3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan
—
302 non-contrast series from six centers across three US states
>1 (board certified cardiologists)
Centerline Extraction
—
True Placement Percentage > 78%
True Placement Percentage 90.4% [88.5%, 92.2%]
3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan
—
116 contrast-enhanced series from six centers across three US states
>1 (board certified cardiologists)
Vessel Labelling
—
F1 Score > 70%
F1 Score 78.4% [76.1%, 80.5%]
3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan
—
116 contrast-enhanced series from six centers across three US states
>1 (board certified cardiologists)
Lumen Wall Segmentation
—
Dice Score > 0.80
Dice Score 0.8996 (0.8938, 0.9055)
3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan
—
63 contrast-enhanced series from six centers across three US states
>1 (board certified cardiologists)
Vessel Wall Segmentation
—
Dice Score > 0.80
Dice Score 0.9016 (0.8962, 0.9070)
3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan
—
63 contrast-enhanced series from six centers across three US states
>1 (board certified cardiologists)
Indications for Use
Salix Central is a web-based software application that is intended to be used for viewing, post-processing, and analyzing cardiac computed tomography (CT) images acquired from a CT scanner in a Digital Imaging and Communications in Medicine (DICOM) Standard format. This software provides tools that can be used for the qualitative and quantitative assessment of physician-identified coronary plaques and stenosis in coronary computed tomography angiography (CCTA) and to perform calcium scoring in non-contrast cardiac CT. Salix Central is intended to complement standard care as an adjunctive tool and is not intended as a replacement to a medical professional's comprehensive diagnostic decision-making process. The software's semi-automated features are intended for an adult population and should only be used by qualified medical professionals experienced in examining and evaluating cardiac CT images.
Device Story
Salix Central is a web-based, cloud-hosted (AWS) SaaS application for cardiac CT image analysis. It ingests DICOM-formatted cardiac CT images from scanners. Using machine learning (ML) algorithms, it performs semi-automated tasks including centerline extraction, vessel labeling, lumen/vessel wall segmentation, and calcium scoring. Physicians use the software to review these outputs, identify coronary plaques and stenosis, and generate reports. It serves as an adjunctive tool to complement standard clinical care. It is operated by radiologists, cardiologists, or clinical specialists in a clinical setting. The device aids in the qualitative and quantitative assessment of CAD, supporting the physician's diagnostic decision-making process.
Clinical Evidence
Bench-only validation using 363 de-identified cardiac CT studies from 6 US centers. Testing evaluated calcium scoring (Pearson correlation 0.958), centerline extraction (90.4% true placement), vessel labeling (78.4% F1 score), and lumen/vessel wall segmentation (Dice scores 0.8996/0.9016). Reference ground truth established by board-certified cardiologists. Validation dataset included diverse scanner manufacturers and CAD-RADS disease severity levels. All performance metrics exceeded pre-defined acceptance criteria.
Technological Characteristics
Web-based SaaS application hosted on AWS. DICOM 3.0 compliant. Employs ML-enabled algorithms for image processing. Features include 2D/3D imaging, multiplanar reformatting (MPR), and manual/semi-automatic centerline extraction and wall segmentation. Connectivity via secure network server integration. Software development follows IEC 62304:2015 and ISO 13485:2016 standards.
Indications for Use
Indicated for adult patients undergoing CCTA or non-contrast cardiac CT for evaluation of coronary artery disease (CAD) or suspected CAD. Used by qualified medical professionals (cardiologists, radiologists, clinical specialists) for viewing, post-processing, and analyzing cardiac CT images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
Artrya Limited Gabriela Quijada Quality and Regulatory Manager 1257 Hay Street West Perth PERTH. WA 6005 AUSTRALIA
March 27, 2025
Re: K243038 Trade/Device Name: Salix Central Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: OIH Dated: February 27, 2025 Received: February 27, 2025
Dear Gabriela Quijada:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
#### K243038
Device Name
Salix Central
#### Indications for Use (Describe)
Salix Central is a web-based software application that is intended to be used for viewing, post-processing, and analyzing cardiac computed tomography (CT) images acquired from a CT scanner in a Digital Imaging and Communications in Medicine (DICOM) Standard format.
This software provides tools that can be used for the qualitative assessment of physicianidentified coronary plaques and stenosis in coronary computed tomography angiography (CCTA) and to perform calcium scoring in non-contrast cardiac CT.
Salix Central is intended to complement standard care as an adjunctive tool and is not intended as a replacement to a medical professional's comprehensive diagnostic decision-making process. The software's semi-automated features are intended for an adult population and should only be used by qualified medical professionals experienced in examining cardiac CT images.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K243038
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Date Prepared: | February 25, 2025 |
|------------------------|-----------------------------------------------------------------------------------------------|
| Submitter: | Artrya Ltd. |
| Official Contact: | Gabriela Quijada<br>1257 Hay Street<br>West Perth, 6005, Australia<br>Phone: +61 405 653 605 |
| Proprietary Name: | Salix Central |
| Common Name: | System, Image Processing, Radiological |
| Classification: | Class II Medical Device<br>Requlation Number: 21 CFR 892.2050<br>Product Code: QIH |
| Predicate Device: | cvi42 Auto Imaging Software Application (K213998) |
| Reason For Submission: | New Traditional 510(k) |
#### Indications for Use
Salix Central is a web-based software application that is intended to be used for viewing, postprocessing, and analyzing cardiac computed tomography (CT) images acquired from a CT scanner in a Digital Imaging and Communications in Medicine (DICOM) Standard format.
This software provides tools that can be used for the qualitative and quantitative assessment of physician-identified coronary plaques and stenosis in coronary computed tomography angiography (CCTA) and to perform calcium scoring in non-contrast cardiac CT.
Salix Central is intended to complement standard care as an adjunctive tool and is not intended as a replacement to a medical professional's comprehensive diagnostic decision-making process. The software's semi-automated features are intended for an adult population and should only be used by qualified medical professionals experienced in examining and evaluating cardiac CT images.
## Device Description
Salix Central is a web-based software application, hosted on AWS cloud services, delivered using a SaaS model. The software provides interactive, post-processing tools for trained radiologists or cardiologists for viewing, and characterizing cardiac computed tomography (CT)
{5}------------------------------------------------
image data obtained from a CT scanner. The physician-driven coronary analysis is used to review CT image data to prepare a standard coronary report that may include the presence and extent of physician-identified coronary plaques (i.e., atherosclerosis) and stenosis, and assessment of calcified plaque (calcium scoring). The Cardiac CT image data are physicianordered and typically obtained from patients who underwent CCTA or CAC CT for evaluation of CAD or suspected CAD.
A reference standard was created to train the machine (ML) enabled algorithms in Salix Central. Data for training and validation sampled from our database based on pre-defined exclusion criteria. Data were also selected to ensure no site was shared by training and validation cohorts. The training dataset included 3835 unique patient cases from 6 unique sites from the USA, Canada, Australia and Japan.
## Substantial Equivalence Discussion
Salix Central is substantially equivalent to the predicate, cvi42 Auto Imaging Software Application (K213998). Both devices and the reference device, iNtuition-Structural Heart Module (K191585), have the same Intended Use/Indications for Use and are similar with respect to key features and performance characteristics. The Predicate Device & Reference Device Comparison Table below provides a comparison between the Salix Central and the predicate device cvi42 Auto Imaging Software Application (K213998) and reference device, iNtuition-Structural Heart Module (K191585):
| Predicate Device & Reference Device Comparison | | | |
|------------------------------------------------|-------------------------|--------------------------------------------|--------------------------------------|
| Topic | Proposed Device | Predicate Device | Reference Device |
| Manufacturer | Artrya Ltd. | Circle Cardiovascular<br>Imaging Inc. | TeraRecon, Inc. |
| Model Name | Salix Central | cvi42 Auto Imaging<br>Software Application | iNtuition-Structural Heart<br>Module |
| 510(k) Number | Subject Device: K243038 | K213998 | K191585 |
{6}------------------------------------------------
| Predicate Device & Reference Device Comparison | | | |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Topic | Proposed Device | Predicate Device | Reference Device |
| Manufacturer | Artrya Ltd. | Circle Cardiovascular<br>Imaging Inc. | TeraRecon, Inc. |
| Model Name | Salix Central | cvi42 Auto Imaging<br>Software Application | iNtuition-Structural Heart<br>Module |
| 510(k) Number | Subject Device: K243038 | K213998 | K191585 |
| Intended Use /<br>Indications for<br>Use | Salix Central is a web-<br>based software application<br>that is intended to be used<br>for viewing, post-<br>processing, and analyzing<br>cardiac computed<br>tomography (CT) images<br>acquired from a CT<br>scanner in a Digital Imaging<br>and Communications in<br>Medicine (DICOM)<br>Standard format.<br>This software provides<br>tools that can be used for<br>the qualitative and<br>quantitative assessment of<br>physician-identified<br>coronary plaques and<br>stenosis in coronary<br>computed tomography<br>angiography (CCTA) and to<br>perform calcium scoring in<br>non-contrast cardiac CT. | cvi42 Auto is intended to<br>be used for viewing, post-<br>processing, qualitative<br>and quantitative<br>evaluation of<br>cardiovascular magnetic<br>resonance (MR) and<br>computed tomography<br>(CT) images in a Digital<br>Imaging and<br>Communications in<br>Medicine (DICOM)<br>Standard format. It<br>enables a set of tools to<br>assist physicians in<br>qualitative assessment of<br>cardiac images and<br>quantitative<br>measurements of the<br>heart and adjacent<br>vessels; perform calcium<br>scoring; and to confirm<br>the presence or absence<br>of physician-identified<br>lesion in blood vessels. | iNtuition-Structural Heart<br>Module is a software<br>solution that is intended to<br>assist Cardiologists,<br>Radiologists and Clinical<br>Specialists with the<br>visualization and<br>measurements of<br>structures of the heart and<br>vessels. iNtuition-<br>Structural Heart Module<br>enables the user to: -<br>Visualize and measure<br>(diameters, lengths,<br>angles, areas and<br>volumes) structures of the<br>heart and vessels for pre-<br>operative planning and<br>sizing for cardiovascular<br>interventions and surgery,<br>and for post- operative<br>evaluation. - Quantify<br>calcium (volume, density)<br>- iNtuition-Structural Heart<br>Module has the following |
| Predicate Device & Reference Device Comparison | | | |
| Topic | Proposed Device | Predicate Device | Reference Device |
| Manufacturer | Artrya Ltd. | Circle Cardiovascular<br>Imaging Inc. | TeraRecon, Inc. |
| Model Name | Salix Central | cvi42 Auto Imaging<br>Software Application | iNtuition-Structural Heart<br>Module |
| 510(k) Number | Subject Device: TBD | K213998 | K191585 |
| | Salix Central is intended to<br>complement standard care<br>as an adjunctive tool and is<br>not intended as a<br>replacement to a medical<br>professional's<br>comprehensive diagnostic<br>decision-making process.<br>The software's semi-<br>automated features are<br>intended for an adult<br>population and should only<br>be used by qualified<br>medical professionals<br>experienced in examining<br>and evaluating cardiac CT<br>images. | The target population for<br>cvi42 Auto's manual<br>workflows is not<br>restricted; however, Salix<br>Central's semi-automated<br>machine learning<br>algorithms are intended<br>for an adult population.<br>cvi42 Auto shall be used<br>only for cardiac images<br>acquired from an MR or<br>CT scanner. It shall be<br>used by qualified medical<br>professionals,<br>experienced in examining<br>and evaluating<br>cardiovascular MR or CT<br>images, for the purpose of<br>obtaining diagnostic<br>information as part of a<br>comprehensive diagnostic<br>decision-making process. | tools and features that<br>facilitate: - Automatic and<br>manual centerline<br>detection. - Segmentation<br>of cardiovascular<br>structures |
| Device Class | II | II | II |
| Device<br>Classification | QIH | LLZ, QIH | LLZ |
| Intended Users | Cardiologists, Radiologists<br>and Clinical Specialists | Qualified Medical<br>Professionals | Cardiologists,<br>Radiologists and Clinical<br>Specialists |
| Operating<br>Platform | Client-Server Google<br>Chrome Application | macOS, Microsoft<br>Windows | Microsoft Windows |
| DICOM<br>Compliant | Yes; DICOM 3.0 or higher | Yes | Yes |
| Image<br>Acquisition | CT | MR and CT | CT, MR, Nuc, PET, Angio,<br>US/Echo, SPECT |
| Secured<br>Network Server<br>Integration | Yes | Yes | Yes |
| Store Images | Yes | Yes | Yes |
| 2D Imaging | Yes | Yes | Yes |
| 3D Imaging | Yes | Yes | Yes |
| Multiplanar<br>Reformat<br>(MPR) | Yes | Yes | Yes |
| Study Analysis<br>- Navigation<br>Tools | Panning<br>Windowing<br>Zooming | Panning<br>Windowing<br>Zooming | Yes |
| Predicate Device & Reference Device Comparison | | | |
| Topic | Proposed Device | Predicate Device | Reference Device |
| Manufacturer | Artrya Ltd. | Circle Cardiovascular<br>Imaging Inc. | TeraRecon, Inc. |
| Model Name | Salix Central | cvi42 Auto Imaging<br>Software Application | iNtuition-Structural Heart<br>Module |
| 510(k) Number | Subject Device: TBD | K213998 | K191585 |
| | Series, slices, and phases | Series, slices, and phases | Series, slices, and phases |
| Study Analysis<br>- Visualization<br>and Editing | Centerline<br>Wall<br>Signal Intensity Overlay | Yes | Yes |
| Measurements | Distance<br>Diameter<br>Area<br>Hounsfield Unit (HU)<br>Lumen % diameter<br>reduction | Distance<br>Perimeter<br>Area<br>Hounsfield Unit (HU)<br>Volume | Yes |
| Centerline<br>Extraction and<br>Wall<br>Segmentation | Manual and semi-automatic<br>/ user-editable | Manual and semi-<br>automatic using Machine<br>Learning technique | Semi-automatic |
| Calcium<br>Scoring | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.