The Tritanium® Peri-Apatite™ Acetabular Shell is a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function.
Device Story
Tritanium® Peri-Apatite™ Acetabular Shell System is a hemispherical, metallic acetabular shell for cementless, biologic fixation in total hip arthroplasty. Device features a Ti-6Al-4V alloy shell with a Tritanium® (CPTi) Particle Sintered Foam (PSF) coating and an overlying Peri-Apatite (precipitated calcium phosphate) coating. Available in solid-backed or cluster screw-hole designs; optional dome hole plugs seal the shell. Compatible with polyethylene Trident® inserts, constrained liners, and modular dual mobility (MDM) liners. Used by orthopedic surgeons in clinical settings for primary or revision hip replacement. Provides a stable acetabular component to restore joint function and alleviate pain. Device is single-use.
Clinical Evidence
No clinical data provided. Substantial equivalence based on non-clinical laboratory testing, including mechanical bond strength, fatigue testing, and coating characterization.
Technological Characteristics
Acetabular shell forged from Ti-6Al-4V alloy (ASTM F136). Tritanium® coating (CPTi) per ASTM F1580. Peri-Apatite™ coating (precipitated calcium phosphate) per ASTM F-1609. Hemispherical form factor, 44-72 mm OD. Cementless, porous-coated design. Single-use. No software or electronic components.
Indications for Use
Indicated for patients with painful, disabling hip joint disease (non-inflammatory degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis, late stage avascular necrosis), revision of failed previous hip procedures, clinical management problems where alternative techniques are less likely to succeed, or poor bone stock/acetabular deficiencies. Constrained liners indicated for high dislocation risk (prior dislocation, bone loss, laxity, neuromuscular disease, instability). Modular dual mobility (MDM) liners indicated for noninflammatory degenerative joint disease, rheumatoid arthritis, functional deformity, revision, nonunion/fractures of proximal femur, and dislocation risks. Cementless use only.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Howmedica Osteonics Tritanium Acetabular Shell System (K081171)
Howmedica Osteonics Vitalock Solid Back Acetabular Shell with Peri-Apatite (K971206)
Aesculap BiContact Hip System with u-CaP (K043079)
Smith & Nephew Hip Systems with HA Coating (K090982)
Submission Summary (Full Text)
{0}------------------------------------------------
APR 1 1 2011
#### 510(k) Summary
Howmedica Osteonics Corp. 325 Corporate Drive Mahwah. NJ 07430
K101072 p1/3
Contact Person
Sponsor
Karen Ariemma Project Manager, Regulatory Affairs/Regulatory Compliance Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 Phone: (201) 831-5718
Date Prepared: April 8, 2011
Tritanium® Peri-Apatite™ Acetabular Shell System Proprietary Name:
Artificial Hip Replacement Components - Acetabular Common Name:
#### Classification Name:
21 CFR §888.3358: Hip joint metal/polymer/metal semi-constrained and porous-coated uncemented prosthesis
21 CFR §888.3310: Hip joint metal/polymer constrained cemented or uncemented prosthesis.
21 CFR §888.3353: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis:
21 CFR §888.3360 - Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis.
21 CFR §888.3350 - Hip joint metal/polymer semi-constrained cemented prosthesis
LPH, MEH, LZO, KWZ, LWJ and JDI Product Codes:
Legally Marketed Device to Which Substantial Equivalence is Claimed: Howmedica Osteonics Tritanium Acetabular Shell System: K081171 Howmedica Osteonics Vitalock Solid Back Acetabular Shell with Peri-Apatite: K971206 Landos Inc. Corail Stem: K953111 Aesculap BiContact Hip System with u-CaP K043079 Smith& Nephew, Smith & Nephew Hip Systems with HA Coating: K090982.
Device Description: The Tritanium® Peri-Apatite ™ acetabular shells are available in both solid backed and cluster screw-hole designs for cementless, biologic fixation. The Howmedica Osteonics Tritanium® Peri-Apatite™ Acetabular Shell consists of a hemispherical, metallic acetabular shell with a Tritanium® (CPTi) coating referred to as a Particle Sintered Foam (PSF) coating with an overlying Peri-Apatite coating (precipitated calcium phosphate coating). The Peri-Apating process is identical to that cleared in K971206 but to a greater thickness range of 35-75 microns. The Tritanium® CPTi PSF coating originally manufactured from material conforming to ASTM F67 was modified to conform to ASTM F1580. The dual coating (Peri-Apatite™ overlying the PSF coating) at
{1}------------------------------------------------
K101072 p 2/3
the thickest specification met the definition of a porous coating per 21 CFR 888.3358 with no additional claims over biologic fixation. The dome hole plugs are optional devices which are available to seal the Howmedica Osteonics Tritanium® Acetabular Shells. The plugs are to be threaded into the dome holes of the shell.
The shells will be available in sizes 44-72 mm outside diameter (OD) in 2 mm increments. The acetabular shells are forged from Ti-6Al-4V alloy per ASTM F136; the Tritanium® coating is fabricated from Commercially Pure Titanium per ASTM F1580; and the Peri-Apatite™ coating is a precipitated calcium phosphate coating per ASTM F-1609.
The Tritanium® Peri-Apatite™ Acetabular Shells are compatible with polyethylene Trident® inserts, constrained liners and modular dual mobility liners. The subject shells have an identical insert locking mechanism to that employed with compatible Tritanium acetabular shell system predicates. All shells are single use devices. The outer surface of the acetabular shell has the identical geometry as the predicate Tritanium® Acetabular Shell System determined substantially equivalent via 510(k) K081171.
Intended Use: The Tritanium® Peri-Apatite™ Acetabular Shell is a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function.
## Indications:
The indications for use of the total hip replacement prostheses include:
- 1. Painful, disabling joint disease of the hip resulting from: non-inflammatory degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis, or late stage avascular necrosis.
- Revision of previous failed femoral head replacement, shell arthroplasty or other procedure. 2.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are 3. less likely to achieve satisfactory results.
- Where bone stock is of poor bone quality or inadequate for other reconstructive techniques as 4. indicated by deficiencies of the acetabulum.
Additional indications for use when using constrained liners: The device is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
Additional indications for use when using modular dual mobility (MDM) liners: The device is indicated for use as a component of a total hip prosthesis and include:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis;
- Correction of functional deformity; 3.
- Revision procedures where other treatments or devices have failed; 4.
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with న్. head involvement that are unmanageable using other techniques; and
- 6. Dislocation risks
The MDM liners are intended for cementless use only. The acetabular shell is intended for cementless use only.
{2}------------------------------------------------
K101072
r3/3
Summary of Technologies: The technology characteristics are the same basic principles as the predicates with minor differences occurring in the material of Tritanium underlying surface and the Peri-Apatite thickness.
Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. Non-clinical testing was provided as outlined in the FDA Guidance Document entitled "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement (28 April 1994)" for the underlying Tritanium surface which included additional testing on the CPT Tritanium conforming to ASTM F1580 in terms of static tensile, static shear and shear fatigue bond strength testing. The dual coating/surface was characterized per the FDA Guidance document entitled "510(k) Information Needed for Hydroxyapatite Coated Orthopaedic Implants (February 20, 1997)." All characterization parameters on the final dual coating/surface were provided. The NIST SRM 2910(a) material was used as a comparator for the Xray diffraction, dissolution rate and solubility product parameters. The dual coating/surface underwent additional characterization to demonstrate that the definition of porosity was met per 21 CFR 888.3358 at the thickest specification. Additionally, four-point bend fatiguing, abrasion, and fatigue strength of the subject acetabular shell were performed. Third body wear information was provided to address the theoretical risk. All of the observed results indicate that the subject Tritanium® Peri-Apatite™ Acetabular Shell System is substantially equivalent to devices currently cleared for marketing.
Clinical Testing: None provided as a basis for substantial equivalence.
Conclusion: The Tritanium® Peri-Apatite™ Acetabular Shell System is substantially equivalent to the predicate devices identified in this premarket notification.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name encircling a stylized eagle-like symbol. The symbol is composed of three angled lines that form the body and wings of the bird, with three curved lines below representing its legs or feet.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Howmedica Osteonics Corp. % Ms. Karen Ariemma, RAC Project Manager, Regulatory Affairs/ Regulatory Compliance 325 Corporate Drive Mahwah, New Jersey 07430
APR 1 1 2011
Re: K101072
Trade/Device Name: Tritanium® Peri-Apatite™ Acetabular Shell System (Tritanium® Peri-Apatite™ Acetabular Solid-backed Shell and Tritanium® Peri-Apatite™ Acetabular Cluster Screw-hole Shell) Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MEH, LZO, KWZ, LWJ, JDI Dated: March 31, 2011 Received: April 04, 2011
Dear Ms. Ariemma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Karen Ariemma, RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): K101072
### Device Name: Tritanium® Peri-Apatite™ Acetabular Shell System
Indications for Use:
The indications for use of the total hip replacement prostheses include:
- Painful, disabling joint disease of the hip resulting from: non-inflammatory degenerative arthritis, ﺘﺴ rheumatoid arthritis, post-traumatic arthritis, or late stage avascular necrosis.
- Revision of previous failed femoral head replacement, shell arthroplasty or other procedure. 2.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less 3. likely to achieve satisfactory results.
- Where bone stock is of poor bone quality or inadequate for other reconstructive techniques as 4. indicated by deficiencies of the acetabulum.
Additional indications for use when using constrained liners: The device is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
Additional indications for use when using modular dual mobility (MDM) liners: The device is indicated for use as a component of a total hip prosthesis and include:
- Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 1.
- Rheumatoid arthritis; 2.
- Correction of functional deformity; 3.
- Revision procedures where other treatments or devices have failed; 4.
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with న్.
- head involvement that are unmanageable using other techniques; and
- Dislocation risks 6.
The MDM liners are intended for cementless use only. The acetabular shell is intended for cementless use only.
Over-The-Counter Use Prescription Use . X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101072
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.