K193015 · Access Vascular, Inc. · LJS · Feb 20, 2020 · General Hospital
Device Facts
Record ID
K193015
Device Name
HydroPICC (PICC-142)
Applicant
Access Vascular, Inc.
Product Code
LJS · General Hospital
Decision Date
Feb 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5mL/sec.
Device Story
HydroPICC is a 4Fr, 55cm peripherally inserted central catheter (PICC) designed for short- or long-term access to the superior vena cava; device features hydrophilic polyol construction to reduce thrombus/platelet accumulation. Kit includes introducer, guidewire, needle-free valve, and fixation device. Used in clinical settings by healthcare professionals for intravenous therapy, blood sampling, and power injection of contrast media (up to 3.5mL/sec). Device provides central venous access via peripheral insertion; enables clinicians to deliver fluids/medications and monitor pressure. Modifications from predicate include updated hydration aid, modified catheter dimensions, removal of radiopaque additive from suture wing, and updated MRI safety labeling.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including power injection flow rate, burst strength, kink/flex resistance, tensile testing, particulate matter (USP <788>), and MRI compatibility (ASTM F2052, F2182, F2213, F2119).
Technological Characteristics
4Fr single-lumen PICC; 55cm length. Materials: Radiopaque hydrophilic polyol catheter, polyurethane extension tube, suture wing (no radiopaque additive). Sterilization: Ethylene Oxide. Standards: EN ISO 10555-1/3, EN ISO 11135, EN ISO 13868, ASTM F2052/F2182/F2213/F2119. Connectivity: None (passive device).
Indications for Use
Indicated for patients requiring short- or long-term peripheral access to the central venous system for intravenous therapy, including fluid/medication/nutrient administration, blood sampling, central venous pressure monitoring, and power injection of contrast media.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
February 20, 2020
Access Vascular, Inc Brian Hanley Sr. Director, Operations 175 Middlesex Turnpike Suite 1A Bedford, Massachusetts 01730
Re: K193015
Trade/Device Name: HydroPICC Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: January 17, 2020 Received: January 21, 2020
Dear Brian Hanley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193015
Device Name
HydroPICC
Indications for Use (Describe)
HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media.
HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5mL/sec.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Access Vascular. The word "ACCESS" is in a dark blue font, and the word "VASCULAR" is in a lighter blue font. The bottom of the "A" in "ACCESS" is connected to the "V" in "VASCULAR" by a blue line.
# K193015 510(k) Summary for the HydroPICC PICC
Date prepared: 18 February, 2020
## Submitter:
Access Vascular, Inc. 175 Middlesex Turnpike Suite 1A Bedford, MA 01730 Tel. 781-538-6594
#### Contact:
Brian Hanley Access Vascular, Inc. Tel. 781-538-6594
## Subject Device
| Trade Name: | HydroPICC |
|-----------------------|-----------------------------------------------------------|
| Common Name: | Intravascular Catheter |
| Regulation Number: | 21CFR§880.5970 |
| Regulation Name: | Percutaneous, Implanted, Long-Term Intravascular Catheter |
| Regulatory Class: | Class II |
| Product Code: | LJS |
| Classification Panel: | General Hospital |
## Primary Predicate Device
| Trade Name: | HydroPICC |
|-----------------------|-----------------------------------------------------------|
| Manufacturer: | Access Vascular |
| 510(k) Reference: | K172885 |
| Common Name: | Intravascular Catheter |
| Regulation Number: | 21 CFR§880.5970 |
| Regulation Name: | Percutaneous, Implanted, Long-Term Intravascular Catheter |
| Regulatory Class: | Class II |
| Product Code: | LJS |
| Classification Panel: | General Hospital |
{4}------------------------------------------------
#### Device Description
The device description of HydroPICC-142 is the same as the predicate device, HydroPICC-141 (K172885).
The HydroPICC is a peripherally inserted central catheter (PICC) intended for short or long-term use. The HydroPICC is a 4Fr, 55cm long catheter suitable for periodic access to the superior vena cava (SVC) while maintaining a reduced level of thrombus accumulation. HydroPICC is supplied in a kit with the necessary accessories to insert and maintain the catheter. The patient population for the proposed and predicate devices are identical, namely patients who require short- and long-term vascular access.
The following accessories are provided with the HydroPICC:
- Tear-away Introducer
- Introducer needle
- Guidewire
- Male Cap Plug
- 60cm. Paper Tape Measure
- Needle-Free Valve
- Luer Lock Syringe
- Adhesive Fixation Device - Bard StatLock
- Scalpel
- Transparent Film Dressing
#### Indications for Use
HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5ml/sec. The Indications for Use statement for the HydroPICC-142 is identical to the predicate device, HydroPICC-141 (K172885).
#### Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the HydroPICC-142 are substantially equivalent to the predicate, the HydroPICC-141 (K172885), in terms of intended use, application, user population, basic design, performance, and labeling. The hydration aid, nominal catheter dimensions, and product packaging were modified in order to reduce the time it takes for the catheter to hydrate during procedural setup. A radiopaque additive was removed from the suture wing overmold. The labeling of the HydroPICC-142 was revised to indicate MRI Safety Information following the evaluation of the device.
Briefly, both the subject and predicate device are,
{5}------------------------------------------------
- intended for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; for central venous pressure monitoring and for power injection of contrast media.
- available in 4 French size
- rated for maximum power injection flow rate up to 3.5mL/sec
- available kitted with a range of procedural accessories for user convenient; and
- demonstrative of enhanced resistance to blood component (platelet and thrombus) accumulation
| | HydroPICC-142<br>Access Vascular<br>Proposed Device | HydroPICC-141<br>Access Vascular<br>K172885 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Specification | | |
| Intended Use | intended for short- or long-term<br>peripheral access to the central<br>venous system for intravenous<br>therapy. | Same |
| Indication for Use | HydroPICC is indicated for short- or<br>long-term peripheral access to the<br>central venous system for<br>intravenous therapy, including but<br>not limited to; the administration<br>of fluids, medications, and<br>nutrients; the sampling of blood;<br>central venous pressure<br>monitoring; and power injection of<br>contrast media.<br>HydroPICC is rated for a Maximum<br>Power Injection Flow Rate of<br>3.5mL/sec. | Same |
| Device Class | Class II | Same |
| Product Code | LJS | Same |
| Regulation Number | 880.5970 | Same |
| Prescription Device | Yes | Same |
| Catheter Type | Peripherally Inserted Central<br>Catheter (PICC) | Same |
| Catheter Outer Diameter | 4 French | Same |
| Catheter Inner Diameter | 1mm | Substantially equivalent |
| Usable Catheter Length | 55cm | Same |
| Guidewire compatibility | 0.018" | Same |
| Catheter Shaft Design | No Taper | Same |
| Number of Catheter Lumens | Single | Same |
#### Table 1 Summary of Technological Characteristics of Proposed and Predicate Devices
{6}------------------------------------------------
| Specification | HydroPICC-142<br>Access Vascular<br>Proposed Device | HydroPICC-141<br>Access Vascular<br>K172885 |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Key Device Components | Catheter Shaft, Suture Wing,<br>Extension Tube, Luer Hub | Same |
| Short or Long Term Access | Yes | Same |
| Use with Power Injection<br>Power Settings Flow Rate | Yes<br>Flow rate: 3.5mL/sec | Same |
| Catheter Materials | Radiopaque hydrophilic polyol<br>catheter with luer lock hub,<br>polyurethane extension tub, and<br>suture wing without radiopaque<br>additive | Same - suture wing contained<br>radiopaque additive |
| X-Ray Confirmation<br>Required | Yes | Same |
| Sterilization Method | Ethylene Oxide | Same |
| Single Use | Yes | Same |
| How supplied | Convenience kit | Same |
| MRI Safety | Evaluated – labeled with MRI<br>Safety Information | Not evaluated – labeled with<br>contraindication to use in a MR<br>environment |
## Performance Data
Verification and validation tests have been performed in accordance with Design Controls per 21 CFR §820.30. The subject devices met all predetermined acceptance criteria derived from the applicable guidance documents, standards and in-house protocols, which were used to determine appropriate methods for evaluating the performance of the device.
The following performance tests were performed in support of the substantial equivalence determination:
## Mechanical and Performance Testing
The testing and test methods that were performed for HydroPICC-142 are substantially equivalent to testing and tests methods that were performed for the predicate, the HydroPICC-141 (K172885). Testing was conducted in accordance with the following FDA Guidance Document and international standards:
- . FDA's "Guidance on Premarket Notification [510(k)] Submissions for Short-Term and Long-Term Intravenous Catheter"
- . EN ISO 10555-1:2013 – "Intravascular Catheters – Sterile and Single-Use Catheters – Part 1: General Requirements"
- . EN ISO 10555-3:2013 – "Intravascular Catheters – Sterile and Single-Use Catheters – Part 3 Central Venous Catheters:
{7}------------------------------------------------
- . EN ISO 11135:2014 "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for The Development, Validation and Routine Control Of A Sterilization Process For Medical Device"
- . EN ISO 13868:2014 "Catheters – Test method for kinking of single lumen catheters and medical tubing"
- . ASTM F2052:2015 "Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2182: 2011 "Standard Test Method for Measurement of Radio Frequency Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging"
- . ASTM F2213: 2017 "Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment"
- . ASTM F2119: 2013 "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
- USP <788> Particulate Matter in Injections (Method 1)
- Evaluation of Humectant change on Thrombus accumulation compared to predicate (internal specifications)
The following tests were conducted in accordance with the standards identified above as applicable:
- . Power Injection Flow Rate
- Static Burst Strength
- Multiple Power Injections
- Catheter Length
- Dimensional Verification (including ID, OD, Length)
- Catheter Kink/Flex Resistance
- . Tensile Testing (of Catheter and Assembly)
- Particulate testing
- Shelf life testing
- Sterility testing
- Packaging distribution testing
- MR Compatibility testing
#### Biocompatibility Testing
Risk assessment evaluation that the proposed modifications did not impact biocompatibility
## Conclusion
Based on the performance testing, intended use, materials, design, and components the HydroPICC-142 is determined to be substantially equivalent to the predicate HydroPICC-141 (K172885).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.