K242272 · Blue Belt Technologies, Inc. · Nov 15, 2024 · OR
Device Facts
Record ID
K242272
Device Name
CORIOGRAPH Pre-Op Planning and Modeling Services
Applicant
Blue Belt Technologies, Inc.
Decision Date
Nov 15, 2024
Decision
SESE
Submission Type
Traditional
Attributes
Software as a Medical Device
Indications for Use
CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide preoperative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for generating the plan are identifiable on patient imaging scans.
Device Story
CORIOGRAPH Pre-Op Planning and Modeling Services provide preoperative surgical planning for orthopedic procedures. Input data includes patient imaging (CT, MRI, X-ray) and surgeon preferences. Smith & Nephew personnel generate patient-specific bone models and preoperative plans, which are then viewable and editable by surgeons via the CORIOGRAPH Modeler software. The device is used in professional healthcare facilities. The output—a preoperative surgical plan and implant position simulation—assists surgeons in clinical decision-making regarding component selection and positioning. Benefits include patient-matched planning to support surgical accuracy.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed software safety and effectiveness. Summative usability testing in a simulated environment demonstrated that surgeons could use the device safely and effectively. Credibility evaluation confirmed the clinical relevance of kinematic models and Activities of Daily Living (ADLs) used in the device.
Technological Characteristics
Software as a Medical Device (SaMD) for orthopedic surgical planning. Utilizes patient imaging (CT, MRI, X-ray) to generate 3D bone models and implant simulations. Compatible with specific Smith & Nephew hip and knee implant systems. Operates in professional healthcare facilities. Connectivity involves viewing/editing via the CORIOGRAPH Modeler software.
Indications for Use
Indicated for use in unicondylar knee replacement (UKR), total knee arthroplasty (TKA), and primary total hip arthroplasty (THA). Target population includes patients able to undergo CT/MRI safely, fitting indications, and without contraindications for the specific hip/knee implant system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
November 15, 2024
Blue Belt Technologies, Inc. Chandler Caufield Senior Regulatory Affairs Specialist 2875 Railroad Street Pittsburgh, Pennsylvania 15222
Re: K242272
Trade/Device Name: CORIOGRAPH Pre-Op Planning and Modeling Services Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PBF Dated: July 31, 2024 Received: August 1, 2024
### Dear Chandler Caufield:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Tejen D. Soni -S
For
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K242272
Device Name
CORIOGRAPH Pre-Op Planning and Modeling Services
Indications for Use (Describe) Intended Use
CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide preoperative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for generating the plan are identifiable on patient imaging scans.
Indications for Use
The CORIOGRAPH Pre-op Planning and Modeling Services are indicated for use for the following procedures:
- · Unicondylar Knee Replacement (UKR)
- · Total Knee Arthroplasty (TKA)
- · Primary Total Hip Arthroplasty (THA)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
Image /page/4/Picture/3 description: The image shows the logo for Smith+Nephew. The logo is in a bright orange color. The words "Smith" and "Nephew" are connected by a plus sign.
#### Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
## 510(k) Summary
| 510(k) Owner | Blue Belt Technologies, Inc.<br>2875 Railroad Street<br>Pittsburgh, Pennsylvania 15222 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Chandler Caufield<br>Senior Regulatory Affairs Specialist<br>Email: chandler.caufield@smith-nephew.com |
| Date Prepared | 31-July-2024 |
| Classification Reference | 21 CFR 888.3030 |
| Product Code | PBF |
| Supported Codes | JDI, KWZ, LPH, LWJ, LZO, MBL, MEH |
| Common/Usual Name | Orthopaedic Surgical Planning and Instrument Guides |
| Trade/Proprietary Name | CORIOGRAPH Pre-Op Planning and Modeling Services |
| Primary Predicate Device | CORIOGRAPH Knee Pre-Op Plan (K240113) |
| Secondary Predicate<br>Device | RI.HIP Modeler (K212040) |
| Reason for Submission | This Traditional 510(k) submission is seeking clearance for a new<br>offering CORIOGRAPH Hip Pre-Op Plan V1.0 and CORIOGRAPH Modeler<br>V1.0, modules of the CORIOGRAPH Pre-Op Planning and Modeling<br>Services V2.0. |
{5}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
T: 412-683-3844
Smith+Nephew
www.smith-nephew.com
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
## Intended Use
CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide pre-operative planning for surgical procedures based on patient imaging, provided that anatomic landmarks generating the plan are identifiable on patient imaging scans.
## Indications for Use
The CORIOGRAPH Pre-op Planning and Modeling Service is indicated for use for the following procedures:
- unicondylar knee replacement (UKR)
- . total knee arthroplasty (TKA)
- . primary total hip arthroplasty (THA)
## Device Description
The subject CORIOGRAPH Hip Pre-Op Plan and CORIOGRAPH Modeler are medical function modules within the CORIOGRAPH Pre-Op Planning and Modeling Services are additional offerings being introduced by Blue Belt Technologies, Inc. to allow for pre-operative planning for surgical procedures based on patient imaging for primary total hip arthroplasty (THA). The CORIOGRAPH Hip Pre-Op Plan system will utilize Smith and Nephew personnel to generate patient specific bone models and preoperative plans for primary THA which will be viewable and editable on CORIOGRAPH Modeler 1.0. Together, CORIOGRAPH Hip Pre-Op Plan and CORIOGRAPH Modeler are the subject of this submission.
{6}------------------------------------------------
T: 412-683-3844 www.smith-nephew.com
Image /page/6/Picture/2 description: The image shows the logo for Smith+Nephew. The logo is in orange and features the company name in a sans-serif font. The plus sign is slightly larger than the other characters and is positioned between "Smith" and "Nephew."
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
## Currently Supported Hip Implants
The CORIOGRAPH Hip Pre-Op Plan is compatible with implant systems provided in Table 1.
TABLE 1. CURRENTLY SUPPORTED SMITH+ NEPHEW HIP IMPLANTS FOR THE CORIOGRAPH HIP PRE-OP PLAN V1.0
| Implant System Name | 510(k) |
|----------------------------------------------------|---------|
| ANTHOLOGY HIP SYSTEM | K211176 |
| CATALYSTEM | K240381 |
| CPCS CEMENTED HIP SYSTEM | K970351 |
| | K211176 |
| POLARSTEM | K203175 |
| | K211176 |
| REDAPT MONOBLOCK REVISION FEMORAL SYSTEM | K211176 |
| SL-PLUS AND SLR-PLUS CEMENTLESS FEMORAL HIP SYSTEM | K211176 |
| SYNERGY HIP SYSTEM | K963509 |
| | K211176 |
| FEMORAL HEADS | K021673 |
| | K211176 |
| OR30 HIP SYSTEM | K220959 |
| | K232667 |
| R3 ACETABULAR SYSTEM - STIKTITE SHELLS | K070756 |
| R3 ACETABULAR SYSTEM | K211176 |
| REDAPT ACETABULAR SYSTEM | K211176 |
## Discussion of Similarities and Differences
The purpose of this Traditional 510(k) submission is to seek clearance for the CORIOGRAPH Hip Pre-Op Plan V1.0 and CORIOGRAPH Modeler V1.0, which are medical function modules of the CORIOGRAPH Pre-Op Planning and Modeling Services V2.0. This Software as a Medical Device (SaMD) is based on CORIOGRAPH Knee Pre-Op Plan V1.0 also used to generate preoperative plans (K240113) as well as the secondary predicate RI.HIP Modeler (K212040).
The software planning tools and processes used for the CORIOGRAPH Hip Pre-Op Plan have similar intended use, indications for use and technological characteristics as the software planning tools used for the preoperative plans generated for surgeon concurrence in the predicate CORIOGRAPH Knee Pre-Op Plan V1.0 (K240113) as well as in the secondary predicate RI.HIP Modeler (K212040).
Smith & Nephew believes that CORIOGRAPH Hip Pre-Op Plan V1.0 is subject to premarket notification requirements under Section 510(k) of the Federal Food, Drug and Cosmetic Act ("FFDCA" or "the Act") and is substantially equivalent to the previously cleared primary predicate CORIOGRAPH Knee Pre-Op Plan V1.0 (K240113) and secondary predicate RI.HIP Modeler (K212040).
{7}------------------------------------------------
T: 412-683-3844
www.smith-nephew.com
Image /page/7/Picture/2 description: The image shows the logo for Smith+Nephew. The logo is in orange color. The words "Smith" and "Nephew" are connected by a plus sign.
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
#### Table 2. Primary Predicate Device
| Manufacturer | Description | Submission Number | Clearance Date |
|----------------------|--------------------------------------|-------------------|----------------|
| Smith & Nephew, Inc. | CORIOGRAPH Knee Pre-<br>Op Plan V1.0 | K240113 | 03/18/2024 |
#### Table 3. Secondary Predicate Device
| Manufacturer | Description | Submission Number | Clearance Date |
|-----------------------------|----------------|-------------------|----------------|
| Blue Belt Technologies Inc. | RI.HIP MODELER | K212040 | 03/11/2022 |
Substantial equivalence analysis for the CORIOGRAPH Knee Pre-Op Plan is provided in Table 4.
{8}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
#### T: 412-683-3844
www.smith-nephew.com
#### Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
#### Table 4. Substantial Equivalence Table
| Design Aspect<br>Reviewed | CORIOGRAPH Hip Pre-Op Plan<br>(Subject Device) | CORIOGRAPH Knee Pre-Op Plan<br>(Primary Predicate Device) | RI.HIP Modeler<br>(Secondary Predicate Device) | Comparison |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K242272 | K240113 | K212040 | N/A |
| Indications for<br>Use | Intended Use<br>CORIOGRAPH Pre-operative Planning<br>Service is intended to<br>provide pre-operative<br>planning for surgical<br>procedures based on<br>patient imaging,<br>provided that anatomic<br>landmarks necessary<br>generating the plan are<br>identifiable on patient<br>imaging scans.<br><br>Indications for Use<br>The CORIOGRAPH Pre-operative Planning<br>Service is indicated for<br>use for the following<br>procedures:<br>unicondylar knee<br>replacement (UKR) total knee<br>arthroplasty (TKA) primary total hip<br>arthroplasty (THA) | Intended Use<br>CORIOGRAPH Pre-operative Planning<br>Service is intended to<br>provide pre-operative<br>planning for surgical<br>procedures based on<br>patient imaging,<br>provided that<br>anatomic landmarks<br>necessary for<br>alignment and<br>positioning of the<br>implant are<br>identifiable on patient<br>imaging scans.<br><br>Indications for Use<br>The CORIOGRAPH Pre-operative Planning<br>Service is indicated for<br>use for the following<br>procedures:<br>unicondylar knee<br>replacement<br>(UKR) total knee<br>arthroplasty (TKA) | Intended Use<br>The RI.HIP MODELER is<br>intended for<br>preoperative planning<br>for primary total hip<br>arthroplasty. RI.HIP<br>MODELER is intended<br>to be used as a tool to<br>assist the surgeon in<br>the selection and<br>positioning of<br>components in primary<br>total hip arthroplasty.<br><br>Indications for Use<br>RI.HIP MODELER is<br>indicated for<br>individuals undergoing<br>primary hip surgery. | Substantially<br>equivalent -<br>New indication<br>added for Total Hip<br>Arthroplasty Introduction of Hip<br>Modules has made<br>the services offered<br>plural. Anatomic<br>landmarks<br>necessary<br>statement has<br>changed from<br>"alignment and<br>positioning of<br>implant" to<br>"generating the<br>plan". |
{9}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
www.smith-nephew.com
#### Smith ew
| | | Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc. | | | | | | | |
|--|--|----------------------------------------------------------------------|--|--|--|--|--|--|--|
|--|--|----------------------------------------------------------------------|--|--|--|--|--|--|--|
| Design Aspect<br>Reviewed | CORIOGRAPH Hip Pre-<br>Op Plan<br>(Subject Device) | CORIOGRAPH Knee<br>Pre-Op Plan<br>(Primary Predicate<br>Device) | RI.HIP Modeler<br>(Secondary Predicate<br>Device) | Comparison |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Target<br>Population | Patients who are able<br>to undergo an CT<br>safely, fit the<br>indications, and do not<br>present any<br>contraindications for<br>the existing hip implant<br>system to which it is<br>designed. | Patients who are able<br>to undergo an MRI<br>safely, fit the<br>indications, and do<br>not present any<br>contraindications for<br>the existing knee<br>implant system to<br>which it is designed. | The patient population<br>is adults only. | Substantially<br>equivalent -<br>CORIOGRAPH Pre-Op<br>Planning and Modeling<br>Services is<br>incorporating new<br>modules that use an<br>alternative imaging<br>modality with potential<br>patient restrictions.<br>Therefore, the Target<br>Population statement is<br>being updated to<br>reflect the broader<br>applications. |
| Target<br>Anatomy | Knee<br>Hip | Knee | Hip | Substantially<br>equivalent -<br>CORIOGRAPH Pre-Op<br>Planning and Modeling<br>Services are expanding<br>the offering to planning<br>for Hip implants.<br>Compatible implants<br>for both primary<br>predicate device and<br>secondary predicate<br>device are implants<br>manufactured by Blue<br>Belt Technologies,<br>Inc.'s parent company<br>Smith+Nephew. |
| Where used | Professional healthcare<br>facility | Professional<br>healthcare facility | Clinical Setting | Same as primary<br>predicate |
| Compatible<br>Total Knee<br>Systems | Smith and Nephew Inc.<br>Implants:<br>Genesis II CR<br>Genesis II PS<br>LEGION CR<br>LEGION PS<br>Journey II BCS<br>Journey II CR<br>Journey II XR<br>JOURNEY II<br>Unicompartmental Knee | Smith and Nephew Inc.<br>Implants:<br>Genesis II CR<br>Genesis II PS<br>LEGION CR<br>LEGION PS<br>Journey II BCS<br>Journey II CR<br>Journey II XR<br>JOURNEY II<br>Unicompartmental Knee | N/A | Same as primary<br>predicate |
| Design Aspect<br>Reviewed | CORIOGRAPH Hip Pre-<br>Op Plan<br>(Subject Device) | CORIOGRAPH Knee<br>Pre-Op Plan<br>(Primary Predicate<br>Device) | RI.HIP Modeler<br>(Secondary Predicate<br>Device) | Comparison |
| Compatible<br>Total Hip<br>Systems | Smith and Nephew Inc.<br>Implants:<br>ANTHOLOGY HIP System<br>CATALYSTEM<br>CPCS Cemented Hip<br>System<br>POLARSTEM Standard<br>Offset Stem<br>POLARSTEM Lateral Offset<br>Stem<br>POLARSTEM Valgus Offset<br>Stem REDAPT Monoblock<br>Revision Femoral System<br>SL-PLUS AND SLR-PLUS<br>Cementless Femoral Hip<br>System<br>Femoral Heads<br>OR30 Hip System<br>R3 Acetabular System<br>R3 Acetabular System<br>Anteverted Liner<br>R3 Constrained Liner<br>SYNERGY Hip System | N/A | Smith and Nephew Inc.<br>Implants:<br>ANTHOLOGY Stem<br>POLARSTEM Standard<br>Offset Stem<br>POLARSTEM Lateral Offset<br>Stem<br>POLARSTEM Valgus Offset<br>Stem<br>Femoral Heads<br>R3 Acetabular System<br>R3 Acetabular System<br>Anteverted Liner<br>R3 Constrained Liner<br>OR30 Liner<br>SYNERGY Hip system | Substantially<br>equivalent to<br>secondary predicate.<br>CORIOGRAPH Pre-Op<br>Planning and Modeling<br>Services are expanding<br>the CORIOGRAPH<br>offering to planning for<br>primary Total Hip<br>Arthroplasty Hip<br>implants. Compatible<br>implants for both<br>primary predicate<br>device and secondary<br>predicate device are<br>implants manufactured<br>by Blue Belt<br>Technologies, Inc.'s<br>parent company<br>Smith+Nephew. |
| Design Type | Patient Matched | Patient Matched | Patient Matched | Same as predicate<br>devices. |
| Input | Surgeon Preferences<br>X-Ray<br>MRI<br>CT Scan | Surgeon Preferences<br>X-Ray<br>MRI | Surgeon Preferences<br>X-Ray | Substantially<br>Equivalent |
| Output | Preoperative surgical<br>plan<br>Simulation of implant<br>position | Preoperative surgical<br>plan | Simulation of implant<br>position | Substantially<br>Equivalent - Hip<br>modules are generating<br>a preoperative plan<br>that the surgeon may<br>use to drive clinical<br>decisions which is<br>similar to the Knee<br>module.<br>Computational<br>modeling simulation is<br>equivalent to<br>Secondary Predicate<br>Device RI.HIP<br>MODELER. |
| Design Aspect<br>Reviewed | CORIOGRAPH Hip Pre-<br>Op Plan<br>(Subject Device) | CORIOGRAPH Knee<br>Pre-Op Plan<br>(Primary Predicate<br>Device) | RI.HIP Modeler<br>(Secondary Predicate<br>Device) | Comparison |
| Common Name | Orthopaedic Surgical<br>Planning and<br>Instrument Guides | Orthopaedic Surgical<br>Planning and<br>Instrument Guides | System, Image<br>Processing, Radiological | Same as primary<br>predicate |
| Device Class | Class II | Class II | Class II | Same as primary<br>predicate and<br>secondary predicate |
| Device<br>Classification<br>Name and<br>Reference | 21 CFR 888.3030 -<br>Single/multiple<br>component metallic<br>bone fixation<br>appliances and<br>accessories. | 21 CFR 888.3030 -<br>Single/multiple<br>component metallic<br>bone fixation<br>appliances and<br>accessories. | 21 CFR 892.2050 -<br>Medical image<br>management and<br>processing system. | Same as primary<br>predicate |
| Panel Code | Orthopaedics/87 | Orthopaedics/87 | Radiology/90 | Same as primary<br>predicate |
| Product Code | PBF | PBF | LLZ | Same as primary<br>predicate |
| Supported<br>Product Code<br>(Knee) | HSX | HSX | N/A | Same as primary<br>predicate |
| | JWH | JWH | N/A | |
| | MBH | MBH | N/A | |
| | OOG | OOG | N/A | |
| Supported<br>Product Code<br>(Hip) | JDI | N/A | JDI | Same as secondary<br>predicate |
| | KWZ | N/A | KWZ | |
| | LPH | N/A | LPH | |
| | LWJ | N/A | LWJ | |
| | LZO | N/A | LZO | |
| | MBL | N/A | MBL | |
| | MEH | N/A | MEH | |
{10}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
### T: 412-683-3844
www.smith-nephew.com
#### Sn W
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
{11}------------------------------------------------
T: 412-683-3844
www.smith-nephew.com
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
## Non-Clinical Testing (Bench)
Verification and validation testing demonstrated the safety and effectiveness of the software applications used in the CORIOGRAPH Pre-Op Planning & Modeling Services V2.0. Summative usability testing demonstrated that participating surgeons were able use the subject device safely and effectively in a simulated use environment. In addition, the credibility evaluation demonstrated that the kinematic models and Activities of Daily Living (ADLS) utilized in the subject device are clinically relevant. Blue Belt Technologies, Inc. has concluded that all design inputs have been met and that the verification and validation testing performed did not raise any new questions of safety or effectiveness.
{12}------------------------------------------------
2875 Railroad Street Pittsburgh, Pennsylvania 15222 USA
Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.
## Conclusion
The software tools and process used for CORIOGRAPH Pre-Op Planning & Modeling Services V2.0 have similar intended use, indications for use and technological characteristics used in the software planning tools and processes used for the preoperative plans generated for the primary predicate CORIOGRAPH Knee Pre-Op Plan (K240113) and the secondary predicate RI.HIP Modeler (K212040).
Verification testing demonstrated that the system meets required design inputs. Summative usability testing results demonstrate that representative users can use the safely and effectively in a simulated use environment. The information presented in this 510(k) premarket notification demonstrates that the CORIOGRAPH Pre-Op Planning & Modeling Services V2.0, specifically the CORIOGRAPH Hip Pre-Op Plan System V1.0 and the CORIOGRAPH Modeler V1.0 may be used to generative plans for surgical procedures based on patient imaging. This information also demonstrates that the CORIOGRAPH Pre-Op Planning & Modeling Services V2.0 are as safe and effective as the software and planning process used in the primary predicate CORIOGRAPH Knee Pre-Op Plan (K240113) and secondary predicate RI.HIP Modeler (K212040). Blue Belt Technologies, Inc. believes that FDA can find CORIOGRAPH Pre-Op Planning & Modeling Services V2.0 to be substantially equivalent to the primary and secondary predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.