RI.HIP MODELER

K212040 · Blue Belt Technologies, Inc. · LLZ · Mar 11, 2022 · Radiology

Device Facts

Record IDK212040
Device NameRI.HIP MODELER
ApplicantBlue Belt Technologies, Inc.
Product CodeLLZ · Radiology
Decision DateMar 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

RI.HIP MODELER is intended for preoperative planning for primary total hip arthroplasty. RI.HIP MODELER is intended to be used as a tool to assist the surgeon in the selection and positioning of components in primary total hip arthroplasty. RI.HIP MODELER is indicated for individuals undergoing primary hip surgery.

Device Story

RI.HIP MODELER is a standalone iPad application for preoperative planning of primary total hip arthroplasty (THA). It assists surgeons in selecting and positioning acetabular cups to reduce implant impingement. Input data includes standing and sitting X-rays, captured via iPad camera, imported from Quentry, or from the camera roll. The software processes these images to determine patient spinopelvic mobility classification. Using this classification and a simulation database of hip kinematics for daily activities, the device calculates implant kinematics and distance to impingement. It provides a baseline cup orientation recommendation. Surgeons use the interface to visualize 3D models, overlay implants, and adjust parameters based on clinical judgment. The device is used in clinical settings by surgeons to manage patient expectations and optimize implant placement, potentially improving patient outcomes by reducing impingement risk.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including software verification (IEC 62304), image capture verification (landmarking accuracy within 2 degrees), and usability engineering validation. Computational models were verified and validated per ASME V&V 40:2018.

Technological Characteristics

Standalone iPad software application. Uses computational modeling to simulate 3D hip anatomy and range of motion based on 2D X-ray inputs. Features include image enhancement (contrast/brightness), cropping, and virtual measuring tools. Complies with ASME V&V 40:2018 for computational modeling. Software developed per IEC 62304.

Indications for Use

Indicated for individuals undergoing primary total hip arthroplasty. Used as a tool to assist surgeons in component selection and positioning.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Blue Belt Technologies, Inc. Amy Winegarden Senior Regulatory Affairs Specalist 2905 Northwest Boulevard. Suite 40 Plymouth, MN 55441 Re: K212040 Trade/Device Name: Ri.hip Modeler Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: [NOTE: Use date of most recent supplement] Received: February 10, 2022 Dear Amy Winegarden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K212040 Device Name RI.HIP MODELER Indications for Use (Describe) RI.HIP MODELER is intended for preoperative planning for primary total hip arthroplasty. RI.HIP MODELER is intended to be used as a tool to assist the surgeon in the selection and positioning of components in primary total hip arthroplasty. RI.HIP MODELER is indicated for individuals undergoing primary hip surgery. Type of Use (Select one or both, as applicable) | <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> | |-----------------------------------------------------------------------------------------------|-----------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | 510(k) Owner | Blue Belt Technologies, Inc.<br>2905 Northwest Blvd Ste. 40<br>Plymouth, MN 55441 USA Tel:<br>(763) 452-4950<br>Fax: (763) 452-4675 | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Michael Gosha<br>Regulatory Affairs Specialist II<br>Email: Michael.Gosha@Smith-Nephew.com<br>Phone: 216-816-6728 | | Date of Submission | June 29, 2021 | | Classification Reference | 21 CFR 892.2050 | | Product Code | LLZ | | Common/Usual Name | Picture archiving and communications system | | Trade/Proprietary Name | RI.HIP MODELER™ | | Predicate Device(s) | OneFit Medical - hipEOS (K173390) | | Reason for Submission | New Device | # Indications for Use RI.HIP MODELER is intended for preoperative planning for primary total hip arthroplasty. RI.HIP is intended to be used as a tool to assist the surgeon in the selection and positioning of components in primary total hip arthroplasty. RI.HIP MODELER is indicated for individuals undergoing primary hip surgery. ### Device Description RI.HIP MODELER is a non-invasive standalone Total Hip Arthroplasty (THA) planning software application intended to provide preoperative planning for hip implant acetabular cup rotational selection and placement. RI.HIP MODELER allows surgeons to visualize and perform analysis of digital images for assessment of spinopelvic mobility of the patient. The app is used to characterize patient conditions, manage patient performance expectations, and help surgeons determine acetabular cup placement based on spinopelvic mobility. The software provides a baseline cup orientation recommendation intended to reduce incidences of implant impingement based on patient condition and implant specifications. The surgeon is reminded to verify and adjust to the parameters based on their clinical judgment. {4}------------------------------------------------ # Discussion of Similarities and Differences RI.HIP MODELER is substantially equivalent to the predicate device hipEOS (K173390). The intended use and the established technologies used for preoperative planning of Total Hip Arthroplasty and placement of THA implants, is the same as the predicate device. The table below compares the RI.HIP MODELER, the subject device, to hipEOS, the predicate device. A discussion of the differences follows. | Parameter | Subject Device<br>RI.HIP MODELER | Predicate Device<br>hipEOS<br>(K173390) | |--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use<br>[Substantially<br>Equivalent]1 | RI.HIP MODELER is intended for preoperative planning<br>for primary total hip arthroplasty. RI.HIP MODELER is<br>intended to be used as a tool to assist the surgeon in<br>the selection and positioning of components in<br>primary total hip arthroplasty.<br><br>RI.HIP MODELER is indicated for individuals<br>undergoing primary hip surgery. | hipEOS software is indicated to be used for assisting<br>healthcare professionals in preoperative planning of<br>total hip arthroplasty (THA). The software allows for<br>overlaying of 3D/2D hip implant models on<br>radiological images or reconstruction of bone and<br>includes tools for performing measurements on the<br>image or model of bones and for selecting and<br>positioning the implant models. Clinical judgments<br>and experience are required to properly use the<br>software. | | Indications for Use<br>[Substantially<br>Equivalent]1 | RI.HIP MODELER is intended for preoperative planning<br>for primary total hip arthroplasty. RI.HIP MODELER is<br>intended to be used as a tool to assist the surgeon in<br>the selection and positioning of components in<br>primary total hip arthroplasty.<br><br>RI.HIP MODELER is indicated for individuals<br>undergoing primary hip surgery. | hipEOS software is indicated to be used for assisting<br>healthcare professionals in preoperative planning of<br>Total Hip Arthroplasty (THA). The software allows for<br>overlaying of 3D/2D hip implant models on<br>radiological images or reconstruction of bone,<br>and includes tools for performing measurements on<br>the image or model of bones and for selecting and<br>positioning the implant models. Clinical judgments<br>and experience are required to properly use the<br>software. | | Classification and<br>Regulation<br>[Same] | LLZ<br>892.2050 | LLZ<br>892.2050 | | Main System<br>Components<br>[Same] | Standalone software | Standalone software | | Operational<br>Environment<br>[Same] | RI.HIP MODELER is intended to be used in a clinical<br>setting by a qualified surgeon. | hip EOS is intended to be used in a clinical setting by a<br>qualified surgeon. | | Imaging Requirements<br>[Same] | Weight bearing standing and sitting x-ray | Weight bearing standing and sitting x-ray | | Image Source<br>[Substantially<br>Equivalent]2 | Images from Brainlab Quentry, iPad Camera Roll, iPad<br>Camera | DICOM | | Primary Device<br>Function<br>[Substantially<br>Equivalent]3 | RI.HIP MODELER uses stand-alone software for<br>computer assisted surgical planning.<br>The primary function of the software is to provide a<br>baseline cup orientation recommendation intended to<br>reduce incidences of implant impingement based on<br>inclination/anteversion and implant specifications,<br>and stem anteversion. | hipEOS uses a standalone software for computer<br>assisted surgical planning.<br>The primary function of the software is to provide a<br>recommendation of stem and cup placement<br>intended to reduce incidences of implant<br>impingement, based on inclination/anteversion,<br>implant specifications, stem torsion and localization of<br>the femoral neck resection plane. | {5}------------------------------------------------ RI.HIP MODELER is intended for preoperative planning for primary Total Hip Arthroplasty. RI.HIP MODELER is intended to be used as a tool to assist the surgeon in the selection and positioning of components in primary Total Hip Arthroplasty. RI.HIP MODELER is indicated for individuals undergoing primary Total Hip Arthroplasty. The indication is substantially equivalent to the indications for use of the predicate, hipEOS (K173390), which is also indicated as a tool to assist healthcare professionals in preoperative planning for Total Hip Arthroplasty. The tools that are included in the predicate device's indications are substantially equivalent to the subject device and compared in the technological characteristic section of the substantial equivalence discussion. ### ²Image Format DICOM images downloaded from Quentry are displayed as a PNG file. RI.HIP MODELER also allows the user to take a picture of an x-ray with the iPad which displays as a PNG file or pull a picture from the camera roll. The software was designed to autodetect and dewarp (if necessary) the specified x-ray. Images can be enhanced using contrast and brightness filters. Additionally, they can be cropped, reset, and retaken. Verification confirmed the camera functionality generated a clinically relevant image with minimal distortion even in worstcase scenario conditions. Training is provided to surgeons on the use of the camera in the application. Human Factors validation confirmed that users could successfully use the camera function to obtain a clinically acceptable image. Validation further showed that those images could be successfully landmarked. The use of the iPad camera to capture an x-ray image does not raise new questions of safety or effectiveness for RI.HIP MODELER. #### 3 Primary Device Function RI.HIP MODELER is substantially equivalent to hipEOS in primary function. hipEOS provides a recommendation for both stem and cup placement in the hip. RI.HIP MODELER provides a baseline cup orientation recommendation. The cup recommendation is a subset of the functionality and does not raise new questions of safety or effectiveness for RI.HIP MODELER. Like hipEOS, RI.HIP MODELER uses spinopelvic mobility of the patient to recommend cup inclination/anteversion. hipEOS derives stem torsion based on the 3D model of the patient's anatomy. RI.HIP MODELER allows the user to input stem anteversion based on clinical judgment. Both applications allow the user to adjust stem anteversion. The difference does not raise new questions of safety or effectiveness as the stem torsion is ultimately identified by the qualified healthcare professional. #### 4Principles of Operation RI.HIP MODELER is substantially equivalent to hipEOS in principles of operation. Both devices use standing and sitting x-rays to ascertain the spinopelvic mobility of the patient. {6}------------------------------------------------ hipEOS reconstructs the patient's anatomy in 3D using frontal and lateral 2D x-rays. The 3D model shows torsion and rotation parameters based on the patient's anatomy. The software provides a baseline stem and cup orientation based on the spinopelvic mobility classification derived from the 3D model and implant specifications. RI.HIP MODELER employs an algorithm using the determined spinopelvic mobility classification and hip kinematics for typical patient activities of daily living (accessed through a simulation database). Implant kinematics and distance to impingement are calculated for each activity, given a set of implant specifications. The software provides a baseline cup orientation recommendation based on the spinopelvic mobility classification and implant specifications. The computational models used in the design of RI.HIP MODELER comply with ASME V&V 40: 2018, Assessing Credibility of Computational Modeling through Verification and Validation; Application to Medical Devices and support the algorithm that provides the baseline cup orientation. There are no new questions of safety or effectiveness in using the RI.HIP MODELER to provide a baseline cup orientation as the computational models that support the design have been verified and validated. ### §Technological Characteristics There are no new concerns for safety or effectiveness associated with the technological characteristics of RI.HIP MODELER. - . RI.HIP MODELER and hipEOS are both standalone software applications with an interactive design that allows the user to overlay 2D x-rays with virtual measuring tools to derive a spinopelvic mobility classification of the patient. - hipEOS uses a frontal and lateral x-ray to develop a 3D model of the patient's hip. RI. HIP MODELER displays a simulated 3D model of the hip based on the spinopelvic mobility classification. The simulated model is derived from validated computational models that effectively display the range of implant motion in the model. - The 3D models in both RI.HIP MODELER and hipEOS allow the user to select and overlay implants and view a simulated range of motion. - . RI.HIP MODELER and hipEOS both provide a range of implant motion simulation that allows the surgeon to see how the implant moves in the hip. - RI.HIP MODELER recommends an acetabular cup placement based on spinopelvic mobility classification. The surgeon selects and compares implants from an application database. hipEOS automatically selects and positions the stem and cup based on 3D reconstruction of the patient anatomy. The surgeon can select and compare the stem and cup from an online database. - . The user can change the inclination/anteversion and stem torsion within the app and visualize the effect of those changes on the 3D model. - RI.HIP MODELER uses computational modeling in its design that allows a surgeon to view distance to implant impingement through a variety of daily activities of the patient, including, but not limited to, sitting, standing, and walking. hipEOS uses a 3D reconstruction that allows a surgeon to plan implant placement for sitting, standing, and theoretical sitting positions. {7}------------------------------------------------ - . RI.HIP MODELER is an iPad application and hipEOS is a web application. Both devices are standalone software. There are no new concerns for safety or effectiveness due to the difference in technological characteristics because RI.HIP MODELER was designed to meet its intended use with the technological characteristics within the app. Verification and Validation from Performance Testing confirms that the RI.HIP MODELER is safe and effective and meets its intended use. Performance Testing confirms that representative users of the software application were able to successfully follow the workflow and use the software application in a safe and effective way in the intended use environment. ### َ User Interface RI.HIP MODELER is standalone software that is intended to be installed on an iPad. The predicate device is standalone software that can be installed on any computer. The subject software was verified and validated to ensure specifications were met. User validation showedsurgeons can use the RI.HIP MODELER application safely and effectively on an iPad to plan for total hip arthroplasty procedures. There are no new questions of safety or effectiveness due tothe different interface as the design of RI.HIP MODELER was specific to the intended interface and allows the device to perform its intended use. ## Non-Clinical Testing (Bench) Design verification and validation testing was performed to demonstrate that RI.HIP MODELER meets all design requirements and is as safe and effective as its predicate device. Comprehensive performance testing demonstrated that the subject devices meets the required design inputs. Performance data consisted of physical performance test for all the software application. Additionally, the following test data was provided: - . Software Verification Testing, including software integration and workflow testing, was completed. Software was developed in accordance with IEC 62304 Medical Device Software -Software Life Cycle Processes, and this submission contains documentation per the requirements of FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. - Image Capture Verification Testing was completed to ensure worst case image capture would not produce distorted images that would affect clinical measurements. Testing demonstrated that even in worst-case conditions accuracy of landmarking measurements was within 2 degrees. - Usability Engineering Validation Testing was completed demonstrating that representative users were able to safely and effectively use RI.HIP MODELER in a simulated use environment. No human clinical testing was required to determine the safety and effectiveness of RI.HIP MODELER. {8}------------------------------------------------ # Conclusion The subject device, RI.HIP MODELER, described in this submission has a similar intended use and primary device function as the predicate device, hipEOS (K173390). The primary difference between the two devices is that RI.HIP MODELER uses 2D x-rays and an algorithm based on computational modeling to determine a baseline acetabular cup placement while hipEOS uses a 3D model of the patient derived from a compilation of x-rays to determine stem and cup placement. The differences do not raise any new questions of safety or effectiveness. The key determining factor in establishing substantial equivalence is whether RI.HIP MODELER can provide a baseline cup orientation that increases distance to implant impingement during activities of daily living, thereby reducing the risk of implant impingement. The results of verification and validation indicate that there are no safety or performance concerns and that RI.HIP MODELER meets its intended use. The usability testing results for Human Factors demonstrate that representative users are able to use the subject device safely and effectively to meet the desired intended use. The information presented in this 510(k) premarket notification demonstrates that RI.HIP MODELER is as safe and effective as the predicate hipEOS system (K173380). Blue Belt Technologies believes that RI.HIP MODELER is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...