HydroPICC Single Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0 mL/s HydroPICC Dual Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 5.0 mL/s HydroMID Single Lumen: Indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products. Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 mL/sec The Hydro Catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter's steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis.
Device Story
HydroPICC and HydroMID are intravascular catheters for venous access; used for fluid/medication administration, blood sampling, and power injection. Devices feature MIMIX® Technology, an anti-thrombogenic bulk hydrogel material acting as a steric barrier to reduce thrombus accumulation and occlusions. Catheters are radiopaque, hydrophilic, and include suture wings, Luer lock hubs, and extension tubing. Used in clinical environments (IR, nursing); placed by clinicians. Output is venous access; healthcare providers use the device to deliver therapy or sample blood. Benefits include reduced risk of thrombotic complications. No clinical trials performed; safety supported by in vitro blood loop and flow models.
Clinical Evidence
Bench testing only. Multiple in vitro studies (blood loop models, flow-induced thrombosis models) compared Hydro-Line catheters to polyurethane devices. Results consistently demonstrated reduced thrombus accumulation and thrombotic occlusions on external surfaces and internal fluid pathways. No human clinical trials conducted.
Indicated for short- or long-term peripheral access to the central venous system (HydroPICC) or short-term (<30 days) peripheral venous access (HydroMID) for intravenous therapy, fluid/medication/nutrient administration, blood sampling, CVP monitoring, and power injection of contrast media. Adult and pediatric populations (HydroMID). Contraindicated for treatment of existing vein thrombosis.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 1, 2025
Access Vascular, Inc.
Scott Blood
Vice President of Quality & Regulatory
749 Middlesex Turnpike
Billerica, Massachusetts 01821
Re: K243458
Trade/Device Name: HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001201); HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire - Basic Kit (70001202); HydroPICC 4Fr Single Lumen Marked catheter - Maximal Barrier kit (70001204); HydroPICC 5Fr Dual Lumen Marked catheter, 130 cm guidewire - Basic Kit (70002201); HydroPICC 5Fr Dual Lumen Marked catheter, 70 cm guidewire - Basic Kit (70002202); HydroPICC 5Fr Dual Lumen Marked catheter - Maximal Barrier kit (70002204); HydroMID 4Fr Single Lumen Marked catheter - Basic kit (70004302); HydroMID 4Fr Single Lumen Marked catheter - Maximal Barrier Kit (70004304); HydroPICC 4Fr Single Lumen Marked catheter - Mobile Maximal Barrier Kit (90001304); HydroPICC 5Fr Dual Lumen Marked catheter - Mobile Maximal Barrier Kit (90002304); HydroMID 4Fr
Regulation Number: 21 CFR 880.5970
Regulation Name: Percutaneous, implanted, long-term intravascular catheter.
Regulatory Class: Class II
Product Code: LJS, FOZ
Dated: July 3, 2025
Received: July 3, 2025
Dear Scott Blood:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243458 - Scott Blood
Page 2
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243458 - Scott Blood
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Wolloscheck -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243458
Device Name
HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001201);
HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire - Basic Kit (70001202);
HydroPICC 4Fr Single Lumen Marked catheter - Maximal Barrier kit (70001204);
HydroPICC 5Fr Dual Lumen Marked catheter, 130 cm guidewire - Basic Kit (70002201);
HydroPICC 5Fr Dual Lumen Marked catheter, 70 cm guidewire - Basic Kit (70002202);
HydroPICC 5Fr Dual Lumen Marked catheter - Maximal Barrier kit (70002204);
HydroMID 4Fr Single Lumen Marked catheter - Basic kit (70004302);
HydroMID 4Fr Single Lumen Marked catheter - Maximal Barrier Kit (70004304);
HydroPICC 4Fr Single Lumen Marked catheter - Mobile Maximal Barrier Kit (90001304);
HydroPICC 5Fr Dual Lumen Marked catheter - Mobile Maximal Barrier Kit (90002304);
HydroMID 4Fr Single Lumen Marked catheter - Mobile Maximal Barrier Kit (90004304)
Indications for Use (Describe)
HydroPICC Single Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0 mL/s
HydroPICC Dual Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 5.0 mL/s
HydroMID Single Lumen: Indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products. Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 mL/sec
The Hydro Catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter's steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
Page 1 of 12
510(k) Summary
K243458
This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92.
Submitter Information
510(k) Sponsor Access Vascular Inc.
Address 749 Middlesex Turnpike Billerica MA 01821
Contact Person(s) Mr. Scott Blood; Vice President, Quality & Regulatory
Phone: 978.729.5978
Email: sblood@accessvascularinc.com
Date Prepared: July 31, 2025
Proposed Device(s)
Common/Usual Name: Intravascular Catheter
Trade/Proprietary Name: HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001201);
HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire - Basic Kit (70001202);
HydroPICC 4Fr Single Lumen Marked catheter - Maximal Barrier kit (70001204);
HydroPICC 5Fr Dual Lumen Marked catheter, 130 cm guidewire - Basic Kit (70002201);
HydroPICC 5Fr Dual Lumen Marked catheter, 70 cm guidewire - Basic Kit (70002202);
HydroPICC 5Fr Dual Lumen Marked catheter - Maximal Barrier kit (70002204);
HydroMID 4Fr Single Lumen Marked catheter - Basic kit (70004302);
HydroMID 4Fr Single Lumen Marked catheter - Maximal Barrier Kit (70004304);
HydroPICC 4Fr Single Lumen Marked catheter– Mobile Maximal Barrier Kit (90001304);
HydroPICC 5Fr Dual Lumen Marked catheter– Mobile Maximal Barrier Kit
{6}
(90002304);
HydroMID 4Fr Single Lumen Marked catheter– Mobile Maximal Barrier Kit (90004304)
Classification Name: Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Device Classification: 880.5970, 880.5200
Product Code: LJS, FOZ
Class: II
Device Panel: General Hospital
## Predicate Device(s)
The predicate devices are the HydroPICC (K193015), HydroPICC-251 Dual Lumen (K213550), and HydroMID Catheters (K220772).
## Device Description
The HydroMID catheters are a family of midline catheters comprised of a radiopaque, hydrophilic catheter with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp. HydroMID has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using an in vitro model. Pre-clinical in vitro evaluations do not necessarily predict clinical performance with respect to thrombus formation.
The HydroPICC catheters are a family of peripherally inserted central catheter (PICC) comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp. HydroPICC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using an in vitro model. Pre-clinical in vitro evaluations do not necessarily predict clinical performance with respect to thrombus formation.
The purpose of this 510(k) is to add an additional statement to the indications for use.
Page 2 of 12
{7}
Page 3 of 12
# Indications for Use
HydroPICC Single Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0 mL/s
HydroPICC Dual Lumen: Indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 5.0 mL/s
HydroMID Single Lumen: Indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products. Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 mL/sec
The Hydro Catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter's steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis.
# Substantial Equivalence Discussion
The following table is a matrix of the intended use and technological characteristics between the subject device and the predicate device. The table also discusses why these differences do not introduce new or different questions of safety and effectiveness.
| Specification | HydroPICC Single Lumen Catheter (K243458)
Subject device | HydroPICC Single Lumen (K193015)
Access Vascular Inc.
Primary Predicate Device | Same / Different between
Proposed and Predicate? |
| --- | --- | --- | --- |
| Intended Use | intended for short- or long-term peripheral access to the central venous system for intravenous therapy | intended for short- or long-term peripheral access to the central venous system for intravenous therapy | Same |
| Indications for Use | Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0mL/s
The following language is proposed to be added to the current indications for use for the HydroPICC Single-Lumen Catheter. There is no | Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0 mL/s | Substantially equivalent. |
{8}
| Specification | HydroPICC Single Lumen Catheter (K243458)
Subject device | HydroPICC Single Lumen (K193015)
Access Vascular Inc.
Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| | change to the remainder of the indications for use for the device:
The Hydro catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on the device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter’s steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis. | | |
| Device Classification | II | II | Same |
| Product Code | LJS | LJS | Same |
| Regulation | 21 CFR 880.5970 | 21 CFR 880.5970 | Same |
| Prescription Device | Yes | Yes | Same |
| Intended population | Adult | Adult | Same |
| Catheter Type | Peripherally Inserted Central Catheter (PICC) | Peripherally Inserted Central Catheter (PICC) | Same |
| Catheter Outer Diameter French Size | 4 French (1.40mm)
(Post Hydrated) | 4 French (1.40mm)
(Post Hydrated) | Same |
| Catheter Outer Diameter as supplied | 1.30 mm
(Supplied dehydrated) | 1.30 mm
(Supplied dehydrated) | Same |
| Catheter Inner Diameter | 1mm | 1mm | Same |
| Useable Length | 55cm | 55cm | Same |
| Priming Volume | < 1.0mL | < 1.0mL | Same |
| Guidewire Compatibility | ∅.018" | ∅.018" | Same |
Page 4 of 12
{9}
| Specification | HydroPICC Single Lumen Catheter (K243458) Subject device | HydroPICC Single Lumen (K193015) Access Vascular Inc. Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Catheter Shaft Design | Taper | Taper | Same |
| Number of Lumens | 1 | 1 | Same |
| Key Device Components | Catheter shaft, suture wing, extension tube, Luer hub | Catheter shaft, suture wing, extension tube, Luer hub | Same |
| Implant Time | Short- or Long-term | Short- or Long-term | Same |
| Use with Power Injection and Specified Flow Rate | Yes 5.0 mL/s | Yes 5.0 mL/s | Same |
| Catheter Materials | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Same |
| X-Ray Confirmation Required | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Single Use | Yes | Yes | Same |
| MRI Safety | MRI Conditional | MRI Conditional | Same |
| How Supplied | Convenience Kit: Basic IR Kit Full Nursing Kit Maximal Barrier Kit | Convenience Kit: Basic IR Kit Full Nursing Kit Maximal Barrier Kit | Same |
| Catheter Markings | Markings (every cm) | Markings (every cm) | Same |
| Specification | Hydro PICC Dual Lumen (K243458) Subject device | HydroPICC-251 5F DL Catheter (K213550) Access Vascular Inc. Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Intended Use | intended for short- or long-term peripheral access to the central venous system for intravenous therapy HydroMID: Intended for short-term access to the peripheral venous system | intended for short- or long-term peripheral access to the central venous system for intravenous therapy HydroMID: Intended for short-term access to the peripheral venous system | Same |
| Indications for Use | Indicated for short- or long term peripheral access to the central venous system for | Indicated for short- or long term peripheral access to the central venous | Substantially equivalent. |
Page 5 of 12
{10}
| Specification | Hydro PICC Dual Lumen (K243458) Subject device | HydroPICC-251 5F DL Catheter (K213550) Access Vascular Inc. Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| | intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 6.0mL/s
The following language is proposed to be added to the current indications for use for the HydroPICC Single-Lumen Catheter. There is no change to the remainder of the indications for use for the device:
The Hydro catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on the device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter’s steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis. | system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for a maximum power injection flow rate of 6.0mL/s | |
| Device Classification | Same | II | Same |
| Product Code | Same | LJS | Same |
| Regulation | 21 CFR 880.5970 | 21 CFR 880.5970 | Same |
| Prescription Device | Yes | Yes | Same |
| Intended population | Adult | Adult | Same |
| Catheter Type | Peripherally Inserted Central Catheter (PICC) | Peripherally Inserted Central Catheter (PICC) | Same |
| Catheter Outer Diameter French Size | 5 French (1.67mm) (Post Hydrated) | 5 French (1.67mm) (Post Hydrated) | Same |
| Catheter Outer Diameter as supplied | 1.60 mm (Supplied dehydrated) | 1.60 mm (Supplied dehydrated) | Same |
| Catheter Inner Diameter | 1mm | 1mm | Same |
| Useable Length | 55cm | 55cm | Same |
Page 6 of 12
{11}
| Specification | Hydro PICC Dual Lumen(K243458)Subject device | HydroPICC-251 5F DL Catheter(K213550)Access Vascular Inc.Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Priming Volume | < 1.0mL per lumen | < 1.0mL per lumen | Same |
| Guidewire Compatibility | ∅.018" | ∅.018" | Same |
| Catheter Shaft Design | Taper | Taper | Same |
| Number of Lumens | 2 | 2 | Same |
| Key Device Components | Catheter shaft, suture wing, extension tube, Luer hub | Catheter shaft, suture wing, extension tube, Luer hub | Same |
| Implant Time | Short- or Long-term | Short- or Long-term | Same |
| Use with Power Injection and Specified Flow Rate | Yes6.0 mL/s | Yes6.0 mL/s | Same |
| Catheter Materials | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Same |
| X-Ray Confirmation Required | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Single Use | Yes | Yes | Same |
| MRI Safety | MRI Conditional | MRI Conditional | Same |
| How Supplied | Convenience Kit:Basic IR KitFull Nursing KitMaximal Barrier Kit | Convenience Kit:Basic IR KitFull Nursing KitMaximal Barrier Kit | Same |
| Catheter Markings | Markings (every cm) | Markings (every cm) | Same |
| Specification | HydroMID Single Lumen(K243458)Subject device | HydroMICC (K220772)Access Vascular Inc.Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Intended Use | Intended for short-term access to the peripheral venous system | Intended for short-term access to the peripheral venous system | Same |
| Use | Not Applicable | Not Applicable | Not Applicable |
{12}
| Specification | HydroMID Single Lumen (K243458) Subject device | HydroMID (K220772) Access Vascular Inc. Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Indications for Use | Indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products.
Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 mL/sec
The following language is proposed to be added to the current indications for use for the HydroPICC Single-Lumen Catheter. There is no change to the remainder of the indications for use for the device:
The Hydro catheter lines incorporate MIMIX® Technology, an anti-thrombogenic bulk hydrogel material, designed to minimize the risk of thrombus formation on the device surfaces. In vitro studies have demonstrated that both the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions. This reduction in thrombus accumulation is achieved due to the catheter’s steric barrier. The clinical impact has not been evaluated in human clinical trials. This device is not intended for the treatment of existing vein thrombosis. | Indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products.
Maximum Power Injection Flow Rate: 4Fr Single Lumen, 20cm: 6 mL/sec | Substantially equivalent. |
| Device Classification | II | II | Same |
| Product Code | FOZ | FOZ | Same |
| Regulation | 21 CFR 880.5200 | 21 CFR 880.5200 | Same |
| Prescription Device | Yes | Yes | Same |
| Intended population | Adult and Pediatric Use | Adult and Pediatric Use | Same |
| Catheter Type | Midline Catheter | Midline Catheter | Same |
| Catheter Outer Diameter French Size | 4 French (1.40mm) (Post Hydrated) | 4 French (1.40mm) (Post Hydrated) | Same |
| Catheter Outer Diameter as supplied | 1.30 mm (Supplied dehydrated) | 1.30 mm (Supplied dehydrated) | Same |
Page 8 of 12
{13}
| Specification | HydroMID Single Lumen (K243458) Subject device | HydroMID (K220772) Access Vascular Inc. Primary Predicate Device | Same / Different between Proposed and Predicate? |
| --- | --- | --- | --- |
| Catheter Inner Diameter | 1mm | 1mm | Same |
| Useable Length | 20 cm | 20 cm | Same |
| Priming Volume | < 1.0mL | < 1.0mL | Same |
| Guidewire Compatibility | ∅.018" | ∅.018" | Same |
| Catheter Shaft Design | Taper | Taper | Same |
| Number of Lumens | 1 | 1 | Same |
| Key Device Components | Catheter shaft, suture wing, extension tube, Luer hub | Catheter shaft, suture wing, extension tube, Luer hub | Same |
| Implant Time | Short term | Short term | Same |
| Use with Power Injection and Specified Flow Rate | Yes
6mL/s | Yes
6mL/s | Same |
| Catheter Materials | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Radiopaque hydrophilic polyol catheter with Luer lock hub, polyurethane extension tubing, and suture wing | Same |
| X-Ray Confirmation Required | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Single Use | Yes | Yes | Same |
| MRI Safety | MRI Conditional | MRI Conditional | Same |
| How Supplied | Convenience Kit:
Basic IR Kit
Full Nursing Kit
Maximal Barrier Kit | Convenience Kit:
Basic IR Kit
Full Nursing Kit
Maximal Barrier Kit | Same |
| Catheter Markings | Markings (every cm) | Markings (every cm) | Same |
Page 9 of 12
{14}
Page 10 of 12
# Discussion of Similarities and Differences
## Indications for Use:
Rationale Supporting 21 CFR 807.92(a)(5) for a Change in Indications for Use
1. Technological Characteristics and Performance Data: The proposed devices, HydroPICC DL and SL, and HydroMID SL, incorporate MIMIX® Technology, which is designed to minimize thrombus formation. This technology is present in the predicate devices but was not highlighted in their indications for use statements. In vitro studies have demonstrated that the external surfaces and internal fluid pathways of the catheter effectively reduce thrombus accumulation and thrombotic occlusions.
2. Safety and Effectiveness: The proposed devices have been shown to be safe and effective for their intended use. The reduction in thrombus formation can potentially lead to fewer complications, such as thrombotic events, which are common concerns with catheter use. Although no direct link between these testing methods and clinical outcomes has been conclusively established, the laboratory data provide a strong basis for the proposed indications.
3. Comparison with Predicate Devices The predicate devices, HydroPICC (K193015), HydroMID (K220772), and HydroPICC-251 5F DL Catheter (K213550), are indicated for similar uses, including short-term and long-term peripheral access to the venous system for intravenous therapy, administration of fluids, medications, and nutrients, and the sampling of blood. The proposed devices offer additional benefits due to the explicit incorporation of MIMIX® Technology in their indications for use statements, which was not previously mentioned for the predicate devices.
4. Intended Use and Indications The intended use of the proposed devices aligns with that of the predicate devices, with the added benefit of reduced thrombus formation. The proposed indications for use are consistent with the current clinical practices and address the same patient population and clinical needs as the predicate devices.
5. Regulatory Compliance The proposed changes comply with the requirements of 21 CFR 807.92(a)(5), which necessitates a comparison of the new device with a legally marketed predicate device to demonstrate that the new device is as safe and effective as the predicate device. The data provided support that the proposed devices meet these criteria.
Conclusion: The proposed HydroPICC DL and SL and HydroMID SL catheters, with their advanced MIMIX® Technology, offer significant improvements in terms of reducing thrombus formation. These enhancements, supported by robust laboratory data, justify the change in indications for use and demonstrate that the proposed devices are as safe and effective as the predicate devices.
## Technological Comparison
The proposed devices, HydroPICC Single Lumen and Dual Lumen and HydroMID Single Lumen have identical technological characteristics as the predicate device(s) identified above. These characteristics include design, material, chemical composition, and principle of operation.
{15}
# Non-Clinical and/or Clinical Tests Summary & Conclusions
The following tests were completed to demonstrate device safety and efficacy with respect to the predicate devices:
| Test Description/Title | Test Summary | Results |
| --- | --- | --- |
| Assessment of PICC Catheter Thrombosis in an in vitro Model | Evaluate the reduction of thrombus on Hydro-Line catheters compared to polyurethane devices | PASS |
| Thrombosis Accumulation Report | Assessment of PICC catheter thrombosis in an in-vitro model aimed to evaluate the reduction of thrombus and bacterial accumulation on HydroPICC catheters compared to traditional devices | PASS |
| In Vitro Thrombosis Study with Saline Conditioning | Assess PICC catheter thrombosis, incorporating preconditioning prior to flow-induced thrombosis using a multi-phase model | PASS |
| In Vitro Assessment of PICC Thrombus | Assess thrombus accumulation on the exterior surface of select PICC devices using an in vitro blood loop model (BLM) | PASS |
| In Vitro Assessment of Catheter Thrombotic Occlusion | Study comparing pressures requires to remove thrombotic occlusion of 4F PICC samples using an in-vitro blood flow model | PASS |
| Blood Loop Analysis of HydroPICC Against Competitors | Determine and compare the quantitative thrombogenicity of subject device relative to existing commercial standard and non-thrombogenic devices using in-vitro blood loop testing | PASS |
| Sterigenics Sterilization Validation | Describes the catheter to have no inhibitory characteristics per ISO 11737-2:2019 | PASS |
| | | |
{16}
| Test Description/Title | Test Summary | Results |
| --- | --- | --- |
| Exhaustive Recovery Assessment of Catheter Thrombosis in an In-Vitro Blood Flow Model | Determine thrombosis differences of catheter materials in whole blood under thrombotic conditions. | PASS |
This 510(k) submission provides comprehensive data demonstrating the safety and efficacy of the Hydro Catheter Line, which includes the HydroMID and HydroPICC devices, available in both single and dual lumen configurations. Rigorous testing protocols were employed to evaluate these devices' performance under various conditions, ensuring they meet all applicable regulatory standards. The submission includes in vitro testing, which substantiates the devices' intended use and their ability to perform as claimed. Additionally, the form, fit, and function of these devices have not changed from their previous clearances. The data presented herein supports the conclusion that the HydroMID and HydroPICC devices are substantially equivalent to the predicate device(s).
Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the Hydro Line of Catheters with the addition to the indications for use is substantially equivalent to existing legally marketed devices.
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.