD Catheter is a ventricular catheter used to drain cerebrospinal fluid (CSF) to manage intracranial pressure. The device is placed into the brain ventricles; CSF enters through inlet holes at the catheter tip and drains into an external collection system. The catheter is impregnated with rifampicin and clindamycin hydrochloride to reduce gram-positive bacterial colonization and incorporates Endexo® polymer additive to reduce protein-biomaterial interactions, thrombus formation, and cellular adhesion. The device includes a barium sulfate stripe for radiopacity and numerical depth markings. It is used in clinical settings (e.g., neurosurgery/ICU) by physicians. The catheter is supplied with accessories including a trocar, stylet, and anchoring clip to facilitate placement. The device is sterilized via electron-beam radiation. The output is the physical drainage of CSF, which allows clinicians to monitor and control intracranial pressure, potentially benefiting patients by preventing or treating elevated pressure states.
Clinical Evidence
Bench testing only. No clinical studies were performed. Performance was validated through mechanical testing (ASTM F647, ISO 7197), MRI compatibility (ASTM F2503), antimicrobial efficacy (Zone of Inhibition), antibiotic content, residual solvents, and biocompatibility (ISO 10993-1). Shelf life and packaging integrity were validated per ASTM F1980 and ISO 11607-1.
Technological Characteristics
Silicone catheter impregnated with rifampicin and clindamycin hydrochloride; Endexo surface-modifying macromolecule additive. Features barium sulfate radiopaque stripe and ink depth markings. Sterilized via electron-beam radiation (SAL 10^-6). Accessories include trocar, stylet, and anchoring clip. Packaging uses PETG trays and nylon/LDPE sterile barriers in low-pressure N2 atmosphere.
Indications for Use
Indicated for patients requiring ventricular access for CSF drainage to reduce intracranial pressure and CSF volume.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set (K233448)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 6, 2025
Integra LifeSciences Production Corporation
Kali Sacco
Manager, Regulatory Affairs
11 Cabot Boulevard
Mansfield, Massachusetts 02048
Re: K243552
Trade/Device Name: Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761); Codman Libertís™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762)
Regulation Number: 21 CFR 882.5550
Regulation Name: Central nervous system fluid shunt and components
Regulatory Class: Class II
Product Code: JXG
Dated: November 15, 2024
Received: July 7, 2025
Dear Kali Sacco:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K243552 - Kali Sacco
Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243552 - Kali Sacco
Page
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yen-chih Lin
-S
Digitally signed by Yen-chih Lin -S
Date: 2025.08.06 16:51:02
-04'00'
For
Adam Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243552
Device Name
Codman Libertis™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761);
Codman Libertis™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762)
Indications for Use (Describe)
The Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Integra LifeSciences-Traditional 510(k)
Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection
# 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a) (1) Submitter Information | |
| --- | --- |
| Name and Address | Integra LifeSciences Production Corporation
11 Cabot Boulevard
Mansfield, MA 02048 |
| Telephone number | (680) 910-4999 |
| Primary Contact | Kali Sacco |
| Date Summary Prepared | August 6, 2025 |
| 807.92(a) (2) Name of Device | |
| Trade or Proprietary Name | Codman Libertís™ EVD Catheter with Bactiseal® and Endexo® Technology |
| Common Name | Ventricular Catheters |
| Classification Name | Central Nervous System Fluid Shunt and Components
(21 CFR 882.5550) |
| Device Class | II |
| Product Code | JXG |
| 807.92(a) (3) Predicate Information | |
| Predicate Device | Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set (K233448) |
| 807.92(a) (4) Device Description | |
| The Codman Libertís™ EVD Catheter with Bactiseal® and Endexo® Technology (Libertís™ EVD Catheters) include a ventricular catheter that is supplied with component accessories that facilitate placement and use of the catheter for reducing and controlling intracranial pressure due to excess cerebrospinal fluid (CSF). The ventricular catheter is subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. Laboratory studies show Bactiseal treated catheters reduce the colonization of gram-positive bacteria on the tubing surface. Additionally, Libertís™ EVD Catheters contain an Endexo® polymer additive; a surface modifying macromolecule (SMM) polymer blended into the catheter's base silicone. The ventricular catheter is placed in the ventricles of the brain and CSF enters the fluid conduit through the inlet holes near the tip of the catheter and drains into the external drainage system connected to the catheter. The catheter contains a barium sulfate stripe for radiopacity and includes numerical depth markings and circumferential bands, made of ink, from the proximal tip. | |
| 807.92(a) (5) Indications for Use | |
| The Codman® Libertís™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume. | |
| 807.92(a) (6) Technological Characteristics Compared to Predicate | |
| The proposed Libertís™ EVD Catheters have the same intended use and functionality as the predicate devices. The proposed changes to the design, materials, and sterility do not raise different questions of safety and/or effectiveness. | |
{5}
Integra LifeSciences-Traditional 510(k)
Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection
| Component Affected | Proposed Modification | Rationale |
| --- | --- | --- |
| Libertis™ EVD Catheters | Addition of Endexo additive | A surface modifying macromolecule (SMM) polymer blended into the catheter’s base silicone. The Endexo technology modifies polymeric medical device surfaces to passively reduce the protein – biomaterial interactions which mediate thrombus formation and cellular adhesion. |
| | The diameter of the catheter inlet hole on the smaller diameter catheter (821761) is larger than the predicate P/N 821745 and there are fewer inlet holes | To ensure the same performance specifications as the predicate by maintaining the same inlet hole opening surface area. |
| | Additional ventricular catheter markings | Addition of numerical markings and circumferential bands is part of continuous improvement in catheter usability |
| | Different material formulations in the ink markings | A different ink formulation was used to ensure adherence to the catheter. |
| Catheter Anchoring Clip | Different material formulation | To accommodate the manufacturing processes |
| | Increased thickness of the catheter anchoring clip | Continuous improvement for usability |
| Trocar | Reduced diameter in the bending region | More bendable as part of continuous improvement |
| | Reduced barb ridges to 4 compared to 6 on the predicate | Continuous improvement for usability to maintain consistency between the trocars used with each Libertis EVD Catheter (821761 and 821762) |
| Stylet for 821761 | Decreased cross-section size of the stylet | The decrease in cross-section is required due to the smaller size of the catheter |
| Catheter and Stylet | Sterilization using electron-beam radiation | Radiation (electron beam) sterilization methods were chosen to reduce thermal degradation of the antibiotics. |
| Packaging of Catheter and Stylet | Change the catheter and stylet tray to PETG | PETG was selected for greater compatibility with electron beam sterilization |
| | Change the sterile barrier for the proposed catheter and stylet to two sheets of nylon/LDPE (low density polyethylene) rather than Tyvek 1073B and polyester/polyolefin. | To facilitate packaging in a low pressure N_{2} atmosphere. |
| | Filling the pouch with low pressure N_{2} | To reduce oxidative degradation of the antibiotics |
| Labeling | Device and package labels:
• New device name
• New reference number (SKU)
• New description to identify contents of each pouch
• Updated symbology for pouch containing Catheter and Stylet | To accurately reflect Libertis EVD Catheter |
Page 2 of 4
{6}
Integra LifeSciences-Traditional 510(k)
Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection
| | to identify the sterilization method | |
| --- | --- | --- |
| Instructions for Use | • Update Description to add information about Endexo and location of catheter marking and inlet holes
• Revised How Supplied section to describe the two separate sterilized pouches and contents
• Added symbol for sterilization method change | Technological differences between the Libertis EVD Catheter and predicate Codman Bactiseal EVD Catheters (K233448) |
| | Added new sections: Intended Clinical Benefits, Intended User, Intended Patient Population, Disposal | To align with EU regulatory labeling requirements |
| | Added statement that the trocar and stylet provided to aid in the placement of the Codman Libertis EVD Catheter are MR-Unsafe | The trocar and stylet are not intended to be used in the MR environment and are labeled MR-unsafe |
# 807.92(b) 1-2: Summary of Nonclinical and Clinical Testing Performed
The following performance testing has been conducted in support of the substantial equivalence determination. The testing utilized FDA recognized consensus standards and internal test methods. All testing was performed on production equivalent devices.
| Performance Bench Test Results | |
| --- | --- |
| Test | Conclusion |
| Evaluation of Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 | Pass |
| Evaluation of Aged Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 (T = 12 months) | Pass |
| Functional/Mechanical Performance of Codman Libertís EVD Catheters | Pass |
| Functional/Mechanical Performance of Aged Codman Libertís EVD Catheters (T = 12 months) | Pass |
| MRI Compatibility Assessment of Codman Libertís EVD Catheters per ASTM F2503 | Pass |
| Antimicrobial Efficacy of Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use | Pass |
| Antimicrobial Efficacy of Aged Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use (T = 12 months) | Pass |
| Antibiotic Content Testing of Codman Libertís EVD Catheters | Pass |
| Residual Solvents Testing of Codman Libertís EVD Catheters | Pass |
{7}
Integra LifeSciences-Traditional 510(k)
Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection
| Libertis EVD Catheters with Bactiseal® and Endexo® Technology – Summative Usability Study | |
| --- | --- |
| Product Shelf Life Testing per ASTM F1980 and ISO 11607-1 | Pass |
| Biocompatibility Testing Results | |
| Test | Conclusion |
| Biocompatibility Testing per ISO 10993-1 | Pass |
| Sterilization/Cleaning The proposed device (Codman Libertis™ EVD Catheter and stylet) is sterilized via electron beam radiation. A sterilization validation was performed and determined that the sterility assurance level achieved (SAL = 10^{-6}) is equivalent to the sterility level achieved in the predicate device, which is sterilized via steam. There has been no change to the sterilization method (steam sterilized) or sterility assurance level (SAL = 10^{-6}) for the Libertis™ EVD Catheter accessories (Trocar, Female LUER-LOK™ Connectors, Male LUER-LOK Cap, and Catheter Anchoring Clip) as compared to the predicate device. Shelf Life There are no changes in shelf life as a result of the proposed changes. The proposed device, Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology shelf life is 12 months which is the same as the predicate. Animal Studies No animal studies were required as appropriate verification and validation testing of the subject device was achieved based on the comparison to the predicate device and from the results of the bench testing. Clinical Studies No clinical studies were required as appropriate verification and validation testing of the subject device was achieved based on the comparison to the predicate device and from the results of the bench testing. | |
| 807.92(b) (3) Conclusion Based upon the intended use, design, comparison to the predicate device, and testing performed, Integra LifeSciences believes that the proposed modifications to the Codman Libertis™ EVD Catheter with Bactiseal® and Endexo® Technology do not raise different questions of safety and effectiveness, and is therefore, substantially equivalent to the predicate Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set. | |
Page 4 of 4
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.