K233857 · JJGC Indústria e Comércio de Materiais Dentários S.A. · NHA · May 2, 2024 · Dental
Device Facts
Record ID
K233857
Device Name
Neodent Implant System – Custom Abutments
Applicant
JJGC Indústria e Comércio de Materiais Dentários S.A.
Product Code
NHA · Dental
Decision Date
May 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center. The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
Device Story
Customized prosthetic abutments; titanium alloy; used for screw-retained single or cemented-retained single/multiple restorations. Input: digital design files from CAD/CAM software. Transformation: milling at validated center to create patient-specific abutment geometry. Output: custom titanium abutment. Used in dental clinics; operated by dental professionals. Healthcare provider uses output to support prosthetic restorations (copings/crowns) on dental implants. Benefits: provides customized support for dental prosthetics, improving fit and function for patients with specific anatomical needs.
Clinical Evidence
No clinical data. Bench testing only. Dynamic fatigue testing per ISO 14801 demonstrated performance equivalent to predicates. Torsion testing confirmed adequate strength (2.0 N-cm over installation torque). Biocompatibility evaluated per ISO 10993-1, -5, -12, and -18. MR compatibility assessed via analysis of material and dimensions relative to reference devices.
Technological Characteristics
Materials: Titanium alloy (ASTM F136). Design: Customized abutment with straight or angled screw channel. Interfaces: Grand Morse, Narrow Grand Morse, HS. Minimum wall thickness: 0.4mm. Sterilization: Non-sterile, validated for user-performed moist heat sterilization (SAL 1x10^-6). Connectivity: Digital workflow via CAD/CAM software.
Indications for Use
Indicated for patients requiring screw-retained single restorations or cemented-retained single or multiple restorations supported by dental implants. Used for Grand Morse (GM), Narrow Grand Morse (NGM), and Helix Short (HS) implant systems.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
MRI Compatibility For Existing Neodent Implant System (K182620)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
JJGC Indústria e Comércio de Materiais Dentários S.A. % Jennifer Jackson Sr. Director, Regulatory Affairs and Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
May 2, 2024
### Re: K233857
Trade/Device Name: Neodent Implant System - Custom Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: April 2, 2024 Received: April 2, 2024
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233857
#### Device Name
Neodent Implant System - Custom Abutments
#### Indications for Use (Describe)
Custom Abutment Ti with Screw (milled):
The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center.
#### Custom Abutment Ti with Angled Screw Channel (milled):
The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------|
| <p>区 Prescription Use (Part 21 CFR 801 Subpart D)</p> |
| <p> Over-The-Counter Use (21 CFR 801 Subpart C)</p> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# K233857 Traditional 510(k) Submission
# Neodent Implant System – Custom Abutments
510(k) Summary
# 510(k) Summary
### Submitter's Contact Information
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| On the behalf of: | |
| | JJGC Indústria e Comércio de Materiais Dentários AS (dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Paraná, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
| Contact Person: | Jennifer M. Jackson, MS, RAC<br>Sr. Director, Regulatory Affairs and Quality<br>Phone Number: +1 978 747-2509<br>Fax Number: +1 978 747-0023 |
| Prepared By &<br>Alternate Contact: | Leticia Milani<br>Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA<br>E-mail: leticia.milani@neodent.com |
| Date Prepared: | May 2, 2024 |
| Name of the Device | |
| Trade Names: | Neodent Implant System – Custom Abutments |
| Trade Names: | Needent Implant System - Custom Abutment |
|------------------------|------------------------------------------|
| Common Name: | Endosseous dental implant abutment |
| Classification Name: | Endosseous dental implant abutment |
| Regulation Number: | 21 CFR 872.3630, Class II |
| Device Classification: | II |
| Product Code(s): | NHA |
| Classification Panel: | Dental Products Panel |
{6}------------------------------------------------
# K233857 Traditional 510(k) Submission
# Neodent Implant System – Custom Abutments
510(k) Summarv
### Predicate Device(s)
Primary Predicate:
- . K203309 – Nuvo CF Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
Reference Devices:
- . K150367 – Neodent Implant System – CM Preface Abutment (JJGC Indústria e Comércio de Materiais Dentários S.A)
- K162890 Straumann SC CARES Abutment (Straumann USA, LLC) ●
- . K182620 – MRI Compatibility For Existing Neodent Implant System
### Device Description
The Custom Abutments subject to this submission are similar to the devices already cleared in previous Neodent and Nuvo Implant System submissions, per the predicate and reference devices described above. This submission intends to expand the portfolio of digital abutments for the Grand Morse (GM), Narrow Grand Morse (NGM) and Helix Short (HS) lines, to provide more treatment options for customers. These abutments are composed of a unique body with two regions: the upper region is the customized portion, while the end region presents the prosthetic interface that fits with the implant, which does not allow customization.
They are intended for single use and provided non-sterilization recommended before installation in the mouth.
All proposed abutments are delivered to the final user already customized in a validated milling center, with a straight channel solution. For the GM line, the abutments could also be milled with a new angled channel solution for the screw access. All subject devices must be used with their corresponding prosthetic interface.
#### Intended Use
The Custom Abutments are used onto dental implants to provide support for customized prosthetic restorations (copings and crowns), indicated for screw-retained or cement-retained single restorations. They are indicated according to the interocclusal space available, existing transmucosal height and three-dimensional position of the implant. The subject devices are available in GM, NGM and HS prosthetics interfaces compatible with the implants of the prosthetic interface.
{7}------------------------------------------------
# K233857 Traditional 510(k) Submission
# Neodent Implant System – Custom Abutments
510(k) Summarv
#### Indications for Use
The Custom Abutment Ti with Screw is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments Ti with Screw are indicated for screw-retained single restorations or cemented-retained single restorations. All digitally designed abutments for use with the Custom Abutment Ti with Screw are intended to be sent to Straumann for manufacturing at a validated milling center.
The Custom Abutment with Angled Screw Channel is a customized prosthetic abutment, manufactured in titanium alloy, placed onto dental implants to provide support for customized prosthetic restorations. All abutments are only intended to be digitally designed and manufactured using specifics CAD/CAM software according to digital dentistry workflow. Custom Abutments with Angled Screw Channel are indicated for screw-retained single restorations or cemented-retained single or multiple restorations. All digitally designed abutments for use with the Custom Abutment Ti with Angled Screw Channel are intended to be sent to Straumann for manufacturing at a validated milling center.
{8}------------------------------------------------
# Neodent Implant System – Custom Abutments
510(k) Summary
### Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE PREDICATE DEVICE | REFERENCE PREDICATE DEVICE |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Neodent Implant System<br>Customizable Abutments | K203309<br>Nuvo CF Implant System | K162890<br>Straumann SC CARES Abutment | K150367<br>Neodent Implant System - CM Preface<br>Abutment |
| Indications<br>for<br>Use | The Custom Abutment Ti with Screw is a customized prosthetic<br>abutment, manufactured in titanium alloy, placed onto dental<br>implants to provide support for customized prosthetic<br>restorations. All abutments are only intended to be digitally<br>designed and manufactured using specific CAD/CAM software<br>according to digital dentistry workflow. Custom Abutments Ti<br>with Screw are indicated for screw-retained single restorations<br>or cemented-retained single or multiple restorations. All digitally<br>designed abutments for use with the Custom Abutment Ti with<br>Screw are intended to be sent to Straumann for manufacturing<br>at a validated milling center.<br>The Custom Abutment with Angled Screw Channel is a<br>customized prosthetic abutment, manufactured in titanium<br>alloy, placed onto dental implants to provide support for<br>customized prosthetic restorations. All abutments are only<br>intended to be digitally designed and manufactured using<br>specific CAD/CAM software according to digital dentistry<br>workflow. Custom Abutments with Angled Screw Channel are<br>indicated for screw-retained single restorations or cemented-<br>retained single or multiple restorations. All digitally designed<br>abutments for use with the Custom Abutment Ti with Angled<br>Screw Channel are intended to be sent to Straumann for<br>manufacturing at a validated milling center. | The CARES® Abutment CF is a<br>customized prosthetic<br>abutment,<br>manufactured in titanium alloy, placed<br>onto dental implants to provide support<br>for customized prosthetic restorations<br>(copings or crowns). It is indicated for<br>screw-retained or cement-retained<br>single-unit restorations. All digitally<br>designed abutments for use with the<br>CARES® Abutment are intended to be<br>sent to Straumann for manufacturing at<br>a validated milling center. | Straumann SC CARES® abutments<br>are indicated for<br>single-tooth<br>replacements and multiple tooth<br>restorations. The<br>prosthetic<br>restoration can be cemented or<br>directly veneered/screw-retained. | PreFace Abutment is a titanium<br>abutment to be used in fabricating a full<br>custom abutment and placed onto<br>Neodent dental implants to provide<br>support<br>for customized prosthetic<br>restorations. PreFace Abutments are<br>indicated<br>for screw-retained single<br>restorations or cement-retained single or<br>multi-unit restorations. |
| Material | Devices: Titanium alloy, according to ASTM F136.<br>Screw: Titanium alloy, according to ASTM F136 or TAN. | Devices: Titanium alloy, according to<br>ASTM F136.<br>Screw: Titanium alloy, according to<br>ASTM F136. | Devices: TAN<br>Screw: TAN | Devices: Titanium alloy, according to<br>ASTM F136.<br>Screw: Titanium alloy, according to ASTM<br>F136. |
| Implant-Abutment<br>Connection | Grand Morse and Narrow Grand Morse<br>HS: Internal Hex | Internal Hex | SC Internal Connection | Cone Morse |
| Channel Solution | Straight and angled | Straight | Straight | Straight |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE PREDICATE DEVICE | |
| K Number | Neodent Implant System<br>Customizable Abutments | K203309<br>Nuvo CF Implant System | K162890<br>Straumann SC CARES Abutment | K150367<br>Neodent Implant System - CM Preface<br>Abutment |
| Minimum Wall<br>Thickness | 0.4mm | 0.4mm | 0.4mm | 0.4mm |
| Gingival Height | NGM Customized abutment 0.6-5.8mm<br>HS Customized abutment 0.2-5.4mm**<br>GM Customized abutment 0.6-5.8mm<br>GM Customized abutment with angled screw channel<br>0.6-5.6mm | Minimum 0.8mm | Minimum 0.1mm | Min 0.6mm |
| Maximum Gingival<br>Height | NGM Customized abutment 5.8mm<br>HS Customized abutment 5.4mm<br>GM Customized abutment with angled screw channel 5.6mm<br>GM Customized abutment 5.8mm | 5.8mm | | 3.5mm |
| Abutment Post<br>Height* | Minimum 4.0mm | Minimum 4.0mm | Minimum 6mm | Minimum 4.0mm |
| Maximum Abutment<br>Angulation<br>Customization | 30° | 30° | 30° | 30° |
| Single Use | Yes | Yes | Yes | Yes |
| Sterilization<br>Method | Provided non-sterile. Terminally sterilized by user via moist<br>heat. Moist heat cycle parameters have been validated to a<br>SAL of 1x10-6. | Provided non-sterile. Terminally<br>sterilized by user via moist heat. Moist<br>heat cycle parameters have been<br>validated to a SAL of 1x10-6. | Provided non-sterile. Terminally<br>sterilized by user via moist heat. Moist<br>heat cycle parameters have been<br>validated to a SAL of 1x10-6. | Provided non-sterile. Terminally sterilized<br>by user via moist heat. Moist heat cycle<br>parameters have been validated to a SAL<br>of 1x10-6. |
{9}------------------------------------------------
# Traditional 510(k) Submission
# Neodent Implant System – Custom Abutments
510(k) Summary
*Post height is the length above the abutment collar/gingival height.
**Helix Short (HS) Implant System, K223638, is a tissue level implant system.
{10}------------------------------------------------
# Traditional 510(k) Submission
# Neodent Implant System – Custom Abutments
Performance Testing
### Performance Testing
#### Bench Testing
Assessment regarding dynamic fatigue testing was conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801 "Dentistry — Implants — Dynamic loading test for endosseous dental implants". For dynamic fatigue tests, the results demonstrated that in identical conditions the subject devices exhibit a level of performance equivalent to that reviewed for the presented predicate devices.
Torsion tests were performed to evaluate the strength of the screw used to fix all subject abutments against maximum twisting forces. The results prove that there is an adequate torsion strength of 2.0 N-cm over the indicated installation torque.
#### Sterilization Validation
Sterilization validation was performed according to ISO 17665-1 "Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices" and ISO 17665-2 "Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1, as provided in K180536. A minimum Sterility Assurance Level (SAL) of 1x10° has been validated for this method. The subject devices are not represented to be "non-pyrogenic".
#### MR Compatibility Testing
The MR compatibility was performed to assess the risk of exposing patients who have implantable medical devices. An assessment was made to demonstrate that the subject devices do not configure a new worst case and can be represented by the previously conducted studies reviewed for reference devices, since both have the same raw material and similar dimensions. The subject devices are therefore MR conditional devices and a patient treated with the subject devices can be safely scanned observing the parameters previously established per reference devices.
{11}------------------------------------------------
#### Biocompatibility Testing
A biological assessment was performed according to ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and to the FDA Guidance document "Use of International Standard ISO 10993- 1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016". Representative samples of the subject devices were subjected to the following:
- Biocompatibility sample preparation was made according to ISO 10993-12.
- · Biological Safety Assessment guided by ISO 10993-1.
- · Chemical characterization was performed per ISO 10993-18.
- · Cytotoxicity testing was performed per ISO 10993-5.
The subject devices are equivalent in material and manufacturing processes to the primary predicate and reference devices, therefore, no new issues regarding biocompatibility were raised and no additional biocompatibility testing was required.
#### Conclusion
The documentation submitted in this premarket notification demonstrates the subject devices are substantially equivalent to the primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.