Sophono Bone Conduction Systems (S) Configuration and (M) Configuration
K153391 · Sophono, Inc. · LXB · Jun 16, 2016 · Ear, Nose, Throat
Device Facts
Record ID
K153391
Device Name
Sophono Bone Conduction Systems (S) Configuration and (M) Configuration
Applicant
Sophono, Inc.
Product Code
LXB · Ear, Nose, Throat
Decision Date
Jun 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Sophono® Sound Processor is intended for use with the Sophono® Headband (no age limitations), or with the Sophono® Magnetic Implant (patients 5 years of age and up) for the following patients and indications: Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz). Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies. Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2, and 3 kHz).
Device Story
Sophono Bone Conduction Systems consist of sound processors and accessories (headbands or magnetic implants) that convert sound into mechanical vibrations. In (S) configuration, processor is held against head via headband; in (M) configuration, processor is held via magnetic attraction to an implanted magnetic spacer. Vibrations are transduced through skin to bone, stimulating the cochlea. Used by patients with conductive/mixed hearing loss or single-sided deafness. Audiologists use software to program the device based on individual patient hearing loss profiles. Device improves hearing by bypassing damaged outer/middle ear structures. Operated by patients or caregivers; clinical adjustments performed by audiologists.
Clinical Evidence
Bench testing only. Testing included sterilization (ISO 11135-1, ISO 11138-1, EN 1422), biocompatibility (ISO 10993-1/7), electrical/EMC (IEC 60601-1/1-2), software verification (IEC 62304), and MR safety (ASTM F2503, F2052, F2213, F2182, F2119). Mechanical integrity verified via ASTM F88/F88M-09 and F1980-07.
Technological Characteristics
Bone conduction hearing system using transcutaneous energy transfer. Configurations: (S) headband-based or (M) magnetic implant-based. Materials evaluated for biocompatibility (ISO 10993). MR safety standards: ASTM F2503, F2052, F2213, F2182, F2119. Electrical safety: IEC 60601-1. Software validated per IEC 62304.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (PTA BC threshold < 45 dB HL) or profound unilateral sensorineural hearing loss with normal hearing in the opposite ear (PTA AC threshold < 20 dB HL). Headband use has no age limit; magnetic implant use is for patients 5 years and older.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
Cochlear Baha® Auditory Osseointegrated Implant System: Model BI300 Implant and Model BA300 Abutment (K100360)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2016
Sophono, Inc. Ms. Rozanne Paciej Regulatory Affairs Project Manager 5744 Central Ave #100 Boulder. CO 80301
Re: K153391
Trade/Device Name: Sophono Bone Conduction Systems (S) Configuration and (M) Configuration Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: LXB Dated: May 13, 2016 Received: May 16, 2016
Dear Ms. Paciej:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
| 510(k) Number (if known) | K153391 |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sophono (S) & (M) Bone Conduction Hearing Systems |
| Indications for Use (Describe) | The Sophono® Sound Processor is intended for use with the Sophono® Headband (no age limitations), or with the Sophono® Magnetic Implant (patients 5 years of age and up) for the following patients and indications: |
| | <ul><li>Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz). </li><li>Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2, and 4 kHz, or less than 15 dB at individual frequencies. </li><li>Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1, 2, and 3 kHz). </li></ul> |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Sohono Bone Conduction System
K153391
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Sophono, a company that provides bone conduction hearing solutions. The logo consists of an orange ear-shaped symbol on the left and the company name "Sophono" in blue on the right. Below the company name is the tagline "bone conduction hearing solutions" in a smaller, lighter font.
510(k) Summary
| Date summary prepared: | 12/17/2015 |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter/Holder: | Sophono Inc.<br>5744 Central Avenue #100<br>Boulder, CO 80301 |
| Contact: | Rozanne Paciej (Primary)<br>Regulatory Affairs Project Manager<br>Telephone: 904-332-8233<br>Fax: 904-296-2386<br>Email: rozanne.paciej@medtronic.com<br>Paul Smolenski (Alternate)<br>Senior Regulatory Affairs Manager<br>Telephone: 978-698-6065<br>Email: paul.d.smolenski@medtronic.com |
| Trade Name: | Sophono® Bone Conduction Systems |
| Systems: | Sophono® (S) Configuration<br>and Sophono® (M) Configuration |
| Common Name: | Bone Conduction Hearing System |
| Classification Name: | Hearing Aid (21 CFR § 874.3300, Class II, LXB). |
## Predicate Devices:
| Trade/Proprietary Name: | Otomag Bone Conduction Hearing System (Primary Predicate) |
|--------------------------|--------------------------------------------------------------|
| Common/Usual Name: | Bone Conduction Hearing System |
| Classification Name: | Hearing Aid |
| Class/Panel: | Class II, LXB, 21 CFR 874.3300 |
| 510(k) Submitter/Holder: | Sophono, Inc.<br>5744 Central Ave. #100<br>Boulder, CO 80301 |
| 510(k) #s: | K132189 (including K100193, K102199, K123962) |
{4}------------------------------------------------
| Trade/Proprietary Name: | Cochlear Baha® Auditory Osseointegrated Implant System: Model BI300 Implant and Model BA300 Abutment (Secondary Predicate) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Bone Conduction Hearing System |
| Classification Name: | Hearing Aid |
| Class/Panel: | Class II, LXB, 21 CFR 874.3300 |
| 510(k) Submitter/Holder: | Cochlear Americas<br>13059 E. Peakview Ave.<br>Centennial, CO 80111 |
| 510(k) #: | K100360 |
## Device Description
The Sophono® Bone Conduction Hearing Systems are a family of sound processors and accessories that operate on the principle of bone conduction of sound vibrations.
The Sophono® (S) configurations are held directly against the head using either a soft band or a headband with the Sophono™ Sound Processor. The vibration from the Sound Processor is transduced through direct contact with the patient's skin and the bone below.
The Sophono® (M) configuration is magnetically attracted to the Sophono® Sound Processor through a Magnetic Implant, and a Magnetic Spacer. The Magnetic Implant attracts a Magnetic Spacer that is held against the head through magnetic attraction forces to the Magnetic Implant, with the Sound Processor also held magnetically against the Magnetic Spacer. The vibration from the Sound Processor is transduced through direct contact with the patient's skin and the bone below to the inner ear.
The Sophono® Systems are designed for use for those patients with conductive hearing loss, those patients who have sensorineural hearing loss up to 45 dB in combination with their conductive loss, and single sided deafness as defined in the indications for use. The prescriptive formula and adjustments available to the audiologist in the software allow for programming the Sophono® Systems for individual patient hearing loss.
#### Indications for Use
The Sophono® Sound Processor is intended for use with the Sophono® Headband (no age limitations), or with the Sophono® Magnetic Implant (patients 5 years of age and up) for the following patients and indications:
- Patients with conductive or mixed hearing losses, who can still benefit from amplification of sound. . The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz).
{5}------------------------------------------------
- . Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5. 1, 2, and 4 kHz, or less than 15 dB at individual frequencies.
- Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5. 1. 2. and 3 kHz).
## Technological Characteristics
The Sophono® Bone Conduction Hearing Systems transmits audio vibrations through the skin into the bone where sound is sensed by the inner ear / cochlea.
In the Sophono® (S) configuration the Sound Processor is held against the head using a soft band or headband.
In the Sophono (M) configurations the Magnetic Implant is magnetically attracted to the Sophono Sound Processor through a Magnetic Spacer. The Magnetic Implant attracts a Magnetic Spacer that is held against the head through magnetic forces to the Magnetic Implant, with the Sound Processor magnetically attracted to the Magnetic Spacer. The vibration from the Sound Processor is transduced through direct contact with the patient's skin and the bone below.
The intended use, indications for use, function, and general fundamental technological operating principles of the devices have not changed.
Transcutaneous Energy Transfer (TETIN) enables sound (vibration, energy) transfer from the Sophono® Sound Processor, through the patient's skin and magnetic bone implant, to the working cochlea.
## Performance
Evidence of safety and effectiveness were obtained from bench testing, which included:
- Sterilization testing in accordance with ISO11135-1 .
- Sterilization testing in accordance with ISO 11138-1
- Sterilization testing in accordance with EN 1422 ●
- Biocompatibility testing in accordance with ISO 10993-1, ISO 10993-7 ●
- Packaging testing in accordance with ISO 11607-2 ●
- Electrical testing in accordance with IEC 60601-1
- Electromagnetic compatibility testing in accordance with IEC 60601-1-2
- Transportation testing in accordance to ISTA 2A
- Software verification and validation in accordance with IEC 62304 ●
- ASTM F88/F88M-09. Standard Test Method for Seal Strength of Flexible Barrier Materials ●
- ASTM F1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical ● Devices
{6}------------------------------------------------
- ASTM F2503-08, Standard Practice for Marking Medical Devices and Other Items for Safety in the ● Magnetic Resonance Environment
- ASTM F2052-06e1, Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- ASTM F2213-06, (Reapproved 2011), Standard Test Method for Measurement of Magnetically ● Induced Torque on Medical Devices in the Magnetic Resonance Environment
- . ASTM F2182-11a, Standard Test Method for Measurement of Radio Frequency Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging
- ASTM F2119-07, Standard Test Method for Evaluation of MR Image Artifacts from Passive ● Implants
- Engineering and functional testing including system, mechanical, electrical and general functional . testing
## Conclusion
The results of the testing demonstrate that the proposed Sophono® Bone Conduction Hearing Systems subject devices operated as intended and are substantially equivalent to the primary predicate subject device, Otomag Bone Conduction Hearing System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.