iNtuition-Structural Heart Module

K191585 · Terarecon,Inc. · LLZ · Jul 12, 2019 · Radiology

Device Facts

Record IDK191585
Device NameiNtuition-Structural Heart Module
ApplicantTerarecon,Inc.
Product CodeLLZ · Radiology
Decision DateJul 12, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Indications for Use

iNtuition-Structural Heart Module is a software solution that is intended to assist Cardiologists, and Clinical Specialists with the visualization and measurements of the heart and vessels. iNtuition-Structural Heart Module enables the user to: - Visualize and measure (diameters, lengths, angles, areas and volumes) structures of the heart and vessels for pre-operative planning and sizing for cardiovascular interventions and for post-operative evaluation. - Quantify calcium (volume, density) iNtuition-Structural Heart Module has the following tools and features that facilitate: - Automatic and manual centerline detection. - Segmentation of cardiovascular structures. - Measurement tools (diameters, lengths, areas, volumes, angles) for the dimensions of vessels and structures. - Calcium quantification and scoring. - Various visualization techniques: 2D/3D/4D visualization, MPR, Stretched MPR, MIP, MinIP, Raysum and MAR. - Capture and Report.

Device Story

Software module for image post-processing; integrates with iNtuition platform. Inputs: medical images (CT, MR, PET, Angio, US, SPECT). Operation: provides tools for automatic/manual centerline detection, segmentation of cardiovascular structures, and quantitative measurements (diameters, lengths, angles, areas, volumes). Visualization: 2D/3D/4D, MPR, MIP, MinIP, Raysum, MAR. Used in clinical settings by cardiologists/radiologists/specialists for pre-operative planning and post-operative evaluation. Output: quantitative analysis, reports, and visualizations. Healthcare providers use output to assess anatomy, identify calcifications/aneurysms, and plan vascular procedures. Benefits: facilitates reliable, guided clinical workflows for cardiovascular interventions.

Clinical Evidence

No clinical data; bench testing only. Verification and validation testing performed according to internal protocols to confirm software requirements, risk management, and functional equivalence to predicates.

Technological Characteristics

Software-only module; runs on Windows-based workstations or servers. Connectivity: DICOM-compliant. Visualization: 2D/3D/4D, MPR, MIP, MinIP, Raysum, MAR. Features: automatic/manual centerline detection, segmentation, calcium scoring, measurement tools. Standards: ISO 14971 (risk management), IEC 62304 (software lifecycle), IEC 62366-1 (usability), NEMA PS 3.1-3.20 (DICOM).

Indications for Use

Indicated for cardiologists, radiologists, and clinical specialists to visualize and measure heart and vessel structures for pre-operative planning, sizing for cardiovascular interventions/surgery, and post-operative evaluation; includes calcium quantification (volume, density).

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo includes a stylized depiction of the Rod of Asclepius, a symbol associated with medicine and healing. The text "U.S. FOOD & DRUG" is prominently displayed, followed by "ADMINISTRATION" in a smaller font size. The logo is clean and professional, reflecting the FDA's role in regulating and overseeing food and drug safety. July 12, 2019 TeraRecon, Inc. % Megha Jain QARA Manager 4000 East, 3rd Ave, Suite 200 FOSTER CITY CA 94402 Re: K191585 Trade/Device Name: iNtuition-Structural Heart Module Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: June 11, 2019 Received: June 14, 2019 Dear Megha Jain: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K 191585 #### Device Name iNtuition - Structural Heart Module 4.4 #### Indications for Use (Describe) iNtuition-Structural Heart Module is a software solution that is intended to assist Cardiologists, and Clinical Specialists with the visualization and measurements of the heart and vessels. iNtuition-Structural Heart Module enables the user to: - · Visualize and measure (diameters, lengths, angles, areas and volumes) structures of the heart and vessels for pre-operative planning and sizing for cardiovascular interventions and for post-operative evaluation. - · Quantify calcium (volume, density) iNtuition-Structural Heart Module has the following tools and features that facilitate: - · Automatic and manual centerline detection. - · Segmentation of cardiovascular structures. - · Measurement tools (diameters, lengths, areas, volumes, angles) for the dimensions of vessels and structures. - · Calcium quantification and scoring. - · Various visualization techniques: 2D/3D/4D visualization, MPR, Stretched MPR, MIP, MinlP, Raysum and MAR. - · Capture and Report. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### K191585 Date of Summary: 09 July 2019 | 510(k) Summary of Safety and Effectiveness<br>(As required by 21 CFR 807.92(c)) | iNtuition-Structural Heart Module | |---------------------------------------------------------------------------------|-----------------------------------| |---------------------------------------------------------------------------------|-----------------------------------| Submitter: TeraRecon, Inc. Applicant: 4000 E 3rd Ave, Suite 200 Sponsor: Foster City, CA 94402 Establishment: Registration#: 2954793 Contact erson: Magha Jain QARA Manager Ph: 650 -372-1100 Fax: 650 -372-1101 Email: mjain@terarecon.com ### Device Information: | Name of Device: | iNtuition-Structural Heart Module | |------------------------|---------------------------------------------------------| | Common Name: | Medical Imaging System | | Classification Name: | § 892.2050, Picture Archiving and Communications System | | Product Code: | LLZ | | Classification Panel: | Radiology | | Device Classification: | Class II | ### Predicate Devices:: - 3mensio Workstation (K153736) (Primary predicate device) - - iNtuition (K121916) (Secondary predicate device) - ### Indications for Use: iNtuition-Structural Heart Module is a software solution that is intended to assist Cardiologists, Radiologists and Clinical Specialists with the visualization and measurements of structures of the heart and vessels. iNtuition-Structural Heart Module enables the user to: - Visualize and measure (diameters, lengths, angles, areas and volumes) structures of . the heart and vessels for pre-operative planning and sizing for cardiovascular interventions and surgery, and for post-operative evaluation. - Quantify calcium (volume, density). . iNtuition-Structural Heart Module has the following tools and features that facilitate: - Automatic and manual centerline detection. . - Segmentation of cardiovascular structures. . - Measurement tools (diameters, lengths, areas, volumes, angles) for the dimensions . of vessels and structures. {4}------------------------------------------------ - . Calcium quantification and scoring. - Various visualization techniques: 2D/3D/4D visualization, MPR, Curved MPR, . Stretched MPR, MIP, MinIP, Raysum and MAR. - . Capture and Report. ## Device Description: iNtuition-Structural Heart Module is an optional image post-processing software module using iNtuition (K121916) standard features and part of its optional features. It is a software device generally used with the off-the-shelf hardware, offered in various configurations, with the simplest configuration being a stand-alone workstation capable of image review, communications, archiving, database maintenance, remote review, reporting and basic 3D capabilities. It can also be configured as a server with some, or with all, or none of its optional features. Whether provided as a workstation or a server, the iNtuition-Structural Heart Module software is designed to provide access by a local user physically sitting at the computer hosting the iNtuition server software, and/or by one or more remote users who concurrently connect to the server using a freely-downloadable thin client application or through a zero-footprint web viewer (with conference capabilities) over local network or internet. iNtuition-Structural Heart Module is iNtuition (K121916) based optional feature and employs all standard features offered by iNtuition such as convenient image manipulation tools like drawing of region of interests, manual and automatic segmentation of structures, image assessment and measurement tools - linear, diameter, angle, area and volume and tools that support the creation of reports, transmitting and storing this report in digital form and tracking historical information about the studies analyzed by the software. iNtuition Vessel analysis and calcium scoring features are utilized to support automatic and manual centerline extraction and analysis and calcium quantification. iNtuition-Structural Heart Module: - . Supports the visualization and quantification of coronary vessels and cardiac structures for anatomic and pre- or post-operative evaluations through guided clinical workflows. - . Enables the assessment and measurement of different structures of the heart, e.g. aorta, aortic valves, mitral valve, pulmonic valve, atria and atrial appendages, and ventricles. - . Provides analysis of the feasibility of a transapical, transfemoral or subclavian approach to structures for replacement or repair procedures via 3D measurements. - Uses the same iNtuition (K121916) Vessel Analysis and Calcium modules.Enables . {5}------------------------------------------------ assessment and measurement of vessels and can help identify calcifications, aneurysms and other anomalies to quickly and reliably prepare for various types of vascular procedures.Supports the creation, transmission and storage of a report in digital form. It can also track historical information about the studies analyzed by the software.Displays results analysis, that can be printed as hardcopy or saved in a variety of formats to a hard disk, network, PACS system or CD/DVD/USB. ## Comparison Technological Characteristics: A comparison of the technological characteristics of the predicates and subject device is given in the table below: | | New Device | Predicate Device<br>(Secondary) | Predicate Device<br>(Primary) | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | iNtuition-Structural Heart<br>Module | iNtuition | 3mensio Workstation | | Manufacturer | Terarecon, Inc. | Terarecon, Inc. | Pie Medical Imaging | | 510(k) Number | K191585 | K121916 | K153736 | | | Indication for Use | | | | Indications for Use | iNtuition-Structural Heart<br>Module is a software solution<br>that is intended to assist<br>Cardiologists, Radiologists and<br>Clinical Specialists with the<br>visualization and measurements<br>of structures of the heart and<br>vessels.<br>iNtuition-Structural Heart<br>Module enables the user to:<br>- Visualize and measure<br>(diameters, lengths, angles,<br>areas and volumes)<br>structures of the heart and<br>vessels for pre- operative<br>planning and sizing for<br>cardiovascular<br>interventions and surgery,<br>and for post- operative<br>evaluation.<br>- Quantify calcium (volume,<br>density)<br>iNtuition-Structural Heart<br>Module has the following tools<br>and features that facilitate:<br>- Automatic and manual<br>centerline detection.<br>- Segmentation of cardiovascular<br>structures. | - To receive, store, transmit,<br>post- process, display and<br>allow manipulation of<br>reports and medical images<br>from acquisition devices,<br>including optical or other<br>non-DICOM format images,<br>DICOM images with<br>modality type XA, US, CR,<br>DR, SPECT, NM and MG,<br>and images from volumetric<br>medical scanning devices<br>such as EBT, CT, PET or<br>MRI. To provide access to<br>images derived data and<br>derived images via client-<br>server software, web<br>browser and mobile<br>technology.<br>- Visualization in 2D/3D and<br>4D are supported for single<br>or multiple datasets. or<br>combinations thereof. Tools<br>are provided to define and<br>edit paths through structures<br>such as center lines, which<br>may be used to analyze<br>cross- sections of structures,<br>or to provide flythrough<br>visualizations rendered<br>along such a centerline. | 3mensio Workstation enables<br>visualization and measurement<br>of structures of the heart and<br>vessels for:<br>- Pre-operational planning<br>and sizing for<br>cardiovascular<br>interventions and surgery<br>- Postoperative evaluation<br>- Support of clinical<br>diagnosis by quantifying<br>dimensions in coronary<br>arteries<br>- Support of clinical<br>diagnosis by quantifying<br>calcifications (calcium<br>scoring) in the coronary<br>arteries<br>To facilitate the above, the<br>3mensio Workstation provides<br>general functionality such as:<br>- Segmentation of<br>cardiovascular structures<br>- Automatic and manual<br>centerline detection<br>- Visualization and image-<br>reconstruction techniques:<br>2D review, Volume<br>Rendering, MPR, Curved<br>MPR, Stretched CMRP, | {6}------------------------------------------------ | Measurement tools (diameters, lengths, areas, volumes, angles) | Segmentation of regions of interest and quantitative analysis tools are provided, for images of vasculature, pathology and morphology, including distance, angle, volume, histogram, ratios thereof, and tracking of quantities over time. A database is provided to track and compare results using published comparison techniques such as RECIST and WHO. Calcium scoring for quantification of atherosclerotic plaque is supported. | Slabbing, MIP, AIP, MinIP | |----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| |----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| {7}------------------------------------------------ | New Device | Predicate Device<br>(Secondary) | Predicate Device<br>(Primary) | |----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | for the dimensions of vessels<br>and structures. | Support is provided for digital<br>image processing to derive | Measurement and annotation<br>tools | | Calcium quantification and<br>scoring. | metadata or new images from input<br>image sets, for internal use or for | Reporting tools | | Various visualization<br>techniques: 2D/3D/4D<br>visualization, MPR, Curved<br>MPR, Stretched MPR, MIP,<br>MinIP, Raysum and MAR.<br>Capture and Report. | forwarding to other devices using<br>the DICOM protocol. Image<br>processing tools are provided to<br>extract metadata to derive<br>parametric images from<br>combinations of multiple input<br>images, such as temporal phases, or<br>images co-located in space but<br>acquired with different imaging<br>parameters, such as different MR<br>pulse sequences, or different CT<br>image parameters (e.g. dual<br>energy). | | | | iNtuition is designed for use by<br>healthcare professionals and is<br>intended to assist the physician in<br>diagnosis, who is responsible for<br>making all final patient<br>management decisions. | | | | Interpretation of mammographic<br>images or digitized film screen<br>images is supported only when the<br>software is used without<br>compression and with an FDA-<br>Approved monitor that offers at<br>least 5Mpixel resolution and meets<br>other technical specifications<br>reviewed and accepted by the FDA. | | | | iNtuitionMOBILE provides<br>wireless and portable access to<br>medical images. This device is not<br>intended to replace full<br>workstations and should be used<br>only when there is no access to a<br>workstation. Not intended for<br>diagnostic use when used via a web | | | Technical Characteristics | | | | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Data Type | - CT, MR, Nuc, PET,<br>- Angio, US/Echo,<br>- SPECT,<br>- CR/DR Review<br>- 2D, 3D, 4D Medical<br>- Image review including<br>- cine play | - CT, MR, Nuc, PET, Angio,<br>- US/Echo, SPECT, CR/DR<br>- Review<br>- 2D, 3D, 4D Medical Image<br>- review including cine play | - CT data in DICOM<br>- format (vendor<br>- independent) | | Input Patient Data | - Manual through<br>- keyboard/mouse<br>- Command line interface | - Manual through<br>- keyboard/mouse<br>- Command line interface | - Manual through<br>- keyboard/mouse<br>- Command line interface | | Study list<br>functionality | Importing<br>Exporting<br>Deleting<br>Search<br>Anonymization | Importing<br>Exporting<br>Deleting<br>Search<br>Anonymization | Exporting<br>Deleting<br>Anonymizing (no<br>automatic deletion of<br>original patient data)<br>Search | | Centerline<br>Extraction and | Automatic and manual<br>centerlines<br>Centerline edits and<br>refinements.<br>Vessel Analysis<br>Automatic and manual<br>segmentation of<br>structures<br>Segmentation editing | Automatic and manual<br>centerlines<br>Centerline edits and<br>refinements.<br>Vessel Analysis<br>Automatic and manual<br>segmentation of structures<br>Segmentation editing | Realign orthogonal<br>MPR's<br>Segmentation toolset:<br>Automatic segmentation<br>Automatic centerline<br>Manual centerline<br>Centerline editing<br>Undo/Redo operations<br>Volume sculpting | | Image Assessment | Linear (length, diameter,<br>perimeter), distance pair,<br>angular and ROI<br>measurements<br>Area measurements<br>Volume measurements<br>including volumetric<br>histogram, VOI and TVA<br>for Time Volume<br>Analysis for heart<br>chamber segmentation<br>and analysis<br>C-arm angulation<br>calculation<br>Text and arrow<br>annotations<br>Anatomy ID (Landmark<br>Label Selection)<br>Calcium scoring for<br>assessment of calcium in<br>the aortic root<br>Calcium scoring for<br>assessment of calcium in<br>the coronary arteries<br>Segmentation and<br>analysis of coronary<br>artery tree centerline<br>Synchronized side-by-<br>side review<br>Synchronized center of<br>rotation viewing<br>Findings workflow for<br>temporal correlative<br>analysis<br>2D/3D Batch movie tool | Linear (length, diameter,<br>perimeter), distance pair,<br>angular and ROI<br>measurements<br>Area measurements<br>Volume measurements<br>including volumetric<br>histogram, VOI and TVA for<br>Time Volume Analysis for<br>heart chamber segmentation<br>and analysis<br>C-arm angulation calculation<br>Text and arrow annotations<br>Anatomy ID (Landmark Label<br>Selection)<br>Calcium scoring for<br>assessment of calcium in the<br>aortic root<br>Calcium scoring for<br>assessment of calcium in the<br>coronary arteries<br>Segmentation and analysis of<br>coronary artery tree centerline<br>Synchronized side-by- side<br>review<br>Synchronized center of<br>rotation viewing<br>Findings workflow for<br>temporal correlative analysis<br>2D/3D Batch movie tool and<br>export | Linear (length and<br>diameter), angular and<br>ROI measurements<br>Volume measurements<br>C-arm angulation<br>calculation<br>Text and arrow<br>annotations<br>Calcium scoring for<br>assessment of calcium in<br>the aortic root<br>Calcium scoring for<br>assessment of calcium in<br>the coronary arteries<br>Segmentation and<br>analysis of coronary<br>artery tree centerline | | | New Device | Predicate Device<br>(Secondary) | Predicate Device<br>(Primary) | | Image Assessment<br>Rendering | Volume rendering, MIP,<br>MPR, MinIP, Raysum<br>(ThickMPR)<br>3D triangulation<br>Perspective endoluminal<br>view<br>Medial Axial Reformat<br>(MAR)<br>Curved Planar Reformat<br>(CPR)<br>Double-oblique MIP and<br>MPR<br>Image enhancement<br>filters<br>Synchronized side-by-<br>side viewing<br>Synchronized center of<br>rotation viewing<br>Cube View<br>Workflow templates<br>Multi-Mask Display<br>(multi- object display)<br>User-defined<br>measurement protocols<br>Editing tools: crop, cut,<br>free- hand | Volume rendering,<br>MIP, MPR, MinIP,<br>Raysum (ThickMPR)<br>3D triangulation<br>Perspective endoluminal view<br>Medial Axial Reformat (MAR)<br>Curved Planar Reformat<br>(CPR)<br>Double-oblique MIP and MPR<br>Image enhancement filters<br>Synchronized side-by-side<br>viewing<br>Synchronized center of<br>rotation viewing<br>Cube View<br>Workflow templates<br>Multi-Mask Display (multi-<br>object display)<br>User-defined measurement<br>protocols<br>Editing tools: crop, cut, free-<br>hand | Orthogonal, oblique,<br>double oblique, curved,<br>cross-curved, stretched<br>MPR rendering<br>MIP, AveIP, MinIP and<br>color volume slabs<br>MIP volume rendering<br>Color volume rendering<br>Grayscale volume<br>rendering<br>2D slice review and stack<br>comparison<br>4D cine<br>Interactive VOI clipping<br>Multi-tissue color and<br>opacity control<br>Active presets<br>User-defined presets | | Storage of Results | Structured reporting with<br>xml, text, xls output<br>Word and html report<br>DICOM SC<br>Workflow scenes: restore<br>saved state | Structured reporting with xml,<br>text, xls output<br>Word and html report<br>DICOM SC<br>Workflow scenes: restore<br>saved state | Printout<br>Session state<br>PDF format<br>DICOM PDF report | | Conferencing and<br>Collaboration | Conferencing and<br>Collaboration | Conferencing and Collaboration | N/A | | Operating System | Microsoft Windows | Microsoft Windows | Microsoft Windows | {8}------------------------------------------------ {9}------------------------------------------------ ### Summary of Clinical Performance Tests: iNtuition-Structural Heart Module did not require clinical studies to demonstrate its safety and effectiveness. # General Safety and Effectiveness Concerns: iNtuition-Structural Heart Module doesn't present any safety concerns or effectiveness. iNtuition- Structural Heart Module has the same standard and part of optional features, technological characteristics, and similar indications for use than its predicates. {10}------------------------------------------------ iNtuition-Structural Heart Module instructions for use are included within the device labeling provided in Volume 013 and in Appendix A, and the information provided in the labeling will enable the user to operate the device in a safe and effective manner. Risk management is ensured via risk analysis, which is used to identify and mitigate potential hazards from early design phase till design software validation and release. These potential hazards are controlled during software development by verification and validation testing. Furthermore, the software operators are healthcare professionals who are responsible for making all final patient management decisions. # Conformance Standards: There are no applicable FDA mandated performance standards for this device. However, voluntary standards such as DICOM, various in-house standard operating procedures are in place and have been utilized in the production of the software. The device complies with the following conformance standards: - ISO 14971:2007- Medical devices Application of risk management to medical devices. - IEC 62304:2006/A1:2015 Medical device software Software life cycle processes. - - -IEC 62366-1:2015 - Medical Devices - Application of usability engineering to medical devices. - NEMA PS 3.1-3.20 (2016) Digital Imaging and Communications in Medicine -(DICOM) set. # Performance Data The verification and Validation tests have been performed to address the indication for use, the technological characteristics claims, requirement specifications and risk management results. ### Substantial Equivalence: In all material aspects, iNtuition-Structural Heart Module is substantially equivalent to the predicate devices. Performance testing was carried out according to internal company procedures. Software testing and validation were done according to written test protocols established before testing was conducted. Test results were reviewed by designated technical professionals before being formalized and after ensuring that the software fully satisfies all expected and previously defined system requirements and features. Test results support the conclusion that iNtuition- Structural Heart Module performance satisfies the design intent and is equivalent to its predicate devices. {11}------------------------------------------------ ## Conclusion: iNtuition-Structural Heart Module, as described in this premarket notification, has similar indication for use and same technical characteristics as the predicate devices, and therefore, is substantially equivalent in terms of basic design, features and intended use to those predicates. In summary, TeraRecon, Inc., believes that iNtuition-Structural Heart Module is an optional feature of iNtuition (K121916) and doesn't present any new safety or effectiveness risks to users and is substantially equivalent to and performs as well as the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...