K191629 · Ismart Marketing Svcs, Ltd. · OHS · Sep 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191629
Device Name
faceLITE
Applicant
Ismart Marketing Svcs, Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Sep 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The faceLITE™ LED mask emits light in the red and near infra-red (NIR) region of the light spectrum and is indicated for the treatment of full face wrinkles.
Device Story
The faceLITE LED mask is a wearable, home-use phototherapy device for treating facial wrinkles. It consists of a flexible silicone mask containing LEDs, a controller, and a rechargeable lithium-ion battery. The device emits red and near-infrared (NIR) light to the face via a non-thermal photobiomodulation mechanism. The user operates the device by wearing the mask, secured with an adjustable strap, and using the controller to initiate a 10-minute treatment session. The controller manages the LED power and treatment duration. The device is intended for self-use by consumers. It does not perform measurements or diagnostic conclusions. The output is light energy, which the user receives passively during the treatment session. The device is not operational while charging.
Clinical Evidence
No clinical data was required. The submission relied on non-clinical performance testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993). Label comprehension testing was conducted with 32 subjects (mean age 35, 59% female) to ensure lay-user safety and effectiveness.
Technological Characteristics
Flexible silicone mask (Methyl vinyl silicone rubber) with integrated LEDs. Controller made of Polycarbonate (PC1100). Emits red (630nm +/- 10nm) and NIR (830nm +/- 10nm) light. Powered by a rechargeable lithium-ion polymer battery. Connectivity via micro USB A-C for charging. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, IEC 62133, ISO 10993-1/5/10, EN 62304, ISO 14971.
Indications for Use
Indicated for the treatment of full-face wrinkles in adult patients using an over-the-counter, home-use device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 12, 2019
iSMART Marketing SVCS Ltd Susan D'Arcy Director/Owner 129 Green Lanes, Sutton Coldfield Birmingham, B735LT GB
Re: K191629
Trade/Device Name: faceLITE Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: June 12, 2019 Received: June 19, 2019
Dear Susan D'Arcy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191629
Device Name faceLITE LED mask
Indications for Use (Describe)
The faceLITE LED mask is an over the counter device that is intended for the use in the treatment of full-face wrinkles.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------------------------|----------------------------------------------|
| <div> <object data="checkbox-checked.svg" height="15" type="image/svg+xml" width="15"></object> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for FaceLITE. The text "faceLITE" is in a sans-serif font, with the "face" portion in a rounded, lowercase style and the "LITE" portion in uppercase. Below the main text, in a smaller font, it reads "FROM THE NATURAL FACELIFT COMPANY".
## Section 5: 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).
Submitter's Name: Ismart Marketing SVCS Ltd.
Submitter's Address: 129 Green Lanes, Sutton Coldfield, Birmingham B735LT
SBDN: SBD195544
Contact Person: Susan D'Arcy iSMART Marketing Services, 129 Green Lanes, Wylde Green, Birmingham B73 5LT. United Kingdom. Telephone +44 (0) 7880313315
Date Prepared: June 12th, 2019
Date Amended: September 9th, 2019
Device Trade Name: faceLITE LED mask
#### Device Classification Information:
| Regulation<br>Number | Device Classification<br>name | Device<br>Class | Product<br>Code | Classification<br>Panel | Type |
|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|-----------------|-----------------|------------------------------|------------------------|
| 21 CFR 878.4810<br>Laser surgical<br>instrument for use<br>in general and<br>plastic surgery and<br>in dermatology. | Light Based Over the<br>Counter Wrinkle<br>Reduction | Class 2 | OHS | General &<br>Plastic Surgery | Traditional<br>510 (k) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "faceLITE" in a simple, modern font. The word "face" is in a rounded, sans-serif typeface, while "LITE" is in a more angular, sans-serif font. Below the logo, in smaller letters, is the tagline "FROM THE NATURAL FACELIFT COMPANY."
#### 5.1. Device Description
#### 5.1.1. Overall Description
The faceLite consists of;
- 1. Silicon flexible face mask
- 2. Controller
- 3. Power supply and country specific adaptors
- 4. USB C to USB A connector
- 5. Head straps
Image /page/4/Figure/10 description: The image shows a white face mask with LED lights on top of a black box. The mask has cutouts for the eyes, nose, and mouth. There is a white controller attached to the mask with three buttons. The image also shows a white power adapter and a black strap. The objects in the image are labeled with numbers 1-5.
faceLite is a home use wearable LED phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment cosmetic indications including facial wrinkles.
The system consists of a flexible silicon mask that contains light emitting diodes (LEDs) and a controller. The LEDs generate the light. The mask is worn on the face and is held in place by an adjustable Velcro strap. The mask compromises of 2 surfaces. An inner surface that contacts the skin and an outer surface. Both surfaces are constructed of the same silicone (Methyl vinyl silicone rubber, Genvan GA 3 series GA9 series).
The inner surface contacts with the user's skin. Contact is restricted to normal intact skin and only for a duration of 10 minutes (treatment time). Methyl vinyl silicone rubber GENVAN GA 3 series GA9 series has been tested to those tests specified under ISO 10993-5:2009 tests for in vitro cytotoxicity and ISO10993-10:2010 Tests for irritation and skin sensitization
The controller (Polycarbonate PC1100) switches the LEDs ON/OFF and controls power to the mask. The controller contains a rechargeable Lithium ion polymer battery. The controller
{5}------------------------------------------------
# K191629
Image /page/5/Picture/1 description: The image shows the logo for FaceLite, a company that specializes in natural facelifts. The logo is in black and white, with the word "faceLITE" in a sans-serif font. Below the company name is the tagline "FROM THE NATURAL FACELIFT COMPANY" in a smaller font. The logo is simple and modern, and it conveys the company's focus on natural beauty.
uses a visible display comprising of 3 micro LEDs to show the user the battery charge status of the device.
The power supply (SK01T-0500100Z) is used to charge the Lithium battery and is connected to a suitable mains outlet via a 2 or 3 pin input socket and wall plug. The power cable is connected to the controller by a standard micro USB A-C connector. The faceLite cannot be operated while charging.
The equipment is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The equipment does not require checks on the light output as the LEDs do not dim with age to any practical extent.
The faceLite Light emitting diode (LED) system emits light energy in the red and near infrared (NIR) region of the light spectrum and is intended to treat facial wrinkles through a nonthermal mechanism called Photobiomodulation.
### 5.2. Intended Use
The faceLITE™ LED mask emits light in the red and near infra-red (NIR) region of the light spectrum and is indicated for the treatment of full face wrinkles.
### 5.3. Substantial Equivalence
The faceLITE LED mask is substantially equivalent to the LG BEAUTY LED MASK (K170984).
The faceLITE LED Mask is predicated against the LG Beauty LED mask because both masks are LED phototherapy devices intended to emit light in the red and near infra-red region of the light spectrum and indicated for the treatment of full-face wrinkles.
| Property | faceLITE LED<br>mask | K170984<br>LG Beauty LED<br>mask | Significant<br>differences |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Device Manufacturer | iSMART Marketing<br>SVCS Ltd | LG Electronics, Inc. | na |
| Device Trade Name | faceLITE™ | LG Beauty™ | na |
| 510(K) Number | - | K170984 | na |
| Device Common<br>Name | faceLITE™ | LG Beauty™ | na |
| Property | faceLITE LED<br>mask | K170984<br>LG Beauty LED<br>mask | Significant<br>differences |
| Device Classification<br>name | Light Based Over the<br>Counter Wrinkle<br>Reduction | Light Based Over<br>the Counter Wrinkle<br>Reduction | Identical |
| Device Product Code | OHS | OHS | Identical |
| Regulation Number | 878.4810<br>Laser surgical<br>instrument for use in<br>general and plastic<br>surgery and in<br>dermatology. | 878.4810<br>Laser surgical<br>instrument for use in<br>general and plastic<br>surgery and in<br>dermatology. | Identical |
| FDA Device<br>Classification | Class II | Class II | Identical |
| Use | Over the Counter | Over the Counter | Identical |
| Intended use and<br>Indications | The faceLITE LED<br>mask is an over the<br>counter device that is<br>intended for the use<br>in<br>the treatment of full-<br>face wrinkles. | The LG BEAUTY<br>LED MASK is an<br>over the counter<br>device that is<br>intended for the use<br>in<br>the treatment of full-<br>face wrinkles. | Identical |
| Intended Location of<br>Use | Face | Face | Identical |
| Energy Type | Light emitting diodes | Light emitting<br>diodes | Identical |
| Peak Wavelength<br>(FWHM) | Red: 630nm+/-10nm.<br>NIR: 830nm+/-10nm | Red: 637nm<br>NIR:854nm | The output of the<br>FaceLITE™ LED<br>mask is within the<br>predicate's<br>wavelength |
| Intensity | 30mw/cm² total | 25 mW/cm² total | Similar |
| Treatment time | 600 seconds | 540 seconds | Similar |
| Treatment protocol<br>(Treatment time) | 5 x weekly, 6 weeks | 5 x weekly, 8 weeks | Similar |
| Property | faceLITE LED<br>mask | K170984<br>LG Beauty LED<br>mask | Significant<br>differences |
| Cumulative dose | 540J/cm² | 540J/cm² | Identical |
| Timers | Device uses a timer<br>and software to<br>control treatment<br>duration. | Device uses a timer<br>and software to<br>control treatment<br>duration | Identical |
| Software Controlled | Yes | Yes | Identical |
### 5.3.1 Summary of Substantial Equivalence
{6}------------------------------------------------
MPANY FROM THE NATURAL FAC
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for "face LITE" in a simple, sans-serif font. The word "face" is written in a rounded, lowercase style, while "LITE" is in uppercase and more angular. Below the main logo, in smaller font, it reads "FROM THE NATURAL FACELIFT COMPANY".
FROM THE NATURAL FACELIFT COMPANY
### 5.3.2. Substantial Equivalency and Comparison of Technological Similarities & Differences
From the comparative table above, the faceLITE LED Mask demonstrates equivalence to the LG Beauty LED mask. The key similarities are;
- The intended use of the faceLITE LED mask is equivalent to the listed predicate; 1. an over the counter device that is intended for the use in the treatment of full-face wrinkles.
- ii. Both devices are phototherapy units utilizing light emitting diodes that emit in the red and near infra-red-light spectrum.
- iii. Both devices have a similar power density and deliver an identical cumulative dose and have similar treatment protocols.
#### 5.4. Non- clinical performance testing
The faceLITE LED system has been thoroughly evaluated for electrical safety and performance and has been found to conform to the following standards;
IEC/EN 60601-1: 2006 + A12:2014 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
IEC/EN 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
EN 60601-1-6: 2010. AMD12013 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability IEC 60601-1-11:2015 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for
{8}------------------------------------------------
# K191629
Image /page/8/Picture/1 description: The image shows the logo for "faceLITE", a company that specializes in natural facelifts. The logo is in black and white, with the word "faceLITE" in a sans-serif font. Below the company name is the tagline "FROM THE NATURAL FACELIFT COMPANY" in a smaller font. The logo is simple and modern, and it conveys the company's focus on natural beauty.
medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC60601-2-57 1st ed:2011 Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use IEC 62471:2006 Photobiological safety of lamps and lamp systems.
IEC62133:2012 2nd ed. Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
ISO 10993-1: 2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management.
EN ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
EN ISO 10993-10:2010 ISO 10993-10 Third Edition 2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
EN 62304: 2006 (ed. 1.0) Medical Device Software - Software Life Cycle Processes.
ISO 14971: 2012 Medical Devices - Application of Risk Management to Medical Devices
In addition to the aforementioned standards the faceLITE labelling was subject to label comprehension testing. With respect to medical devices available without the intervention of a physician, termed 'Over the Counter' (OTC), the labelling and instructions for use (IFU or User Guide)) must convey to the consumer enough information to allow them to safely operate the device to attain its purpose. To determine the effectiveness of labelling pertaining to a medical device, the labelling and device was tested with an appropriate random sample of users.
A study was conducted and is appended to this section demonstrating comprehension of the faceLITE labelling. 32 subjects took part in the study. The average age of the participants was 35 years of age. Fifty nine percent (59%) of subjects were female (13:19, M:F ratio). In terms of ethnicity 18 subjects classed themselves as Caucasian and 14 non-Caucasian (Black: 4. Hispanic: 6. Indian: 2. Asian: 2). Seven (22%) of subjects indicated that English was their second language. REALM reading tests were conducted.
The comprehension and use test demonstrated that the faceLITE labelling could be used by lay persons to safely and effectively operate the device to attain its intended use and purpose.
{9}------------------------------------------------
# K191629
Image /page/9/Picture/1 description: The image shows the logo for "faceLITE" in a simple, sans-serif font. The word "face" is in a bolder font than "LITE". Below the logo, in a smaller font, is the text "FROM THE NATURAL FACELIFT COMPANY".
#### Clinical Performance
Since the faceLITE LED mask raises no new questions in terms of safety and efficacy, clinical data is not required.
#### Statement of Substantial Equivalence:
513(i) of the FD&C Act (21 U.S.C. 360c(i) states that for substantial equivalence a proposed device is required to have the same intended use and similar technological characteristics as the predicate device. Where there are differences in technological characteristics, these can be negated by appropriate clinical or scientific data demonstrating that the proposed device is as safe and effective as the predicate device, and that the proposed device does not raise any different questions of safety and effectiveness than the predicate device for the same intended use. iSMART Marketing SVCS Ltd has demonstrated that the faceLITE LED mask has an identical intended use, have the same generic classification and basic principles and technologies as the predicate device. Both devices utilize red and NIR wavelengths of light with similar power densities and identical cumulative dose. iSMART Marketing SVCS Ltd has conducted non-clinical performance testing
applicable to those general controls deemed necessary by the agency for this product classification and has determined that the faceLITE LED mask raises no new questions relating to safety and therefore has demonstrated that the faceLITE LED mask is substantially equivalent to the referenced predicate LG Beauty mask (K170984).
June 12th, 2019
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.