Field testing/clinical use data from five patients
Field testing was conducted with five patients to compare the performance of the contact forte Alpha with the predicate device (Sophono Alpha 2) in real-world conditions, evaluating audiometric thresholds, speech perception, and subjective patient experience.
Field testing; Real-world conditions; Patient-reported outcomes; Audiometric thresholds
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Field testing; Head-to-head comparison of subject and predicate device in actual users
Five patients (users of bone conduction systems); Sample Size: 5
Sophono Alpha 2 (predicate device)
Audiometric thresholds, speech perception thresholds, and subjective patient hearing experience
Indications for Use
The contact forte Alpha is a Processor and is intended for use with the Sophono Headband or Softband (no age limitations), or with the Sophono Magnetic Implant (patients 5 years of age and up) for the following patients and indications: - Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1,2, and 3 kHz). - Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies. - Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an Air Conduction Contralateral Routing of Signal (AC CROS). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5,1, 2 and 3 kHz).
Device Story
The contact forte Alpha is a bone-conduction sound processor; converts ambient sound into mechanical vibrations. Input via microphones; signal processed by internal DSP chip; output drives electromechanical transducer. Device magnetically attaches to Sophono Headband, Softband, or Magnetic Implant. Vibrations transfer through skin/skull to cochlea for auditory perception. Used in clinical or home settings; operated by patient or caregiver. Healthcare providers use fitting software to program device to patient-specific audiometric needs. Benefits patients with conductive, mixed, or single-sided deafness by providing amplification where traditional air-conduction aids are unsuitable or ineffective. Symmetrical oval housing allows for left or right-side application.
Clinical Evidence
Field testing conducted with five patients. Head-to-head comparison between contact forte Alpha and Sophono Alpha 2. Primary endpoints included audiometric thresholds and speech perception thresholds. Results showed no measurable difference in performance between devices; patients reported no subjective difference in hearing. Non-clinical testing included biocompatibility (ISO 10993), mechanical performance, and EMC (IEC 60601-1-2).
Technological Characteristics
Bone-conduction sound processor; plastic (ABS, PA) housing. 16-band, 8-channel DSP. Max output 114 dBOFL; max gain 51 dB. Powered by size 13 (PR 48) Zinc-Air battery. Programmable with 4 user-selectable programs. Biocompatible per ISO 10993. EMC compliant per IEC 60601-1-2.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (PTA BC threshold <45 dB HL) or profound unilateral sensorineural hearing loss with normal hearing in the opposite ear (PTA AC threshold <20 dB HL). Suitable for bilateral fitting in symmetric conductive/mixed loss. Compatible with Sophono Headband/Softband (all ages) or Sophono Magnetic Implant (age 5+).
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Predicate Devices
Sophono Bone Conduction Systems (S) Configuration and (M) (K153391)
Reference Devices
Sophono Alpha 2 Sound Processor
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 25, 2025
BHM-Tech Produktionsgesellschaft mbH
Markus Hirtzi
Quality Control, Regulatory Affairs
Grafenschachen 242
Grafenschachen, 7423
Austria
Re: K243041
Trade/Device Name: contact forte Alpha
Regulation Number: 21 CFR 874.3302
Regulation Name: Bone-Conduction Hearing Aid
Regulatory Class: Class II
Product Code: LXB
Dated: March 28, 2025
Received: March 28, 2025
Dear Markus Hirtzi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243041 - Markus Hirtzi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243041 - Markus Hirtzi
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243041
Device Name
contact forte Alpha
Indications for Use (Describe)
The contact forte Alpha is a Processor and is intended for use with the Sophono Headband or Softband (no age limitations), or with the Sophono Magnetic Implant (patients 5 years of age and up) for the following patients and indications:
- Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1,2, and 3 kHz).
- Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies.
- Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an Air Conduction Contralateral Routing of Signal (AC CROS). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5,1, 2 and 3 kHz).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K243041 | 510(k) Summary | Prepared on: 2025-04-25 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | BHM-Tech Produktionsgesellschaft mbH | |
| Applicant Address | Grafenschachen 242 Grafenschachen 7423 Austria | |
| Applicant Contact Telephone | +43 3359 200 78 | |
| Applicant Contact | Mr. Markus Hirtzi | |
| Applicant Contact Email | mhirtzi@bhm-tech.at | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | contact forte Alpha | |
| Common Name | Bone-conduction hearing aid | |
| Classification Name | Hearing Aid, Bone Conduction | |
| Regulation Number | 874.3302 | |
| Product Code(s) | LXB | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K153391 | Sophono Bone Conduction Systems (S) Configuration and (M) | LXB |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The contact forte Alpha Processor is a sound processor that converts sound into mechanical vibrations. The sound processor is magnetically connected to a Sophono Headband, Sophono Softband or Sophono Attract Magnetic Spacer by the integrated metal disk. Sound enters the audio processor through the microphones, which then goes through a DSP chip for signal processing. The output of the DSP chip drives an electromechanical transducer, which converts the signal into mechanical vibrations. These vibrations are transferred to the patient's skin through either the Sophono Headband, Sophono Softband or the Sophono Attract magnetic spacer. The vibrations are carried through the patient's skin and skull to the cochlea where they are perceived as sound. The device is designed for use on either the right or left side application. The device has a "oval" housing with a symmetrical design (for right or left side application). The material used: Plastic - contact with intact skin; | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The contact forte Alpha is a Processor and is intended for use with the Sophono Headband or Softband (no age limitations), or with the Sophono Magnetic Implant (patients 5 years of age and up) for the following patients and indications: • Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1.2, and 3 kHz). • Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies. • Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an Air Conduction Contralateral Routing of Signal (AC CROS). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2 and 3 kHz). | | |
{5}
Indications for Use Comparison
21 CFR 807.92(a)(5)
Same intended use. Both - Sophono (Medtronic) and contact forte Alpha (BHM) - are sound processors that convert sound into mechanical vibrations. The vibrations are carried through the patient's skin and skull to the cochlea where they are perceived as sound.
The contact forte Alpha is developed and manufactured by BHM-Tech Produktionsgesellschaft mbH in Austria. The contact forte Alpha is only a Sound processor and is intended to be used with Sophono Bone Conduction Systems from METRONIC XOMED, INC. This combination reflects the same application possibility as with the Sophono Alpha 2 sound processor, which was initially part of the overall MEDTRONIC-Sophono Bone Conduction System.
Note that the MEDTRONIC system is discontinued and no new sound processors are produced. The combination with the BHM processor allows patient to upgrade to a new sound processor and/or replace their defective processor without having to change the implant.
# Technological Comparison
21 CFR 807.92(a)(6)
Same technological characteristics:
- converting sound into mechanical vibrations
- battery type: 13 (PR 48, Zinc-Air)
- max. output: 114 dBOFL
- full on gain (max): 51dB
- programmable
- frequency processing: 16 band, 8 channels
- number of programs: 4
- individual patient fitting via fitting software
- housing - material: plastic (ABS, PA)
- contact with intact skin only (short term only, e.g. cleaning)
- no direct skin contact when used as intended
1. Intended Use & Indications for Use
Both the contact forte Alpha Sound Processor and the Sophono Alpha 2 Sound Processor are bone conduction hearing devices intended to amplify sound for individuals with conductive, mixed, or single-sided deafness. They are designed to be used with an implantable or non-implantable bone conduction system to enhance auditory perception. There is no significant difference in their intended use, supporting substantial equivalence.
2. Technological Characteristics
Bone Conduction Technology: Both devices use direct bone conduction to transmit sound vibrations to the inner ear.
Sound Processing: The digital signal processing (DSP) in both devices provide sound amplification and frequency adjustments based on patient needs.
Attachment Mechanism: Both models are designed to attach magnetically to an implanted or non-implanted-based coupling system.
Power Source: Both devices are powered by standard zinc-air batteries (Sophono Alpha2 may also be powered by rechargeable batteries), ensuring similar operational performance.
Materials & Biocompatibility: The external casing and skin-contacting materials are biocompatible and conform to ISO 10993 standards.
3. Performance Data & Safety Considerations
Acoustic Performance: Comparative testing has shown that both devices deliver identical gain, output, and speech intelligibility performance.
Electromagnetic Compatibility (EMC): Both processors comply with IEC 60601-1-2 for EMC requirements.
Safety & Effectiveness: No new safety concerns are introduced, and risk assessment (per ISO 14971) confirms that the risk profile remains equivalent between the two devices.
Field testing was conducted with five patients to assess whether the subject device performed as intended under real-world conditions. Results showed that performance with the contact forte Alpha was comparable to that of the predicate device.
# Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Magnetic Retention forces in different clinically significant configurations as well as Output Level Measurements with contact forte Alpha and Sophono Alpha 2 was done.
Good concordance was seen between device outputs across the various clinically relevant configurations. Measurements have shown that there is no difference between the devices which affects the safety and performance of the devices. The acceptance criteria for all defined tests were met.
Field tests: The field measurements showed that Sophono Alpha2 as well as contact forte Alpha are delivering comparable gain and output levels. A head to head comparison of the subject and predicate device showed that audiometric thresholds and speech
{6}
perception thresholds for the two devices were not measurably different for five different users. The patients reported that they did not notice any subjective difference in hearing between the two devices.
Based on the results of nonclinical and clinical testing, the contact forte Alpha has been demonstrated to be as safe, as effective, and to perform as well as the legally marketed predicate device, the Sophono Alpha 2. The nonclinical tests, including biocompatibility, mechanical performance, and electromagnetic compatibility demonstrate compliance with applicable safety standards.
These findings provide reasonable assurance that the contact forte alpha is substantial equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.