The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stems for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: - Femoral neck and trochanteric fractures of the proximal femur; - . Osteonecrosis of the femoral head; - . Revisions procedures where other devices or treatments for these indications have failed
Device Story
iNSitu Bipolar Hip System is a hemiarthroplasty prosthesis; consists of factory-assembled UHMWPE liner, cobalt chrome outer shell, and UHMWPE retention ring with Ti-6Al-4V ELI spring. Device functions as a self-aligning (eccentric) bipolar head; mates with iNSitu Total Hip System Femoral Stems. Used by orthopedic surgeons in clinical settings for hip hemiarthroplasty. Provides articulation for hip joint; restores joint function in patients with femoral neck/trochanteric fractures or osteonecrosis. Benefits include mechanical stability and anatomical fit via range of outer diameters (38-60 mm).
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included range of motion testing, push-out strength, lever-out strength, and assembly forces, demonstrating equivalence to the predicate BioPro Bipolar Head.
Technological Characteristics
Outer shell: Cobalt chromium (ASTM F75). Liner/Retention ring: UHMWPE (ASTM F648) and Ti-6Al-4V (ASTM F136). Sterilization: Ethylene Oxide. Design: Bipolar, self-aligning (eccentric) head. Outer diameters: 38-60 mm. Inner diameters: 22.225 mm or 28 mm. Assembly: Snap-fit liner-head.
Indications for Use
Indicated for patients requiring uncemented primary or revision hemiarthroplasty of the hip due to femoral neck and trochanteric fractures of the proximal femur, osteonecrosis of the femoral head, or failed previous treatments.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
{0}------------------------------------------------
September 17, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Theken Companies, LLC % Bob Poggie President BioVera 65 Promenade Saint-Louis Notre-Dame-de-L'lle-Perrot, QC, J7V 7P2 CANADA
Re: K191297
Trade/Device Name: iNSitu Bipolar Hip System Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY Dated: August 19, 2019 Received: August 20, 2019
Dear Bob Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|--|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
Indications for Use
| 510(k) Number (if known) | K191297 |
|--------------------------------|---------------------------|
| Device Name | iNSitu Bipolar Hip System |
| Indications for Use (Describe) | |
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stems for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
- Femoral neck and trochanteric fractures of the proximal femur;
- . Osteonecrosis of the femoral head;
- . Revisions procedures where other devices or treatments for these indications have failed
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over - The - Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary 1.
The following 510(k) Summary is provided in accordance with 21 CFR 807.92.
# 1.1 510(k) Owner and Registration
| Owner's Name: | Theken Companies, LLC<br>Subsidiary: NextStep Arthropedix | |
|------------------------------------|-----------------------------------------------------------|--|
| Address: | 1800 Triplett Blvd., Akron, OH 44306 | |
| Phone Number: | (330) 733-7600 | |
| Fax Number: | (330) 733-7602 | |
| Date Summary Prepared: | August 12, 2019 | |
| Establishment Registration Number: | 3002498892 | |
# 1.2 510(k) Contact
| Contact: | BioVera, Inc. |
|-----------------------|---------------------------------------------|
| Address: | Notre-Dame-de-L'Ile-Perrot (QC) J7V 7P2, CA |
| Phone and Fax Number: | 514-901-0796 |
| Contact Person: | Robert A Poggie, PhD |
# 1.3 Device Name and Classification
| Device Trade Name: | iNSitu Bipolar Hip System |
|------------------------------------|--------------------------------|
| Device Common Name: | Hemi-hip prothesis, uncemented |
| Regulation Number and Description: | 21 CFR 888.3390 |
| Device Class: | Class II |
| Product Codes: | KWY |
| Advisory Panel: | 87 (Orthopedic) |
# Legally Marketed Predicate 1.4
NextStep Arthropedix is utilizing the BioPro Bipolar Head as a predicate device (K082705/K100761). The iNSitu Bipolar Hip System features component designs, materials, indications and manufacturing methods that are identical to the BioPro Bipolar Head.
| | iNSitu Bipolar Hip System | BioPro Bipolar Head |
|--------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(K) Number | K191297 | K082705/K100761 |
| FDA Product Code | KWY | KWY |
| DESIGN | | |
| Head-Liner-Shell | CoCr-UHMWPE-CoCr | CoCr-UHMWPE-CoCr |
| Head Outer Diameter | 38 to 60 mm in 1 mm increments | 38 to 60 mm in 1mm increments |
| Self-aligning (eccentric head) | Yes | Yes |
| Liner Inner Diameter | 22.225 or 28 mm | 22.225 or 28mm |
| Liner-Head Assembly | Head snap-fit into bipolar liner | Head snap-fit into bipolar liner |
| UHMWPE retention | Yes | Yes |
| MATERIALS | | |
| Outer Shell | Cobalt chromium (ASTM F75) | Cobalt chromium (ASTM F75) |
| Liner and Retention Ring | UHMWPE (ASTM F648), Ti-6Al-4V<br>(ASTM F136), EO sterilized (not<br>highly crosslinked) | UHMWPE (ASTM F648) and Ti-6Al-4V<br>(ASTM F136), EO sterilized (not highly<br>crosslinked) |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide |
| Hip System Compatibility | iNSitu Total Hip System<br>(K161184/K172501) | BioPro PSL Hip System (K922500) |
{4}------------------------------------------------
#### 1.5 Device Description
The iNSitu Bipolar Hip System consists of a factory assembled UHMWPE (ASTM F648) liner in a cobalt chrome (ASTM F75) outer shell, and UHMWPE (ASTM F648) retention ring with a Ti-6Al-4V ELI (ASTM F136) spring. These Bipolar Heads include outer diameters ranging from 38 to 60 mm, in 1 mm increments, to properly fit the patient anatomy. The smaller Bipolar Heads (38 to 43 mm) have an inner diameter that mates with the subject 22mm diameter Femoral Head; the larger Bipolar Heads (44 to 60 mm) have an inner diameter that mates with the previously cleared 28 mm diameter Femoral Head. The iNSitu Bipolar Hip System may be used in conjunction with an iNSitu Total Hip System Femoral Stem (K161184/K172501) for hemiarthroplasty.
# 1.6 Intended Use
The iNSitu Bipolar Hip System is intended for use in combination with the iNSitu Total Hip System Femoral Stems for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
- Femoral neck and trochanteric fractures of the proximal femur;
- Osteonecrosis of the femoral head;
- Revisions procedures where other devices or treatments for these indications have failed
# Summary of Technological Characteristics 1.7
The iNSitu Bipolar Hip System has an identical design and is manufactured from the same materials as the predicate device system. The iNSitu Bipolar Hip System components are packaged and sterilized using similar processes. The subject system is substantially equivalent to the predicate based on comparisons of intended use, design features, and technological characteristics.
# 1.8 Performance Testing
Extensive preclinical performance testing was conducted, and substantial equivalence determined per K082705 and K100761. The components of the subject device are identical to the predicate BioPro Bipolar Head device, and therefore the predicate device testing demonstrates substantial equivalence for the results confirm that all components of the iNSitu Bipolar Hip System exhibit the appropriate mechanical characteristics for hip hemi-arthroplasty and are substantially equivalent to the predicate devices. Range of motion testing was performed using the iNSitu Bipolar Head and the iNSitu Total Hip System Femoral Stems (K161184/K172501).
The following tests were performed using the BioPro Bipolar Head, which is identical to the iNSitu Bipolar Hip System.
- Push-Out and Lever-Out Strength
- . Assembly Forces
# Conclusions 1.9
The iNSitu Bipolar Hip System is identical to the predicate BioPro Bipolar Head (K082705/K100761). The subject device has the same design features, materials and indications for use as the predicate devices. The testing performed for the predicate and subject devices indicate that the subject iNSitu Bipolar Hip System is substantially equivalent to the predicate BioPro Bipolar Head.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.