EMPHASYS Acetabular Shell with RapiTite HA
K242084 · Depuy Ireland UC · LPH · Oct 7, 2024 · Orthopedic
Device Facts
| Record ID | K242084 |
| Device Name | EMPHASYS Acetabular Shell with RapiTite HA |
| Applicant | Depuy Ireland UC |
| Product Code | LPH · Orthopedic |
| Decision Date | Oct 7, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3358 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. The EMPHASYS Acetabular Shell with RapiTite HA is intended for single use only.
Device Story
Modular cementless porous-coated acetabular shell; features hydroxyapatite (HA) coating applied conformally throughout porous coating. Available in No-Hole, 3-Hole, and Multi-Hole configurations. Used in total hip arthroplasty to replace damaged hip joint articulation; intended to increase patient mobility and reduce pain. Implanted by orthopedic surgeons in clinical settings. Device provides stable fixation via cementless interface with patient bone; HA coating facilitates bone ingrowth. Benefits include restoration of joint function and pain relief.
Clinical Evidence
Bench testing and animal study. Bench testing included chemical analysis (ASTM F1609, ISO 13779), dissolution (ASTM F1926), solubility (ISO 13779), morphological/particle characterization, interfacial strength (shear/tensile per ASTM F1044/F1147), fatigue (ASTM F1160, F3090), stereological evaluation (ASTM F1854), friction (ASTM D4518), biocompatibility (ISO 10993-1), and MRI safety (ASTM E2503, E2182, E2052). Bone ingrowth animal study performed. No clinical human data reported.
Technological Characteristics
Modular cementless acetabular shell; porous-coated metal substrate with conformal hydroxyapatite (HA) coating. Materials and design identical to predicate EMPHASYS system. Testing per ASTM F1609, F1926, F1044, F1147, F1160, F1854, D4518, F3090, and ISO 13779. Biocompatibility per ISO 10993-1. MRI safety evaluated per ASTM E2503, E2182, E2052.
Indications for Use
Indicated for patients requiring total hip replacement due to: osteoarthritis, traumatic arthritis, rheumatoid arthritis, congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head or neck, failed previous hip surgery (reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement), or ankylosis. Indicated for cementless use only.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
- DePuy EMPHASYS Acetabular System (K221636/K231873)
- DePuy Pinnacle Duofix HA Acetabular Cup (K031495)
- DePuy Pinnacle with Gription Acetabular Cups (K090998/K093646)
Reference Devices
- Stryker Tritanium® Peri-Apatite™ Acetabular Shell System (K101072)
Related Devices
- K990666 — HYDROXYAPATITE (HA) REFLECTION ACETABULAR SHELL, HYDROXYAPATITE (HA) REFLECTION INTERFIT SHELL · Smith & Nephew, Inc. · Aug 6, 1999
- K972393 — INTER-OP HA POROUS ACETABULAR SYSTEM (HA/CSTI) · Sulzer Orthopedics, Inc. · Sep 19, 1997
- K071784 — DEPUY PINNACLE WITH GRIPTON ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jul 25, 2007
- K090998 — DEPUY PINNACLE 100 WITH GRIPTION ACETABULAR CUPS · DePuy Orthopaedics, Inc. · Jun 12, 2009
- K043099 — LINEAGE HA ACETABULAR SHELLS · Wrightmedicaltechnologyinc · Sep 29, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 7, 2024
DePuy Ireland UC Elaine Pears Regulatory Affairs Manager Loughbeg, Ringaskiddy Cork, Ireland
Re: K242084
Trade/Device Name: EMPHASYS Acetabular Shell with RapiTite HA Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: July 16, 2024 Received: September 12, 2024
Dear Elaine Pears:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the text "Limin Sun -S" in a large, sans-serif font. The text is black and is set against a white background. The words are arranged horizontally, with a hyphen separating "Sun" and "S".
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242084
Device Name
EMPHASYS Acetabular Shell with RapiTite HA
Indications for Use (Describe)
Total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
The EMPHASYS Acetabular Shell with RapiTite HA is intended for single use only.
INDICATIONS
Total hip replacement is indicated in the following conditions:
1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis,
- or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head or neck.
- 4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis,
hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
- 5. Certain cases of ankylosis.
EMPHASYS Acetabular Shell with RapiTite HA is indicated for cementless use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) SUMMARY – EMPHASYS ACETABULAR SHELL WITH RAPITITE HA
(As required by 21 CFR 807.87(h))
| Submitter Information | | |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Name | DePuy Ireland UC | |
| Address | Loughbeg, Ringaskiddy<br>Co. Cork, IRELAND | |
| Establishment Registration Number | 3015516266 | |
| Submission Prepared by | Elaine Pears | |
| e-mail address | epears@its.jnj.com<br>DePuySynthesJointsRegulatoryAffairs@its.jnj.com | |
| Work mobile | +44 (0)7876 217532 | |
| Alternative Contact Person | Clare Hill | |
| e-mail address | chill7@its.jnj.com | |
| Work mobile | +44 7795 389956 | |
| Date prepared | 16 July 2024 | |
| Name of device | | |
| Trade or proprietary name | EMPHASYS Acetabular Shell with RapiTite HA | |
| Common or usual name | Total Hip Arthroplasty Prosthesis | |
| Classification name | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous<br>Uncemented<br>Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer,<br>Cemented Or Non-Porous, Uncemented | |
| Class | II | |
| Classification panel | 87 Orthopedics | |
| Regulation | 21 CFR 888.3358; 21 CFR 888.3353 | |
| Product Codes | LPH; LZO | |
| Legally marketed devices<br>to which equivalence is<br>claimed | Predicate Devices | Reference Device |
| | DePuy EMPHASYS Acetabular<br>System (K221636/K231873)<br>(primary predicate) | |
| | DePuy Pinnacle Duofix HA<br>Acetabular Cup (K031495) | Stryker Tritanium® Peri-<br>Apatite™ Acetabular Shell<br>System (K101072) |
| | DePuy Pinnacle with Gription<br>Acetabular Cups<br>(K090998/K093646) | |
| Reason for 510(k)<br>submission | The purpose of this 510(k) submission is to obtain market clearance<br>for the DePuy EMPHASYS Acetabular Shell with RapiTite HA. | |
| Device description | The EMPHASYS Acetabular Shell with RapiTite HA is a modular<br>cementless porous-coated acetabular shell, with a hydroxyapatite (HA)<br>coating applied conformally throughout the porous coating, offered in<br>three configurations (No-Hole, 3-Hole, and Multi-Hole). | |
| Intended use of the device | Total hip arthroplasty is intended to provide increased patient mobility and<br>reduce pain by replacing the damaged hip joint articulation in patients<br>where there is evidence of sufficient sound bone to seat and support the<br>components.<br>The EMPHASYS Acetabular Shell with RapiTite HA is intended for<br>single use only. | |
| Indications for use | Total hip replacement is indicated in the following conditions:<br>1. A severely painful and/or disabled joint from osteoarthritis,<br>traumatic arthritis, rheumatoid arthritis, or congenital hip<br>dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Failed previous hip surgery including joint reconstruction,<br>internal fixation, arthrodesis, hemiarthroplasty, surface<br>replacement arthroplasty, or total hip replacement.<br>5. Certain cases of ankylosis.<br>EMPHASYS Acetabular Shells with RapiTite HA are indicated for<br>cementless use only. | |
| Comparison of<br>Technological<br>Characteristics with<br>predicate devices | The EMPHASYS Acetabular Shell with RapiTite HA is identical to the<br>acetabular shells in the primary predicate EMPHASYS Acetabular<br>System in terms of intended use, indications, design and materials of<br>construction, with the only exception being that the subject devices have<br>an additional layer of HA coating. The HA coating is equivalent in<br>structure and composition to that used in the secondary predicate (DePuy<br>Pinnacle Duofix HA Acetabular Cup), and is applied in a way similar<br>to that used for the reference device (Stryker Tritanium® Peri-<br>Apatite™ Acetabular Shell System). | |
| Animal & Performance<br>Testing | The EMPHASYS Acetabular Shell with RapiTite HA was tested to<br>demonstrate its substantial equivalence to the identified predicate<br>devices. Testing and analyses included: | |
| | Chemical Analysis (per ASTM F1609-08(2014) & ISO 13779-<br>3:2018) Dissolution (per ASTM F1926-14) Solubility (per ISO 13779-6:2015) SEM Morphological Characterization Abraded Particle Characterization FTIR Analysis XRD Analysis (per ISO 13779-3:2018) Static interfacial shear strength (per ASTM F1044-05(2017)e1 Static interfacial tensile strength (per ASTM F1147-<br>05(2017)e1) Interfacial shear fatigue (per ASTM F1160-14 (2017)e1) Stereological evaluation (per ASTM F1854-15) Static friction testing (per ASTM D4518-91) Fatigue testing (per ASTM F3090-20) Biocompatibility (per ISO 10993-1:2018) Bone Ingrowth Animal Study Characterization of the Physical and Mechanical Properties of<br>the HA coated porous surface per FDA guidance documents<br>“Guidance Document for Testing Orthopedic Implants With<br>Modified Metallic Surfaces Apposing Bone Or Bone Cement”<br>(Apr. 1994) and “510(K) Information Needed for<br>Hydroxyapatite Coated Orthopedic Implants (Feb 1997)” MRI Safety Evaluation Testing of Total Hip Systems - ASTM<br>E2503-23, ASTM E2182 -19e2, ASTM E2052-21, ASTM | |
| Substantial Equivalence | The EMPHASYS Acetabular Shell with RapiTite HA is substantially<br>equivalent to the identified predicate devices with respect to intended<br>use, indications, materials of construction, geometry, range of sizes,<br>and method of fixation. Results of animal and performance testing<br>and analyses demonstrate that the EMPHASYS Acetabular Shell with<br>RapiTite HA performs as well as the predicate devices. | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------