K132189 · Sophono, Inc. · LXB · Nov 22, 2013 · Ear, Nose, Throat
Device Facts
Record ID
K132189
Device Name
OTOMAG BONE CONDUCATION HEARING SYSTEM
Applicant
Sophono, Inc.
Product Code
LXB · Ear, Nose, Throat
Decision Date
Nov 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Otoman Alpha Sound Processor is intended for use with the Otomag Headbard or Otomag Softband (no age limitations), or with the Otomon Magnetic Implant (patients 5 years of age and up). This is the same intended use, and for the same patient population as the current legally marketed device.
Device Story
Otomag Bone Conduction Hearing System consists of sound processors and accessories (Headband, Softband, or Magnetic Implant). Device operates via bone conduction of sound vibrations; sound processor attaches magnetically to headband/softband or implanted magnet. Vibration transduced through direct contact with skin and underlying bone. Used by patients with conductive, mixed, or single-sided deafness. Audiologist uses software to program prescriptive formulas and adjustments for individual hearing loss. System benefits patients by providing amplification for those unable to use traditional air conduction devices.
Clinical Evidence
Bench testing only. Validation included packaging process, shipping, sterilization (SAL 10^-6), sterilant residuals (ISO 10993-7), seal integrity, and shelf life. Biocompatibility of materials confirmed per ISO 10993-1.
Technological Characteristics
Bone conduction hearing system. Components: Sound processor, headband, softband, magnetic implant, surgical/implant templates. Materials: HDPE (templates). Sterilization: Ethylene Oxide (EO). Standards: ISO 14971, ISO 15223-1, ISO 11135-1, ISO 11607-1/2, ISO 10993-1/7, ASTM F1980, ASTM F1929, ASTM F88. Connectivity: Software-based programming via audiologist interface.
Indications for Use
Indicated for patients with conductive or mixed hearing loss (PTA BC threshold better than 45 dB HL at 0.5, 1, 2, 3 kHz) or profound sensorineural hearing loss in one ear with normal hearing in the opposite ear (PTA AC threshold better than 20 dB HL at 0.5, 1, 2, 3 kHz). Bilateral fitting indicated for symmetric conductive/mixed loss. Magnetic implant indicated for patients 5 years and older; headband/softband has no age limitation.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
KI32189
Image /page/0/Picture/2 description: The image shows the logo for Sophono. The logo consists of a stylized ear shape on the left, followed by the word "Sophono" in a bold, sans-serif font. Below the word "Sophono" is a tagline in a smaller font, but the text is not legible. The logo is black and white.
5744 Central Avenue #100, Boulder, CO 80301 Telephone: +1 720-407-5160 • Fax: +1 720-407-5168 • www.sophomo.com.
# 510(k) Summary: Otomag Bone Conduction Hearing System
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and title of the Code of Federal Requiations Part 807 (21 CFR §807), and in particular §807.92, this summary of substantial equivatence information is being submitted.
Submitted By: Sophono, Inc. 5744 Central Avenue #100 Boulder, Colorado 80301
Establishment Registration Number: 3008514292
Contact Person:
Krista B. Travnor, M.A., RAC Director of Regulatory Affairs Ph: 720 407-5160 Fax: 720 407-5168 Email: krista.traynor@sophono.com
Date Prepared: July 12, 2013
Trade or Proprietary Name: Otomag Bone Conduction Hearing System
Regulation Number: 21 CFR §874.3300
Regulation Name: Hearing Aid, Bone Conduction
Regulatory Class: Class II
Product Code: LXB
Panel: Ear Nose and Throat Specialty Panel
Predicate Device: Otomag Bone Conduction Hearing System - K102199
Device Description:
The subject of this Traditional 510(k) is to obtain clearance for a modification in the method of sterifization and packaging for the Alpha (M) Magnetic Implant, cleared under 510(k) K102199, and the Surgical and Implant Templates, via a Letter to File under 510(k) K102199, and to supply these components of the system as sterile. The predicate device components are provided nonsterile and must be sterifized prior to implantation by gravity steam (autoclaving). The proposed method of sterilization for these components is by Ethylene Oxide (EO).
The Otomag™ Bone Conduction Hearing System is a family of sound processors and accessories that operate on the principle of bone conduction of sound vibrations.
The Otomag™ System is configured in either of two configurations. The first configuration is the Alpha (S), where the Otomag™ Sound Processor is attached magnetically to a Headband or Softband. The second configuration is the Alpha (M), where the Otomag™ Sound Processor is attached magnetically to an implanted magnet. The Headband, Softband, or Magnetic Implant holds the sound processor against the head, and vibration is transduced through direct contact with the patient's skin and the bone below.
The Otomag™ System is designed for use for those patients with conductive hearing loss, those patients who have sensorineural hearing toss up to 45 dB in combination with their conductive loss, and single sided deafness as defined in the indications for use. The prescriptive formula and adjustments avoilable to the audiologist in the software allow for programming the Ottomage System for individual patient hearing loss.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Sophono. The logo consists of a stylized ear shape on the left, followed by the word "Sophono" in a bold, sans-serif font. Below the word "Sophono" is some smaller text that is difficult to read. The logo appears to be for a company that specializes in hearing solutions.
#### Intended Use:
The Otoman Alpha Sound Processor is intended for use with the Otomag Headbard or Otomag Softband (no age limitations), or with the Otomon Magnetic Implant (patients 5 years of age and up). This is the same intended use, and for the same patient population as the current legally marketed device.
#### Technological Characteristics:
Technological characteristics of the modified device that are equivalent to throse of the predicate device are:
- Overall device performance and operating principle ●
- . Fundamental technology of the device
Technological characteristics that are different between the modified device and the predicate are:
- Packaging and expiration dating: The Magnetic Implant, Surgical Template and Implant . Template are supplied on a high-density polyethylene (HDPE) card in double Tyvelt pouches.
- Method of Sterilization: The method is by Ethylene Oxide (EO) .
- Patient Contact Materials: The Surgical Template is now composed of high-density . polyethyiene (HDPE) plastic instead of anodized aluminum.
#### Performance Standards:
The changes reported in this submission for the Alpha (M) Magnetic Implant and accessories conform to applicable requirements of the following National and International Standards.
ISO 14971 Second edition 2007-03-01, Medical devices - Application of risk management to medical devices
ISO 15223-1 Second Edition 2012-07-01, Medical devices - Symbols to be used with medical device labels, labeling, and information to be supplied - Part 1: General requirements
DIN EN ISO 11135-1:2007 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices
EN 556-1:2006 Sterifisation of Medical Devices - Requirements for medical devices to be designated "STERILE" Part 1: Requirements for medical devices that have been sterlised in their final packaging
DIN EN ISO 11138-1:2008 Sterilisation of Health Care Products Biological Indicators- Part 1: General
EN 1422:2009 Ethylene Oxide Sterilisers- Requirements and Test Procedures
AAMI ANSI ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
DIN EN ISO 10993-7:2009 Birological Evaluation of Medical Devices - Part 7: Ethylene Oxide-Sterilisation Residuals
AAMI ANSI ST72:2011 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the word "Sophono" in a stylized font. To the left of the word is a symbol that resembles a question mark. Below the word "Sophono" is some smaller text that is difficult to read due to the image quality.
DIN 58948-7 2010 Sterilisation- Low Temperature Sterilisers - Part 7: Installation Requirements and Service Supply Requirements for Ethylene Oxide Sterilisers
ISO 11607-1: 2006 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
ISO 11607-2 First edition 2006-04-15 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
DIN EN 868-5-2009 - Packaging for terminally sterilized medical devices - Part 5: Sealable pouches and reels of porous materials and plastic film construction - Requirements and test methods
ISTA 2A 2011 Preshipment Test Procedures
ASTM F1980-07 (Reapproved 2011). Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
ASTM F1929-98 (Reapproved 2004), Standard Test Method for Detecting Seal Leaks in Porqus Medical Packaging by Dye Penetration
ASTM F88F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials
## Non-Clinical Performance Testing:
Testing included the following:
- Packaging Process Validation: to assure that the sealing of the Tyvels pouches maintains . package integrity
- . Shipping Validation: to assure product and package meet requirements after exposure to environmental conditions
- . Sterilization Validation: to assure a stertlity assurance level (SAL) of 10 " using one half cycle
- . Sterilant Residuals: to assure that the residuals that remain on the device are considered acceptable per ISO 10993-7:2009. Bictogical Evaluation of Medical Devices - Part 7: Ethylene Oxide-Sterilization Residuals.
- . Packaging Validation: to assure seal integrity of the pouches that they would withstand the rigors of the intended sterilization and distribution environments
- Shelf Life Validation: to validate that product and packaging meet requirements after . defined shelf life
The modified device has been subjected to extensive safety and vertication testing. The testing information presented in this submission demonstrates the device is as safe and effective and substantially equivalent to the predicate device.
### Conclusion:
The EO sterilization method, change in patient contact materials, and modified packaging and labeling does not alter the intended use, indications for use or fundamental scientific technological characteristics of the device and all processes have been validated. The materials that compose the Maanetic Implant Template and Surgical Template meet the requirements of ISO 10993-1 for biocompatibility. Based on non-clinical testing, the changes have been appropriately validated and raise no new questions of safety and effectiveness. The Otomag Bone Conduction Hearing System performs the same and is as safe and effective and considered to be substantially equivalent to the previously 510(k) cleared predicate device (K102199).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The logo is presented in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2013
Sophono, Inc. c/o Krista Traynor, M.A., RAC Director of Regulatory Affairs 5744 Central Avenue #100 Boulder, CO 80301
Re: K132189
Trade/Device Name: Otomag Bone Conduction Hearing System Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid, Bone Conduction Regulatory Class: Class II Product Code: LXB Dated: August 26, 2013 Received: August 27, 2013
Dear Ms. Traynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device ' referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Krista Traynor, M.A., RAC
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K132189
Device Name: Otomag Bone Conduction Hearing System
Indications for Use:
The Otomag™ Alpha Scund Processor is intended for use with the Otomag™ Headband or Otomag™ Softband (no age limitations), or with the Otomag™ Magnetic Implant (patients 5 years of age and up) for the following patients and indications:
- Patients with conductive or mixed hearing losses, who can still benefit from amplification of . sound. The pure tose average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2, and 3 kHz).
- Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed . hearing loss. The difference between the left and right sides BC thresholds should be less than 100B on average measured at 0.5, 1, 2, and 4 kHz. or less than 15 dB at individual frequencies.
- Patients who have a profound sensorineural hearing loss in one ear and normal bearing in the ● opposite ear, who for some reason will not or cannot use an AC CROS. The pure tome average (PTA) air conduction (AC) threshold of the hearing ear should be better than 20 dB HL (measured at 0.5, 1. 2 and 3 kHz).
Prescription Uze AND/OR Over-The-Counter Use 1 (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PACE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Image /page/5/Picture/10 description: The image shows the text "Cherish R. Giusto -S". The letters "R" and "G" are stylized with a geometric pattern. The text is in black and the background is white.
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.