K241933 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OHS · Sep 27, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241933
Device Name
HIGHERDOSE Red and Infrared Light Mask (MK66-L)
Applicant
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Sep 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HIGHERDOSE Red And Infrared Light Mask (Model: MK66-L) is an over the counter device that is intended for the use in the treatment of full-face wrinkles.
Device Story
The HIGHERDOSE Red and Infrared Light Mask (MK66-L) is a wearable, home-use phototherapy device. It utilizes light-emitting diodes (LEDs) to deliver two specific wavelengths: red (630nm) and near-infrared (830nm). The device is constructed with silicone surfaces and is secured to the face via head straps. It is powered by an internal rechargeable lithium battery, charged via an external adapter; the device cannot operate while charging. The device incorporates a timer and software to control treatment duration. It is intended for consumer self-use to stimulate skin rejuvenation processes for the treatment of full-face wrinkles. The device does not perform measurements or provide diagnostic conclusions. Users apply the mask to the face to deliver light energy to the skin surface.
Clinical Evidence
No clinical data. The safety and effectiveness of the device were supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic disturbances (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), non-laser light source safety (IEC 60601-2-57), photobiological safety (IEC 62471), shelf-life testing, and software verification/validation.
Indicated for the treatment of full-face wrinkles in adult patients using an over-the-counter device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 27, 2024
Shenzhen Kaiyan Medical Equipment Co., Ltd Alain Dijkstra Regulatory Affairs Engineer Building #3 and Building #5, 40th of Fuxin Street Huaide Community Fuyong Town, Baoan District Shenzhen, Guangdong 518103 China
Re: K241933
Trade/Device Name: HIGHERDOSE Red and Infrared Light Mask (MK66-L) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: June 12, 2024 Received: July 2, 2024
Dear Alain Dijkstra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
# K241933 - Alain Dijkstra
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE - ○ Date: 2024.09.27
23:22:48 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241933
Device Name
HIGHERDOSE RED AND INFRARED LIGHT MASK (MK66-L)
Indications for Use (Describe)
The HIGHERDOSE Red And Infrared Light Mask (Model: MK66-L) is an over the counter device that is intended for the use in the treatment of full-face wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K241933
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
Sponsor Name: SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD Establishment Registration Number: 3011644607 Address: Building #3, and Building#5 40th of Fuxin street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China Contact Person (including title): Alain Dijkstra Tel: +86 755 82129361 Fax: +86 755 25024651 E-mail: alaindijkstra@kaiyanmedical.com
### Distributor
Company Name: HIGHERDOSE LLC Address: 42 Broadway, 12th Floor, #212, NewYork, NY 10004 Establishment Registration Number: 3019734322 Contact Person: Sumish Khadka E-mail: sumish@higherdose.com
#### Manufacturer:
Manufacturer Name: SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD Address: Building #3, and Building#5 40th of Fuxin street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China Contact Person (including title): Alain Dijkstra Tel: +86 755 82129361 Fax: +86 755 25024651 E-mail: alaindijkstra@kaiyanmedical.com
#### Application Correspondent:
Contact Person: Alain Dijkstra Company: SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD Address: Building #3, and Building#5 40th of Fuxin street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China Tel: +86 755 82129361
{5}------------------------------------------------
Fax: +86 755 25024651 E-mail: regulation@kaiyanmedical.com
### 2. Date of the summary prepared: September 26, 2024
### 3.Subject Device Information
Classification Name: Light Based Over-The Counter Wrinkle Reduction (OHS) Trade Name: HIGHERDOSE Red And Infrared Light Mask Model Name: MK66-L Review Panel: General & Plastic Surgery Product Code: OHS Regulation Number: 878.4810 Regulatory Class: II
### 4. Predicate Device Information
### Predicate Device 1 Information
Sponsor: Light Tree Ventures Europe B.V Trade Name: LED Light Therapy Mask Classification Name: Light based over the counter wrinkle reduction 510(K) Number: K221775 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Requlation Class: II
### Predicate Device 2 Information
Sponsor: Harpar Grace International Trade Name: Shani Darden LED light therapy mask Classification Name: Over-the-counter powered light based laser for acne 510(K) Number: K214103 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Regulation Class: II
### Predicate Device 3 Information
Sponsor: ISMART Marketing Svcs Ltd Trade Name: FaceLITE
{6}------------------------------------------------
Classification Name: Light based over the counter wrinkle reduction 510(K) Number: K191629 Review Panel: General & Plastic Surgery Product Code: OHS Regulation Number: 878.4810 Regulation Class: II
### Predicate Device 4 Information
Sponsor: SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD Trade Name: Aduro Light Beauty Mask Classification Name: Light based over the counter wrinkle reduction 510(K) Number: K202390 Review Panel: General & Plastic Surgery Product Code: OHS,OLP Regulation Number: 878.4810 Requlation Class: II
### 5. Device Description
The HigherDose Red and Infrared Light Mask is a home wearable light-emitting diode phototherapy device with two proven wavelengths of light 630nm and 830nm(near infrared light ) for treatment of fullface wrinkles,
The face mask combine Red (630nm)and near infrared light (830nm).Harnessed at the perfect measurement to rejuvenate the skin, these two wavelengths work below the skin's surface to stimulate the natural rejuvenation process ,creating a complexion that is visibly growing and noticeably healthier looking.
The mask is worn on the face and is held in place by head straps , The mask compromises of 2 surfaces. An inner surface that contacts the skin and an outer surfaces are constructed of silicone.
The controller contains a rechargeable Lithium battery, the power supply (adaptor) is used to charge the Lithium battery and is connected to a suitable mains outlet via a 2 or 3 pin input socket and wall plug. The HIGHERDOSE Red And Infrared Light Mask cannot be operated while charging.
The device is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The device does not require checks on the light output as the LEDs do not dim with age to any practical extent.
{7}------------------------------------------------
### 6. Intended Use / Indications for Use
For MK-66L
The HIGHERDOSE Red And Infrared Light Mask (Model: MK66-L) is an over the counter device that is intended for the use in the treatment of full-face wrinkles.
### 7. Comparison to predicate device and conclusion
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of | Subject Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Remark |
|---------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Company | HigherDOSE LLC | Light Tree<br>Ventures<br>Europe B.V. | Harpar Grace<br>International | ISMART<br>Marketing<br>Svcs Ltd | Shenzhen Kaiyan<br>Medical<br>Equipment Co.,<br>Ltd | -- |
| Trade Name | HigherDose Light<br>Face Mask | LED Light<br>Therapy<br>Mask | Shani Darden<br>LED light<br>therapy mask | FaceLITE | Aduro Light<br>Beauty Mask | -- |
| Classification<br>Name | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction | Light Based<br>Over The<br>Counter<br>Wrinkle<br>Reduction | Over-the-<br>counter<br>powered light<br>based laser<br>for acne | Light Based<br>Over The<br>Counter<br>Wrinkle<br>Reduction | Light Based<br>Over-The-Counter<br>Wrinkle Reduction | -- |
| 510(k)<br>Number | TBD | K221775 | K214103 | K191629 | K202390 | -- |
| Product<br>Code | OHS | OHS, OLP | OHS, OLP | OHS | OHS,OLP | Same |
| FDA Device<br>Classification | Class II | Class II | Class II | Class II | Class II | Same |
| Use | Over the Counter | Over the<br>Counter | Over the<br>counter | Over the<br>counter | Over the counter | Same |
| Intended<br>Use /<br>Indications<br>for Use | The<br>HIGHERDOSE<br>Red And Infrared<br>Light Mask<br>(Model: MK66-L)<br>is an over the | The LED<br>Light Therapy<br>Mask<br>(Models: MK-<br>78, MK-04) is<br>an over the | The Shani<br>Darden LED<br>light therapy<br>mask is an<br>over-the-<br>counter | The faceLITE<br>LED mask is<br>an over the<br>counter device<br>that is<br>intended for | The Aduro Light<br>Beauty Mask<br>(Model: MK-660,<br>MK-66USBO, MK-<br>020) emits energy<br>in the red and<br>blue region of the<br>spectrum, | Same |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Remark |
| | counter device<br>that is intended for<br>the use in the<br>treatment of full-<br>face wrinkles. | counter<br>device that is<br>intended for<br>the use in the<br>treatment of<br>full-face<br>wrinkles. | device<br>intended to<br>emit energy in<br>the red and<br>blue region of<br>the light<br>spectrum,<br>specifically<br>indicated to<br>treat mild to<br>moderate<br>acne vulgaris<br>of the face.<br>The Shani<br>Darden LED<br>light therapy<br>mask is an<br>over-the-<br>counter<br>device<br>intended to<br>emit energy in<br>the red and<br>Near Infra-red<br>spectrum and<br>is intended for<br>the use in the<br>treatment of<br>full-face<br>wrinkles. | the use in the<br>treatment of<br>full-face<br>wrinkles. | indicated to treat<br>full face wrinkles<br>and/or mild to<br>moderate acne. | |
| | | The LED<br>Light Therapy<br>Mask<br>(Models:<br>MK66-H,<br>EL00003) is<br>an over the<br>counter<br>device<br>intended to<br>emit energy<br>in the red and<br>blue region of<br>the light<br>spectrum,<br>specifically<br>indicated to<br>treat mild to<br>moderate<br>acne vulgaris<br>of the face.<br>The LED<br>Light Therapy<br>Mask<br>(Models:<br>MK66-H, | | | | |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Remark |
| | | EL00003) is<br>an over the<br>counter<br>device<br>intended to<br>emit energy<br>in the red and<br>Near Infra-red<br>spectrum and<br>is intended<br>for the use in<br>the treatment<br>of full-face<br>wrinkles. | | | | |
| Intended<br>location of<br>use | Face | Face | Face | Face | Face | same |
| Energy Type | Light emitting<br>diodes | Light emitting<br>diodes | Light emitting<br>diodes | Light emitting<br>diodes | Light emitting<br>diodes | Same |
| Wavelengths | Red:<br>630nm±10nm<br>NIR:<br>830nm±10nm | 1.MK-78, MK-<br>04:<br>Red: 630±5<br>nm<br>NIR: 830nm<br>2.MK66-H,<br>EL00003:<br>Blue: 415nm,<br>Red: 630nm<br>+/- 5nm,<br>NIR: 830nm | Blue: 415nm<br>+/- 10nm,<br>Red: 630nm<br>+/- 10nm,<br>NIR: 830nm<br>+/-10nm. | Red:<br>630nm±10nm<br>NIR:<br>830nm±10nm | Red:640nm±10nm<br>Blue:465nm±10nm | Same |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Remark |
| Total<br>Intensity<br>(mW/cm²) | Red 20mw/cm²,<br>Infrared Red :<br>30mw/cm² | 1.MK-78: 20-<br>30 mw/cm²<br>2.MK-04:<br>30mw/cm²<br>3.MK66-H,<br>EL00003:<br>(1)Blue/Red<br>44 mw/cm²<br>(2)Red/NIR<br>30 mw/cm² | Blue/Red 44<br>mW/cm²<br><br>Red/NIR 29<br>mW/cm² | 30mw/cm²<br>total | 30mw/cm² | same |
| | | Treatment<br>Time | 10 minutes | 10 minutes | 10 minutes | 600 seconds |
| Dose | Red 12, Infrared<br>Red: 18 J/cm² | 1.MK-78: 12-<br>18 J/cm²<br>2.MK-04: 18<br>J/cm²<br>3.MK66-H,<br>EL00003:<br>(1)Blue/Red:<br>26.4 J/cm²<br>(2)Red/NIR:<br>18 J/cm² | Blue 16.8<br>J/cm²<br>Red 9.6 J/cm²<br><br>Red 11 J/cm²<br>NIR 7 J/cm² | 540 J/cm²<br>(cumulative<br>does) | | Same |
| | | Treatment<br>protocol | 3-5 times /week | Acne: 4 x<br>weekly, 6<br>weeks;<br>Wrinkles: 5 x<br>weekly, 6<br>weeks | Acne: 4 x<br>weekly, 6<br>weeks;<br>Wrinkles: 5 x<br>weekly, 6<br>weeks. | 5 x weekly, 6<br>weeks |
| Software<br>controller | Device uses a<br>timer and software | Device uses<br>a timer and | Device uses a<br>timer and | Yes | Device uses a<br>timer and software | Same |
| Elements of<br>Comparison | Subject Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Remark |
| | to control<br>treatment duration | software to<br>control<br>treatment<br>duration | software to<br>control<br>treatment<br>duration | | to control<br>treatment duration | |
| Power<br>supply | Rechargeable<br>Lithium battery | Rechargeable<br>Lithium<br>battery | 100-240V | Rechargeable<br>Lithium battery | Rechargeable<br>Lithium battery | Same |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
### Final Conclusion:
The subject device is the same or similar to the legally marketed predicate device K221775, K214103, K191629 and K202390.
### 8. Test Summary
### 8.1 Summary of Non-Clinical Performance Testing
### 1) Performance Testing Summary
The HIGHERDOSE Red And Infrared Light Mask (Model: MK-66L) has been evaluated the safety and performance by lab bench testing as following:
| Title of the test | Device<br>Description<br>/Sample<br>Size | Test<br>Method/Ap<br>plicable<br>Standards | Acceptance<br>criteria | Unexpec<br>ted<br>Results/<br>Significa<br>nt<br>Deviatio<br>ns | Test<br>results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-----------------|
| General<br>requirements for<br>basic safety and<br>essential<br>performance | The test<br>sample is<br>the final,<br>finished<br>product. | IEC 60601-<br>1:2005/AMD<br>1:2012/AMD<br>2:2020 | The test is<br>carried out<br>under the test<br>method<br>specified in the<br>standard, and<br>the test result is<br>within the test<br>acceptance<br>range of the<br>standard. | NA | Pass |
| Electromagnetic<br>disturbances | The<br>test<br>sample is<br>the<br>final,<br>finished<br>product. | IEC 60601-<br>1-<br>2:2014+A1:<br>2020 | No degradation<br>of performance<br>was<br>found<br>during test or<br>Lower<br>than<br>limits<br>of<br>measurement | NA | Pass |
| Requirements<br>for medical<br>electrical | The<br>test<br>sample is<br>the<br>final | IEC 60601-<br>1-<br>11:2015/AM | The<br>device<br>operates<br>normally and | NA | Pass |
| equipment and<br>medical<br>electrical<br>systems used in<br>the home<br>healthcare<br>environment. | finished<br>product. | D1:2020 | can provide<br>basic safety<br>and essential<br>performance. | | |
| Particular<br>Requirements<br>for The Basic<br>Safety And<br>Essential<br>Performance Of<br>Non-Laser Light<br>Source<br>Equipment<br>Intended For<br>Therapeutic,<br>Diagnostic,<br>Monitoring And<br>Cosmetic/Aesth<br>etic Use | The test<br>sample is<br>the final,<br>finished<br>product. | IEC 60601-<br>2-57:2011 | The test is<br>carried out<br>under the test<br>method<br>specified in the<br>standard, and<br>the test result is<br>within the test<br>acceptance<br>range of the<br>standard. | NA | Pass |
| Photobiological<br>safety of lamps<br>and lamp<br>systems. | The test<br>sample is<br>the final,<br>finished<br>product. | IEC<br>62471:2006 | The test is<br>carried out<br>under the test<br>method<br>specified in the<br>standard, and<br>the test result is<br>within the test<br>acceptance<br>range of the<br>standard. | NA | Pass |
| Shelf Life Test | The test<br>sample is<br>the final,<br>finished<br>product. | The Shelf Life<br>Test Report<br>performs the<br>following tests<br>on the product<br>before and<br>after<br>accelerated<br>aging, and<br>after use:<br>Performance<br>test;<br>Power Density<br>Test;<br>Leakage<br>current test. | The device can<br>meet the<br>requirement of the<br>performance test,<br>Power Density<br>test and Leakage<br>current test. | NA | Pass |
{12}------------------------------------------------
{13}------------------------------------------------
### 2) Biocompatibility testing
The component materials of the subject device are identical to the corresponding component materials of the previously cleared devices ( K202390) in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents).
Here is no change in biocompatibility since the previously cleared devices. Therefore, based on this information, the subject device can comply with the biocompatibility requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-10 (Irritation).
### 3) Usability Testing
Usability testing was conducted on the HIGHERDOSE Red And Infrared Light Mask, the device complies with IEC 62366-1 and IEC 60601-1-6.
### 4) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
#### 8.2 Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
#### 9. Final Conclusion:
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K221775, K214103, K191629 and K202390.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.