K242608 · Varian Medical Systems, Inc. · KRD · Jan 17, 2025 · Cardiovascular
Device Facts
Record ID
K242608
Device Name
Embozene Color-Advanced Microspheres
Applicant
Varian Medical Systems, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Jan 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Embozene Color-Advanced Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH). The device is not intended for neurovascular use.
Device Story
Embozene Color-Advanced Microspheres are spherical, non-resorbable hydrogel microspheres coated with a perfluorinated polymer (Polyzene-F). Supplied in pre-filled syringes, they are delivered via microcatheter into the bloodstream to permanently occlude vessels. Used by physicians to stop bleeding in non-healing lesions or to induce necrosis/shrinkage in hypervascular tumors. The device is a sterile, single-use implant. Clinical benefit includes targeted vessel occlusion for therapeutic management of specified conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included mechanical performance, biocompatibility (ISO 10993-1), sterilization (ISO 17665-1), and packaging integrity (ISO 11607-1/2) to demonstrate equivalence to the predicate.
Technological Characteristics
Spherical, non-resorbable hydrogel microspheres coated with inorganic perfluorinated polymer (Polyzene-F). Sizes: 40-1300 µm. Supplied in 20 ml pre-filled syringes. Steam sterilized per ISO 17665-1. Biocompatible per ISO 10993-1. Packaging per ISO 11607-1/2.
Indications for Use
Indicated for embolization of arteriovenous malformations, hypervascular tumors (including uterine fibroids and hepatoma), and prostatic arteries for symptomatic benign prostatic hyperplasia (BPH). Not for neurovascular use.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2025
Varian Medical Systems, Inc. Lynn Allman Senior Director, Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304
Re: K242608
Trade/Device Name: Embozene Color-Advanced Microspheres Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: II Product Code: KRD, NOY, NAJ Dated: December 19, 2024 Received: December 20, 2024
Dear Lynn Allman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242608
Device Name
# Embozene Color-Advanced Microspheres
Indications for Use (Describe)
Embozene Color-Advanced Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH). The device is not intended for neurovascular use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Page 1 of 3
Image /page/4/Picture/2 description: The image shows three blue circles in a row. The circles are evenly spaced and appear to be the same size. The background is white, which makes the blue circles stand out. The image is simple and clean, with no other objects or details present.
Image /page/4/Picture/3 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller font. The logo is black and white.
Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304-1038
Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com
# PREMARKET NOTIFICATION
510(k) Summary Embozene Color-Advanced Microspheres The following information is provided as required by 21 CFR 807.92
#### l. Submitter's Information:
| Name and Address: | Varian Medical Systems Inc.<br>3100 Hansen Way<br>Palo Alto, CA 94304-1038 |
|-------------------|----------------------------------------------------------------------------------|
| Manufacturer: | Varian Medical Systems, Inc.<br>9825 Spectrum Drive, Building 2 Austin, TX 78717 |
| Contact Name: | Lynn Allman, Senior Director Regulatory Affairs |
| E-mail: | submissions.support@varian.com |
| Date Prepared: | August 31, 2024 |
#### Device Information: ll.
| Proprietary Name: | Embozene Color-Advanced Microspheres |
|-----------------------|----------------------------------------------------------------------------------------------------------------|
| Common/ Usual Name: | Microspheres for embolization |
| Classification Name: | Device, Vascular, for Promoting Embolization<br>Agents, Embolic, for Treatment of Benign Prostatic Hyperplasia |
| Classification Panel: | Cardiovascular<br>Gastroenterology/Urology Gastroenterology/Urology |
| Regulation Number: | 21 CFR Part 870.3300 (KRD, NAJ)<br>21 CFR Part 876.5550 (NOY) |
| Product Codes: | KRD, NOY, NAJ |
#### lll. Predicate Device:
Embozene Color-Advanced Microspheres (K180102)
#### IV. Subject Device Description:
Embozene Color-Advanced Microspheres (hereafter may be referenced as Embozene Microspheres or Embozene) are spherical, tightly calibrated, biocompatible, non-resorbable, hydrogel microspheres coated with an inorganic perfluorinated polymer shell (Polyzene-F). The microspheres are suspended in liquid and then injected into the bloodstream to permanently occlude blood vessels. They are used to stop bleeding when the underlying lesion is not likely to heal or block arteries supplying a tumor to cause tumor necrosis and/or shrinkage.
{6}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo is simple and modern, and it conveys the message that Varian is a part of the Siemens Healthineers family.
Embozene Color-Advanced Microspheres are sterile, single use devices and are supplied in pre-filled 20 ml syringes containing 2ml of microspheres in approximately 7 ml of transport solution. The Embozene Microspheres are available in 40-1300 µm sizes.
#### V. Intended Use/ Indications of Use Statement:
## Embozene Color-Advanced Microspheres
Embozene Color-Advanced Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH). The device is not intended for neurovascular use.
#### VI. Substantial Equivalence Discussion:
Embozene Color-Advanced Microspheres are substantially equivalent to the legally marketed Embozene Color-Advanced Microspheres, cleared by FDA (K180102). Both devices have the same indications for use and the same technological characteristics. Both devices are spherical, non-resorbable polymer microspheres which are delivered through a microcatheter to embolize a target vessel. While there are no technological differences, the subject device incorporates an updated syringe due to end of life of the previously used syringe. The potential impacts of this change, including mechanical performance, biocompatibility, sterility, packaging and shelf-life, have been evaluated and demonstrated that the subject device meets all device specifications of the predicate device.
#### VII. Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
# 1. Non-clinical Testing:
Embozene underwent non-clinical testing to demonstrate the design and performance met the established design criteria and is substantial equivalent to the predicate device. The subject device successfully completed mechanical testing to ensure the device with the new syringe met all related specifications.
# 2. Sterilization:
Embozene Microspheres are steam sterilized to ensure a minimum sterility assurance level of 10 in accordance with ISO 17665-1, ISO 11138-1, ISO 11737-1 and ISO 11737-2.
#### Biocompatibility: 3.
Biocompatibility testing of Embozene Microspheres in accordance with ISO 10993-1 demonstrated that the patient-contacting components are biocompatible.
{7}------------------------------------------------
Page 3 of 3
Image /page/6/Picture/2 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font.
#### Packaging Integrity and Shelf Life: 4.
Packaging integrity and shelf-life testing for Embozene Microspheres was performed in accordance with ISO 11607-1 and ISO 11607-2.
# VIII. Conclusion:
The performance data for Embozene Color-Advanced Microspheres support the safety of the devices and the verification and validation demonstrate that the subject device should perform as intended in the specified use conditions. Varian considers Embozene Color-Advanced Microspheres to be as safe, effective as the predicate device and therefore, substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.