Common Type Insulin Pen Needle, Safety Type Insulin Pen Needle
K161950 · Promisemed Hangzhou Meditech Co., Ltd. · FMI · May 11, 2017 · General Hospital
Device Facts
Record ID
K161950
Device Name
Common Type Insulin Pen Needle, Safety Type Insulin Pen Needle
Applicant
Promisemed Hangzhou Meditech Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
May 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.
Device Story
Sterile, single-use hypodermic needles for subcutaneous insulin delivery via pen injectors. Common Type: needle hub, tube, shield. Safety Type: includes spring-loaded retractable shield that automatically covers and locks needle after withdrawal to prevent needlestick injuries. Operated manually by patient or clinician. Input: insulin pen injector; Output: subcutaneous insulin delivery. Safety mechanism provides passive protection post-injection. Benefits: ease of insulin administration and reduced risk of sharps injury.
Clinical Evidence
Bench testing only. Performance verified against ISO 7864, ISO 9626, ISO 11608-2, and ISO 23908. Biocompatibility testing (ISO 10993) confirmed no cytotoxicity, irritation, sensitization, systemic toxicity, or hemolysis. Simulated clinical use testing of 600 Safety Type devices showed zero needlestick injuries and zero protective feature failures.
Technological Characteristics
Materials: 304 Stainless Steel (tube), Polypropylene (hub), ABS (safety shield). Dimensions: 29G-33G, 4mm-12mm lengths. Energy: Manual. Connectivity: None. Sterilization: Gamma irradiation (Safety) or EO (Common). Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 23908, ISO 10993.
Indications for Use
Indicated for patients requiring subcutaneous insulin injection via pen injector devices. Safety Type includes integrated shield to prevent accidental needlestick injuries.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 11, 2017
Promisemed Hangzhou Meditech Co., Ltd. % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd No. 55 Shizhou Middle Road, Nanshan District Shenzhen, GD755 Guangdong CHINA
Re: K161950
Trade/Device Name: Verifine® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: April 5, 2017 Received: April 10, 2017
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". The signature is a complex, looping design to the left of the name. Below the name is the character sequence "-s", which may be part of the signature or a separate element.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161950
Device Name
Verifine® Common Type Insulin Pen Needle and Verifine® Safety Type Insulin Pen Needle
#### Indications for Use (Describe)
The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.
The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Production Use (Part 21 CFR 221.Subpart D) | <span></span> |
|-------------------------------------------------------------|---------------|
| <span></span> Test Strip Control Use (21 CFR 221.Subpart C) | <span></span> |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Contact Details 1.
1.1 Applicant information
| Applicant Name | Promisemed Hangzhou Meditech Co.,Ltd. |
|-------------------------|-----------------------------------------------------------------------------|
| Address | No.18 Haishu Road Future Sci-Tech Park, Yuhang District, Hangzhou,<br>China |
| Phone No. | +86-571-86775619 |
| Fax No. | +86-571-88772985 |
| Contact person | Keshin Hu |
| Contact person's e-mail | keshin@promisemed.ca |
| Company e-mail | admin@promisemed.ca |
| Date Prepared | April 05, 2017 |
| Website | http://www.promisemed.ca/ |
### 1.2 Consultant information
-
| Name | Shenzhen Joyantech Consulting Co., Ltd<br>Image: [logo]<br>Image: [Chinese characters] |
|-------------------------|-----------------------------------------------------------------------------------------------------------------|
| Address | Room 1122, International Mayors Communication Centre, NO. 55<br>Shizhou middle road, Nanshan District, Shenzhen |
| Phone No. | +86-755-86069197 |
| Contact person | Elly Xu; Field Fu |
| Contact person's e-mail | elly@cefda.com; info@cefda.com |
| Website | http://www.cefda.com |
#### Device information 2.
| Trade name | Verifine® Common Type Insulin Pen Needle; Verifine® Safety Type<br>Insulin Pen Needle |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | Insulin Pen Needle |
| Model | Common Type:IPN-29-12;IPN-30-8;IPN-31-4;IPN-31-5;IPN-31-6;<br>IPN-31-8;IPN-32-4;IPN-32-5; IPN-32-8; IPN-33-4<br>Safety Type:SPN-29-6;SPN-29-8;SPN-30-4;SPN-30-5;SPN-30-6;<br>SPN-30-8;SPN-31-4;SPN-31-5;SPN-31-6;SPN-31-8 |
| Classification | II |
| Classification name | Needle, Hypodermic, Single Lumen |
| Product code | FMI |
| Regulation No. | 880.5570 |
#### 3. Legally Marketed Predicate Device
{4}------------------------------------------------
| Trade Name | Insulin Pen Needle |
|---------------|------------------------------------------------------------------------------------------------|
| 510(k) Number | K133059 |
| Product Code | FMI |
| Manufacturer | Wenzhou Beipo Science & Technology Co., Ltd |
| Trade Name | Safe Block Pen Needle |
| 510(k) Number | K151311 |
| Product Code | FMI |
| Manufacturer | Artsana S.p.A Unit Production N°30, Via Saldarini Catelli 6/10, 22070<br>Grandate COMO - ITALY |
#### Device Description 4.
The proposed product, insulin pen needle, is manufactured by Promisemed Hangzhou Meditech Co., Ltd, which is designed for use with a pen injector for the subcutaneous injection of insulin. The products have two types, common type and safety type. Both of the two types are sterile with a Sterility Assurance Level (SAL) of 104, non-pyrogenic and single-use devices. Each type has several models. Different models are distinguished by needle gauge and length.
The Common Type Insulin Pen Needle consists of needle shield, needle shield, needle tube, needle hub, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.
The Safety Type Insulin Pen Needle consists of needle shield, spring, housing, needle hub, needle container, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.
The user proceeds with inserting the needle into the skin manually. Safety Type Insulin Pen Needle is designed to reduce occurrence of accidental needle sticks from patient end of the needle by providing a shield that covers and locks the needle after use. As the user proceeds with inserting the needle into the skin the shield will retract. After the injection is completed and needle is removed from the skin, the shield will automatically extend to cover the needle and lock in place. Once the Safety Type Insulin Pen Needle is in the locked mode, it can no longer be used.
Both the Common Type Insulin Pen Needle and the Safety Type Pen Needle are for OTC use, and they are external communication, blood indirect devices. The contact duration for both subjected devices is within 24h, and they belong to limited contact device according ISO 10993-1.
The dimension for both Common Type Insulin Pen Needle and the Safety Type Pen Needle are shown as below.
| Type | Model | Needle OD | Needle available<br>length |
|----------------|-----------|-----------|----------------------------|
| Common<br>Type | IPN-29-12 | 0.33mm | 12mm |
| | IPN-30-8 | 0.30mm | 8mm |
| | IPN-31-4 | 0.25mm | 4mm |
| | IPN-31-5 | 0.25mm | 5mm |
| | IPN-31-6 | 0.25mm | 6mm |
| | IPN-31-8 | 0.25mm | 8mm |
| | IPN-32-4 | 0.23mm | 4mm |
{5}------------------------------------------------
-
| | IPN-32-5 | 0.23mm | 5mm |
|----------------|----------|--------|-----|
| | IPN-32-8 | 0.23mm | 8mm |
| | IPN-33-4 | 0.20mm | 4mm |
| Safety<br>Type | SPN-29-6 | 0.33mm | 6mm |
| | SPN-29-8 | 0.33mm | 8mm |
| | SPN-30-4 | 0.30mm | 4mm |
| | SPN-30-5 | 0.30mm | 5mm |
| | SPN-30-6 | 0.30mm | 6mm |
| | SPN-30-8 | 0.30mm | 8mm |
| | SPN-31-4 | 0.25mm | 4mm |
| | SPN-31-5 | 0.25mm | 5mm |
| | SPN-31-6 | 0.25mm | 6mm |
| | SPN-31-8 | 0.25mm | 8mm |
#### 5. Intended Use/Indication for Use
The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.
The Safety Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.
Additionally, after withdrawal of the Safety Type Insulin Pen Needle from the body, the attached needle safety shield automatically covers the needle to minimize the risk of accidental needlestick.
#### 6. Substantial Equivalence Comparison
### 6.1 Comparison between Common Type Insulin Pen Needle and Predicate Device(K133059)
| Item | Proposed Device:<br>Common Type Insulin Pen<br>Needle | | Predicate Device:<br>Insulin Pen Needle<br>(K133059) | | Comments |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------|
| Product Code | FMI | | FMI | | Same |
| Intended Use | The Common Type Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. | | The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. | | Same |
| Operating<br>Principle | The user proceeds with inserting the needle into the skin manually. The patient end and the cartridge end of the tube are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration. | | The user proceeds with inserting the needle into the skin manually. The patient end and the cartridge end of the tube are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration. | | Same |
{6}------------------------------------------------
### SEC 005:001 510K Summary
Version:A/0
| Configuration | Needle Tube, Needle shield (Tube Sheath), Needle hub (Hub), Needle container (Hub Sheath), Sealing dialysis paper (Sealed Paper). | Needle Tube, Hub, Tube Sheath, Hub Sheath and Sealed Paper. | Same |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Length | 4mm, 5mm, 6mm, 8mm, 12mm | 4mm, 5mm, 6mm, 8mm, 10mm, 12mm | Similar (* 1) |
| Needle Gauge | 29G/30G/31G/32G/33G | 29G/30G/31G/32G | Same |
| Tip configuration | Per ISO 11608-2, section 4.5, visually sharp at 2.5X magnification, designed to minimize coring and fragmentation | Per ISO 11608-2, section 4.5, visually sharp at 2.5X magnification, designed to minimize coring and fragmentation | Same |
| Material | Tube: 304 Stainless Steel<br>Needle hub: Polypropylene | Tube: 304 Stainless Steel<br>Hub: Polypropylene | Same |
| Performance | Complied with ISO 7864, ISO 9626, and ISO 11608-2 | Complied with ISO 7864, ISO 9626, and ISO 11608-2 | Same |
| Sterilization | SAL: 10-6<br>Method: EO Sterilized | SAL: 10-6<br>Method: EO Sterilized | Same |
| Shelf Life | 5 years | 5 years | Same |
| Single Use | Yes | Yes | Same |
| Biocompatibility | Complied with ISO10993 series standards, and the following tests are performed,<br>- Cytotoxicity: No cytotoxicity<br>- Skin Irritation: No evidence of skin irritation<br>- Skin Sensitization: No evidence of sensitization<br>- Acute Systemic Toxicity: No systemic toxicity<br>- Hemolysis: No evidence of hemolysis<br>- Pyrogen: No pyrogen | Complied with ISO10993 series standards, and the following tests are performed,<br>- Cytotoxicity: No cytotoxicity<br>- Skin Irritation: No evidence of skin irritation<br>- Skin Sensitization: No evidence of sensitization<br>- Acute Systemic Toxicity: No systemic toxicity<br>- Hemolysis: No evidence of hemolysis<br>- Pyrogen: No pyrogen | Same |
*1: The needle tube lengths of the proposed device are covered by the predicate device. Compare to the predicate device, the proposed device has an additional 33G model insulin pen needle. 33G model for Insulin Pen Needle is a widely used in the market and had been approved in FDA (Such as K152410). On the other hand, the bench tests of Needle with 33G demonstrated conformances to ISO 7864, ISO 9626, and ISO 11608-2.
6.2 Comparison between Safety Type Insulin Pen Needle and Predicate Device(K151311)
{7}------------------------------------------------
# SEC_005:001_510K Summary
| Item | Proposed Device:<br>Safety Type Insulin Pen<br>Needle | Predicate Device:<br>Insulin Pen Needle (K151311) | Comments |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Product Code | FMI | FMI | Same |
| Intended Use | The Safety Type Insulin Pen<br>Needle is intended for use with<br>pen injector device for<br>subcutaneous injection of<br>insulin.<br>Additionally, after withdrawal of<br>the Safety Type Insulin Pen<br>Needle from the body, the<br>attached needle safety shield<br>automatically covers the<br>needle to minimize the risk of<br>accidental needlestick. | It is intended for use with pen<br>injector devices for the<br>injection of fluids, including<br>insulin.<br>Additionally the attached safety<br>shield automatically locks in<br>place and reduces the<br>occurrence of accidental sticks<br>from the patient end of the<br>needle. The shield also serves<br>to conceal the needle before<br>and after injection. | Similar(* 2) |
| Operating<br>Principle | As the user proceeds with<br>inserting the needle into the<br>skin the shield will retract. After<br>the injection is completed and<br>needle is removed from the<br>skin, the shield will<br>automatically extend to cover<br>the needle and lock in place.<br>Once the Safety Type Insulin<br>Pen Needle is in the locked<br>mode, it can no longer be<br>used. | As the user proceeds with<br>inserting the needle into the<br>skin the shield will retract. After<br>the injection is completed and<br>needle is removed from the<br>skin, the shield will<br>automatically extend to cover<br>the needle and lock in place.<br>Once the Safety Type Insulin<br>Pen Needle is in the locked<br>mode, it can no longer be<br>used. | Same |
| Length | 4mm, 5mm, 6mm,<br>8mm(tolerance as per ISO<br>11608-2) | 5mm, 8mm(tolerance as per<br>ISO 11608-2) | Similar (* 3) |
| Gauge<br>Sharps Injury<br>Prevention<br>Features | G29, G30, G31<br>Safety shield | G31<br>Safety shield | Same |
| Configuration<br>and Material | Tube: Stainless Steel<br>Needle Hub: Polypropylene<br>Spring: Stainless Steel<br>Needle container(Primary<br>Container):PP<br>Needle/Safety shield: ABS | Cannula: Stainless Steel<br>Needle Hub: Polypropylene<br>Spring: Stainless Steel<br>Primary Container:HDPE<br>Shield:HDPE | Same<br>Same<br>Same<br>Different (* 4) |
| | | | |
| Performance | Complied with ISO 7864, ISO 9626, ISO 11608-2 and ISO 23908 | Complied with ISO 7864, ISO 9626, ISO 11608-2 and ISO 23908 | Same |
| Sterilization | SAL: 10-6<br>Method: Irradiation Sterilized | SAL: 10-6<br>Method: EO Sterilized | Different (* 5) |
| Shelf Life | 5 years | 5 years | Same |
| Single Use | Yes | Yes | Same |
| Biocompatibility | Complied with ISO10993 series standards, and the following tests are performed,<br>• Cytotoxicity: No cytotoxicity<br>• Skin Irritation: No evidence of skin irritation<br>• Skin Sensitization: No evidence of sensitization<br>• Acute Systemic Toxicity: No systemic toxicity<br>• Hemolysis: No evidence of hemolysis<br>Pyrogen: No pyrogen | Complied with ISO10993 series standards, and the following tests are performed,<br>• Cytotoxicity: No cytotoxicity<br>• Skin Irritation: No evidence of skin irritation<br>• Skin Sensitization: No evidence of sensitization<br>• Acute Systemic Toxicity: No systemic toxicity<br>• Hemolysis: No evidence of hemolysis<br>Pyrogen: No pyrogen | Same |
{8}------------------------------------------------
### Version:A/0
* 2: The proposed device is intended to use with pen injector device for subcutaneous injection of insulin. Compare to proposed device, the predicate device is used for the injection of fluids, including insulin. The usage range of proposed device is included in the predicate device.
* 3: The length and the gauge is the specification of needle tube that has no relationship with the safety prevention features. The Insulin Pen Needle (K151311) has the tube diameter of 29G and 30G, it also has the tube length of 4mm and 6mm. These models of insulin pen needle are widely used in the market. Moreover, the bench tests of Needles with 29G, 30G and 4mm, 6mm demonstrated conformances to ISO 7864. ISO 9626. ISO 11608-2 and ISO 23908. Therefore the differences do not raise new concerns to establish substantial equivalence to the predicate.
* 4: Although the materials of safety shield and needle container between the proposed device and the predicate device are different, the ABS and PP are two kinds of materials that are widely used on medical device. Moreover, the biocompatibility test report of the proposed device demonstrate that subject device is biocompatible and the performance as intended.
* 5: Although the sterilization methods between the proposed device and the predicate device are different, but the radiation sterilization effect of the proposed device is proved by the Microbiological validation report of gamma radiation sterilization dose audit test report in VOL 014. The validation report showed that the sterilization effect of the proposed device can achieve a Sterility Assurance Level (SAL) of 10°, and the radiation dose audit test reports showed that the minimum gamma radiation dose to achieve a 10° SAL was acceptable according to ISO 11737-2:2012.
{9}------------------------------------------------
#### Non-clinical Testing 7.
All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing are demonstrating the standards and matching the performance of new devices to the predicate devices.
The following performance data were provided in support of the substantial equivalence determination.
| Test | Requirements | Results | |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Materials | The needle shall be made of tubing materials specified in ISO 9626. | Passed | |
| Dimensions | The needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2. | Passed | |
| Flow rate through the needle | The needles were tested in accordance with Annex A to ISO 11608-2. | Passed | |
| Bond between hub and needle tube | The union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2. | Passed | |
| Needle points | The needle points shall fulfill the 4.5 of ISO 11608-2. | Passed | |
| Freedom from defects | The needle points shall fulfil the requirements of ISO 7864:1993, 11.3. | Passed | |
| Lubrication | The needle tube should be lubricated at both patient end and the cartridge end. The lubrication shall fulfil the 4.7 of ISO 11608-2. | Passed | |
| Dislocation of measuring point patient end | Dislocation of the cannula point at the patient end shall be in accordance with Table 2 (ISO 11688-2) when tested in accordance with Clause 8 of ISO 11608-2. | Passed | |
| Functional compatibility with needle-based injected systems | Compatibility with any NIS shall be claimed only after testing in accordance with Clause 11 of ISO 11608-2. | Passed | |
| Easy of assemble and disassembly | Attachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11 of ISO 11608-2. | Passed | |
| Biocompatibility | Cytotoxicity | ISO 10093-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity | Passed |
| | Sensitization | ISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation | Passed |
| | Intracutaneous<br>Reactivity | ISO 10993-10 Biological evaluation of<br>medical devices -Part 10:Tests for irritation<br>and skin sensitization | Passed |
| | | and skin sensitization | |
| | Haemocompatibility | ISO 10993-4 Biological evaluation of medical<br>devices Part 4: Selection of tests for<br>interactions with blood<br>ASTM F765 Standard Practice for<br>Assessment of Hemolytic Properties of<br>Materials | Passed |
| | | | |
| | | | |
| | | | |
| | | | |
| | System toxicity<br>(acute) | ISO 10993-11:2006/(R)2010, Biological<br>evaluation of medical devices - Part 11: Tests<br>for systemic toxicity | Passed |
| | | | |
| | | | |
{10}------------------------------------------------
### Simulated Clinical Use Testing
As for Safety Type Insulin Pen Needle, a simulated clinical use testing was conducted to verify the safety feature per FDA guidance document of Medical Devices with Sharps Injury Prevention Features. Based on the guidance, 600 subject devices were chosen in the simulated clinical use testing. There was no needlestick injuries occurred and zero failure of the protective feature in the testing.
#### 8. Clinical Testing
Substantial equivalence does not depend on the clinical test data.
#### Conclusions 9.
Based on device comparison information and non-clinical bench testing, the proposed device is substantially equivalent to legally marketed predicate devices (K133059, K151311).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.