Customized Abutment and Screw

K242768 · Chengdu Besmile Medical Technology Co., Ltd. · NHA · Sep 2, 2025 · Dental

Device Facts

Record IDK242768
Device NameCustomized Abutment and Screw
ApplicantChengdu Besmile Medical Technology Co., Ltd.
Product CodeNHA · Dental
Decision DateSep 2, 2025
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

Customized Abutment and Screw is used as a complement component for dental implants placed in the jawbone after tooth loss, to connect, support and retain the restoration or implant superstructure. Customized Abutment and Screw is indicated for single unit restorations. Customized Abutment and Screw is compatible with following Implant Systems: [Table of compatible systems omitted for brevity]. All digitally designed abutments for use with Customized Abutment and Screw are intended to be manufactured at a BESMILE validated milling center.

Device Story

Customized Abutment and Screw serves as an interface between endosseous dental implants and prosthetic restorations. Input consists of patient-specific digital design data; device is manufactured from Ti-6Al-4V (abutment) and Ti-6Al-4V ELI (screw) via milling at a validated center. Device is installed by dental professionals in clinics to connect, support, and retain single-unit restorations. Output is a physical, patient-specific abutment that stabilizes the prosthesis. Clinical benefit includes restoration of function and aesthetics for patients with tooth loss. Healthcare providers use the device to secure the final restoration to the implant platform, directly impacting the stability and longevity of the dental prosthesis.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including dimensional analysis, reverse engineering, static and dynamic fatigue testing per ISO 14801, biocompatibility evaluation per ISO 10993, sterilization validation, surface cleanliness analysis (SEM/EDS), and MRI safety assessment based on scientific literature.

Technological Characteristics

Materials: Ti-6Al-4V (ASTM F1472) for abutment, Ti-6Al-4V ELI (ASTM F136) for screw. Internal connection type. Dimensions: 4.0-6.5mm diameter, 1-5mm transgingival height, 4-12mm post height. Non-sterile, single-use. Steam sterilization. Digitally designed and milled at validated centers.

Indications for Use

Indicated for single unit restorations in patients with tooth loss requiring dental implant-supported prosthetics. Compatible with specific Straumann and Nobel Biocare implant systems.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 09, 2025 Chengdu Besmile Medical Technology Co., Ltd. Moushan Liu Manager No.9,Sec.2, Shengwucheng North Rd., Chengdu Tianfu International Bio-Town, Shuangliu District, Chengdu, Sichuan Chengdu, 610200 CHINA Re: K242768 Trade/Device Name: Customized Abutment and Screw Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 21, 2025 Received: August 1, 2025 Dear Moushan Liu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K242768 - Moushan Liu Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K242768 - Moushan Liu Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Sherrill Lathrop Blitzer for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K242768 Device Name Customized Abutment and Screw Indications for Use (Describe) Customized Abutment and Screw is used as a complement component for dental implants placed in the jawbone after tooth loss, to connect, support and retain the restoration or implant superstructure. Customized Abutment and Screw is indicated for single unit restorations. Customized Abutment and Screw is compatible with following Implant Systems: | No. | Implant System Compatibility | Implant Body Diameter(mm) | Platform Diameter(mm) | | --- | --- | --- | --- | | 1 | Straumann Bone Level RC 4.1, RC 4.8 (composed of pure titanium) | 4.1,4.8 | 4.1,4.8 | | 2 | Straumann Bone Level NC 3.3 (composed of pure titanium) | 3.3 | 3.3 | | 3 | Nobel Biocare Active NP3.5 | 3.5 | 3.5 | | 4 | Nobel Biocare Active RP4.3/5.0 | 4.3, 5.0 | 3.9, 3.9 | All digitally designed abutments for use with Customized Abutment and Screw are intended to be manufactured at a BESMILE validated milling center. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} 510(k) Summary for K242768 # 1 510(k) Submitter Device Submitter: Chengdu Besmile Medical Technology Co., Ltd. No.9,Sec.2, Shengwucheng North Rd., Chengdu Tianfu International Bio-Town, Shuangliu District, Chengdu, Sichuan 610200, P.R.China Contact Person: Moushan Liu (Mr.) Manager Phone: +86-28-81501188 E-mail: xuetaohu@cdbesmile.com # 2 Device Trade Name of Device: Customized Abutment and Screw Classification Name: Endosseous Dental Implant Abutment Regulation Number: 21 CFR 872.3630 Device Classification: II Product Code: NHA Classification Panel: Dental Date prepared: April 28, 2025 # 3 Predicate Devices ## Primary Predicate Device K234112, Customized Abutment ## Reference Devices K193425, Pre-milled Blank # 4 Device Description The Customized Abutment and Screw is composed of two parts: the customized abutment and screw. Among them, the customized abutment is machined from Ti-6Al-4V titanium alloy material in accordance with ASTM F1472, and the screw is machined from Ti-6Al-4V ELI titanium alloy material in accordance with ASTM F136. And the surface of product is not modified. The lower part of the product has a fixed interface shape, and the upper part is machined according to the needs of the patient. In the process of dental implant surgery, as the upper structure of the implant, it is installed on the implant platform Page 1 of 12 {5} anchored in the bone, which plays the role of supporting, retaining and stabilizing the prosthesis. Customized Abutment and Screw is created via use of the Pre-milled Blank. The information of the Implant Systems compatible with the product is detailed in Table 1. Table 1 Compatible Implant Systems | No. | Model | Implant System Compatibility | Implant Body Diameter (mm) | Platform Diameter (mm) | 510(k) | Implant Name | Company Name | | --- | --- | --- | --- | --- | --- | --- | --- | | 1 | QU-02 | Straumann Bone Level RC 4.1, RC 4.8 (composed of pure titanium) | 4.1,4.8 | 4.1,4.8 | K140878 | Straumann® Bone Level Tapered Implants | Straumann USA, LLC | | 2 | QU-03 | Straumann Bone Level NC 3.3 (composed of pure titanium) | 3.3 | 3.3 | K140878 | Straumann® Bone Level Tapered Implants | Straumann USA, LLC | | 3 | HE-04 | Nobel Biocare Active NP3.5 | 3.5 | 3.5 | K142260 K083205 | NobelActive® NobelActive 8.5 mm & 18.0 mm | Nobel Biocare AB | | 4 | HE-05 | Nobel Biocare Active RP4.3/5.0 | 4.3,5.0 | 3.9,3.9 | K142260 K083205 | NobelActive® NobelActive 8.5 mm & 18.0 mm | Nobel Biocare AB | # 5 Intended Use The product is used as a complement component for dental implants implanted in the jaw after tooth loss to attach, support and retain the prosthesis or implant superstructure. # 6 Intended users The target users of the product are hospitals, clinics or dental technicians. Users need to acquire relevant skills, knowledge and professional training before using these products. # 7 Indications for use Customized Abutment and Screw is used as a complement component for dental implants placed in the jawbone after tooth loss, to connect, support and retain the restoration or implant superstructure. Customized Abutment and Screw is indicated for single unit restorations. Customized Abutment and Screw is compatible with following Implant Systems: {6} | No. | Implant System Compatibility | Implant Body Diameter(mm) | Platform Diameter(mm) | | --- | --- | --- | --- | | 1 | Straumann Bone Level RC 4.1, RC 4.8 (composed of pure titanium) | 4.1,4.8 | 4.1,4.8 | | 2 | Straumann Bone Level NC 3.3 (composed of pure titanium) | 3.3 | 3.3 | | 3 | Nobel Biocare Active NP3.5 | 3.5 | 3.5 | | 4 | Nobel Biocare Active RP4.3/5.0 | 4.3, 5.0 | 3.9, 3.9 | All digitally designed abutments for use with Customized Abutment and Screw are intended to be manufactured at a BESMILE validated milling center. Page 3 of 12 {7} 8 Substantial Equivalence Discussion | Table 2 Comparison of Indications for Use | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Parameter | Subject Device Customized Abutment and Screw K242768 | | | | Primary Predicate Device Customized Abutment K234112 | | | | Reference Device Pre-milled Blank K193425 | | | Indications for Use | Customized Abutment and Screw is used as a complement component for dental implants placed in the jawbone after tooth loss, to connect, support and retain the restoration or implant superstructure. Customized Abutment and Screw is indicated for single unit restorations. Customized Abutment and Screw is compatible with following Implant Systems: | | | | The Customized Abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Customized Abutments are indicated for screw retained single restorations or cement-retained single or multi-unit restorations. | | | | ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw. ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table: Compatibility Table | | | | No. | Implant System Compatibility | Implant Body Diameter (mm) | Platform Diameter (mm) | No. | Proprietary Name | Implant diameter size (mm) | Restorative Platform diameter (mm) | | | | | 1 | Straumann Bone Level RC 4.1, RC 4.8 (composed of pure titanium) | 4.1,4.8 | 4.1,4.8 | 1 | Zimmer Tapered Screw-vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | ARUM Pre-Milled Blank | Implant Platform | | | | | | | 2 | Straumann® Bone Level Tapered Implant | 3.3, 4.1, 4.8 | 3.1, 3.7, 4.4 | | | | | | | | | All digitally designed abutments for use with | | | | | | Page 4 of 12 {8} | | 2 | Straumann Bone Level NC 3.3 (composed of pure titanium) | 3.3 | 3.3 | Customized Abutment are intended to be manufactured at an ARUM DENTISTRY validated milling center. | Ø10 mm | Ø14 mm | m compatibility | m diameter [mm] | diameter [mm] | w | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 3 | Nobel Biocare Active NP3.5 | 3.5 | 3.5 | | CIHE 037 | CIHE 038 | Nobel Active NP | 3.5 | 3.5 | CST O00 1 | | | 4 | Nobel Biocare Active RP4.3/5.0 | 4.3,5.0 | 3.9, 3.9 | | CIHE 039 | CIHE 040 | Nobel Active RP | 3.9 | 4.3/5.0 | | | | All digitally designed abutments for use with Customized Abutment and Screw are intended to be manufactured at a BESMILE validated milling center. | CIHE 135 | CIHE 136 | Nobel Active WP | | 5.1 | 5.5 | CST O00 2 | | | | | | | | | | | All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture. | | | | | | | | | | | | | | | | | | | # Discussion on differences in substantive equivalence: The Subject Device is compared with the Primary Predicate Device, the Primary Predicate Device can be used for single and multiple teeth, and the Subject Device is used for single teeth only. Therefore, the intended use of the Subject Device is included within the intended use of the Primary Predicate Device. The difference between the subject device and Primary Predicate Device is the compatible implant bodies. This comparison is for similarity of device, not for implant's compatibility. {9} The Subject Device, in contrast to the Reference Device (K193425) which can be used for single and multiple tooth restorations, is used only for single tooth restorations. Therefore, the intended use of the Subject Device is within the intended use of the Reference Device. | Table 3 Technological Characteristics' Comparison | | | | | | --- | --- | --- | --- | --- | | Parameter | Subject Device | Primary Predicate Device | Reference Device | Discussion | | K number | K242768 | K234112 | K193425 | / | | Trade Name | Customized Abutment and Screw | Customized Abutment | Pre-milled Blank | / | | Manufacturer | Chengdu Besmile Medical Technology Co., Ltd. | ARUM DENTISTRY Co., Ltd. | ARUM Dentistry Co., Ltd. | / | | Product Code(s) | NHA | NHA | NHA | Identical | | Device Class | Class II | Class II | Class II | Identical | | Regulation Number | 21 CFR 872.3630 | 21 CFR 872.3630 | 21 CFR 872.3630 | Identical | | Categorization by nature of body contact | Implant medical device: Tissue/bone-Medical devices principally contacting bone | Implant medical device: Tissue/bone-Medical devices principally contacting bone | Implant medical device: Tissue/bone-Medical devices principally contacting bone | Identical | | Contact duration | Long-term exposure | Long-term exposure | Long-term exposure | Identical | {10} | categories | | | | | | --- | --- | --- | --- | --- | | Restoration | Single-unit | Single-unit, Multi-unit | Single-unit & Multi-Unit | Different Comment 1 | | Diameter | 4.0-6.5mm | 3.3-6.0mm | / | Similar Comment 2 | | Transgingival height | 1-5mm | 0.5-4.0mm | 0.5-4.0mm | | | Post height | 4-12mm | 4-13mm | 4-13mm | | | Abutment Angle | 0-30° | 0-30° | 0-30° | | | Wall thickness | ≥0.5mm | 0.5-6.0mm | ≥0.5mm | Identical | | Material | Ti-6Al-4V (ASTM F1472)Ti-6Al-4VELI (ASTM F136) | Ti-6AL-4V ELI (ASTM F136) | Titanium Ti-6Al-4V ELI | Similar Comment 3 | | Surface Treatment | Untreated | / | Untreated | Identical | | Connection type | Internal connection | / | Internal connection | Identical | | Reusable | No, single-use | No, single-use | No, single-use | Identical | | Delivery | Non sterile | Non sterile | Non sterile | Identical | | Sterilization method | Steam Sterilization | Steam Sterilization | Steam Sterilization | Identical | {11} Page 8 of 12 # Comment 1 The Subject Device is used for single-unit restorations only, and the Primary Predicate Device (K234112) can be used for single-unit and multi-unit restorations. The Primary Predicate Device override the Subject Device. The difference do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. The Subject Device, in contrast to the Reference Device (K193425) which can be used for single and multiple tooth restorations, is used only for single tooth restorations. Therefore, the type of restoration for the Subject Device is within the restoration type of the Reference Device. # Comment 2 Subject Device Diameter ranges from 4.0-6.5mm, Primary Predicate Device(K234112) ranges from 3.3-6.0mm. Transgingival height range of Subject Device(1-5mm) exceeds the Primary Predicate Device((0.5-4.0mm). Primary Predicate Device's post height (4-13mm) contains post height range of the Subject Device (4-12mm). Primary Predicate Device and Subject Device have the same abutment angle range. There are slight differences in Abutment Design Parameters for the Subject Device, Primary Predicate Device. According to ISO 14801: 2016 Dentistry - Implants - Dynamic loading test for endosseous dental implants, the worst-case static fatigue test and dynamic fatigue test were performed on dental implants connected with the Subject Device. The test results show that the worst-case scenario of the Subject Device has sufficient strength for the intended clinical application. Therefore, slight differences in Abutment Design Parameters do not affect the intended use and do not adversely affect the safety and effectiveness of the Subject Device. Compared with Reference Device (K193425), Subject Device has the different post height and transgingival height. Subject Device is within the post height range of Reference Device; Subject Device is outside the transgingival height range of Reference Device. Abutments made of Pre-milled Blank are designed according to the patient's condition. We performed static fatigue tests and dynamic fatigue tests in accordance with ISO 14801:2016 Dentistry - Implants - Dynamic loading test for endosseous dental implants, static {12} fatigue test and dynamic fatigue test for the worst-case scenario. The test results show that the worst-case scenario of the Subject Device has sufficient strength for the intended clinical application. Therefore, differences in Abutment Design Parameters do not change the intended use of the device and do not raise safety effectiveness issues. ## Comment 3 The material of Subject Device abutment is Ti-6AL-4V, screw’s material is Ti-6AL-4V ELI. The material of Primary Predicate Device (K234112) is Ti-6AL-4V ELI. Both Ti-6AL-4V and Ti-6AL-4V ELI are common fabrication materials for abutment and implant with a long history of clinical use. According to ISO 14801:2016 Dentistry - Implants - Dynamic loading test for endosseous dental implants, the worst-case static fatigue test and dynamic fatigue test were performed on dental implants connected with Subject Device. The test results show that the worst-case scenario of the Subject Device has sufficient strength for the intended clinical application. Therefore, Subject Device is made of the same material as the commercially available products, both of which are titanium alloys, so the materials do not adversely affect the safety and effectiveness of the Subject Device. The abutment material of Subject Device is Ti-6AL-4V, screw’s material is Ti-6AL-4V Eli, and Reference Device’s (K193425) material is Ti-6AL-4V Eli. Both Ti-6AL-4V and Ti-6AL-4V ELI are common fabrication materials for abutment and implant with a long history of clinical use. According to ISO 14801:2016 Dentistry - Implants - Dynamic loading test for endosseous dental implants, we performed the static fatigue testing and dynamic fatigue testing of worst-case abutments made of Subject Device. The test results indicate that the worst case of the Subject Device has sufficient strength for the intended clinical application. Therefore, differences in materials do not change the intended use of the device and do not raise questions of safety and effectiveness. Page 9 of 12 {13} Page 10 of 12 # Substantial Equivalence Conclusion: The corresponding data show that the technical characteristics of the Subject Device and the Predicate Device(K234112) are basically identical. The minor difference between the Subject Device and the Predicate Device(K234112) is the difference in the compatible implant systems, but we compare the similarity of the devices rather than the compatibility of the implant systems, so it does not affect the intended use. We have also analyzed the compatibility of our products with the OEM implants, and the results show that our products are compatible with the OEM implants, and we have taken effective measures to ensure that our products maintain continuous compatibility with the appropriate OEM implants. Based on the above, we consider the Subject Device and the Predicate Device(K234112) to be essentially equivalent. Subject Device is as safe and effective as the reference device (K193425) when used as instructed by knowledgeable and trained dental personnel. The product is identical to its reference device (K193425) concerning to the indications for use. The corresponding data show that the technical characteristics of the Subject Device and the reference device (K193425) are basically identical. The minor difference between the Subject Device and the reference device (K193425) is the difference in the compatible implant systems. However, we compare the similarity of the devices, not the compatibility of the implant systems, so it does not affect the intended use. Based on the above points, we consider that Subject Device is essentially equivalent to reference device (K193425). {14} Page 11 of 12 ## 9 Performance Data The following performance data were provided in support of the substantial equivalence determination. ## 9.1 Non-Clinical Performance Data (Bench) Performance testing was conducted on the Subject Device according to the technical specification, and the test results meet the standard. ## 9.2 Engineering Analysis Dimensional analysis and reverse engineering of the implant-to-abutment connection platform were performed, including an assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible OEM implant body, OEM abutment, and OEM fixation screw. The testing demonstrated implant to abutment compatibility and has established substantial equivalency of the proposed device with predicate devices. ## 9.3 Fatigue Testing According to the FDA guidance document: Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments and ISO 14801 Dentistry - Implants - Dynamic loading test for endosseous dental implants, the static fatigue tests and dynamic fatigue tests were performed on the worst case for each implant which is compatible with, of the Subject Device. The test results show that the Subject Device has sufficient strength for the intended clinical application. ## 9.4 Biocompatibility Testing We conducted a biocompatibility evaluation in accordance with ISO 10993-1 Medical devices - Biological evaluation - Part 1: Evaluation and testing in the context of risk management. Biological test results comply with ISO 10993 series standards. ## 9.5 Sterilization validation Validation of moist heat sterilization showed that it could achieve a sterility assurance level (10⁻⁶). ## 9.6 Surface cleanliness analysis Scanning electron microscopy (SEM) and Energy dispersive X-ray spectroscopy (EDS) confirmed that the Subject Device surface was clean, with no chemical residues or particles remaining. {15} Page 12 of 12 ## 9.7 MR Environment Condition Non-Clinical worst-case MRI review was performed to evaluate the Subject Device in the MRI environment using scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan. (2019). Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783795), based on the entire system including all variations and material composition. Rationale addressed parameters per the FDA guidance “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment,” including magnetically induced displacement force and torque. ## 9.8 Shelf-life The shelf-life is not applicable because of low likelihood of time-dependent product degradation. ## 10 Clinical Test Conclusion No clinical study is included in this submission. ## 11 Conclusion The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The documentation submitted in this premarket notification demonstrates that the subject devices are substantially equivalent to the primary predicate and reference devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%