The clinical study evaluated the effectiveness of HeartFocus guidance in enabling non-specialist medical professionals (RNs) to acquire diagnostic-quality echocardiographic images in a real-world clinical environment. Retrospective data was also used to tune and validate AI/ML algorithm performance.
The HeartFocus software is intended to assist and guide medical professionals in the acquisition of cardiac ultrasound images. HeartFocus software is an accessory to compatible general-purpose diagnostic ultrasound systems. Heartfocus guides the acquisition of two-dimensional transthoracic echocardiography (2D-TTE). Specifically, in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (A4C), Apical 5-Chamber (A5C), Apical 2-Chamber (A2C), Apical 3-Chamber (A3C), Subcostal 4-Chamber (SC-4C), and Subcostal inferior Vena Cava (SC-IVC). HeartFocus software is indicated for use in adult patients who require a cardiac ultrasound exam.
Device Story
HeartFocus is an offline software accessory for general-purpose ultrasound systems; assists medical professionals in acquiring 10 standard 2D-TTE cardiac views. Input: real-time ultrasound image stream. Operation: AI/ML models emulate sonographer expertise by providing live probe-positioning cues (translational, tilt, rotation) and detecting diagnostic-quality frames. Output: real-time visual guidance UI, automated clip recording (Auto Record), and retrospective best-clip selection (Best-Effort Record). Used in clinical settings by trained medical professionals; data stored locally on tablet. Benefits: enables non-expert users to acquire diagnostic-quality cardiac images; improves consistency and efficiency of echocardiographic exams.
Clinical Evidence
Prospective multicentric study of 240 adults (120 US, 120 France) comparing exams by 8 trained RNs using HeartFocus vs. control exams by experts. Primary endpoints (qualitative visual assessment of LV size, LV function, RV size, and pericardial effusion) met with 100% success [95% CI 98.4-100]. Secondary endpoints (RV function, LA/RA size, valve assessment, segmental kinetics) showed high diagnostic quality. 99.7% of Auto-Record clips were diagnostic. Bench testing of AI algorithms on 1.4M+ images showed Cohen's kappa > 0.6 for view detection and PPV > 0.8 for guidance cues.
Technological Characteristics
Software-based AI/ML guidance system; operates offline on tablets. Compatible with Clarius ultrasound scanners. Uses deep learning/machine learning algorithms for image classification and probe guidance. Requires 15+ fps stream, 2-5 MHz frequency, 30 cm depth, and 80-degree field of view. No cloud connectivity; local storage only.
Indications for Use
Indicated for adult patients requiring 2D-TTE cardiac ultrasound exams. Used by medical professionals to assist in acquiring 10 standard cardiac views (PLAX, PSAX-AV, PSAX-MV, PSAX-PM, A4C, A5C, A2C, A3C, SC-4C, SC-IVC).
Regulatory Classification
Identification
A radiological acquisition and/or optimization guidance system is a device that is intended to aid in the acquisition and/or optimization of images and/or diagnostic signals. The device interfaces with the acquisition system, analyzes its output, and provides guidance and/or feedback to the operator for improving image and/or signal quality.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed, technical device description, including a detailed description of the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, and the associated inputs and outputs.
(ii) A detailed, technical report on the non-clinical performance testing of the subject device in the intended use environments, using relevant consensus standards when applicable.
(iii) A detailed report on the clinical performance testing, obtained from either clinical testing, accepted virtual/physical systems designed to capture clinical variability, comparison to a closely-related device with established clinical performance, or other sources that are justified appropriately. The choice of the method must be justified given the risk of the device and the general acceptance of the test methods. The report must include the following:
(A) A thorough description of the testing protocol(s).
(B) A thorough, quantitative evaluation of the diagnostic utility and quality of images/data acquired, or optimized, using the device.
(C) A thorough, quantitative evaluation of the performance in a representative user population and patient population, under anticipated conditions and environments of use.
(D) A thorough discussion on the generalizability of the clinical performance testing results.
(E) A thorough discussion on use-related risk analysis/human factors data.
(iv) A detailed protocol that describes, in the event of a future change, the level of change in the device technical specifications or indications for use at which the change or changes could significantly affect the safety or effectiveness of the device and the risks posed by these changes. The assessment metrics, acceptance criteria, and analytical methods used for the performance testing of changes that are within the scope of the protocol must be included.
(v) Documentation of an appropriate training program, including instructions on how to acquire and process quality images and video clips, and a report on usability testing demonstrating the effectiveness of that training program on user performance, including acquiring and processing quality images.
(2) The labeling required under § 801.109(c) of this chapter must include:
(i) A detailed description of the device, including information on all required and/or compatible parts.
(ii) A detailed description of the patient population for which the device is indicated for use.
(iii) A detailed description of the intended user population, and the recommended user training.
(iv) Detailed instructions for use, including the information provided in the training program used to meet the requirements of paragraph (b)(1)(iv) of this section.
(v) A warning that the images and data acquired using the device are to be interpreted only by qualified medical professionals.
(vi) A detailed summary of the reports required under paragraphs (b)(1)(ii) and (iii) of this section.
(vii) A statement on upholding the As Low As Reasonably Achievable (ALARA) principle with a discussion on the associated device controls/options.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 4, 2025
DESKi
Saray Ugidos Seman
Quality and Regulatory Manager
2 Place de la Bourse
Bordeaux, 33000
FRANCE
Re: K242807
Trade/Device Name: HeartFocus (V.1.1.1)
Regulation Number: 21 CFR 892.2100
Regulation Name: Radiological Acquisition And/Or Optimization Guidance System
Regulatory Class: Class II
Product Code: QJU
Dated: February 28, 2025
Received: March 3, 2025
Dear Saray Ugidos Seman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242807 - Saray Ugidos Seman
Page 2
premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}
K242807 - Saray Ugidos Seman
Page 3
803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242807 | |
| Device Name HeartFocus (V.1.1.1) | |
| Indications for Use (Describe) | |
| The HeartFocus software is intended to assist and guide medical professionals in the acquisition of cardiac ultrasound images. HeartFocus software is an accessory to compatible general-purpose diagnostic ultrasound systems. Heartfocus guides the acquisition of two-dimensional transthoracic echocardiography (2D-TTE). Specifically, in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (A4C), Apical 5-Chamber (A5C), Apical 2-Chamber (A2C), Apical 3-Chamber (A3C), Subcostal 4-Chamber (SC-4C), and Subcostal inferior Vena Cava (SC-IVC). | |
| HeartFocus software is indicated for use in adult patients who require a cardiac ultrasound exam. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
K242807
| CONTACT DETAILS | |
| --- | --- |
| Applicant Name | DESKi |
| Applicant Address | 2 Place de la Bourse, 33000 France |
| Applicant contact Telephone | +34686429657 |
| Applicant Contact | Ms. Saray Ugidos Seman |
| Applicant Contact Email | saray.seman@deski.ai |
| Secondary Contact Email | bertrand.moal@deski.ai |
| Device Trade Name | HeartFocus (v.1.1.1) |
| --- | --- |
| Common Name | Radiological acquisition and/or optimization guidance system |
| Classification Name | Image Acquisition And/Or Optimization Guided By Artificial Intelligence |
| Regulation Number | 892.2100 |
| Product Code | QJU |
| Legally Marketed Predicate Device | DEN190040 Caption Guidance |
| --- | --- |
# DEVICE DESCRIPTION SUMMARY
The HeartFocus software is a radiological computer-assisted acquisition guidance system that provides real-time user guidance during echocardiography to assist the user in acquiring anatomically standard diagnostic-quality 2D echocardiographic views. HeartFocus software is an accessory to compatible general-purpose diagnostic ultrasound systems.
{5}
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
HeartFocus is intended to be used by medical professionals who have received appropriate training on ultrasound basics and training on using the HeartFocus software, provided by either DESKi or by a trained medical professional while using approved training materials.
It supports the acquisitions of 10 echocardiographic views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (A4C), Apical 5-Chamber (A5C), Apical 2-Chamber (A2C), Apical 3-Chamber (A3C), Subcostal 4-Chamber (SC-4C), and Subcostal Inferior Vena Cava (SC-IVC).
To allow the acquisition of these views, HeartFocus can connect to an ultrasound system, allowing it to receive a stream of ultrasound images.
The standard views acquired with HeartFocus may be assessed by qualified medical professionals to support their decision-making regarding patient care. The collected exams can be transferred notably using DICOM protocols.
HeartFocus is an application that operates entirely offline, without requiring a cloud server to provide its functionalities. All collected medical data is stored locally on the tablet. This data is never transferred to a server controlled by DESKi.
HeartFocus uses artificial intelligence (AI) to emulate the expertise of sonographers in positioning the probe on the patient's chest and in identifying and recording diagnostic-quality clips.
It proposes 4 major functionalities to assist healthcare professionals in the acquisition of cardiac ultrasound images: Live guidance, Diagnostic-quality view detection, Auto record, and Best-effort record.
# Live Guidance
This feature provides real-time cues to the user on how to move the probe on the patient's chest to obtain diagnostic-quality images. The guidance UI is generated by a machine learning model that infers the probe's position relative to the target position, i.e. that would provide
{6}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
diagnostic-quality images. It emulates how a sonographer would manipulate the ultrasound probe to acquire a diagnostic-quality clip.
# Diagnostic-Quality View Detection
This feature detects real-time diagnostic-quality images and informs the user to hold the position and perform automatic recording. This detection is performed by a machine learning model that also allows the grading of the recorded clips. When the probe is well positioned and the image is of diagnostic quality, a triangle surrounding the ultrasound image appears in color shifting from blue (lower quality) to green (higher quality). The user shall hold the position so that an automated recording can be performed.
# Auto Record
This feature triggers an automatic recording of a clip when the quality is predicted to be of diagnostic quality for a sufficient amount of time.
When the ultrasound frames are consecutively classified as diagnostic quality for a preset period, the "auto record" feature automatically captures a clip.
The auto-record user interface (UI) element fills itself to encourage the user to hold its position. When the entire auto-record UI element is full, the record is saved and the UI indicates it to the user.
This feature emulates how a sonographer would identify diagnostic-quality clips and record them.
# Best-Effort Record (BER)
This feature continually assesses clip quality while the user is scanning and, if the user cannot obtain a clip sufficient for an Auto-Record, the software allows the user to record the highest quality clip obtained so far retrospectively.
Once recorded, the clip is available for a visual review where the user can either save the clip or cancel his action and go back to the acquisition. It emulates how a sonographer would identify the best diagnostic-quality clips retrospectively.
{7}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
# INTENDED USE/INDICATIONS FOR USE
The HeartFocus software is intended to assist and guide medical professionals in the acquisition of cardiac ultrasound images. HeartFocus software is an accessory to compatible general-purpose diagnostic ultrasound systems. Heartfocus guides the acquisition of two-dimensional transthoracic echocardiography (2D-TTE). Specifically, in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (A4C), Apical 5-Chamber (A5C), Apical 2-Chamber (A2C), Apical 3-Chamber (A3C), Subcostal 4-Chamber (SC-4C), and Subcostal Inferior Vena Cava (SC-IVC).
HeartFocus software is indicated for use in adult patients who require a cardiac ultrasound exam.
# INDICATIONS FOR USE COMPARISON
The indications for use are the same.
# TECHNOLOGICAL COMPARISON
Table 1: Predicate table
| Parameter | Caption Guidance (DEN 190040) | Proposed HeartFocus | Conclusion |
| --- | --- | --- | --- |
| Classification Name | Image Acquisition and/Or Optimization Guided by Artificial Intelligence | Image Acquisition and/Or Optimization Guided by Artificial Intelligence | Same |
| Product Code | QJU | QJU | Same |
| Intended Use | The Caption Guidance software is intended to assist medical professionals in the | The HeartFocus software is intended to assist healthcare professionals in the acquisition of cardiac | Same |
{8}
DESKi
HeartFocus 510(k) Summary
Apr 4, 2025
| | acquisition of cardiac ultrasound images. Caption Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems. | ultrasound images. The HeartFocus software is an accessory to compatible general purpose diagnostic ultrasound systems. | |
| --- | --- | --- | --- |
| Indications for use | The Caption Guidance software is indicated for use in two-dimensional transthoracic echocardiography (2D-TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC). | The HeartFocus software is indicated for use in two-dimensional transthoracic echocardiography (2DTTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscles (PSAX-PM), Apical 4-Chamber (A4C), Apical 5-Chamber (A5C), Apical 2-Chamber (A2C), Apical 3-Chamber (A3C), Subcostal 4-Chamber (SC-4C), and Subcostal Inferior Vena Cava (SC-IVC). | Same
*A4C, A5C, A2C, A3C, SC-4C are the American Society of Echocardiography Acronyms: Guidelines for Performing a Comprehensive Transthoracic Echocardiographic Examination in Adults; Recommendations from the American Society of Echocardiography (asecho.org) |
| Intended User | Medical professionals (including expert sonographers) | Medical professionals (including expert sonographers) | Same |
| Compatible Ultrasound System and Probe | The uSmart 2300t Plus ultrasound system with the 2300t-compatible Terason 4v2A linear phased array probe | Clarius Scanner PAHD3 (wireless), PAHD, PAL, and Clarius Ultrasound Scanner Software | Substantially Equivalent: Both the predicate devices and the proposed device function as a |
5
{9}
6
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
| | | | software accessory to a cleared-to-market ultrasound device. Both ultrasound devices are cleared for the diagnostic ultrasound clinical application of acquiring cardiac images of adult patients. Since both ultrasound devices are cleared for the required clinical use, the use of different ultrasound devices does not raise new questions of safety and/or effectiveness. |
| --- | --- | --- | --- |
## Clinical Features
| Image Acquisition Guidance | The prescriptive guidance feature in Caption Guidance provides direction to the user to emulate how a sonographer would manipulate the transducer to acquire the optimal view | HeartFocus algorithm provides live-guidance to guide the user on how to position the probe on the patient's chest to obtain diagnostic-quality images | Substantially Equivalent: While the user interface is slightly different, the functionality and type of guidance provided to the user will be the same. Specifically, both devices provide to the user instructions on how to manipulate the probe (translational, tilt and rotation guidance cues) together with real-time feedback on the expected diagnostic quality of the resulting clip to |
| --- | --- | --- | --- |
{10}
DESKi
HeartFocus 510(k) Summary
Apr 4, 2025
| | | | direct the user to a probe position that will enable acquisition of a diagnostic quality clip. Hence, this difference does not raise new questions of safety and/or effectiveness. |
| --- | --- | --- | --- |
| Real Time Feedback on Image Quality | Quality Meter: real-time feedback from the Quality Meter advises the user on the expected diagnostic quality of the resulting clip, such that the user can make decisions to further optimize the quality, for example by following the prescriptive guidance feature | HeartFocus algorithm provides analysis of the diagnostic-quality of the image and indicates to the user when the current frame is diagnostic-quality to hold the position for a recording Further, a score on the expected diagnostic quality of the Auto-Record and Best-Effort-Record clips is indicated such that the user can make decisions to further optimize the quality | Substantially Equivalent |
| Manual Image Recording | Manual recording | Manual recording if prompted by the user | Same |
| Automatic Capture of Clips and Predicted Diagnostic Quality | Auto-Capture: The Caption Guidance Auto-Capture feature triggers an automatic capture of a clip when the quality is predicted to be diagnostic, emulating the way in which a sonographer knows when an image is of sufficient quality to be diagnostic and records it. | Auto-Record: automates the capture of diagnostic-quality recordings, emulating how a sonographer knows when an image is of sufficient quality to be diagnostic and records it. | Substantially Equivalent |
| Retrospectively Recording of Highest Quality Clip | Save Best Clip: This feature continually assesses clip quality while the user is scanning and, if the user is not able to obtain a clip sufficient for | Best-Effort-Record: this feature stores the clip that has been evaluated as having the best expected diagnostic quality during an exam for a given reference view. If the user | Substantially Equivalent |
7
{11}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
| | Auto-Capture, the software allows the user to retrospectively record the highest quality clip obtained so far, mimicking the choice a sonographer might make when recording an exam. | does not manage to perform an auto-record for the reference view, the Best-Effort-Record clip is proposed to the user. | |
| --- | --- | --- | --- |
| Deep Learning Based Algorithm | Yes | Yes | Same |
| Machine Learning – Based Algorithm | Yes | Yes | Same |
# NON-CLINICAL AND CLINICAL TESTS SUMMARY
# 1- Human Factors Validation Study
A human factors validation study on HeartFocus was conducted following the FDA Guidance Document, « Applying Human Factors and Usability Engineering to Medical Devices » and IEC 62366 1:205 standard. This study aimed to demonstrate that the HeartFocus user interface is safe and effective for the intended users, uses, and use environments. The HeartFocus user interface was developed through a series of preliminary evaluations during its development process to find the best and most effective design. The final version of the device has then been assessed in a summative evaluation.
During this evaluation, the characteristics of the user interface that could affect safety were identified by conducting a task analysis and recorded as use scenarios. 9 use scenarios, representing the primary functions of HeartFocus have been established. Potential errors leading to potential risks in these use scenarios have then been identified. The 7 scenarios concerned were considered "hazard-related" and selected for the usability summative
{12}
DESKi
HeartFocus 510(k) Summary
Apr 4, 2025
evaluation, but no risk was identified regarding the health and safety of users.
The summative evaluation was conducted with 31 participants representing the intended users of HeartFocus and included novices (registered nurse) and trained medical professionals (Cardiologists, Sonographers, Intensivists, Emergency Physicians).
User tests were carried out to observe the execution of all hazard-related scenarios. They were performed in an environment representative of the intended use conditions of HeartFocus.
The user tests achieved a high success rate, with 100% success in 4 of the 7 scenarios and 97% success in the remaining 3, providing robust evidence that critical tasks can be performed with high accuracy and minimal risk.
## 2- Algorithms Performance Testing
Performance testing of the AI/ML algorithms has been conducted to evaluate the effectiveness of the AI/ML features of the HeartFocus software.
In total, AI/ML algorithms were trained and tuned on 1,483 patients and 1,204,113 ultrasound images. 290 patients and 361,104 ultrasound images were used for performance testing. Data collected as part of the clinical trial, containing 240 patients (120 US + 120 non-US), were also employed for additional retrospective evaluation. Both training/tuning data and test data were collected on patients of varying body mass index (BMI), age, and sex.
The main features of the system were tested independently using the primary endpoints described in Table 2. Secondary endpoints were also evaluated to demonstrate the ability of the algorithms to:
- accurately detect anatomical structures of the heart in ultrasound images to indicate whether they can be used to guide the user,
- maximize the likelihood of capturing high-quality records,
- provide a relevant estimate of recording quality,
- automatically record clips identified to be of diagnostic quality by experts,
{13}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
- guide from one reference view to another.
Subgroup analyses were performed across varying BMI (< 25, ≥ 25 and < 30, ≥ 30), age (< 65, ≥ 65), and sex (male, female).
Table 2: Primary objectives, endpoints, and success criteria
| Feature | AI/ML Algorithm | Objective | Endpoint | Success Criteria |
| --- | --- | --- | --- | --- |
| Diagnostic-quality view detection | View Classification | Ability to classify ultrasound images with similar accuracy as experts | Cohen's kappa score between the model's predictions and the ground truth labels made by experts (by frame) | Lower bound of the 95% confidence interval of the Cohen's kappa score > 0.6 (substantial agreement) for the 10 reference view |
| Live guidance | Guidance | Ability to provide successful guidance cues on ultrasound frames | Positive predictive value of successful guidance cues (by frame) | Lower bound of the 95% confidence interval of the positive predictive value > 0.8 for the 10 reference views |
| Auto record | View Classification + Recording | Ability to save high-quality records according to experts | Positive predictive value of high-quality records among auto-records (by clip) | Lower bound of the 95% confidence interval of the positive predictive value > 0.6 and point estimate of the positive predictive value > 0.8 for the 10 reference views |
# Diagnostic-quality view detection
Experts (cardiologists and/or experienced sonographers) annotated the ultrasound clips to identify diagnostic-quality images. First annotation was performed by an expert (expert annotator) and was reviewed by a second expert (expert reviewer). When necessary, disagreements were resolved either through direct reconciliation by the 2 experts or by a third expert. The ground truth (or gold standard) was defined from the consensus between the first expert annotator and the expert reviewer(s). For each ultrasound frame, the View Classification model predicts whether it is of diagnostic quality. The evaluation consists of comparing the
{14}
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
agreement between the model predictions and the experts' annotations using Cohen's kappa score.
The evaluation was performed on 30,361 images from 14 patients. Cohen's kappa scores range from 0.699 [0.673, 0.724] to 0.873 [0.861, 0.884], meeting the success criteria of Cohen's kappa score $>0.6$ on the lower bound of the $95\%$ CI for each reference view.
# Live guidance
Live guidance was evaluated on ultrasound acquisitions where the frame and probe position were collected simultaneously. For a given reference view, each ultrasound frame is at a certain position (called the actual position) regarding the target position (where diagnostic-quality frames are captured). The Guidance model predicts guidance cues to help navigate toward this target position. The evaluation consists of computing the positive predictive value (PPV), which represents the proportion of ultrasound frames where the final position, after following the guidance cues, is closer to the target position than the actual position.
The evaluation was performed on 270,582 images from 20 patients. Guidance cues PPV ranges from 0.810 [0.804, 0.816] to 0.953 [0.951, 0.955], satisfying the success criteria of PPV $>0.8$ on the lower bound of the $95\%$ CI for each reference view.
# Auto record and Best-effort record
For each reference view, long-duration clips where an operator moves the ultrasound probe and obtains both diagnostic-quality and non-diagnostic-quality frames were collected. Based on the View Classification model's output, the Recording algorithms aim to automatically record diagnostic-quality clips (short duration) that could be saved in the clinical practice. The positive predictive value (PPV) of the Recording algorithms is evaluated for each reference view.
The evaluation was performed on 211 long-duration clips from 34 patients. While using the Auto record feature solely, the PPV ranges from 0.846 [0.665, 0.938] to 1.000 [0.908, 1.000], meeting the success criteria of PPV $>0.6$ on the lower bound of the $95\%$ CI and PPV $>0.8$ on the point estimate for each reference view. While using both the Auto record and Best-Effort record, the PPV ranges from 0.816 [0.666, 0.908] to 1.000 [0.914, 1.000].
{15}
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
# 3- Clinical Study
A prospective multicentric clinical study was conducted to evaluate the use of HeartFocus by medical professionals without prior echocardiography training.
# Study Design
Eight (8) registered nurses (RNs) were trained and evaluated on their performance to acquire a 10-view 2D-TTE protocol (two-dimensional transthoracic echocardiography). Participants were scanned by the RN (study exam) and 10 standard views were obtained using a Clarius ultrasound system with HeartFocus software: Parasternal Long Axis (PLAX), Parasternal Short Axis at the Aortic Valve (PSAX-AV), Parasternal Short Axis at the Mitral Valve (PSAX-MV), Parasternal Short Axis at the Papillary Muscle (PSAX-PM), Apical-4-Chamber (A4C), Apical-5-Chamber (A5C), Apical-2-Chamber (A2C), Apical-3-Chamber (A3C), Subcostal-4-Chamber (SC-4C), and Subcostal Inferior Vena Cava (SC-IVC). The study continued enrollment until the eight RNs had completed scans of 30 patients each. For comparison, participants were also scanned to obtain the same 10 views by a trained medical professional without cardiac guidance and using the same Clarius ultrasound system (control exam).
Following the study and control exams, a panel of five (5) expert cardiologist readers (not the same as the ones scanning the patients) independently provided assessments of whether the patient study, in its totality, provided sufficient information to assess 12 clinical parameters. In addition, this panel of five (5) expert cardiologist readers also independently evaluated if the diagnostic image quality per clip was sufficient for clinical interpretation using the ACEP scale (ACEP $\geq$ 3, Liu, R. B. et al. Emergency Ultrasound Standard Reporting Guidelines); the cardiologists graded each clip. The readers were blinded to assessments from other panel members as well as to which site the images were obtained and whether the images were obtained by an RN or a trained medical professional (sonographer or cardiologist). The results from the expert panel reads were used for the statistical analysis.
To reduce possible sources of bias in the design, the RNs, sonographers, and cardiologists were all blinded to results determined by others.
{16}
Four (4) prospectively defined primary endpoints were evaluated sequentially for the study, all of which assessed whether the patient study exam conducted by the RN, taken as a whole, was of sufficient image quality to visually make these clinical assessments. Specifically, the endpoints assessed whether, in the judgment of expert cardiologists, the studies permitted qualitative visual assessment of left ventricular (LV) size, LV function, right ventricular (RV) size, and the presence of non-trivial pericardial effusion. In addition, expert cardiologist readers performed a qualitative and quantitative assessment of the ultrasound measurements. They visually determined the presence of LV or RV hypertrophy, dilation of the left or right ventricle or atrium, abnormal LV or RV function, abnormal mitral or tricuspid or aortic valve (i.e. structurally normal, abnormal, suspected device), pericardial effusion, dilatation of the IVC or any other abnormality. The cardiologists also measured the LV end-systolic and end-diastolic volumes and function. On parasternal analysis, they measured the septal and posterior wall thickness, the internal diameter of the LV (systole and diastole), as well as the aortic root. The diameter of the IVC was also evaluated. The acquisition time for the limited ultrasound exam for the novices was also collected.
# Results
A total of 240 adults aged 22 years and older who were scheduled for a clinically indicated echocardiography examination at one of the two investigation centers were included in this study, with 120 patients at Site 1 (France) and 120 patients at Site 2 (USA).
The four primary endpoints were satisfied and demonstrated the clinical utility of HeartFocus guidance for medical professionals without specialized echocardiography training. Specifically, the eight (8) RNs acquired echocardiographic exams of sufficient image quality to make clinical assessments in the proportion of study exams conducted, as shown below.
Table 3: Primary endpoint results
| Endpoint | Percent of diagnostic quality % [95% Wilson CI*] |
| --- | --- |
| Qualitative Visual Assessment of LV Size | 100 [98.4;100] |
| Qualitative Visual Assessment of LV Function | 100 [98.4;100] |
{17}
DESKi HeartFocus 510(k) Summary Apr 4, 2025
| Qualitative Visual Assessment of RV Size | 100 [98.4;100] |
| --- | --- |
| Qualitative Visual Assessment of Non-Trivial Pericardial Effusion | 100 [98.4;100] |
*Wilson Score method is used due to the point estimate is at or near borderline of 100%
Secondary endpoints and additional analyses presented below were not evaluated based on a specific hypothesis. Since the evaluation of secondary endpoints and additional analyses did not allow for control of Type I error, the study results are presented as a descriptive demonstration of the use of HeartFocus guidance for the specific secondary endpoints and additional analyses.
Additional secondary endpoints were evaluated and demonstrated the robustness of the data, including eight (8) additional patient-level clinical parameters evaluated and each had a high proportion of scans considered to be of sufficient image quality to make each of the eight (8) additional patient-level clinical parameter assessments, i.e., qualitative visual assessment of RV function, inferior vena cava (IVC) size, left atrial (LA), right atrial (RA) size, aortic valve (AV), mitral valve (MV), tricuspid valve (TV) and segmental kinetics of the LV.
Specifically, the eight (8) RNs acquired echocardiographic exams of sufficient image quality to visually make clinical assessments in the proportion of study exams conducted, as shown below.
Table 4: Secondary endpoint results
| Endpoint | Percent of diagnostic quality % [95% Wilson CI*] |
| --- | --- |
| Qualitative Visual Assessment of RV Function | 99.6 [97.7;99.9] |
| Qualitative Visual Assessment of LA Size | 100 [98.4;100] |
| Qualitative Visual Assessment of RA Size | 98.8 [96.4;99.6] |
| Qualitative Visual Assessment of Segmental Kinetics of the LV | 95.4 [92.0;97.4] |
| Qualitative Visual Assessment of AV | 98.8 [96.4;99.6] |
{18}
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
| Qualitative Visual Assessment of MV | 100 [98.4;100] |
| --- | --- |
| Qualitative Visual Assessment of TV | 95.4 [92.0;97.4] |
| Qualitative Visual Assessment of IVC Size | 78.3 [72.7;83.1] |
*Wilson Score method is used due to the point estimate is at or near borderline of 100%
In addition to assessing if image quality was sufficient to make assessments, cardiologists also made specific qualitative visual assessments based on the study and control exams (e.g., presence or absence of non-trivial pericardial effusion). The proportion of scans in which the diagnostic decision was the same between study and control exams was very high, further demonstrating the usability of HeartFocus guidance. For primary clinical parameters, the range was $87.5\%$ to $98.3\%$ . Similarly, for secondary clinical parameters, the range was $87.1\%$ to $99.6\%$ .
To provide a robust assessment of the performance of HeartFocus guidance, subjects enrolled in the study included a broad range of patient characteristics representative of the intended use population. The standard of care echo revealed a high proportion of cardiac abnormalities (70.4%). In addition, many of the study patients were inpatients or had other challenging characteristics, such as high BMI (59.6% with BMI > 25) and cardiac implantable devices (18.8%).
Subgroup analyses were conducted to evaluate the impact of specific baseline and demographic characteristics (i.e., BMI, presence of known cardiac abnormalities, scan sequence number within each acquiring nurse, and study site) on the outcomes of the primary and secondary endpoints. The results demonstrated consistent performance across subgroups.
Furthermore, it was evaluated whether the RN users were able to obtain a high proportion of clips that were considered diagnostic quality. Specifically, the eight (8) RNs acquired echocardiographic clips of diagnostic-image quality for each of the standard views in the following proportion of study exams conducted.
{19}
Table 5: Study results: diagnostic-quality clips
| View | Percent of diagnostic quality % [95% Wilson CI*] |
| --- | --- |
| PLAX | 97.5 [94.7;98.8] |
| PSAX-AV | 89.2 [84.6;92.5] |
| PSAX-MV | 90.8 [86.5;93.9] |
| PSAX-PM | 97.1 [94.1;98.6] |
| A4C | 96.2 [93.0;98.0] |
| A5C | 93.3 [89.4;95.9] |
| A2C | 82.5 [77.2;86.8] |
| A3C | 90.0 [85.6;93.2] |
| SC-4C | 89.2 [84.6;92.5] |
| SC-IVC | 77.5 [71.8;82.3] |
*Wilson Score method is used due to the point estimate is at or near borderline of 100%
The objective was to acquire 10 clips per patient, so in total, 2400 clips. Novices recorded 2362 clips and missed 38 clips. Among the recorded clips, $67.4\%$ of the clips were saved with the "Auto record" feature, $21.3\%$ with the "Best-effort record" feature, and $11.3\%$ with the "manual record" feature.
99.7% of the clips recorded with the "Auto record" feature were of diagnostic quality, demonstrating the high specificity of this feature. 93.9% of the clips recorded with the "best-effort record" feature were of diagnostic quality, and 40.5% of the clips recorded with the "manual record" feature were of diagnostic quality.
The average acquisition time per limited exam for novices in both centers was $23.6 \pm 10.6$ minutes.
{20}
DESKi
HeartFocus 510(k) Summary
Apr 4, 2025
# PREDETERMINED CHANGE CONTROL PLAN
HeartFocus could expand its compatibility to additional ultrasound systems and additional operating systems. This will be addressed through 3 modifications in the Predetermined Change Control Plan (PCCP).
- M1 - Retraining of the core algorithms: This modification's scope is the retraining of the AI/ML models in the perspective of validating a new manufacturer prior to an M2. The modification is limited to retraining with additional data without changing the models' architecture or training procedure. It notably describes the data collection to retrain the models and the evaluation plan to validate the performance of the retrained models on the already-cleared ultrasound systems.
- M2 - Extend the use of the ML-DSF with new ultrasound systems: This modification's scope is the validation of HeartFocus on new ultrasound systems. It notably describes the data collection and evaluation plan to validate the performance of the AI/ML models on the new ultrasound system.
- M3 - Extend the use of the ML-DSF with new operating systems: This modification's scope is the extension of the compatibility with new operating systems.
M1 and M2 will both be triggered when DESKi makes a partnership with a new ultrasound manufacturer, and will always be performed together.
Table 6 describes, for each modification, its rationale and how it will be tested to ensure the safety and effectiveness of the device after the modification.
{21}
Table 6: Rationale and testing for each modification
| Modification | Rationale | Testing |
| --- | --- | --- |
| M1 - Retraining of the core algorithms | Enhance the performance of the AI models in the perspective of extending the use of the ML-DS with new ultrasound systems | Execute AI performance on the reference test set and system-level validation tests |
| M2 - Extend the use of the ML-DSF with new ultrasound systems | Enable support for a broader range of ultrasound systems | Execute AI performance tests on new system test set and system-level validation test |
| M3 - Extend the use of the ML-DSF with new operating systems | Enable support for a broader range of devices | Execute system-level validation test |
Table 7 describes the requirements for the ultrasound systems that can be cleared through the PCCP (M1+M2).
Table 7: Ultrasound system requirements
| Requirement | Description |
| --- | --- |
| FDA-cleared ultrasound system compatible with HeartFocus' intended use | ·Ultrasound system shall be FDA-cleared ·Validated for cardiac use ·Validated for use on adults |
| Integration features | ·Technical integration shall allow HeartFocus to access live ultrasound stream ·Compatible for an integration on HeartFocus' operating systems |
| Minimal specifications | ·The ultrasound system shall provide an image stream at a minimum of 15 frames per second. ·Center or nominal frequency within 2–5 MHz, suitable for adult transthoracic echocardiography ·At least 30 cm depth, to allow full visualization of adult cardiac structures ·2D B-mode imaging (grayscale) ·Cone-shaped field of view with an opening angle of at least 80 degrees |
{22}
For each M1+M2, at least 20 new patients (~270k frames) collected with the new ultrasound system will be added to the train set to retrain the AI/ML models. The hyperparameters will be tuned for the retraining of the AI/ML algorithms (if necessary). Additional testing data will be collected on the new ultrasound system to validate it. 40 US patients (~400 clips) and 40 non-US (~540k frames) patients will be included, ensuring that at least $25\%$ of patients have a BMI $\geq 30~\mathrm{kg / m2}$ .
The performance of the retrained algorithms will be validated on: the reference test set (used in the initial 510(k) submission), the test set(s) collected for the validation of ultrasound systems through previous M1+M2 (if applicable), and the new test set collected on the new ultrasound system. Performance testing on all these data should meet the success criteria defined in Table 8. Performance testing will integrate subgroup analysis on the ultrasound system.
Table 8: Primary objectives, endpoints, and success criteria for M1+M2 modifications.
| Feature | AI/ML Algorithm | Objective | Endpoint | Success Criteria |
| --- | --- | --- | --- | --- |
| Diagnostic-quality view detection | View Classification | Ability to classify ultrasound images with similar accuracy as experts | Cohen's kappa score between the model's predictions and the ground truth labels made by experts (by frame) | Lower bound of the 95% confidence interval of the Cohen's kappa score > 0.6 (substantial agreement) for the 10 reference view |
| Live guidance | Guidance | Ability to provide successful guidance cues on ultrasound frames | Positive predictive value of successful guidance cues (by frame) | Lower bound of the 95% confidence interval of the positive predictive value > 0.8 for the 10 reference views |
| Auto record | View Classification + Recording | Ability to save high-quality records according to experts | Positive predictive value of high-quality records among auto-records (by clip) | Lower bound of the 95% confidence interval of the positive predictive value > 0.6 and point estimate of the positive predictive value > 0.8 for the 10 reference views |
{23}
| DESKi | HeartFocus 510(k) Summary | Apr 4, 2025 |
| --- | --- | --- |
In case M1+M2 fails (success criteria are not met), a single second iteration of M1+M2 will be performed. For the second iteration, the algorithm will be retrained using the existing training dataset and the non-US testing data (40 patients from the new ultrasound system) collected during the first M1+M2 iteration. A new test set of 40 non-US patients and 40 US patients will be collected using the new ultrasound system. The algorithm will then be evaluated using this new test set as well as the reference test set used in the initial 510(k) submission (34 patients from the standalone study and 240 patients from the clinical study, using retrospective analysis) and the test set(s) collected for the validation of previously cleared ultrasound systems (40 US + 40 non-US patients for each previously cleared ultrasound systems). If the second iteration of M1+M2 is not successful, the extension to the new ultrasound system will not be carried out through this PCCP. Each iteration of M1+M2 will be documented.
# CONCLUSION
The HeartFocus software is similar in its technological features to its predicate device, Caption Guidance (DEN 190040). Both systems are intended as an assistive tool to aid medical professionals in the acquisition of cardiac ultrasound images, citing the same echocardiographic views. Both systems apply deep learning based algorithms to emulate the expertise of a sonographer by providing real-time guidance on how to position the transducer on the patient's body. Both are installed on a third-party, previously cleared ultrasound device. There are minor technological differences between the subject and the predicate device, such as the compatible scanner probe and minor differences in how guidance is provided to the user, however, these differences do not raise new or different questions of safety or effectiveness since the principal technology on which the device is based is similar. Thus, Heartfocus is substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.