K242619 · Apira Science, Inc. · OAP · Dec 2, 2024 · Physical Medicine
Device Facts
Record ID
K242619
Device Name
GroWell BT Hair Growth System (GW10042)
Applicant
Apira Science, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Dec 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MEN - The GroWell BT Hair Growth System ("GroWell") is designed to promote hair growth in males with androgenetic alopecia who have Norwood Hamilton Classifications of Ila to V. WOMEN - The GroWell is designed to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin classifications of I-II. The GroWell is approved for Fitzpatrick skin types I-IV.
Device Story
GroWell BT Hair Growth System is an OTC wearable device for home use; consists of 24 red light diode lasers and 39 red light super-luminescent diodes (655nm) mounted in a flexible panel inside a baseball-style cap. Device provides full scalp coverage; powered by internal lithium battery. Operation: user wears cap; device automatically activates light array; includes safety feature to pause/turn off if removed during use. Session duration is 25 minutes; device emits two audible beeps upon completion. Therapeutic effect achieved via direct irradiation of scalp. Benefits include promotion of hair growth in specified androgenetic alopecia populations.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), photobiological safety (IEC 62471), laser safety (IEC 60825-1), and biocompatibility (ISO 10993 series).
Technological Characteristics
Light source: 24 laser diodes and 39 LEDs (655nm). Form factor: baseball-style cap with flexible internal light panel. Power: lithium battery. Biocompatibility: tested per ISO 10993-1, -5, -10, -23. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, IEC 60825-1.
Indications for Use
Indicated for males with androgenetic alopecia (Norwood Hamilton IIa-V) and females with androgenetic alopecia (Ludwig-Savin I-II) with Fitzpatrick skin types I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2024
Apira Science Inc % Tracy Che Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Oianhai Road Shenzhen, Guangdong 518052 China
Re: K242619
Trade/Device Name: GroWell BT Hair Growth System (GW10042) Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: August 30, 2024 Received: September 3, 2024
Dear Tracy Che:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L Digitally signed by
TANISHA L. TANISHA L. HITHE -S HITHE -S ---------------------------------------------------------------------------------------------------------------------------------------------------------------------15:18:45 -05'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Submission Number (if known)
K242619
Device Name
GroWell BT Hair Growth System (GW10042)
Indications for Use (Describe)
MEN - The GroWell BT Hair Growth System ("GroWell") is designed to promote hair growth in males with androgenetic alopecia who have Norwood Hamilton Classifications of Ila to V. WOMEN - The GroWell is designed to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin classifications of I-II.
The GroWell is approved for Fitzpatrick skin types I-IV.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
| 510(k) #: | | | 510(k) Summary | | Prepared on: 2024-12-02 |
|------------------------------------|----------------------------------------------------------|--|------------------------------------------------------------------------------------------------------------|--|-------------------------|
| Contact Details | | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | | | Apira Science Inc | | |
| Applicant Address | | | 409 Edmon Rd. NE Suite C Albuquerque New Mexico 87107 United<br>States | | |
| Applicant Contact Telephone | | | 1-360-4804564 | | |
| Applicant Contact | | | Mr. Ray Pizinger | | |
| Applicant Contact Email | | | rpizinger@gmail.com | | |
| Correspondent Name | | | FEIYING DRUG & MEDICAL CONSULTING TECHNICAL SERVICE<br>GROUP | | |
| Correspondent Address | | | Rm 2401 Zhenye International Business Center, No. 3101-90,<br>Qianhai Road Shenzhen Guangdong 518052 China | | |
| Correspondent Contact Telephone | | | +86 18027206248 | | |
| Correspondent Contact | | | Ms. Tracy Che | | |
| Correspondent Contact Email | | | 1273487536@qq.com | | |
| Device Name | | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | | GroWell BT Hair Growth System (GW10042) | | |
| Common Name | | | Laser, Comb, Hair | | |
| Classification Name | | | Infrared lamp | | |
| Regulation Number | | | 890.5500 | | |
| Product Code(s) | | | OAP | | |
| Legally Marketed Predicate Devices | | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | | Product Code |
| K140931 | iGrow-II Hair Growth System | | | | OAP |
| K141567 | iGrow-II Hair Growth System | | | | OAP |
| K173729 | Revian Red | | | | OAP |
| Device Description Summary | | | | | 21 CFR 807.92(a)(4) |
The GroWell BT Hair Growth System consists of 24 proprietary red light diode lasers and 39 proprietary red light super-luminescent diodes configured in a flexible panel which is inserted in an adult, baseball-style cap. The GroWell contains 655nm laser and LED diodes, which generates light after being excited by the internal power supply, the light is controlled and output by the control device, and directly and effectively irradiated to the treatment site, so as to achieve therapeutic effect. The GroWell's conbination
{5}------------------------------------------------
of diode lasers and non-laser LED's provides for a full coverage of the head: i.e, the scalp area roughly from the top of the head to the top of the ears. The GroWell session will be "paused" and the light array will automatically turn off if the user removes the GroWell during use, and will resume when it is replaced on the 25 minute session, the lights will turn off and the GroWell will emit two audible beeps to signal that the therapy is complete.
### Intended Use/Indications for Use
MEN - The GroWell BT Hair Growth System ("GroWell") is designed to promote hair growth in males with androgenetic alopecia who have Norwood Hamilton Classifications of Ila to V.
WOMEN - The GroWell is designed to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin classifications of I-II.
The GroWell is approved for Fitzpatrick skin types I-IV.
## Indications for Use Comparison
The subject device and predicate devices have the same indications for use, both are mainly used to prowth.
## Technological Comparison
The subject device and predicate devices have similar technological characteristics:
1) They use the same laser diodes and LEDs as the light source;
2) They all adopt the design of a cap/helmet which provides for a full coverage of the upper 1/3 of the head;
3) They have the same key performance parameters such as the wavelength and energy per diode.
The main differences between the subject devices are power supply, dimensions and weight, and number of diodes.
1) Power supply: The power supply of the subject device is lithium battery, while the power supply of the predicate devices (K140931 and K141567) are main power. However the lithium battery of the subject device has passed IEC62133 test and the power adapter has been assessed for electrical safety test along with the main unit, so this difference does not raise any safety/effectiveness questions.
2) Number of diodes: The subject device has 24 lasers and 39 LEDs while the predicate devices (K140931 and K141567) has 21 lasers and 30 LEDs. It basically means the subject device can cover more area than the predicate devices.
3) Dimensions and weight: The dimensions of the subject device are similar to the predicate devices. The weight of K140931 and K141567, both of which use a helmer than K173729, which uses a textile cap to mount the light panel. The weight is in between the weights of predicate devices. The difference in weight does not raise safety or effectiveness issues, because the core function, light output, is not affected by device dimension or weight.
Based on the above discussion, these differences do not raise safety and the subject device has passed corresponding international standards tests.
## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical testing have been conducted to verify that the GroWell meets all design specifications which supports the conclusion that it's substantially equivalent to the predicate demonstrate that the subject device complies with the following standards:
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests;
IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment;
IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems;
IEC 60825-1: 2014 Safety of laser products - Part 1: Equipment classification, and requirements
The device has been evaluated for biocompatibility as per ISO 10993-1 Fifth edition 2018-08 and tested as per ISO 10993-5 Third edition 2009-06-01, ISO 10993-10 Fourth edition 2021-11 and ISO 10993-23 First edition 2021-01.
The device software has been evaluated as per FDA guidance " Content of Premarket Submissions for Device Software Functions" . In order to verify and assure the performance, function and quality of the producted performance verification. The clinical test is not applicable, there's no clinical data.
Based on the above analysis and tests, the subject device is substantially equivalent to the predicate devices.
# 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.