Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III)

K241832 · Boston Aesthetics, Inc. · GEI · Feb 13, 2025 · General, Plastic Surgery

Device Facts

Record IDK241832
Device NameUnicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III)
ApplicantBoston Aesthetics, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 13, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation.

Device Story

Unicorn+ RF System consists of host, trolley, footswitch, power cord, and various handpieces/electrodes. Device delivers RF energy to tissue via bipolar or monopolar modes to achieve electrocoagulation, hemostasis, or therapeutic heating. Operated by clinicians in clinical settings; user controls parameters via touchscreen. System utilizes three electrode types (MicroRF, Artist, Pure+) acting at varying skin depths. Energy output is a function of power and duration. Device provides thermal effects for pain relief and circulation improvement. Clinical benefit derived from controlled thermal tissue interaction; safety verified through bench testing and thermal damage comparisons against predicates.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including biocompatibility (ISO 10993 series), electrical safety (IEC 60601-1, 60601-2-2), EMC (IEC 60601-1-2), software validation, and ex vivo thermal testing on tissue samples to confirm therapeutic temperature maintenance (40-45°C).

Technological Characteristics

Radiofrequency electrosurgical system; 1MHz frequency; 50W max output power. Bipolar and monopolar modes. Components: host, trolley, footswitch, handpieces, electrodes. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993. Electrical safety: IEC 60601-1, IEC 60601-2-2. Shelf-life: ASTM F1980-16.

Indications for Use

Indicated for patients requiring dermatologic and general surgical procedures for electrocoagulation and hemostasis, or relief of minor muscle aches, pain, muscle spasms, and temporary improvement of local blood circulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ February 13, 2025 Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that. Boston Aesthetics Inc Hongmei Cao General Manager 1521 Concord Pike Suite 201 Wilmington, Delaware 19803 Re: K241832 Trade/Device Name: Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: November 20, 2024 Received: January 17, 2025 Dear Hongmei Cao: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, James H. Digitally signed by James H. Jang -S Date: 2025.02.13 Jang -S 17:23:19-05'00' For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Submission Number (if known) K241832 Device Name Unicorn+ RF System (Unicorn+); Unicorn+ RF System (Unicorn+ I); Unicorn+ RF System (Unicorn+ II); Unicorn+ RF System (Unicorn+ III) ### Indications for Use (Describe) The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. ### General Information 1. | Submitter: | Boston Aesthetics INC<br>1521 Concord Pike Suite 201 Wilmington DE 19803.US | |---------------------------|------------------------------------------------------------------------------------------| | Contact Person: | Hongmei Cao<br>General Manager<br>Tel: 001 949-792-816<br>Email: bsnaesthetics@gmail.com | | Summary Preparation Date: | June 14,2024 | ### 2. Device Name and Code Trade/Common Name: Unicorn+ RF System Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Classification: Class II Product Code: GEI, PBX Regulation Number: 21 CFR 878.4400 Review Panel: General & Plastic Surgery ### Predicate Device 3. Predicate Device 1:Potenza, cleared under Traditional 510(k) K192545 Predicate Device 2: AGNES, cleared under Traditional 510(k) K160469 Predicate Device 2: Capenergy C Equipment RF System, cleared under Traditional 510(k) K222260 ### 4. Device Description Unicorn+ RF System is consist of a host, a trolley, a footswitch, a power cord, three kinds of handpieces and matched electrodes. #### Indications for Use 5. The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation. ### Technological Characteristics in Comparison to the Predicate Device 6. Although this subject device has three Predicate devices, they have same fundamental technology theory. Operators control host through touch screen, outputs different amounts of RF energy to handpieces and matched electrodes. These three kinds of electrodes(including invasive or non-invasive electrodes) act on different skin depth to achieve different treatment modes. The subject device has the same intended use and indications for use and the same fundamental scientific technology as the Predicate devices. The subject device design, technology, and the {5}------------------------------------------------ principles of operation are the same as the predicate devices. Although subject device output has minor differences with some predicate devices, and RF Energy= RF Output power × On Time, subject device conducts treatment duration to achieve same treatment with predicates devices. Therefore, the minor differences do not raise any new safety and effectiveness results because the parameters of the Subject device are similar to those of the predicate device. #### Substantial Equivalence Table 7. A comparison of the intended use and technological characteristics of the subject device and predicate device is provided in the table below: | Items | Subject device | Predicate device<br>(K192545) | Predicate device<br>(K160469) | Predicate device<br>(K222260) | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Description | Unicorn+ RF<br>System | POTENZA | AGNES | Capenergy C<br>Equipment RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 | | Manufacturer | Boston Aesthetics<br>INC | Jeisys Medical<br>Incorporated | Gowoonsesang<br>Cosmetics Co.,<br>Ltd | Capenergy Medical<br>S.L. | | Classification<br>Product Code | GEI,PBX | GEI | GEI,KCW | PBX | | Indication for<br>Use | The Unicorn+<br>RF System is<br>intended for use<br>in dermatologic<br>and general<br>surgical<br>procedures for<br>electrocoagulatio<br>n and hemostasis.<br>The RF System is<br>used for the relief<br>of minor muscle<br>aches and pain,<br>muscle spasm,<br>and temporary<br>improvement of<br>local blood<br>circulation | The POTENZA is<br>intended for use in<br>dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation<br>and hemostasis. | AGNES is<br>indicated for use<br>in dermatological<br>and general<br>surgical<br>procedures for<br>electrocoagulation<br>and hemostasis. | Capenergy C<br>equipments RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 are intended to<br>provide topical heating<br>for the purpose of<br>elevating tissue<br>temperature for<br>treatment of selected<br>medical conditions<br>such as: relief of pain,<br>muscle<br>spasms, increase in<br>local circulation, and<br>dermatological<br>procedures.<br>The massage device<br>provided is intended to<br>provide a temporary<br>reduction in the<br>appearance of cellulite. | {6}------------------------------------------------ | | | | | Capenergy C<br>equipments RF System<br>- C25, C50, C100,<br>C200, C300, C400,<br>C500 are indicated for<br>use in dermatological<br>and general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | |------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Prescription/Ov<br>er-the Counter<br>Use | Prescription | Prescription | Prescription | Prescription | | System Type | Bipolar RF<br>(Radiofrequency)<br>and Monopolar<br>RF<br>(Radiofrequency) | Bipolar RF<br>(Radiofrequency)<br>and Monopolar<br>RF(Radiofrequency) | Monopolar RF<br>(Radiofrequency) | Monopolar RF<br>(Radiofrequency) | | Frequency | 1MHz | 1MHz, 2Mhz | 1MHz | 1MHz | | Max Output<br>Power | 50W | 50W | 46W | 45W | | Treatment<br>Duration(Time) | MicroRF:10-600<br>ms<br>Artisit:<br>50ms-5000ms<br>Pure+B1:1-30min | 5-500ms | Min 50ms / Max<br>1,000ms | 15-660 sec | | Tips | single electrode:<br>ArtistD2.0,<br>2.0mm<br>ArtistD3.5,<br>3.5mm<br>ArtistD4.5,<br>4.5mm<br>ArtistD6.0,<br>6.0mm<br>Pure+B1, 3cm^2<br>dual electrodes:<br>MicroRF 49,<br>0.5-3.5mm<br>MicroRF 25,<br>0.5-3.5mm | Depth: Maximum of<br>4.0mm<br>Thickness : 0.25mm | Needle length :<br>0.8/1.25/1.5/ 2.0<br>mm<br>Thickness :<br>0.2mm | N.A. | | | MicroRF 9N,<br>0.5-3.5mm<br>MicroRF 25N,<br>0.5-3.5mm | | | | | Single Use<br>or Reusable | Single<br>Use/Resusable | Single<br>Use/Resusable | Single Use | Reusable | | Sterilization | EO gas | EO gas | EO gas | N.A. | {7}------------------------------------------------ Unicorn+ RF System belongs Product Code GEI and PBX in Classification,and does not own Needle-type epilator,so that not belong to KCW. Therefore, Unicorn+ RF System Classification code are all same as three predicate devices code. The Unicorn+ RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. This description is same as POTENZA and AGNES. The RF System is used for the relief of minor muscle aches and pain, muscle spasm, and temporary improvement of local blood circulation. This description is same as Capenergy C Equipment RF System statement, which Capenergy C equipments RF System -C25, C50, C100, C200, C300, C500 are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation, and dermatological procedures. The Unicorn+RF system owns 3 modes matched 3 different types of tips. MicroRF, is Bipolar RF (Radiofrequency) .which is same as POTENZA Bipolar tips in 1Mhz mode, Moreover, Artist&Pure+ tips are Monopolar RF (Radiofrequency) same as others. Treatment depends on device supplied energy , the energy is affected by treatment duration and power supply. Micro RF tips treatment duration is minimal different from POTENZA ones. Artist tips treatment is 50ms-5000ms, much longer than AGNES treatment duration, however, Artist tips max output power is designed 20W, and AGNES is 46W, thus the influence on energy with tissue is also minimal. Finally on Pure+ tips is not a point of use, this tip is effect on skin area, thus the treatment duration depends on treatment area size. MicroRF tips length is no more than POTENZA tips maximum length; Some of Artist tips lengths are longer than AGNES, but tips lengths is effect on treatment target tissue area under the skin, because all tips divided into non-insulated and insulated part, only non-insulated release energy to target tissue. Conclusion, Unicorn+RF system is designed similar with predicate devices, they are same indications for use, and the operating principle and technological characteristics. Although there are some differences as output power, treatment duration and tip physical dimensions. the thermal damage tests which is compare with the predicate devices and performance test reports are supported to the safety and effectiveness of the subject device. #### 8. Performance Date The following performance data were provided in support of the substantial equivalence determination: ## Biocompatibility The biocompatibility evaluation for electrode was conducted in accordance with the International Standard ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing {8}------------------------------------------------ within a Risk Management Process, as recognized by FDA. The subject device is contacted with patient during the treatment for duration of less than 24 hours. The biocompatibility testing includes the following: - · In Vitro Cytotoxicity Test: ISO 10993-5: 2009 - · Skin Sensitization Test: ISO 10993-10:2021 - · Intracutaneous Reactivity Test: ISO 10993-23:2021 - · Acute Systemic Toxicity Test: ISO 10993-11:2017 - · Pyrogen Test: ISO 10993-11:2017 Sterilization Validation Sterilization validation of the electrode has been conducted per the standard ISO 10993-7:2008 for Ethylene Oxide(standard ISO 11135:2014). # Shelf-life Validation The shelf-life validation of the device has been conducted using the accelerated aging method in accordance to ASTM F1980-16. ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety testing and EMC testing were conducted on the Radio frequency System complying with the IEC 60601-1 standard, the IEC 60601-2-2 standard ,the IEC 60601-1-6 standard for safety and the IEC 60601-1-2 standard for EMC. # Software Verification and Validation Testing Software Verification and Validation Testing were conducted and documentation was provided as recommended by FDA relevant Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device was considered as a 'moderate' level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator. # Ex Vivo testing Ex Vivo testing conducted on four types of tissue-Live, Kidney, Skin and Muscle under GLP Thermal testing in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014 ## Summary discussion Non-clinical tests data Based on bench test, the device demonstrated ability to reach and maintain therapeutic temperature 40-45°C on the surface of skin, and the actual skin temperature test was conducted. No preclinical and clinical data was provided in this submission #### 9. Conclusion In comparing between the subject device and the predicate devices, there are the same indications for use, the operating principle and technological characteristics .Although there are some differences as output power, treatment duration and tip physical dimensions. the thermal damage test which is compare with the predicate devices and performance test reports are supported to the safety and effectiveness of the subject device. {9}------------------------------------------------ In this regard, we conclude that the subject device is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...