The CanMINI Hand and Foot System is indicated for fixation of small bones and small bone fragments in the hand and foot. The devices are intended to be single-use and sterilized before use.
Device Story
Orthopedic fixation system for stabilization of fractures or osteotomies of small bones in hand and foot. System comprises various bone plates and screws; provides mechanical support to maintain alignment during bone healing. Used by orthopedic surgeons in clinical scenarios including calcaneal fractures, fracture displacements, and non-union fractures. Benefits include critical stabilization, enhanced healing, and reduced risk of malunion or delayed union. Components are manufactured from pure titanium (ASTM F67) and titanium alloy (ASTM F1472). Supplied non-sterile; requires sterilization before use.
Clinical Evidence
No clinical data. Bench testing only. Mechanical performance evaluated via axial pullout strength (ASTM F543-23), single cycle bend testing (ASTM F382-24), and driving torque/torsional properties (ASTM F543-23). Material composition verified per ASTM F67-13 and ASTM F1472-23.
Technological Characteristics
Materials: Pure titanium (ASTM F67) and titanium alloy (ASTM F1472). Components: Bone plates (locking/non-locking) and bone screws (locking/non-locking/cannulated). Mechanical performance standards: ASTM F382 (plates), ASTM F543 (screws). Sterilization: Required before use (supplied non-sterile).
Indications for Use
Indicated for fixation of small bones and small bone fragments in the hand and foot. Intended for single-use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes (USA) 1.5mm Mini Fragment LCP System (K090047)
Double Engine Bone Plate And Screw Systems (K130108)
Single/multiple Component Metallic Bone Fixation Appliances and Accessories (K033639)
Synthes (USA) Modular Mini Fragment LCP System (K063049)
PERI-LOC Periarticular Locked Plating System (K092015)
Arthrex Mini Comprehensive Fixation System - 2.0mm & 2.4mm Module (K191344)
Synthes (USA) 3.5mm and 4.5mm Locking Compression Plate (LCP) System (K082807)
Related Devices
K050657 — MONDEAL CONTOUR HO SYSTEM · Kapp Surgical Instrument, Inc. · Aug 8, 2005
K180808 — Omni Foot Plating System · Extremity Medical, LLC · May 22, 2018
K133536 — EMERGE MEDICAL BONE PLATE SYSTEM · Emerge Medical · Feb 26, 2014
K241815 — Protean Fragment Plating System · Skeletal Dynamics, Inc. · Sep 26, 2024
K152974 — ORTHOLOC® 3Di Foot Plating Reconstruction System · Wrightmedicaltechnologyinc · Dec 11, 2015
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2025
Canwell Medical Co., Ltd.
% Jiangfeng Ren
Regulatory Affairs
Shanghai Ling Fu Technology Co., Ltd.
Room 1115, Block A, Xuhui Vanke Centre, No. 55 Ding'an Road,
Xuhui District
Shanghai, Shanghai 200030
China
Re: K242973
Trade/Device Name: CanMINI Hand and Foot System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: September 15, 2024
Received: September 26, 2024
Dear Jiangfeng Ren:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242973 - Jiangfeng Ren
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242973 - Jiangfeng Ren
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242973 | |
| Device Name CanMINI Hand and Foot System | |
| Indications for Use (Describe) The CanMINI Hand and Foot System is indicated for fixation of small bones and small bone fragments in the hand and foot. The devices are intended to be single-use and sterilized before use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) summary
# K242973
510(k) Summary
## I. Submitter
CANWELL MEDICAL CO., LTD.
Address: NO.466 South Xianhua Street, High-Tech Industrial Zone, Jinhua, Zhejiang Province, 321016, P R. of China
## Contact Person
Hui CHEN
Position: Regulatory Affairs Director
Phone: +86 13868997004
E-mail: hui.chen@canwell.com.cn
## II. Correspondent
Company: Shanghai Ling Fu Technology Co., Ltd.
Room 1115, Block A, Xuhui Vanke Centre, No. 55 Ding'an Road, Xuhui District, Shanghai, P.R.China
Contact person: Jiangfeng REN Regulatory affairs
Phone: 86-21-33561520
Email: jiangfeng.ren@llins-tech.com
## III. Proposed Device
Trade Name of Device: CanMINI Hand and Foot System
Common Name: Plate, Fixation, Bone; Screw, Fixation, Bone
Regulation Number: 21 CFR 888.3030 (primary), 21 CFR 888.3040
Regulatory Class: Class II
Product code: HRS (primary), HWC
Classification Name: Single/multiple component metallic bone fixation appliances and accessories; Smooth or threaded metallic bone fixation fastener
Review Panel: Orthopedic
## IV. Predicate Devices
### IV.1. Primary Predicate Device
Trade name: Synthes (USA) 1.5mm Mini Fragment LCP System
Common name: /
Classification: Class II, 21 CFR 888.3030
Product Code: HRS
Premarket Notification: K090047
Manufacturer: Synthes (USA)
### IV.2. Additional Predicate Devices
Trade name: Double Engine Bone Plate And Screw Systems
Common name: /
CANWELL MEDICAL CO., LTD.
{5}
510(k) summary
Classification: Class II, 21 CFR 888.3030
Product Code: HRS, HWC
Premarket Notification: K130108
Manufacturer: XIAMEN DOUBLE ENGINE MEDICAL MATERIAL CO., LTD
| Trade name: | Single/multiple | Component | Metallic | Bone | Fixation |
| --- | --- | --- | --- | --- | --- |
| Common name: | Appliances and Accessories | | | | |
| Classification: | Plate, Fixation, Bone | | | | |
| Product Code: | Class II, 21 CFR 888.3030 | | | | |
| Premarket Notification: | HRS | | | | |
| Manufacturer: | K033639 | | | | |
| | Acumed LLC | | | | |
| Trade name: | Synthes (USA) Modular Mini Fragment LCP System |
| --- | --- |
| Common name: | / |
| Classification: | Class II, 21 CFR 888.3030 |
| Product Code: | HRS |
| Premarket Notification: | K063049 |
| Manufacturer: | Synthes (USA) |
| Trade name: | PERI-LOC Periarticular Locked Plating System |
| --- | --- |
| Common name: | Bone Plates and Bone Screws |
| Classification: | Class II, 21 CFR 888.3030 |
| Product Code: | HRS |
| Premarket Notification: | K092015 |
| Manufacturer: | Smith & Nephew, Inc. |
| Trade name: | Arthrex Mini Comprehensive Fixation System - 2.0mm & 2.4mm Module |
| --- | --- |
| Common name: | Single/multiple component metallic bone fixation appliances and accessories |
| Classification: | Smooth or threaded metallic bone fixation fastener |
| Class: | Class II, 21 CFR 888.3030, 21 CFR 888.3040 |
| Product Code: | HRS; HWC |
| Premarket Notification: | K191344 |
| Manufacturer: | Arthrex Inc. |
CANWELL MEDICAL CO., LTD.
{6}
510(k) summary
Common name: /
Classification: Class II, 21 CFR 888.3040
Product Code: HWC
Premarket Notification: K930834
Manufacturer: ACUMED, INC.
Trade name: Synthes (USA) / 3.5 nun and 4.5 nun Locking Compression Plate (LCP) System with Expanded Indications
Common name: /
Premarket Notification: Class II, 21 CFR 888.3030
Product Code: HRS; HWC
Premarket Notification: K082807
Manufacturer: Synthes (USA)
# V. Device Description Summary
The CanMINI Hand and Foot System, manufactured by CANWELL MEDICAL CO. LTD., is an orthopedic fixation system designed to stabilize fractures or osteotomies of the small bones in the hand and foot. The system includes a variety of bone plates and screws made from high-quality titanium and titanium alloy, ensuring both strength and biocompatibility. The primary materials used in the CanMINI Hand and Foot System are pure titanium, conforming to ASTM F67 standards, and titanium alloy, conforming to ASTM F1472 standards.
The CanMINI Hand and Foot System consists of multiple components, each designed to address specific clinical needs in orthopedic surgery. The key components are bone plates and bone screws. The bone plates include straight type bone plates such as models FZSQ01 I, FZSQ12, FZSQ38, FZSQ39 I, FZSQ39 II, FZSQ39 III, FZSQ40 I, and FZSQ40 II, as well as supporting type non-locking bone plates like models FBC54 I, FBC54 II, FBC55, FBC56, and FBC58. Additionally, the system features supporting type locking bone plates including models FBCS29, FBCS57 I, FBCS57 II, FBCS59, FBCS60, FBCS86, FBCS87, FBCS88, FBCS89, FBCS96 I, FBCS96 II, FBCS99 I, FBCS99 II, FBCS132, FBCS134, and FBCS135. The bone screws are also categorized into different types, including metal non-locking bone screws with models FHAQ02 I, FHAQ03 I, FHAQ06 I, metal locking screws with models FHAS04 I, FHCS04 I, and cannulated screws with model FGJYD X. The bone plates and screws in the CanMINI Hand and Foot System come in a range of sizes to accommodate different anatomical and clinical requirements. For example, the 6-hole 1.5mm phalangeal metacarpal locking plate has dimensions of 29.7 mm in length, 4.7 mm in width, and 0.95 mm in thickness, while the 8-hole 2.4mm condylar locking plate has dimensions of 59.0 mm in length, 6.5 mm in width, and 1.4 mm in thickness.
The CanMINI Hand and Foot System is intended for the stabilization of fractures and
CANWELL MEDICAL CO., LTD.
{7}
510(k) summary
osteotomies in the small bones of the hand and foot. It provides mechanical support to the bone during the healing process by maintaining proper alignment and stabilizing the fracture site. The system is designed to be used by orthopedic surgeons in various clinical scenarios involving bone fractures, including but not limited to, calcaneal fractures, fracture displacements, and non-union fractures. The clinical benefits of the system include providing critical stabilization and support during the healing of orthopedic fractures. This system aids in the proper alignment of fractured bones, enhances the healing process, and reduces the likelihood of complications such as malunion or delayed union.
The CanMINI Hand and Foot System is supplied in a non-sterile state and must be sterilized before use. It is intended for single use only.
Table 1 List of product types and specifications
| Model | Product Description | Specifications |
| --- | --- | --- |
| FZSQ38 | 1.5mm Phalangeal Metacarpal Locking Plate | 6-12 holes |
| FZSQ39 I | 1.5mm Locking Plate | 4-10 holes |
| FBCS86 | 1.5mm Mini Locking Plate (T-Shape) | 4-8 holes |
| FBCS87 | 1.5mm Locking Support Plate | 8 holes; R, L |
| FBCS88 | 1.5mm Mini Locking Plate(Y-shape) | 4-8 holes |
| FBCS89 | 1.5mm Mini Condylar Locking Plate | 4-8 holes |
| FZSQ01 I | 2.0mm Tubular Plate | 2-7 holes |
| FZSQ40 I | 2.0mm Adaption Locking Plate | 12 holes |
| FBCS57 I | 2.0mm Condylar Locking Plate | 3-7 holes |
| FBCS96 I | 2.0mm Locking Y-Plate | 4-7 holes |
| FBCS99 I | 2.0mm Locking T-Plate | 4-7 holes |
| FBC54 II | 2.0mm T-Plate | 2-5 holes |
| FBC55 | 2.0mm L-Plate | 4, 5 holes; R, L |
| FBC56 | 2.0mm L-Plate, Oblique-angled | 4, 5 holes; R, L |
| FBC58 | 2.0mm T-Plate(head 3 holes) | 4-8 holes |
| FZSQ39 II | 2.0mm Locking Plate | 4-10 holes |
| FZSQ12 | 2.4mm Reconstruction Plate | 4-6 holes |
| FZSQ39III | 2.4mm Locking Plate | 4-10 holes |
| FZSQ40 II | 2.4mm Adaption Locking Plate | 12 holes |
| FBCS57 II | 2.4mm Condylar Locking Plate | 3-7 holes |
| FBC54 I | 2.4mm T-Plate | 2-5 holes |
| FBCS96 II | 2.4mm Mini Locking Plate(Y-shape) | 4-7 holes |
| FBCS99 II | 2.4mm Mini Locking Plate(T-shape) | 4-7 holes |
| FBCS59 | Posterior Tuberosity Calcaneal Locking Plate | 3, 4; R, L |
| FBCS60 | Anterior Process Calcaneal Locking Plate | 3, 4; R, L |
| FBCS132 | Calcaneal Locking Plate I | Small, Medium, Large; R, L |
| FBCS29 | Calcaneal Locking Plate II | 15 holes; 69mm,76 mm; R, L |
| FBCS134 | Calcaneal Locking Plate III | Small, Medium, Large; R, L |
| FBCS135 | Calcaneal Locking Plate IV | 53mm-68 mm; R, L |
| FHAS04 I | Locking screw, Self-Tapping | Φ1.5, 6~20 mm |
| | | Φ2.0, 6~30 mm |
| FHCS04 I | Locking screw, Self-Tapping | Φ2.4, 6~30 mm |
| | | Φ3.5, 10~55 mm |
| FHAQ02 I | Cortex Screw (Phillips), Self-Tapping | Φ2.0, 6~38 mm |
| | | Φ2.4, 6~40 mm |
CANWELL MEDICAL CO., LTD.
{8}
510(k) summary
| FHAQ06 I | Cortex Screw, Self-Tapping | Φ1.5, 6~20 mm |
| --- | --- | --- |
| | | Φ2.0, 6~38 mm |
| | | Φ2.4, 6~40 mm |
| FHAQ03 I | Cortex Screw, Self-Tapping | Φ3.5, 12~60 mm |
| FGJYDX | Headless Compression Screw | Φ2.5, 8~30 mm |
| | | Φ3.0, 10~50 mm |
| | | Φ3.5, 16~50 mm |
| | | Φ4.0, 16~60 mm |
| | | Φ4.5, 20~60 mm |
| | | Φ5.0, 25~80 mm |
| | | Φ6.5, 40~120 mm |
VI. Indications for use
The CanMINI Hand and Foot System is indicated for fixation of small bones and small bone fragments in the hand and foot. The devices are intended to be single-use and sterilized before use.
VII. Comparison of technological characteristics with the predicate devices
VII.1.Indications for Use Comparison
Both the subject device and the predicate devices are intended for the stabilization of fractured or surgically cut bones, providing structural support as the bone heals. The differences in the exact wording of the indications for use do not introduce any deviations in clinical application or safety profile. The intended use remains consistent across all devices, ensuring that the subject device is suitable for its designed purpose, in full alignment with the predicates.
VII.2. Technological Comparison
The subject device meets the FDA's Safety and Performance Based Pathway by adhering to ASTM standards (F67 and F1472 for materials, F382 and F543 for mechanical performance) and demonstrates substantial equivalence to the predicate devices. The minor differences between the subject and predicate devices, such as different dimensions, variations in hole numbers and specific materials, do not raise new questions of safety or effectiveness. Importantly, since the subject device complies with the FDA-identified performance criteria, no further direct SE comparison is necessary.
VIII. Non-Clinical Testing
Key tests include:
1. Axial Pullout Strength (per ASTM F543-23): This test evaluates the axial extraction strength of the screws, confirming that the device meets the minimum performance criteria for bone screws.
2. Single Cycle Bend Testing (per ASTM F382-24): This test assesses the bending strength and stiffness of the 2.4mm locking plates.
3. Driving Torque and Torsional Properties (per ASTM F543-23): Evaluations were conducted on metallic bone screws and locking screws to ensure they meet the required torsional strength and torque limits.
In addition to the mechanical performance tests, the nonclinical evaluation of the CanMINI Hand and Foot System also included material characterization tests to ensure compliance with ASTM material
CANWELL MEDICAL CO., LTD.
{9}
510(k) summary
standards. The materials used in the device, Pure Titanium (Grade 3) and Titanium Alloy Ti6Al4V, were evaluated in accordance with the following standards:
- Material Composition (per ASTM F67-13 for unalloyed titanium and ASTM F1472-23 for titanium alloy): These tests ensure that the chemical composition of the materials complies with established limits for elements such as nitrogen, carbon, and oxygen. The materials are certified to meet the requirements for surgical implants.
IX. Clinical Testing
Not Applicable.
X. Conclusion
The subject device meets the FDA's Safety and Performance Based Pathway by adhering to ASTM standards (F67 and F1472 for materials, F382 and F543 for mechanical performance) Therefore, it can be concluded that the subject device as safe and effective as the predicate devices.
CANWELL MEDICAL CO., LTD.
Page 6 of 6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.