Verifine® Pen Needles

K242632 · Promisemed Hangzhou Meditech Co., Ltd. · FMI · Sep 26, 2024 · General Hospital

Device Facts

Record IDK242632
Device NameVerifine® Pen Needles
ApplicantPromisemed Hangzhou Meditech Co., Ltd.
Product CodeFMI · General Hospital
Decision DateSep 26, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic

Intended Use

Verifine® Pen Needle is intended for use with pen injector device for the subcutaneous injection of drugs.

Device Story

Verifine® Pen Needles are single-use, non-pyrogenic, sterile hypodermic needles designed for manual attachment to pen injectors for subcutaneous drug administration. The device consists of a needle tube (stainless steel), hub, shield, and container. The user manually inserts the needle into the skin. The device is intended for over-the-counter (OTC) use. It facilitates drug delivery by providing a fluid path from the pen cartridge to the subcutaneous tissue. The device is available in various gauges and lengths to accommodate different clinical needs and pen injector models. It does not include integrated safety features.

Clinical Evidence

No clinical data was necessary. Substantial equivalence was established through bench testing, including dimensional compliance, performance testing (dose accuracy, removal torque per ISO 11608-2), and biocompatibility testing (cytotoxicity, skin irritation, intracutaneous reactivity, and sensitization per ISO 10993-1).

Technological Characteristics

Single lumen hypodermic needle; stainless steel (X5CrNi18-10) tube; polypropylene (PP) hub/container; polyethylene (PE) needle shield; silicone oil lubricant; UV glue jointing; sterile (SAL 10^-6); manual operation; various gauges (29G-34G) and lengths (3.5mm-12mm); 3-bevel or 5-bevel tip options; compliant with ISO 11608-2.

Indications for Use

Indicated for use with pen injector devices for the subcutaneous injection of drugs in patients requiring such therapy.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 26, 2024 Promisemed Hangzhou Meditech Co., Ltd. % John Beasley Founding Member and Senior Consultant MedTech Review, LLC 257 Garnet Garden Street Henderson, Nevada 89015 Re: K242632 Trade/Device Name: Verifine® Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: September 3, 2024 Received: September 3, 2024 Dear John Beasley: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Shruti N. Mistry -S Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K242632 Device Name Verifine® Pen Needles Indications for Use (Describe) Verifine® Pen Needle is intended for use with pen injector device for the subcutaneous injection of drugs. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## K242632 510(k) Summary ## Verifine® Pen Needles #### 1. Submission Sponsor Promisemed Hangzhou Meditech Co., Ltd. No. 1388 Cangxing Street Cangqian Community, Yuhang District Hangzhou City Zhejiang 311121 China Contact: Mr. Zearou Yang Office Phone: +8657188772985 Email: jim@nemoto-do.co.jp #### 2. Submission Correspondent MedTech Review, LLC 257 Garnet Garden Street Henderson NV 89015 United States Contact: John Beasley Office Phone: +7027236889 Email:john@medtechreview.com #### 3. Date Prepared September 03, 2024 #### 4. Device Identification | Trade/Proprietary Name: | Verifine® Pen Needles | |-------------------------|----------------------------------| | Classification Name: | Needle, Hypodermic, Single Lumen | | Regulation Number(s): | 880.5570 | | Product Code(s): | FMI | | Class: | II | | Classification Panel: | General Hospital | #### 5. Legally Marketed Predicate Device(s) Device name: Verifine ® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle 510(k) number: K161950 Manufacturer: Promisemed Hangzhou Meditech Co., Ltd. {5}------------------------------------------------ #### 6. Device Description Verifine® Pen Needles, manufactured by Promisemed Hangzhou Meditech Co., Ltd, are designed for use with pen injectors for the administration of drugs. These are standard pen needles without integrated safety features. The devices are single-use, non-pyrogenic, and maintain a Sterility Assurance Level (SAL) of 10 °. The differences between the PromiseMed current cleared common type needle pen and the new design primarily involve changes to the trade name, the addition of new models, and modifications in some materials and dimensions. Key updates include a change in the trade name, the introduction of new models (FPN and NPN), the addition of new needle sizes and a 5-bevel needle tip option, and a shift from a specific intended use (administering insulin) to a general intended use (administering drugs). The materials used have also been slightly modified, such as changing the needle shield material from polypropylene (PP) to polyethylene (PE) for certain models, which only contact intact skin, ensuring no increased risk. These devices are intended for over-the-counter (OTC) use and require manual insertion by the user during drug administration. #### 7. Indication for Use Statement Verifine® Pen Needle is intended for use with pen injector device for the subcutaneous injection of drugs. #### 8. Substantial Equivalence Discussion The following table compares the Verifine® Pen Needles to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device. | SE Comparison | |---------------| |---------------| be provides a detailed comparison of the modified device, ensuring that no new concerns regarding safety o effectiveness have been introd | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------| | Common name | Insulin Pen Needles | No change | N/A | No new concerns. The<br>common name remains<br>unchanged, ensuring<br>consistency with the intended<br>use. | | | Trade name | Verifine® Common Type Insulin Pen Needle | IPN<br>FPN, NPN | Verifine® Pen Needles<br>Verifine® Plus Pen Needles | It is only for common type pen<br>needle. Only the trade name is<br>change. | No new concerns. The trade<br>name change does not impact<br>safety or effectiveness as the<br>device's function remains the<br>same. | | Classification name | Needle, Hypodermic, Single Lumen (21 CFR 880.5570) | No change | N/A | No new concerns. The<br>classification name remains<br>consistent. | | | Device class | Class II | No change | N/A | No new concerns. The device<br>remains in Class II, indicating<br>no change in risk level. | | | Product Code | FMI | No change | N/A | No new concerns. The product<br>code remains the same,<br>reflecting no change in device<br>functionality. | | | General description | The proposed product, insulin pen needle, is manufactured<br>by Promisemed Hangzhou Meditech Co.,<br>Ltd, which is designed for use with a pen injector for the<br>subcutaneous injection of insulin.<br>The products have two types, common type and safety<br>type. Both of the two types are sterile with a Sterility<br>Assurance Level (SAL) of 10-6, non-pyrogenic and single-use<br>devices. Each type has several models. Different models are<br>distinguished by needle gauge and length.<br>The Common Type Insulin Pen Needle consists of needle<br>container, needle shield, needle tube, needle hub, UV glue | The proposed product, Verifine® Pen Needles, is<br>manufactured by Promisemed Hangzhou Meditech Co.,<br>Ltd, which is designed for use with a pen injector for the<br>subcutaneous injection of insulin.<br>The product is common type without safety protective<br>feature. It is a sterile with a Sterility Assurance Level (SAL)<br>of 10-6, non-pyrogenic and single-use devices. Each type<br>has several models. Different models are distinguished by<br>needle gauge and length.<br>It is consisting of needle container, needle shield, needle<br>tube, needle hub, UV glue and silicone oil. UV glue is used | It is only for common type pen<br>needle.<br>Description about common type<br>is no change. | No new concerns. The change<br>to a broader intended use<br>does not introduce new risks<br>as the fundamental technology<br>and operation remain the<br>same. | | {6}------------------------------------------------ | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | | | | | | | | | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | | and silicone oil. UV glue is used to glue needle tube and<br>needle hub and the silicone oil is used to needle tube<br>lubrication.<br>The user proceeds with inserting the needle into the skin<br>manually.<br>Both the Common Type Insulin Pen Needle and the Safety<br>Type Pen Needle are for OTC use, and they are external<br>communication, blood indirect devices. The contact<br>duration for both subjected devices is within 24h, and they<br>belong to limited contact device according ISO 10993-1. | to glue needle tube and needle hub and the silicone oil is<br>used to needle tube lubrication.<br>The user proceeds with inserting the needle into the skin<br>manually.<br>It is intended for OTC use, and they are external<br>communication, blood indirect devices. The contact<br>duration for both subjected devices is within 24h, and they<br>belong to limited contact device according ISO 10993-1. | | | | | | | | | | | | Principles of operation | The user proceeds with inserting the needle into the<br>skin manually. The patient end and the cartridge end of the<br>tube are lubricated using a silicone based lubricant for ease<br>of injection and rubber septum penetration. | No change | N/A | No new concerns. The<br>principles of operation remain<br>unchanged, ensuring<br>consistent performance. | | | | | | | | | | Indication of use | The Common Type Insulin Pen Needle is intended for use<br>with pen injector device for subcutaneous injection of<br>insulin. | Verifine® Pen Needle is intended for use with pen injector<br>device for the subcutaneous injection of drugs. | Change in the trade name and<br>from specific use (insulin) to<br>general use (drugs). | No new concerns. The broader<br>indication for use does not<br>introduce new safety or<br>effectiveness risks as the<br>device's mechanism remains<br>unchanged. | | | | | | | | | | Model | IPN-29-12,<br>IPN-30-8,<br>IPN-31-4, IPN-31-5, IPN-31-6, IPN-31-8,<br>IPN-32-4, IPN-32-5, IPN-32-8,<br>IPN-33-4 | Model Specification code IPN Not defined. FPN Not defined. NPN Not defined. | | | | | | | | | New two models of FPN and NPN<br>were added.<br>Specification code was not<br>defined that the needle was<br>identified through by metric size. | No new concerns. The addition<br>of new models and sizes<br>expands options without<br>altering the device's safety or<br>effectiveness. | | Needle size | IPN<br>+<br>0.33mm×12mm,<br>0.30mm×8mm,<br>0.25mm×4mm, 0.25mm×5mm, 0.25mm×6mm, 0.25mm×<br>8mm,<br>0.23mm×4mm, 0.23mm×5mm, 0.23mm×8mm,<br>0.20mm× 4mm | IPN, FPN, NPN<br>+<br>0.33mm×12mm,<br>0.30mm×8mm,<br>0.25mm×4mm, 0.25mm×5mm, 0.25mm×6mm, 0.25mm×<br>8mm,<br>0.23mm×4mm, 0.23mm×5mm, 0.23mm×6mm, 0.23mm×<br>8mm,<br>0.20mm× 3.5mm, 0.20mm× 4mm, 0.20mm× 5mm,<br>0.18mm× 3.5mm, 0.18mm× 4mm, 0.18mm× 5mm | Some size was added. | No new concerns. The addition<br>of new needle sizes does not<br>impact safety or effectiveness;<br>the new sizes comply with<br>standard specifications. | | | | | | | | | | Item of description | Cleared device | Modified device | | | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | |----------------------------------|------------------------------------|---------------------------------|---------------|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Bevel | 3 | Bevel | Model | Needle size | Bevel of needle tip is defined.<br>Addition of 5 bevel design. | No new concerns. The<br>introduction of a 5-bevel<br>design is intended to enhance<br>user experience without<br>affecting safety or<br>effectiveness. | | | | 3 bevels | IPN, FPN, NPN | All sizes | | | | | | 5 bevels | IPN, FPN, NPN | All sizes | | | | Needle<br>Gauge | 29G, 30G, 31G, 32G, 33G | 29G, 30G, 31G, 32G, 33G, 34G | | | Outer diameter of 34G (0.18mm)<br>was added, per K210059. | No new concerns. The added<br>dimensions meet regulatory<br>standards, ensuring safety and<br>effectiveness. | | Dimension | Length<br>4mm, 5mm, 6mm, 8mm, 12mm | 3.5mm, 4mm, 5mm, 6mm, 8mm, 12mm | | | Nominal length of 3.5mm was<br>added. | | | Color | Image: Color chart | Not defined. | | | There is no color designated to<br>each size that metric size is<br>identified through the gauge and<br>length, and there is no color<br>requirement for pen needle in<br>ISO 11608-2. | No new concerns. The lack of a<br>defined color does not impact<br>the safety or effectiveness of<br>the device. | | Schematic diagram<br>and picture | Image: Schematic diagram of IPN | No change | | | There is only the difference about<br>shape of needle shield and hub<br>for each model. | No new concerns. The design<br>variations do not introduce<br>new safety or effectiveness<br>risks, as they are minor<br>adjustments to the device's<br>form. | {7}------------------------------------------------ | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | |---------------------|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-----------------------------------| | / | | FPN<br>Fig. 2—Schematic representation of FPN<br>Key<br>1 Needle container 4 Jointing medium<br>2 Needle shield 5 Needle hub<br>3 Needle tube | | | | / | | NPN<br>Fig. 3—Schematic representation of NPN<br>Key<br>1 Needle container 4 Jointing medium<br>2 Needle shield 5 Needle hub<br>3 Needle tube | | | | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | | | |---------------------|-----------------------------------|---------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------| | Product picture | / | Image: Side view<br>Image: Front view | There is only the difference about<br>shape of needle shield and hub<br>for each model. | No new concerns. The design<br>variations do not introduce<br>new safety or effectiveness<br>risks, as they are minor<br>adjustments to the device's<br>form. | | | | | / | | | | | | | Material | Image: IPN symbol IPN | Image: IPN symbol IPN | Material of needle shield is<br>changed to PE which is only<br>contact intact skin. | No new concerns. The material<br>change for the needle shield is<br>minor and does not affect<br>safety as it only contacts intact<br>skin. | | | | | Component | Material | | | Component | Material | | | Needle tube | Stainless steel (X5CrNi18-10) | | | Needle tube | Stainless steel (X5CrNi18-10) | | | Needle Shield | Polypropylene (PP) | | | Needle Shield | Polyethylene (PE) | | | Needle Container | Polypropylene (PP) | | | Needle Container | Polypropylene (PP) | | | Needle Hub | Polypropylene (PP) | | | Needle Hub | Polypropylene (PP) | | | Glue (Joint medium) | UV Glue | | | Glue (Joint medium) | UV Glue | | | Lubricant | Silicone oil | Lubricant | Silicone oil | | | | | Dialyzing paper<br>(Seal) | Medical grade glue dialyzing<br>paper | Dialyzing paper<br>(Seal) | Medical grade glue dialyzing<br>paper | | | | | | Image: FPN symbol FPN | | | Material of needle shield is<br>changed to PE which is only | | | | | | Component | Material | contact intact skin. | | | | | | Needle tube | Stainless steel (X5CrNi18-10) | | | {8}------------------------------------------------ | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | | | | | | | | | | | | | | | | | |-----------------------------|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|------------|--| | | / | Needle Shield<br>Needle Container<br>Needle Hub<br>Glue (Joint medium)<br>Lubricant<br>Dialyzing paper<br>(Seal) | | | | | | | | | | | | | | | | | | | | | | Needle Shield Polyethylene (PE)<br>Needle Container Polypropylene (PP)<br>Needle Hub Polypropylene (PP)<br>Glue (Joint medium) UV Glue<br>Lubricant Silicone oil<br>Dialyzing paper Medical grade glue dialyzing paper<br>(Seal) | | | | | | | | | | | | | | | | | | | | | | NPN<br>Component Material Needle tube Stainless steel (X5CrNi18-10) Needle Shield Polypropylene (PP) Needle Container Polypropylene (PP) Needle Hub Polypropylene (PP) Glue (Joint medium) UV Glue Lubricant Silicone oil Dialyzing paper (Seal) Medical grade glue dialyzing paper | | | | | | | | | | | | | | | | | No change. | | | Performance<br>requirements | Remain no change | No change | N/A | No new concerns.<br>Performance requirements<br>remain unchanged, ensuring<br>consistent functionality. | | | | | | | | | | | | | | | | | | Labeling | Unit<br>package | Remain no change | No change | N/A | | | | | | | | | | | | | | | | | | | User<br>package | Remain no change | Blow contents were changed:<br>1. Indication;<br>2. warning;<br>3. direction for use;<br>4. compatible pens; | These changes are meeting for<br>requirement of ISO 11608-<br>2:2022. | No new concerns. Labeling<br>updates meet current<br>standards and enhance the<br>device's safety and usability. | | | | | | | | | | | | | | | | | | Instruction<br>for use | Remain no change | Blow contents were changed:<br>1. Indication;<br>2. warning;<br>3. direction for use; | These changes are to be met for<br>requirement of ISO 11608-<br>2:2022. | | | | | | | | | | | | | | | | | | Item of description | Cleared device | Modified device | Comments on<br>Similarities/Differences | Safety/Effectiveness<br>Statement | | |-----------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------| | | / | Needle Shield<br>Needle Container<br>Needle Hub<br>Glue (Joint medium)<br>Lubricant<br>Dialyzing paper<br>(Seal) | | | | | | | Needle Shield Polyethylene (PE)<br>Needle Container Polypropylene (PP)<br>Needle Hub Polypropylene (PP)<br>Glue (Joint medium) UV Glue<br>Lubricant Silicone oil<br>Dialyzing paper (Seal) Medical grade glue dialyzing paper | | | | | | | NPN<br>Component Material<br>Needle tube Stainless steel (X5CrNi18-10)<br>Needle Shield Polypropylene (PP)<br>Needle Container Polypropylene (PP)<br>Needle Hub Polypropylene (PP)<br>Glue (Joint medium) UV Glue<br>Lubricant Silicone oil<br>Dialyzing paper (Seal) Medical grade glue dialyzing paper | No change. | | | | Performance<br>requirements | Remain no change | No change | N/A | No new concerns.<br>Performance requirements<br>remain unchanged, ensuring<br>consistent functionality. | | | Labeling | Unit<br>package | Remain no change | No change | N/A | No new concerns. Labeling<br>updates meet current | | | User<br>package | Remain no change | Blow contents were changed:<br>1. Indication;<br>2. warning;<br>3. direction for use;<br>4. compatible pens; | These changes are meeting for<br>requirement of ISO 11608-<br>2:2022. | standards and enhance the<br>device's safety and usability. | | | Instruction<br>for use | Remain no change | Blow contents were changed:<br>1. Indication;<br>2. warning;<br>3. direction for use; | These changes are to be met for<br>requirement of ISO 11608-<br>2:2022 | | The following changes were identified between the subject and predicate devices: - Trade Name: Changes in the subject device trade name from the predicate does not raise new or ● different questions of safety and effectiveness as the device's functions remains the same. {9}------------------------------------------------ - Indications for Use: The broadened indications for use in the subject device trade name from the predicate does not raise new or different questions of safety and effectiveness as the device's principle of operation remains the same. - Introduction of FPN and NPN Models and Needle Sizes (dimensions, bevel): The addition of new models and sizes expands options without altering the device's safety or effectiveness when compared to the predicate device. Additionally, the new needle sizes comply with standard specifications. Finally, unlike with the insulin needle in the predicate device, the subject device uses metric size instead of color, as there is no color requirement for pen needles in ISO 11608-2. - Material: The material of the subject device (IPN and FPN) needle shield has been changed to polyethylene (PE) from polypropylene (PP), which only contacts the skin. Biocompatibility testing was performed in accordance with ISO 10993 to support this change and does not raise new or different questions of safety and effectiveness. - Labeling: Subject device indications for use, warnings, directions for use, and compatible pens have been updated. This does not raise new or different questions of safety and effectiveness, as they are based on results/ review from the points identified above. That is, the device's functions remain the same. #### 9. Non-Clinical Performance Data To support substantial equivalence to the predicate device, Promisemed Hangzhou Meditech Co., Ltd. completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. #### Verification and Validation Activities - Design Changes: Needle Sizes: Additional sizes ranging from 0.33mm×12mm to 0.18mm×3.5mm were introduced to meet clinical needs, focusing on reducing outer diameter for increased patient comfort and ensuring safety by minimizing the risk of intramuscular injections. Compatible Pens: The pen needle compatibility was expanded to include additional pen models, such as ClikSTAR®, SoloSTAR®, and FlexTouch®, to address the evolving market and clinical requirements. Needle Shield Material: The needle shield material was changed from Polypropylene (PP) to Polyethylene (PE) to improve manufacturing efficiency. The new material was tested and found to be biocompatible and non-toxic as per ISO 10993-1. #### Testing Performed: Labeling and Dimensional Compliance: Visual inspections were conducted on unit packaging to ensure metric size labeling compliance with ISO 11608-2. The needle's dimensions, including the container, shield, and hub, were measured, and found to comply with drawing requirements. Performance Testing: The pen needles underwent performance testing (reports VR-20230818-02 and VR-20210528-03) to assess dose accuracy and needle removal torque. The testing confirmed that the needles meet the performance criteria outlined in ISO 11608-2. {10}------------------------------------------------ Biocompatibility Testing: The new Polyethylene (PE) material for the needle shield was tested for cytotoxicity, skin irritation, intracutaneous reactivity, and sensitization according to ISO 10993-1 standards. The results confirmed that the material is non-toxic and safe for use in its intended application. #### Test Results: All verification and validation tests passed without deviations, confirming that the Verifine® Pen Needles meet the necessary design specifications and regulatory requirements. The tests demonstrated that the product modifications did not introduce any new risks related to safety or effectiveness when compared to the predicate device. #### Conclusions: Based on the results of the non-clinical verification and validation activities, it can be concluded that the modified Verifine® Pen Needles are substantially equivalent to the predicate device K161950. The changes, including the addition of new needle sizes, compatible pens, and the material change of the needle shield, have been thoroughly evaluated and verified to meet all applicable performance and safety standards. Therefore, the device is as safe and effective as the predicate device for its intended use, and no new risks have been introduced. #### 10. Clinical Performance Data No clinical data was necessary to determine the substantial equivalence of this device. ### 11. Statement of Substantial Equivalence The Verifine® Pen Needles have the same indications for use as the predicate devices listed above. Any minor differences in the technological characteristics of the subject device when compared to the predicate devices have been successfully evaluated through appropriate safety and performance testing which demonstrates that the subject device, when compared to the predicate device, does not raise any new questions of safety and effectiveness. Therefore, the Verifine® Pen Needles have been determined to be substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...