K241854 · Anhui Jbh Medical Apparatus Co., Ltd. · ITI · Oct 4, 2024 · Physical Medicine
Device Facts
Record ID
K241854
Device Name
Power Wheelchair
Applicant
Anhui Jbh Medical Apparatus Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Oct 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
The D25 power wheelchair is a four-wheeled, motor-driven personal mobility device for indoor and outdoor use. It features an aluminum alloy frame, two electric motors, an electromagnetic brake, and a rechargeable 24V Li-ion battery. The passenger controls speed and direction via a joystick on a tiller console. The device is intended for disabled or elderly individuals with mobility limitations. It provides mobility on flat or obstacle-laden surfaces but is not for road or highway use. The electromagnetic brake engages automatically when the device is not in use. The device is operated by the user; it does not require clinical supervision. The output is physical movement, which directly assists the user in navigating their environment, thereby improving personal mobility and independence.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including static/dynamic stability, brake effectiveness, energy consumption, obstacle climbing, and electromagnetic compatibility (IEC 60601-1-2). Biocompatibility was verified per ISO 10993.
Technological Characteristics
Aluminum alloy frame; 24VDC brush motors (200W); 24VDC 11.6Ah Li-ion battery; PU solid tires; electromagnetic brake system. Dimensions: 1030x610x1010 mm. Connectivity: None. Sterilization: N/A. Standards: ISO 7176 series, ISO 10993, IEC 60601-1-2.
Indications for Use
Indicated for disabled or elderly persons with mobility difficulties who are limited to a seated position.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 4, 2024
Anhui JBH Medical Apparatus Co., Ltd. % Ivv Wang Technical Manager Shanghai SUNGO Management Consulting Co., Ltd. Room 1 401, Dongfang Building, 1500# Century Ave. Shanghai. 200122 China
Re: K241854
Trade/Device Name: Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: September 6, 2024 Received: September 6, 2024
### Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241854
Device Name
Power Wheelchair
Indications for Use (Describe)
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### l. SUBMITTER
Name: Anhui JBH Medical Apparatus Co., Ltd.
Address: No. 116 Qicang Road, Industrial District, Ming Guang City, Chuzhou, Anhui, China
Name of contact person: Wei Yuanwei
Telephone: +86-0550-8108866
Fax: +86-0550-8106688
Email: +86-0550-8106688
Date prepared: 2024-6-26
#### II. Device
Device trade name: Power wheelchair Model: D25 Classification name: Powered wheelchair Regulation class: 2 Regulation number: 21CFR 890.3860 Panel: Physical Medicine Product code: ITI
### III. Predicate device
K113463 Power Wheelchair, PL001
SUZHOU KD Medical Appliance Co. Ltd.
### IV. Device description
The D25 power wheelchair is a four wheeled personal mobile device with a complementary chair support system that is powered by two motors. The traveling speed is controlled by the motor, and the traveling direction is controlled by the passenger. This product is a device suitable for disabled people with mobility difficulties and elderly people and it is intended to provide mobility to a disabled or elderly person limited to a seated position. The power wheelchair can be travelled on flat and obstacle ground surface, and direction and speed of
{5}------------------------------------------------
the wheelchair can be controlled by the passenger's hand with the help of the joystick. The device can be used to provide indoor and outdoor mobility at a certain distance but not allowed to be travelled on the road or highway.
The wheelchair (Model D25) has a base with aluminum alloy frame, two front wheels, two rear wheels, a seat, an adjustable steering column, a tiller console, two electric motors, an electromagnetic brake, 1 rechargeable Lithium-lon Battery with an off-board charger. The movement of the wheelchair is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the wheelchair is not in use and the brake cannot be used manually.
#### V. Indication for use
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Anhui JBH Medical<br>Apparatus Co., Ltd. | SUZHOU KD Medical<br>Appliance Co. Ltd. | / |
| Proprietary name, model | Power Wheelchair, D25 | power wheelchair, PL001 | / |
| 510(k) number | K241854 | K113463 | / |
| Device classification name | Class II | Class II | Same |
| Classification regulations | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| Product code | ITI | ITI | Same |
| Similarities | | | |
| Indication for use | The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | They are motor driven indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | Same |
| Intended | disabled or elderly person | disabled or elderly person | Same |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| user | limited to a seated position | limited to a seated position | |
| Use condition | indoor and outdoor use | indoor and outdoor use | Same |
| Number of wheels | 4, including two front wheels and two rear wheels | 4, including two pivoting casters and two rear drive wheels | Same |
| Function of wheels | Front wheels: driven wheels suitable for rotation, acceleration, retrograde<br>Rear wheels: driving wheels to control the speed and direction | two pivoting casters: driven wheels suitable for rotation, acceleration, retrograde<br>two rear drive wheels: driving wheels to control the speed and direction | Same |
| Movement control method | By Joystick control | By Joystick control | Same |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | Same |
| Brake system | Intelligent electromagnetic brake system | Intelligent regenerative Electromagnetic brake | Same |
| Max speed forward | Up to 6 km/h (3.75 mph), continuously adjustable | Up to 6 km/h (3.75 mph), variable | Same |
| Maximum distance of travel on the fully charged battery | 20 km | 20 km | Same |
| Main frame material | aluminum alloy, front and rear folding | aluminum alloy, front and rear folding | Same |
| Armrest cushion | Polyurethane (PU) | PU | Same |
| Differences | | | |
| Overall dimensions (LxWxH) | 1030 × 610 × 1010 mm | 880 mm×570 mm×890 mm | larger size is designed for bearing more loading weight. All safety and performance have been validated with the maximum rated weight dummy. |
| Folded dimensions (LxWxH) | 610 × 380 × 840 mm | 720 mm×570 mm×400 mm | safety and performance have been validated with the maximum rated weight dummy. |
| Ground clearance | 120 mm | 80 mm | larger ground clearance provides |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Front wheel size/type | 7.9"x2" /PU Solid tire | 6"x 2"/PU Solid tire | larger size of driven wheel will provide better safety performance. |
| Rear wheel size/type | 12.4"×2.4" /PU Solid tire | 8" x 2.4"/PU Solid tire | better stability and safety performance |
| Battery | li-ion battery pack; rechargeable, 24 VDC 11.6Ah | Li-ion, Rechargeable; 24 VDC 20Ah | Same rated voltage with different power capacity will not affect the safety and performance of the subject device. |
| Max Speed backward | 0.8 m/s (2.88km/h) | 2.4 mph (3.84 km/h) | minor difference on max. backward speed will not cause different performance. lower speed will be more safety. |
| Minimum braking distance from maximum speed | Forward: 0.9 m | Forward:1.5 m (59") | subject device has shorter braking distance than the predicate device, such differences will not affect safety and performance of the device. |
| Minimum braking time from maximum speed | 0.6 second | Not available | The minimum braking time for our device is 0.6 second, which will ensure the device can be stopped on time as required, and such differences will not affect safety and performance of the device. |
| Max loading weight | 120kg (264 lbs) | 114 kg (251 lbs) | minor difference on loading weight will not cause different performance. more loading weight provide more convenient and stable performance for |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Maximum<br>safe<br>operational<br>incline<br>degree | 8 ° | 9 ° | minor difference on<br>safe operational<br>incline degree will not<br>cause new safety and<br>effectiveness<br>concerns. Both the<br>static and dynamic<br>stability under specific<br>inclining degree have<br>been evaluated<br>according to standard<br>ISO 7176 series. |
| Battery<br>charger | Off-board charger<br>Input: 100-240V, 50/60Hz,<br>1.5A<br>Output: 24 Vdc, 2A | Off-board, Automatic<br>Type<br>Input: 110-220 V / 50-60<br>Hz,<br>Output: 24 Vdc, 2A; | More wide range of<br>input voltage in the<br>device which will not<br>cause new safety and<br>effectiveness<br>concerns raised. |
| Motor | Brush motor; 24VDC;<br>200W; 2pcs | Brushless DC motor; 24<br>VDC; 180 W; 2 pcs | minor difference on<br>motor power will not<br>cause different<br>performance. larger<br>power will provide<br>more driving force, no<br>safety and<br>effectiveness<br>concerns raised. |
| Electronic<br>controller | brush power wheelchair<br>controller | Brushless dual-drive<br>rocker controller | Both of the control<br>systems are evaluated<br>according to standard<br>ISO 7176-14 and<br>software validation<br>requirement and there<br>are no new safety and<br>effectiveness<br>concerns due to the<br>difference. |
| Turning<br>Radius | 650 mm (25.5") | 31.5" (800 mm) | The minor difference<br>in the turning radius<br>will not raise any new<br>safety and<br>effectiveness |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Maximum obstacle climbing | 20 mm | 1.2" (30 mm) | Minor difference in the obstacle climbing will not impact the safety and effectiveness of the subject device. |
| seat cushion | Sponge foam covered Denim fabric | PU foam covered by nylon fabric cloth | different material used for parts in contact |
| back cushion | Sponge foam covered Denim fabric | PU foam covered by nylon fabric cloth | with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series. |
#### VI. Comparison of technological characteristics with the predicate device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
### VII. Summary of substantial equivalence discussion
The power wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2012, ISO 7176-13:1989, ISO 7176-14:2022, ISO 7176-15:1996, ISO 16840-10:2021, ISO 7176-21:2009, ISO 7176-22:2014, ISO10993-5:2009, ISO 10993-10:2021, ISO 10993-23: 2021, IEC60601-1-2: 2020.
The indications for use for both devices are the same. Mainframes of two devices are and the frame materials are the same aluminum alloy. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3. Turning radius, Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2 and ISO 7176-10. The biocompatibility of the subject device is evaluated according to
{10}------------------------------------------------
standard ISO 10993-1 and meet the requirements accordingly.
The flame-retardant test of the seat cushion/ backrest of both subject device and predicate device is carried out according to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
#### VIII. Summary of non-clinical testing
#### A Performance testing-bench
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- Risk Analysis developed in accordance with ISO 14971: 2019.
- Software validation
- . ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
- ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- . ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
- . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
{11}------------------------------------------------
- . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
- ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
- ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
- . ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method.
- . ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
- . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020.
#### A Biocompatibility of patient-contacting material
Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021) and irritation (ISO 10993-23: 2021). Details of the test summary see biocompatibility summary.
### IX. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
#### X. Conclusions
The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K113463.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.