The iNSitu Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to: - A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia; - Acute traumatic fracture of the femoral head or neck; - Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement. The iNSitu Total Hip System femoral stems are intended for cementless fixation. The iNSitu Total Hip System acetabular cup is intended for cementless fixation. The porous structured surfaces provide biological fixation in a cementless application. The Ray Collared and Collarless Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to: - A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia; - Acute traumatic fracture of the femoral head or neck; - Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement. The Ray Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application. The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System femoral stem for cemented or uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: - Femoral neck and trochanteric fractures of the proximal femur; - Osteonecrosis of the femoral head; - Revision procedures where other devices or treatments for these indications have failed.
Device Story
System comprises lateralized acetabular liners and femoral stems with plasma-sprayed CPTi coating; intended for cementless total hip replacement or hemiarthroplasty. Lateralized liners add 5mm UHMWPE at apical dome for increased offset; femoral stems utilize plasma-sprayed CPTi coating for biological fixation. Used by orthopedic surgeons in clinical settings; provides structural support and joint articulation. Output is physical prosthetic assembly; aids in restoring joint function and reducing pain. Benefits include improved surgical options for offset adjustment and established biological fixation methods.
Clinical Evidence
Bench testing only. Dimensional analysis performed to ensure liner thickness is not worst-case compared to predicate. Range of motion analysis conducted per ISO 21535. Porous structure characterization (physical, chemical, mechanical) for plasma-sprayed coating referenced via Master File MAF 628.
Technological Characteristics
Materials: UHMWPE (liners), CPTi (coating). Principle: Cementless biological fixation via porous structure. Design: Lateralized acetabular liners (5mm offset), femoral stems with plasma-sprayed CPTi. Standards: ISO 21535 (range of motion). Sterilization: Identical to predicate. Connectivity: N/A.
Indications for Use
Indicated for skeletally mature individuals requiring total hip replacement or hemiarthroplasty due to osteoarthritis, traumatic/rheumatoid arthritis, avascular necrosis, congenital hip dysplasia, acute femoral head/neck fractures, or failed previous hip surgeries.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
TheRay Collared and Collarless Hip System (K241875)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 5, 2026
Nextstep Arthopedix, LLC
Garrett Spurgeon
VP - Compliance, Regulatory, and Strategy
1800 Triplett Blvd
Akron, Ohio 44306
Re: K243192
Trade/Device Name: iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi Coating
Regulation Number: 21 CFR 888.3358
Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Regulatory Class: Class II
Product Code: LPH, OQG, LZO, OQI, KWY, MEH
Dated: May 6, 2026
Received: May 6, 2026
Dear Garrett Spurgeon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243192 - Garrett Spurgeon
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243192 - Garrett Spurgeon
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**LIMIN SUN -S**
Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint
Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K243192
Device Name
iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi Coating
Indications for Use (Describe)
The iNSitu Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
- Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
The iNSitu Total Hip System femoral stems are intended for cementless fixation. The iNSitu Total Hip System acetabular cup is intended for cementless fixation. The porous structured surfaces provide biological fixation in a cementless application.
The Ray Collared and Collarless Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
- Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
The Ray Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application.
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System femoral stem for cemented or uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
- Femoral neck and trochanteric fractures of the proximal femur;
- Osteonecrosis of the femoral head;
- Revision procedures where other devices or treatments for these indications have failed.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740 EF
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*'An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.'*
---FORM FDA 3881 (8/23)
Page 2 of 2
{5}
K243192
### 510(k) Summary
The following 510(k) Summary is provided in accordance with 21 CFR 807.92.
### 510(k) Owner and Registration
| Owner's Name: | Theken Companies, LLC Subsidiary: NextStep Arthropedix |
| --- | --- |
| Address: | 1800 Triplett Blvd., Akron, OH 44306 |
| Phone Number: | (330) 733-7600 |
| Fax Number: | (330) 733-7602 |
| Date Summary Prepared: | 6/4/2026 |
| Establishment Registration Number: | 3002498892 |
### 510(k) Contact
| Contact: | Theken Companies |
| --- | --- |
| Address: | 1800 Triplett Blvd., Akron, OH 44306 |
| Phone Number: | 330-733-7600 |
| Fax Number: | 330-733-7600 |
| Contact Person: | Garrett Spurgeon |
### Device Name and Classification
| Device Trade Name: | iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi Coating |
| --- | --- |
| Common Name: | Acetabular Liner, Femoral Stem |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis |
| Regulation Number: | 21 CFR 888.3358 21 CFR 888.3353 |
| Product Codes: | LPH OQG LZO OQI KWY MEH |
| Advisory Panel: | 87 (Orthopedic) |
### Legally Marketed Predicate
The iNSitu Total Hip System (K161184 / K172501 / K191936) is the primary predicate device and included for indications for use and performance comparisons. The iNSitu Bipolar Hip System (K191297) is an additional predicate.
Page 1 of 3
{6}
K243192
# Device Description
# iNSitu Lateralized Acetabular Liners:
The subject iNSitu Lateralized Acetabular Liners are manufactured from the identical material used to manufacture devices cleared in K161184. The devices are also cleaned, packaged, and sterilized in an identical fashion. The subject devices are a line extension to the existing Neutral, Facechanging, and Hooded Acetabular Liner configurations. The subject iNSitu Lateralized Acetabular Liners feature an additional 5mm of UHMWPE material at the apical dome, which allows the surgeon to add offset to the joint reconstruction from the acetabular side of the prosthesis.
# iNSitu Femoral Stems with Plasma Sprayed CPTi Coating:
The subject iNSitu Femoral Stems with plasma sprayed CPTi coating are geometrically identical to the devices reviewed by FDA in K161184 and K172501. The devices are also cleaned, packaged, and sterilized in an identical fashion. The only difference is the application method of the CPTi coating which allows NextStep Arthropedix to manufacture the iNSitu Femoral Stems with either asymmetric CPTi sintered bead or plasma sprayed CPTi coatings.
# Indications for Use:
The iNSitu Total Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
- Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
The iNSitu Total Hip System femoral stems are intended for cementless fixation. The iNSitu Total Hip System acetabular cup is intended for cementless fixation. The porous structured surfaces provide biological fixation in a cementless application.
TheRay Collared and Collarless Hip System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
- Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.
TheRay Collared and Collarless Total Hip System femoral stems are intended for cementless fixation. The acetabular cup is intended for cementless fixation. The porous structured surfaces on the acetabular cup provide biological fixation in a cementless application.
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System femoral stem for cemented or uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
- Femoral neck and trochanteric fractures of the proximal femur;
- Osteonecrosis of the femoral head;
- Revision procedures where other devices or treatments for these indications have failed.
# Indications for Use Comparison:
The subject iNSitu Femoral Stems with plasma-sprayed CPTi coating and Lateralized Acetabular Liners) have identical indications for use to the predicates: iNSitu Total Hip System (K161184, K172501, K191936), the iNSitu Bipolar Hip System (K191297), and the TheRay Collared and Collarless Hip System (K241875).
# Summary of Technological Characteristics
Page 2 of 3
{7}
K243192
The subject iNSitu Femoral Stems with plasma-sprayed CPTi coating feature the identical geometry, sizes, indications, packaging, and sterilization as the femoral stems cleared in K161184/K172501. Only the coating application method is different.
The subject Lateralized Acetabular Liners feature the identical locking mechanism, material, indications, ackaging, and sterilization as the acetabular liners cleared in K161184/K191936. The addition of the material at the apical dome provides surgeons with additional offset options.
## Performance Testing
Preclinical performance testing was conducted on the subject iNSitu Lateralized Liner and iNSitu Femoral Stem w/TPS coating to evaluate the device and to demonstrate substantial equivalence. A dimensional analysis was performed for the subject Lateralized Liners to demonstrate the liner thicknesses were not worst-case when compared to the predicate device (neutral liners K161184). Range of motion analysis was performed in accordance with ISO 21535 for the subject Lateralized Liners. Tests related to the porous structure characterization (physical, chemical, and mechanical) of the coating applied to the subject iNSitu Femoral Stems w/TPS are in Master File MAF 628 (owned by APS Materials).
## Conclusions
The subject Lateralized Liners and iNSitu Femoral Stems w/TPS have the same indications for use as predicate hip systems. A comparison of technological characteristics and performance testing demonstrates that iNSitu Total Hip System and iNSitu Bipolar System is substantially equivalent to the predicate systems.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.