K250330 · Pie Medical Imaging BV · QIH · Nov 3, 2025 · Radiology
Device Facts
Record ID
K250330
Device Name
3mensio Workstation
Applicant
Pie Medical Imaging BV
Product Code
QIH · Radiology
Decision Date
Nov 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K250330 · Nov 3, 2025
3mensio Workstation
Pie Medical Imaging BV
Retrospective MSCT clinical datasets
Retrospective clinical MSCT datasets from 65 patients were used to evaluate the accuracy and performance of the device's automated segmentation and measurement algorithms against expert-derived ground truth.
Multi-centric observational cohort study: 65 patients indicated for valvular intervention.
>1 (experts)
Indications for Use
Standalone diagnostic bioimaging software is intended to measure and visualize cardiovascular structures.
Device Story
3mensio Workstation is a software package for advanced visualization and quantitative analysis of cardiovascular images (CT, X-Ray, ultrasound). It processes DICOM images to segment cardiovascular structures, detect centerlines, and perform measurements (diameters, lengths, areas, volumes, angles). The device supports pre-operational planning, postoperative evaluation, and clinical diagnosis. It is used by cardiologists and radiologists in clinical settings, deployed as a standalone PC application or a web-based solution (network/cloud). The software includes machine learning features for automated segmentation and measurement. Healthcare providers use the output—visualizations, measurements, and reports—to assist in clinical decision-making for cardiovascular interventions and surgery. Benefits include reliable, efficient preparation for vascular and structural heart procedures through automated quantification and visualization.
Clinical Evidence
Bench testing only. Multi-centric observational cohort study using 65 MSCT datasets (age 38-99, 50% female) from diverse CT manufacturers. Primary endpoints: segmentation accuracy (Dice score) and measurement agreement (Bland-Altman, ICC) against expert manual ground truth. Results: Dice scores 0.94 ± 0.07 (mitral annulus), 0.87 ± 0.12 (tricuspid annulus); ICC values 0.99 (aortic annulus), ≥0.94 (mitral), ≥0.97 (tricuspid). Performance met all predefined acceptance criteria, showing high accuracy and reliability comparable to the predicate device.
Technological Characteristics
Software as a medical device (SaMD). Deployment: standalone PC (Windows) or web-based (network/cloud). Data: DICOM (CT, X-Ray, 2D/3D Ultrasound). Features: segmentation (AI/ML and manual), centerline detection, visualization (MPR, Volume Rendering, MIP, etc.), and measurement tools. Standards: ISO 13485, ISO 14971, IEC 62304, IEC 62366-1, IEC 82304-1. Cybersecurity measures implemented for web/cloud access.
Indications for Use
Indicated for advanced visualization and quantitative analysis for diagnostics in cardiology or radiology, enabling visualization and measurement of heart and vessel structures for pre-operational planning/sizing for cardiovascular interventions/surgery, postoperative evaluation, and support of clinical diagnosis by quantifying coronary artery dimensions and calcifications (calcium scoring).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 3, 2025
Pie Medical Imaging B.V.
Annemiek Bouts
Regulatory Affairs Coordinator
Philipsweg 1
Maastricht, Limburg 6227 AJ
Netherlands
Re: K250330
Trade/Device Name: 3mensio Workstation
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: September 29, 2025
Received: September 29, 2025
Dear Annemiek Bouts:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250330 - Annemiek Bouts
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250330 - Annemiek Bouts
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Assistant Director
Imaging Software Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250330
Device Name
3mensio Workstation
Indications for Use (Describe)
Standalone software for medical image analysis intended for advanced visualization and quantitative analysis for diagnostics in the field of cardiology or radiology by means of enabling visualization and measurement of structures of the heart and vessels for:
- Pre-operational planning and sizing for cardiovascular interventions and surgery
- Postoperative evaluation
- Support of clinical diagnosis by quantifying dimensions in coronary arteries
- Support of clinical diagnosis by quantifying calcifications (calcium scoring)
To facilitate the above, the 3mensio Workstation provides general functionality such as:
- Segmentation of cardiovascular structures
- Automatic and manual centerline detection
- Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMRP, Slabbing, MIP, AIP, MinIP
- Measurement and annotation tools
- Reporting tools
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
PIE MEDICAL IMAGING
510(k) Summary – K250330 3mensio Workstation
510(k) Summary
K250330 3mensio Workstation
Pie Medical Imaging BV
# I - General Information
Submitter/Owner Name: Pie Medical Imaging BV
Address: Demertdwarsstraat 8A01, 6227 AK Maastricht, The Netherlands
Phone Number: +31 43 32 81 328
Contact Person: Annemiek Bouts, Regulatory Affairs Coordinator
Email Address: regulatory@pie.nl
Trade Name: 3mensio Workstation
Classification:
- Regulation Name: Medical Image Management and Processing System
- Regulation Class: Class II
- Regulation Number: 21 CFR 892.2050
- Classification Product Code: QIH, LLZ
Predicate Devices:
- Primary predicate device: 3mensio Workstation (K153736, 21 CFR 892.2050, LLZ)
- Secondary predicate device: TOMTEC-ARENA (K213544, 21 CFR 892.2050, LLZ, QIH)
# II - Device Description
3mensio Workstation is an image post-processing software package for advanced visualization and analysis for diagnostics in the field of cardiology and radiology and offers functionality to view CT/X-Ray angiographic and ultrasound images, to segment cardiovascular structures in these images and to analyze and quantify these cardiovascular structures and to present the results in different formats.
3mensio Workstation can be deployed as a web-based solution intended for usage in a network or cloud environment or as a standalone package and runs on a PC with a Windows operating system. It can read DICOM images from an accessible file system, hard disk (local directory), or (indirectly) received from the CT or PACS system. 3mensio Workstation provides the functionality to import the DICOM images and to organize the loaded DICOM images into patients, studies, and series.
3mensio Workstation contains two analysis modules: 3mensio Vascular and 3mensio Structural Heart:
- 3mensio Vascular enables assessment of vessels and can help to measure calcifications, aneurysms and other anomalies to quickly and reliably prepare for various types of vascular procedures.
- 3mensio Structural Heart enables assessment and measurement of different structures of the heart, e.g., the heart valves, coronary arteries and the ventricles. It provides analysis of different approach routes to cardiovascular structures for replacement or repair procedures. In addition, it can help in the quantification of calcifications.
Results can be displayed on the screen, printed, or saved in a variety of formats to hard disk, network or PACS system. Results and clinical images with overlay can also be printed as a hardcopy and exported in various electronic formats.
# III - Intended Use and Indications for Use
## Intended Use
Standalone diagnostic bioimaging software is intended to measure and visualize cardiovascular structures.
## Indications for Use
Standalone software for medical image analysis intended for advanced visualization and quantitative analysis for diagnostics in the field of cardiology or radiology by means of enabling visualization and measurement of the heart and vessels for:
- Pre-operational planning and sizing for cardiovascular interventions and surgery
- Postoperative evaluation
- Support of clinical diagnosis by quantifying dimensions of coronary arteries
- Support of clinical diagnosis by quantifying calcifications (calcium scoring)
Pie Medical Imaging BV
{5}
PIE MEDICAL IPRIGHTS
510(k) Summary – K250330 3mensio Workstation
To facilitate the above, the 3mensio Workstation provides general functionality such as:
- Segmentation of cardiovascular structures
- Automatic and manual centerline detection
- Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMPR, Slabbing, MIP, AIP, MinIP
- Measurement and annotation tools
- Reporting tools
## IV - Substantial equivalence comparison
The devices 3mensio Workstation (K153736) and TOMTEC-ARENA (K213544) have been selected as predicate devices for 3mensio Workstation. The selected devices have technological features and characteristics comparable to 3mensio Workstation and are intended to be used by or under supervision of a cardiologist or radiologist to support clinical decision making of cardiovascular conditions.
The primary predicate device, 3mensio Workstation (K153736), is an earlier version of the current 3mensio Workstation product. The indications for use of the two devices are similar. The two products provide the same functionalities for advanced visualization and measurements of structures of the heart, heart valves and vessels to assist in pre-operational planning and sizing for cardiovascular interventions and surgery and postoperative evaluation. Additionally they both feature functionality for segmentation of cardiovascular structures, quantify dimensions and calcifications in coronary arteries and have measurement and annotation tools. All these functionalities are supported in both the new 3mensio Workstation product and the predicate device 3mensio Workstation (K153736). Therefore, the predicate device 3mensio Workstation (K153736) is selected as the primary predicate device.
The differences in technological characteristics between the predicate device K153736 and the device 3mensio Workstation are that the new 3mensio Workstation runs from a web browser intended for usage in a network or cloud environment. Predicate device K153736 is only a standalone usage device. Furthermore, 3mensio Workstation includes calcium quantification of more cardiovascular structures than the predicate device K153736 which only included calcium quantification in the coronary arteries. In addition, the 3mensio Workstation includes machine learning (ML) features for some of the functionalities already available in the primary predicate device.
Another difference between the previous generation of the 3mensio Workstation product, 3mensio Workstation (K153736) and the current device is a workflow for mitral valve analysis on 3D ultrasound images (3D US mitral valve). The selected secondary predicate device, TOMTEC-ARENA (K213544), offers similar functionality and is therefore selected as a secondary predicate device. Both software applications use the same types of data and operating principles for the user and technology regarding data import, analysis, image display and storage of results.
A summary of the technological characteristics of the subject device and predicate devices is demonstrated in the Table below.
| | Subject Device | Primary predicate device | Secondary predicate device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | 3mensio Workstation | 3mensio Workstation | TOMTEC | / |
| 510(k) Number | K250330 | K153736 | K213544 | / |
| Product Code | QIH, LLZ | LLZ | LLZ, QIH | Similar, the subject device implements AI/ML algorithms |
| Classification | Medical image management and processing system | Medical image management and processing system | Medical image management and processing system | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Intended Use | Standalone diagnostic bioimaging software is intended to measure and visualize cardiovascular structures. | 3mensio Workstation is a software solution that is intended to provide Cardiologists, Radiologists and Clinical Specialists additional information to aid them in reading and interpreting DICOM compliant medical images of structures of the heart and vessels. | TOMTEC-ARENA software is a clinical software package designed for review, quantification and reporting of structures and function based on multi-dimensional digital medical data acquired with different modalities. TOMTEC-ARENA is not intended to | Same, both the subject device and primary predicate device generally share the intended use to enable visualization and measurement of cardiovascular structures. |
Pie Medical Imaging BV
{6}
PIE MEDICAL IMAGING
510(k) Summary – K250330 3mensio Workstation
| | | 3mensio Structural Heart enables the user to: • Visualize and measure (diameters, lengths, areas, volumes, angles) structures of the heart and vessels • Quantify calcium (volume, density) 3mensio Vascular enables the user to: • Visualize and assess stenosis, aneurysms and vascular structures • Measure the dimensions of vessels (diameters, lengths, areas, volumes, angles) | be used for reading of mammography images. | |
| --- | --- | --- | --- | --- |
| Indications for Use | Standalone software for medical image analysis intended for advanced visualization and quantitative analysis for diagnostics in the field of cardiology or radiology by means of enabling visualization and measurement of the heart and vessels for: • Pre-operational planning and sizing for cardiovascular interventions and surgery • Postoperative evaluation • Support of clinical diagnosis by quantifying dimensions of coronary arteries • Support of clinical diagnosis by quantifying calcifications (calcium scoring) To facilitate the above, the 3mensio Workstation provides general functionality such as: • Segmentation of cardiovascular structures • Automatic and manual centerline detection • Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMPR, Slabbing, MIP, AIP, MinIP • Measurement and annotation tools • Reporting tools | 3mensio Workstation enables visualization and measurement of structures of the heart and vessels for: • Pre-operational planning and sizing for cardiovascular interventions and surgery • Postoperative evaluation • Support of clinical diagnosis by quantifying dimensions in coronary arteries To facilitate the above, the 3mensio Workstation provides general functionality such as: • Segmentation of cardiovascular structures • Automatic and manual centerline detection • Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMPR, Slabbing, MIP, AIP, MinI • Measurement and annotation tool • Reporting tools | Indications for use of TomTec-Arena TTA2 software are quantification and reporting of cardiovascular, fetal, abdominal structures and function of patients with suspected disease to support the physician in the diagnosis. | Same for the subject device and primary predicate device. Additionally, the subject device includes functionality for calcium quantification in in other cardiovascular structures then only coronary arteries. For the secondary predicate device, only quantification and reporting of cardiovascular structures is applicable for this comparison. Other anatomical structures are not relevant for the comparison. |
| Anatomical Site | Segmentation, quantification, review, and reporting of cardiovascular structures | Segmentation, quantification, review, and reporting of cardiovascular structures | Segmentation, quantification, review, and reporting of cardiovascular, fetal and abdominal structures. | Same for the subject device and primary predicate device. For the secondary predicate device, only quantification and reporting |
Pie Medical Imaging BV
Page 3 of 5
{7}
PIE MEDICAL IMAGING
510(k) Summary – K250330 3mensio Workstation
| | | | | of cardiovascular structures is applicable for this comparison. Other anatomical structures are not relevant for the comparison. |
| --- | --- | --- | --- | --- |
| Design | Software as a medical device: standalone and web-based (on-premise and in the cloud) | Software as a medical device (standalone) | Software as a medical device (standalone) | The subject device can be deployed either standalone or web-based (on-premise and in the cloud) but this difference does not impact the intended use, indications for use, safety and performance of the device. Any risks related to web-based availability (both on-premise and in the cloud) are mitigated by cybersecurity measures. |
| Data Import and Type | • Vendor independent
• CT data in DICOM format
• X-Ray data in DICOM format
• Ultrasound data (2D and 3D) in DICOM format | • Vendor independent
• CT data in DICOM format
• X-Ray data in DICOM format
• Ultrasound data (2D) in DICOM format | • Vendor independent
• Ultrasound data (2D and 3D) in DICOM format | Same, the subject device includes data types of both predicate device combined. |
| Data Management | • Study list overview
• Deleting
• Exporting
• Anonymizing
• Search | • Study list overview
• Deleting
• Exporting
• Anonymizing
• Search | Note: features not mentioned since they are not relevant for this comparison. | Same, the subject device and primary predicate device use the same types of data management functionality. |
| Image Processing and Contour Definition | • Realign orthogonal MPR's
• Segmentation toolset:
- Automatic segmentation (both AI and non-AI)
- Manual segmentation
- Automatic centerline
- Manual centerline
- Growing centerline
- Centerline editing
• Volume sculpting | • Realign orthogonal MPR's
• Segmentation toolset:
- Automatic segmentation (non-AI)
- Manual segmentation
- Automatic centerline
- Manual centerline
- Growing centerline
- Centerline editing
Volume sculpting | Note: features not mentioned since they are not relevant for this comparison. | Both the subject and primary predicate device include automated and manual segmentation functionality and both devices include the option for the physician to review and edit the segmentation. The segmentation toolset in the subject device additionally includes ML features. |
| Image Assessment | • Linear (length and diameter), area and angle measurements
• Volume measurements
• C-arm angulation calculation
• Text, arrow and 3D annotations
• Calcium scoring
• Valve assessment | • Linear (length and diameter), area and angle measurements
• Volume measurements
• C-arm angulation calculation
• Text, arrow and 3D annotations
• Calcium scoring
• Valve assessment | • Linear (length and diameter), area and angle measurements
• Volume measurements
• Text, arrow and 3D annotations | Same, the subject device and primary predicate device use the same image assessment functionality. Also the secondary predicate device uses relevant image assessment types for analysis of 3D ultrasound images. |
| Image Display | • Orthogonal, oblique, double-oblique, curved, cross-curved, stretched MPR, curved MPR views
• MIP, AIP, MinIP and color volume slabs
• Volume rendering
• 2D slice review and stack comparison
• 4D cine
• Multi-tissue color and opacity control
• Virtual device displays | • Orthogonal, oblique, double-oblique, curved, cross-curved, stretched MPR, curved MPR views
• MIP, AIP, MinIP and color volume slabs
• Volume rendering
• 2D slice review and stack comparison
• 4D cine
• Multi-tissue color and opacity control
Virtual device displays | • Oblique, double-oblique views
• Volume rendering
• 4D cine | Same, both the subject device and primary predicate device use the same types of image display. The secondary predicate device uses relevant image display features for analysis of 3D ultrasound images. |
| Storage and Export of Results | Session states
PDF reports | Session states
PDF reports | Not relevant for comparison. | Same. |
Pie Medical Imaging BV
Page 4 of 5
{8}
PIE MEDICAL IPRIGHTS
510(k) Summary – K250330 3mensio Workstation
# V - Performance Data
Verification and validation of the 3mensio Workstation showed that the system requirements – derived from the indications for use and defined user needs – as well as risk control measures were implemented correctly, and that the device meets its specifications including conformance to the international recognized process standards (i.e., ISO 13485, ISO 14971, IEC 62304, IEC 62366-1, and IEC 82304-1).
Documentation is provided as recommended by FDA’s Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices” and verification and validation was performed per the FDA Guidance “Technical Performance Assessment of Quantitative Imaging in Radiological Device Premarket Submission”.
Performance testing was conducted where the results of automatically determined segmentation and measurements was evaluated against ground-truth values manually obtained by experts with the primary predicate device. The results showed high accuracy and agreement with the predicate device, meeting all predefined acceptance criteria. The performance tests included statistical methods such as Bland-Altman analysis, intraclass correlation coefficient (ICC) calculations and Dice coefficient.
## Testing Characteristics
The performance testing was based upon a multi-centric observational cohort study involving MSCT datasets of 65 patients indicated for valvular intervention. The cohort comprised of at least 50% datasets from USA. The age range of the studied cohort was 38-99 years, with 50% female subjects. The cohort included CT scans from different CT manufacturers and covered imaging parameters such as different slice thickness and contrast enhancement. No datasets were included that were used for training the machine learning models.
## Testing Methodology and Statistical Approach
The accuracy of the automatically determined segmentations was evaluated using Dice scores. The segmentations covered various cardiovascular structures including the aortic nadir positions, mitral annulus plane tricuspid annulus. The accuracy of automatically determined measurements was evaluated using Bland-Altman analysis and the intraclass correlation coefficient (ICC).
## Segmentation Results
The resulting Dice scores ranged from 0.94 ± 0.07 for the mitral annulus, 0.87 ± 0.12 for the tricuspid annulus. The average distance from the automatically detected nadir positions towards the ground truth annular plane was -/- 0.64 ± 1.07 mm. These values demonstrate that a high accuracy is achieved for the automatically determined segmentations meeting the pre-defined acceptance criteria.
## Measurements Results:
The ICC values were 0.99 for the measurements on the aortic annulus plane derived from the aortic nadir positions, 0.94 and above for the measurements on the mitral annulus and 0.97 and above for the measurements on the tricuspid annulus indicating excellent agreement with the ground-truth values obtained by experts. Bland-Altman analysis shows excellent agreement between the automatically determined measurements and the measurements obtained by experts.
Based on the above it can be concluded that the automatically determined segmentations and measurements are as accurate and reliable as those obtained using the predicate device. Overall, comparable performance was achieved on all variability factors including geographic location, scanners, slice thickness, age, and gender.
The verification and validation results demonstrate the safety and effectiveness of 3mensio Workstation in relation to its intended use and indications for use and therefore 3mensio Workstation can be considered as safe and effective as its predicate devices.
# VI - Conclusion
Based on the application of risk management and performance testing inherent to PMI’s QA system (compliant with recognized standards as stated above) the conclusion is that 3mensio Workstation is as safe and effective as its predicate device in terms of indications for use, technological characteristics, measurements, and operating environment and does not raise any new issues related to safety and effectiveness compared to the predicate devices.
Pie Medical Imaging BV
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.